A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals
Open accessThe DOI matches its Crossref record; 63 of the 63 references it cites were found in a registry or an index; 5 of the 50 sources its claims cite were checked for support.
Abstract
Seismocardiography (SCG) is a measure of chest vibration associated with heartbeats. While skin soft electronic tattoos (e‐tattoos) have been widely reported for electrocardiogram (ECG) sensing, wearable SCG sensors are still based on either rigid accelerometers or non‐stretchable piezoelectric membranes. This work reports an ultrathin and stretchable SCG sensing e‐tattoo based on the filamentary serpentine mesh of 28‐µm‐thick piezoelectric polymer, polyvinylidene fluoride (PVDF). 3D digital image correlation (DIC) is used to map chest vibration to identify the best location to mount the e‐tattoo and to investigate the effects of substrate stiffness. As piezoelectric sensors easily suffer from motion artifacts, motion artifacts are effectively reduced by performing subtraction between a pair of identical SCG tattoos placed adjacent to each other. Integrating the soft SCG sensor with a pair of soft gold electrodes on a single e‐tattoo platform forms a soft electro‐mechano‐acoustic cardiovascular (EMAC) sensing tattoo, which can perform synchronous ECG and SCG measurements and extract various cardiac time intervals including systolic time interval (STI). Using the EMAC tattoo, strong correlations between STI and the systolic/diastolic blood pressures, are found, which may provide a simple way to estimate blood pressure continuously and noninvasively using one chest‐mounted e‐tattoo.
Overview
CoChat Summary
TL;DR
This paper reports an ultrathin, stretchable chest-laminated e-tattoo using filamentary serpentine 28-µm PVDF for seismocardiogram (SCG) sensing and integrated Au electrodes for synchronous ECG/SCG (EMAC) sensing. It demonstrates high-fidelity SCG/ECG, motion-artifact reduction via dual PVDF subtraction, and a strong negative correlation between the systolic time interval (RAC) and blood pressure across four subjects.
Key findings
- The FS PVDF e-tattoo is 28 µm thick, has 112.9% stretchability and 0.4 mV per µε sensitivity, and an effective modulus of 8.5 MPa; the integrated EMAC tattoo measures 63.5 mm × 38.1 mm × 0.122 mm, weighs 150 mg, and has about 100% stretchability.
- 3D DIC mapped chest vibration: S1 was ≈5 µm around the left nipple area and S2 was ≈3 µm at a higher location; the middle e-tattoo was best for measuring both S1 and S2. FEM estimated substrate-dependent outputs of 2.74 mV (no substrate), 2.20 mV (Tegaderm), and 0.86 mV (PET).
- Dual PVDF subtraction reduced motion artifacts, making motion-period SNRs comparable to rest-period SNRs; a single PET-substrate sensor had negative SNR under motion.
- In four subjects, RAC negatively correlated with blood pressure: average Pearson R was −0.772 for SBP–RAC, −0.858 for DBP–RAC, and −0.538 for PBP–RAC; individual calibration was necessary.
Methods
- The device was fabricated by a cut-and-paste method to pattern FS PVDF and Au/PET electrodes on Tegaderm; characterization used DMA, electrical measurements, and ABAQUS FEM. 3D DIC used two high-speed cameras at 500 Hz for 16 s to map chest deformation.
- Human experiments were IRB-approved. Synchronous ECG and SCG were measured by the EMAC tattoo and compared with an accelerometer; blood pressure was measured by SOMNOtouch NIBP during Valsalva maneuvers in four subjects. Signals were filtered with 2–40 Hz (ECG) and 12–40 Hz (SCG) Butterworth filters, and RAC intervals were extracted from R and AC peaks.
- Analysis included Pearson correlations between BP and RAC, plus FEM modeling of substrate effects and electromechanical output.
Limitations
- Stated: fitting parameters and correlation coefficients varied by individual and posture; no master fitting equations could be derived, so per-individual calibration is necessary. Valsalva data were omitted because both SOMNO and SCG signals suffered excessive noise from body shivering, and mismatches were likely from motion or slight posture change.
- Stated: wireless NFC or Bluetooth integration remains future work for practical ambulatory applications.
- Evident from the text: BP validation involved only four subjects, and the main DIC mapping was performed on an adult male; no long-term ambulatory validation is reported.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
4 Valid · 4 Not valid · 31 Not checked · 114 Own results and interpretations
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“In contrast, there has been little progress in the development of soft SCG sensors.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Introduction
What the sources say
- Could not verifycites [9] D.‐H. Kim, N. Lu, R. Ma, Y.‐S. Kim, R.‐H. Kim, S. Wang, J. Wu, S. M. Won, H. Tao, A. Islam, K. J. Yu, T.‐i. Kim, R. Chowdhury, M. Ying, L. Xu, M. Li, H.‐J. Chung, H. Keum, M. McCormick, P. Liu, Y.‐W. Zhang, F. G. Omenetto, Y. Huang, T. Coleman, J. A. Rogers, Science 2011, 333, 838. 21836009 sourceExperimental — accuracy not yet measured
The Abstract says that many skin-mounted physiological devices still use conventional designs, but it does not assess progress in soft SCG sensors. Searches for SCG and seismocardiography found passages about ECG and other sensors, not the claimed lack of progress in SCG sensors. (LLM, conf=0.75)
- Not checked — no full text of the sourcecites [10] S. X. Yang, Y. C. Chen, L. Nicolini, P. Pasupathy, J. Sacks, B. Su, R. Yang, D. Sanchez, Y. F. Chang, P. L. Wang, D. Schnyer, D. Neikirk, N. S. Lu, Adv. Mater. 2015, 27, 6423. 26398335 source
- Not checked — no full text of the sourcecites [11] S. Kabiri Ameri, R. Ho, H. Jang, L. Tao, Y. Wang, L. Wang, D. M. Schnyer, D. Akinwande, N. Lu, ACS Nano 2017, 11, 7634. 28719739 source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Physical strapping or taping is also required to reduce the acoustic mismatch caused by the air gap between the rigid accelerometer and the skin.7”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Introduction
What the sources say
- Partly supportedcites [7] Y. Liu, J. J. Norton, R. Qazi, Z. Zou, K. R. Ammann, H. Liu, L. Yan, P. L. Tran, K.‐I. Jang, J. W. Lee, Sci. Adv. 2016, 2, e1601185. 28138529 PMC5262452 sourceExperimental — accuracy not yet measured
Materials and Methods (Device characterization; page not provided) describes attaching a sensor in a rigid acrylic box to the throat with adhesive and medical tape. Results (Device characterization; page not provided) says mechanical impedance matching with skin matters. The paper does not say that strapping or taping is generally required to reduce an acoustic mismatch caused by an air gap. (LLM, conf=0.80)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Several wearable sensors and energy harvesters have employed electrospun PVDF fibers, but electrodes could only be placed at the ends of the fiber bundles instead of along the full length of the fiber,22 which compromises the electrical output and, subsequently, the sensitivity.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Introduction
What the sources say
- Partly supportedcites [22] a) L. Persano, C. Dagdeviren, Y. W. Su, Y. H. Zhang, S. Girardo, D. Pisignano, Y. G. Huang, J. A. Rogers, Nat. Commun. 2013, 4, 1633; 23535654 b) Y. Q. Duan, Y. A. Huang, Z. P. Yin, N. B. Bu, W. T. Dong, Nanoscale 2014, 6, 3289; 24509570 c) C. E. Chang, V. H. Tran, J. B. Wang, Y. K. Fuh, L. W. Lin, Nano Lett. 2010, 10, 726. 20099876 sourceExperimental — accuracy not yet measured
The Results (page unavailable) describe electrospun P(VDF-TrFe) sensors with electrical contacts at the ends of fibre arrays. The paper does not say electrodes could only be placed there or that end contacts compromise output and sensitivity; it reports a large response to minute pressures from those devices. (LLM, conf=0.86)
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In Figure 5d, we find that the R peak of the ECG and the MC peak of the SCG are almost aligned, but the R peak is much easier to extract as it is the highest peak in ECG whereas the MC peak is neither aligned with S1 nor S2.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > EMAC Sensing E‐Tattoo
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Here the result of the dual sensor noise cancellation with PET substrate was provided instead of with Tegaderm substrate to prove the dual sensor system can restore the SCG signal even for the severe motion noise due to the stiff substrate.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > Dual‐PVDF‐Sensor for Motion Artifact Cancellation
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 3 assumptions not shown here.
90 more claims
Create a free account to see all 153 claims and to verify the 7 references still open.
- It may state two separate claims as one. 40
- Probably not a substantial claim. 20
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“There have been multiple previous works that have delved into using dual-mode ECG and SCG sensing for cardiac time intervals measurements. [26-29]”
From A Chest‐Conformable, Wireless Electro‐Mechanical E‐Tattoo for Measuring Multiple Cardiac Time Intervals (2023) · cites this paper as [29]
In the citing paper
“Analogue and digital signal processing methods are often applied to the captured waveforms to remove motion artefacts and improve the SNR; multisensor configurations can also be used for active noise-cancelling and localization by beamforming118,119.”
From End-to-end design of wearable sensors (2022) · cites this paper as [119]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > EMAC Sensing E‐Tattoo
“f) Continuous SBP and DBP estimated by SOMNOtough (black) and EMAC tattoo (red) during Valsalva maneuver.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > Correlation between Hemodynamic Parameters and BP
“Data during the Valsalva maneuver was omitted as both the SOMNO and SCG signals suffered from excessive noise induced by body shivering, as evident in Figure S11 (Supporting Information).”
Citations
Cites
Global status report on noncommunicable diseases 2010.
Alwan, A. (2011) · Cited by 8,494
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationAge-predicted maximal heart rate revisited
Tanaka, H., Monahan, K. D., & Seals, D. R. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01054-8 · Cited by 3,775
Direct-Write Piezoelectric Polymeric Nanogenerator with High Energy Conversion Efficiency
Chang, C., Tran, V. H., Wang, J., Fuh, Y.-K., & Lin, L. (2010) · Nano Letters · doi:10.1021/nl9040719 · Cited by 1,379
Persano, L., Dagdeviren, C., Su, Y., Zhang, Y., Girardo, S., Pisignano, D., Huang, Y., & Rogers, J. A. (2013) · Nature Communications · doi:10.1038/ncomms2639 · Cited by 1,271
Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0287-x · Cited by 895
Fundamental Understanding of Piezoelectric Strain Sensors
Sirohi, J., & Chopra, I. (2000) · Journal of Intelligent Material Systems and Structures · doi:10.1106/8bfb-gc8p-xq47-ycq0 · Cited by 777
Self-citationBallistocardiography and Seismocardiography: A Review of Recent Advances
Inan, O. T., Migeotte, P.-F., Park, K.-S., Etemadi, M., Tavakolian, K., Casanella, R., Zanetti, J., Tank, J., Funtova, I., Prisk, G. K., et al. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2361732 · Cited by 775
Flexible and Stretchable Electronics for Biointegrated Devices
Kim, D.-H., Ghaffari, R., Lu, N., & Rogers, J. A. (2012) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071811-150018 · Cited by 749
Self-citationGraphene Electronic Tattoo Sensors
Kabiri Ameri, S., Ho, R., Jang, H., Tao, L., Wang, Y., Wang, L., Schnyer, D. M., Akinwande, D., & Lu, N. (2017) · ACS Nano · doi:10.1021/acsnano.7b02182 · Cited by 669
Self-citation
Research works printed but not linked: 15
- Castiglioni P., 2007, Proc. of the Annual Int. IEEE EMBS, 3954
- Hendriks F. M., 2005, Technische Universiteit Eindhoven, 106
- Specialties M., 2013, Measurement, 3
- Lionel W. Y. L., 2016, IEEE Eng. Med. Biol., 38, 4816
- Ameri S. K., 2016, Int. Electron Devices Meeting
- Paukkunen M., 2013, J. Bioeng. Biomed. Sci., 3, 1
- Singh A., 2013, J. Dental Mater. Tech., 4, 682
- Obropta E. W., 2016, Aerospace Conf. Proc.
- Obropta E. W., 2015, IEEE Aerospace Conf.
- 10.1063/1.2817234 · DOI · Paper page
- Pandia K., 2010, 2010 IEEE Int. Conf. on Acoustics Speech and Signal Processing (ICASSP)
- Di Rienzo M., 2012, Annual Int. Conf. of the IEEE on Engineering in Medicine and Biology Society (EMBC), 5050
- Korzeniowska‐Kubacka I., 2006, Folia Cardiol., 13, 319
- Imtiaz M. S., 2013, IEEE Canadian Conf. on Electrical and Computer Engineering (CCECE), 501
- Aubert X. L., 2007, Proc. of the Annual Int. IEEE EMBS, 1707
Cited by
Items 101–200 of 218
Parylene C-Based, Breathable Tattoo Electrodes for High-Quality Bio-Potential Measurements
Spanu, A., Mascia, A., Baldazzi, G., Fenech‐Salerno, B., Torrisi, F., Viola, G., Bonfiglio, A., Cosseddu, P., & Pani, D. (2022) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2022.820217 · Cited by 30
Zhao, F., Li, T., Gu, Y., Lei, B., Jiang, W., Wang, X., Wang, L., Jiang, W., & Liu, H. (2024) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c16628 · Cited by 26
Fully rubbery Schottky diode and integrated devices
Jang, S., Shim, H., & Yu, C. (2022) · Science Advances · doi:10.1126/sciadv.ade4284 · Cited by 27
Flexible micro-strain graphene sensors enhanced by laser-induced cracks for health monitoring
Yang, J., Wu, S., Yu, J., Deng, Y., Qiao, F., & Zhang, K. (2024) · Diamond and Related Materials · doi:10.1016/j.diamond.2024.111401 · Cited by 28
Sayyad, P. W., Al-Nashash, H. A., Al–Othman, A., & Al-Sayah, M. H. (2025) · IEEE Transactions on Components Packaging and Manufacturing Technology · doi:10.1109/tcpmt.2025.3539978 · Cited by 29
Robust Skin-Conformal Nano-Electrodes for Sustainable Health and Performance Monitoring
Kim, J., Park, S., Jeon, J., Kang, D.-H., Bryan, G. M., Broderick, T. J., Stone, M. O., & Tsukruk, V. V. (2025) · ACS Nano · doi:10.1021/acsnano.5c08540 · Cited by 33
Sustainable Liquid Metal Composites for Soft Electronics and E‐Waste Reduction
Hajalilou, A., Parvini, E., & Majidi, C. (2025) · Advanced Materials · doi:10.1002/adma.202512968 · Cited by 36
Biodegradable Flexible Electronic Device with Controlled Drug Release for Cancer Treatment
Li, H., Gao, F., Wang, P., Yin, L., Ji, N., Zhang, L., Zhao, L., Hou, G., Lu, B., Chen, Y., et al. (2021) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c04653 · Cited by 27
Self-citationFlexible Hybrid Piezoelectric‐Electrostatic Device for Energy Harvesting and Sensing Applications
Lu, Z., Zhang, C., Kwon, S. H., Jiang, Z., & Dong, L. (2023) · Advanced Materials Interfaces · doi:10.1002/admi.202202173 · Cited by 29
How Far for the Electronic Skin: From Multifunctional Material to Advanced Applications
Chen, Y., Sun, Y., Wei, Y., & Qiu, J. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202201352 · Cited by 26
MXene-Based Flexible Electrodes for Electrophysiological Monitoring
Alex, M., Khan, K. R. B., Al–Othman, A., Al-Sayah, M. H., & Al-Nashash, H. A. (2024) · Sensors · doi:10.3390/s24113260 · Cited by 24
“Effortless monitoring of bioelectric signals such as electrocardiogram (ECG) [4,5], electroencephalogram (EEG) [6–8], and electromyogram (EMG) [9] necessitates the use of surface electrodes capable of signal capture from skin surfaces with minimal discomfort.”
BackgroundHigh-Sensitivity Flexible Pressure Sensor-Based 3D CNTs Sponge for Human–Computer Interaction
Cui, J., Nan, X., Shao, G., & Sun, H. (2021) · Polymers · doi:10.3390/polym13203465 · Cited by 26
Quan, L., Tie, J., Wang, Y., Mao, Z., Zhang, L., Zhong, Y. P., Sui, X., Feng, X., & Xu, H. Y. (2022) · Polymers for Advanced Technologies · doi:10.1002/pat.5643 · Cited by 25
3R Batteries: Resilient, Repairable, and Recyclable Based on Liquid Gallium Electrode
Parvini, E., Hajalilou, A., Lopes, P. A., Silva, A. F., Tiago, M. S. M., Fernandes, P. M. P., de Almeida, A. T., & Tavakoli, M. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202301189 · Cited by 24
Joshi, S. R., Lee, M., Lee, H. Y., Lee, M.-E., & Kim, S. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.154160 · Cited by 25
Flexible Sensors for IoT-Based Health Monitoring
Majumder, S., Roy, A. K., Mondal, T. K., & Deen, M. J. (2025) · IEEE Journal on Flexible Electronics · doi:10.1109/jflex.2025.3538808 · Cited by 31
“An ultrathin and stretchable e-tattoo was introduced in [102] to monitor chest vibrations induced by the heartbeat.”
Leogrande, E., Filosa, M., Ballanti, S., Cicco, L. D., Mazzoleni, S., Ackerley, R., Oddo, C. M., & Dell’Olio, F. (2025) · Sensors and Actuators Reports · doi:10.1016/j.snr.2025.100312 · Cited by 32
Recent Advances in Materials and Flexible Sensors for Arrhythmia Detection
Guess, M., Zavanelli, N., & Yeo, W. (2022) · Materials · doi:10.3390/ma15030724 · Cited by 27
“Ha et al. showed a stretchable e-tattoo SCG based on polyvinylidene fluoride (PVDF) [16].”
BackgroundInfluentialYuan, X., Won, S. M., Han, M., Wang, Y., Rogers, J. A., Huang, Y. Y., & Wang, H. (2021) · Journal of the Mechanics and Physics of Solids · doi:10.1016/j.jmps.2021.104400 · Cited by 21
Ganti, V. G., Carek, A. M., Jung, H., Srivatsa, A. V., Cherry, D., Johnson, L. N., & Inan, O. T. (2021) · JMIR mhealth and uhealth · doi:10.2196/27466 · Cited by 25
Interfacial Polarization Locked Flexible β‐Phase Glycine/Nb2CTx Piezoelectric Nanofibers
Zheng, W., Li, T., Jin, F., et al. (2024) · Small · doi:10.1002/smll.202308715 · Cited by 22
Materials Advances in Devices for Heart Disease Interventions
Jalandhra, G. K., Srethbhakdi, L., Davies, J., Nguyen, C. C., Phan, P. T., Och, Z., Ashok, A., Lim, K. S., Phan, H., Do, T. N., et al. (2025) · Advanced Materials · doi:10.1002/adma.202420114 · Cited by 24
Contactless, Battery-free, and Stretchable Wearable for Continuous Recording of Seismocardiograms
Hesar, M. E., Khan, D., Seyedsadrkhani, N. S., & Ingebrandt, S. (2020) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.0c00768 · Cited by 22
Ye, Z., Li, Q., Zhang, R., Zhang, P., & Gui, L. (2022) · RSC Advances · doi:10.1039/d1ra09394k · Cited by 21
Fluid Microchannel Encapsulation to Improve the Stretchability of Flexible Electronics
Jin, P., Wang, P., Pang, W., Wang, H., Jiao, Y., Zhang, J., Lu, B., Liu, B., Chen, Y., Ma, Y., et al. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202101126 · Cited by 19
Self-citationRecent Advances in Self-Powered Wearable Flexible Sensors for Human Gaits Analysis
Hu, X., Ma, Z., Zhao, F., & Guo, S. L. (2024) · Nanomaterials · doi:10.3390/nano14141173 · Cited by 24
Liu, S., Ha, T., & Lu, N. (2019) · Journal of Applied Mechanics · doi:10.1115/1.4042570 · Cited by 19
“The two piezoelectric coefficients d31 and d32, the dielectric constant ε3, and the Young’s modulus Y of the PVDF thin film have been measured in our previous study [38].”
MethodSelf-citationChest-scale self-compensated epidermal electronics for standard 6-precordial-lead ECG
Yin, L., Wang, Y., Zhan, J., Bai, Y., Hou, C., Wu, J., Huang, R. T., Wang, Y., & Huang, Y. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00159-7 · Cited by 24
Highly Robust and Strain-Resilient Thin Film Conductors Featuring Brittle Materials
Chen, K., Zhang, L., Wu, K., Yang, C., Wang, R., Xu, C., Zhang, J., Liu, G., & Sun, J. (2023) · Nano Letters · doi:10.1021/acs.nanolett.3c01781 · Cited by 19
Self-citationHighly Conformal Polymers for Ambulatory Electrophysiological Sensing
Hu, M., Zhang, J. P., Liu, Y., Zheng, X., Li, X., Li, X., & Yang, H. (2022) · Macromolecular Rapid Communications · doi:10.1002/marc.202200047 · Cited by 20
A Wrist-worn Respiration Monitoring Device using Bio-Impedance
Sel, K., Brown, A., Jang, H., Krumholz, H. M., Lu, N., & Jafari, R. (2020) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc44109.2020.9176367 · Cited by 19
“We previously shared the cut-and-paste manufacturing method of the tattoos as well as the acquisition of certain signal modalities such as ECG, SCG, and capacitive sensing [20], [21].”
MethodSelf-citationRecent Advance in Stretchable Self‐Powered Piezoelectric Sensors: Trends, Challenges, and Solutions
Huang, X. M., & Zhang, X. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202301226 · Cited by 19
Next-generation tattoo-like-electronics with promising fabrication and wider application scenarios
Zhou, Y., Qamar, O. A., Hwang, G. B., Knapp, C. E., Li, G., Lubineau, G., & Tai, Y. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.157336 · Cited by 20
Jiang, Q., Zhang, S., Jiang, J., Fei, W., & Wu, Z. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202000966 · Cited by 18
A Flexible PVDF Sensor for Forcecardiography
Parlato, S., Centracchio, J., Cinotti, E., Gargiulo, G. D., Esposito, D., Bifulco, P., & Andreozzi, E. (2025) · Sensors · doi:10.3390/s25051608 · Cited by 18
“The flexibility of PVDF sensors allows them to be integrated into wearable devices, making continuous, long-term monitoring feasible [36,45].”
Shao, R., Wang, C., Zhao, J., Yang, H., & Sun, S. (2021) · Advanced Science · doi:10.1002/advs.202101876 · Cited by 15
Shin, G., Baek, J. Y., Kim, J. H., Lee, J. H., Kim, H. J., So, B. J., Choi, Y., Yun, S., Kim, T., Jeon, J. G., et al. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202412524 · Cited by 33
Smart epidermal electrophysiological electrodes: Materials, structures, and algorithms
Ye, Y., Wang, H., Tian, Y., Gao, K., Wang, M., Wang, X., Liang, Z., You, X., Gao, S., Shao, D., et al. (2023) · Nanotechnology and Precision Engineering · doi:10.1063/10.0019678 · Cited by 14
Zhao, J., Feng, J., Jiang, Y., Wang, Z., Zhang, J., Mei, S., Yang, G., Gu, Z., Tan, C., Qin, Y., et al. (2023) · Small · doi:10.1002/smll.202309470 · Cited by 14
Smart Tattoo Sensors 2.0: A Ten-Year Progress Report through a Narrative Review
Pirrera, A., & Giansanti, D. (2024) · Bioengineering · doi:10.3390/bioengineering11040376 · Cited by 19
“Ha et al. [50] introduced a stretchable, ultrathin seismocardiography (SCG) sensing e-tattoo based on a 28-µ m-thick piezoelectric polymer, PVDF.”
BackgroundMethodInfluentialFlexible Wearable Heart Rate Monitoring System and Low-Power Design: A Review
Zheng, C., Yong, C., Wei, Q., & Qiao, F. (2025) · Sensors · doi:10.3390/s25164913 · Cited by 20
“For instance, Lu et al. [61] reported an electronic tattoo sensor based on PVDF, realizing a self-powered and high-precision sensing module.”
InfluentialLiao, Z., Golparvar, A. J., Bathaei, M. J., Cardoso, F. A., & Boutry, C. M. (2025) · Advanced Science · doi:10.1002/advs.202509934 · Cited by 20
Hesar, M. E., Seyedsadrkhani, N. S., Khan, D., Naghashian, A., Piekarski, M., Gall, H. J., SCHERMULY, R. T., Ghofrani, H. A., & Ingebrandt, S. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115693 · Cited by 18
Cardiac Time Intervals Under Motion Using Bimodal Chest E-Tattoos and Multistage Processing
Bhattacharya, S., Santucci, F., Jankovic, M., Huang, T. C., Basu, J., Tan, P., Schena, E., & Lu, N. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3454067 · Cited by 14
“These seismic vibrations generated by the heartbeat can be recorded on the chest surface using accelerometers, gyroscopes, fiber optic sensors, or piezoelectric films [1]– [6].”
BackgroundSelf-citationBlood pressure monitoring techniques in the natural state of multi-scenes: A review
Liu, Z., Zhou, C., Wang, H., & He, Y. (2022) · Frontiers in Medicine · doi:10.3389/fmed.2022.851172 · Cited by 15
A Flexible Near-Field Biosensor for Multisite Arterial Blood Flow Detection
Mohammed, N., Cluff, K., Sutton, M., Villafana-Ibarra, B., Loflin, B. E., Griffith, J. L., Becker, R. A., Bhandari, S., Alruwaili, F., & Desai, J. (2022) · Sensors · doi:10.3390/s22218389 · Cited by 14
Highly Sensitive Piezoelectric E‐Skin Design Based on Electromechanical Coupling Concept
Yang, X., Zhang, M., Xie, M., Sun, M., Luo, H., Li, Q., Chen, X., & Pang, W. (2023) · Advanced Electronic Materials · doi:10.1002/aelm.202201339 · Cited by 14
Non-invasive Cardiac Output Monitoring in Congenital Heart Disease
Tandon, A., Bhattacharya, S., Morca, A., Inan, O. T., Munther, D. S., Ryan, S. D., Latifi, S. Q., Lu, N., Lasa, J. J., Marino, B. S., et al. (2023) · Current Treatment Options in Pediatrics · doi:10.1007/s40746-023-00274-1 · Cited by 12
“Recent work on developing ultra-thin ( 200 μm) thick, soft, chest-conformable non-invasive e-tattoos has the potential to improve the wearability and signal quality of mobile CO/CI sensors for children [18, 19].”
Motion Artifact‐Resilient Zone for Implantable Sensors
Jeong, C., Koirala, G. R., Jung, Y. H., Ye, Y. S., Hyun, J. H., Kim, T. H., Park, B., Ok, J., Jung, Y., & Kim, T. (2022) · Advanced Functional Materials · doi:10.1002/adfm.202206461 · Cited by 12
Yang, X., Zhang, M., Gong, S., Sun, M., Xie, M., Niu, P., & Pang, W. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202201043 · Cited by 12
Emerging epidermal electrodes towards digital health and on-skin digitalization
Wang, Y. (2023) · Soft Science · doi:10.20517/ss.2023.44 · Cited by 14
Epidermal Bioelectronics for Management of Chronic Diseases: Materials, Devices and Systems
Qiao, Z., Chen, S., Fan, S., Xiong, Z., & Lim, C. T. (2023) · Advanced Sensor Research · doi:10.1002/adsr.202200068 · Cited by 13
Ke, Y., Li, T., Li, J., Pei, M., Wang, X., Xie, W., Zhuang, S., Ye, X., Li, Z., Wang, Z., et al. (2025) · eScience · doi:10.1016/j.esci.2025.100436 · Cited by 15
Seismocardiography-based detection of heartbeats for continuous monitoring of vital signs
Cocconcelli, F., Mora, N., Matrella, G., & Ciampolini, P. (2019) · Computer Science and Electronic Engineering Conference · doi:10.1109/ceec47804.2019.8974343 · Cited by 11
“stretchable piezoelectric sensor for acquiring SCG and ECG waveforms [22].”
MethodShao, R., Ma, R., An, X., Wang, C., & Sun, S. (2022) · Journal of Materials Chemistry C · doi:10.1039/d1tc04384f · Cited by 12
Han, L., Zhang, W., Fan, C., Qi, Z., Liu, Y., Wang, Y., Zhao, Y., Zhang, B., Chen, X., Dong, Y., et al. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202512101 · Cited by 10
Optimisation and material considerations of piezoelectric implants for cardiac applications
Wang, Y., & Dargusch, M. S. (2025) · Current Opinion in Solid State and Materials Science · doi:10.1016/j.cossms.2024.101211 · Cited by 9
Citizenship as the exception to the rule: an addendum
Jaynes, T. L. (2020) · AI & Society · doi:10.1007/s00146-020-01105-9 · Cited by 12
“…international judiciaries and is arguably one of the most pressing topics that needs to be addressed given the progress of technological implantations worldwide (Saal and Bensmaia 2015; Saadi, Touhami and Yagoub 2017; Ha et al. 2019; Jeong et al. 2019; Zhuang et al. 2019; Bumbaširević et al. 2020).”
BackgroundElectromagnetic based flexible bioelectronics and its applications
Pan, S., Zhou, M., Liu, L., & Shen, H. (2024) · Frontiers in Electronics · doi:10.3389/felec.2024.1240603 · Cited by 9
Hu, X., He, Q., Ma, H., Li, J., Jiang, Y., & Wang, K. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c21230 · Cited by 9
Flexible and wearable bioelectronics for electrocardiography monitoring: a review
Zheng, C., Wang, Z., Li, L., Ma, Y., & Feng, X. (2025) · Science China Materials · doi:10.1007/s40843-025-3688-9 · Cited by 13
Validation of a Novel Tattoo Electrode for ECG Monitoring
Baldazzi, G., Spanu, A., Mascia, A., Viola, G., Bonfiglio, A., Cosseddu, P., & Pani, D. (2021) · Computing in cardiology · doi:10.23919/cinc53138.2021.9662799 · Cited by 8
“In fact, one of the main characteristics of tattooable systems is the extreme thinness of the employed substrates [7], [8], which also allows for an optimal skin/electrode interface without the use of conductive hydrogels, thus resulting in extremely unobtrusive recording electrodes.”
BackgroundCuffless Blood Pressure Monitoring: Academic Insights and Perspectives Analysis
Li, S., Zhang, C., Xu, Z., Liang, L., Tian, Y., Li, L., Wu, H., & Zhong, S. (2022) · Micromachines · doi:10.3390/mi13081225 · Cited by 10
“However, with the development of PWA, the research on pressure pulse wave signals and ultrasonic pulse wave signals will be more concerned with offering BP measurements in a cuffless and wearable manner [36,56,68,69].”
Yang, X., Zhang, M., Niu, P., Guo, W., Sun, C., Pang, W., Cui, G., & Liu, Q. (2024) · SmartMat · doi:10.1002/smm2.1271 · Cited by 7
A Mechanical–Electrochemical Dual-Model E-Skin for the Monitoring of Cardiovascular Healthcare
Fang, J., Jia, Y., Liao, Z., Qi, B., & Huang, T. (2024) · Biosensors · doi:10.3390/bios15010005 · Cited by 7
Electrically Stable and Reliably Adhered E-Skin for Heart Health Monitoring
Zhao, X., Zhao, Z., Wen, X., et al. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.5c02596 · Cited by 7
Conforming strategies for bioelectronics on arbitrary surfaces
Liang, X., Yu, B., Yokota, T., & Someya, T. (2025) · npj Biosensing · doi:10.1038/s44328-025-00064-9 · Cited by 15
“2b), to the relatively fl at surfaces of the chest or back 39,40 (Fig.”
A Portable Cardiac Dynamic Monitoring System in the Framework of Electro-Mechano-Acoustic Mapping
Gao, Z., Wang, Y., Xu, X., Zhang, C., Dai, Z., Zhang, H., Zhang, J., & Yang, H. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3307188 · Cited by 10
“CombinedsignalanalysisofECGandSCG[15],[16],[17],[23], [24],ECGandPCG[12],[13],[14],[18],[19],[20],[21],ECG and PCG and SCG [22], PCG and SCG [30], [31], [32], [33], and SCG and GCG [29] were performed to reflect cardiac activity and hemodynamic changes.”
InfluentialMann, A. J., Rahman, M. M., Vanga, V., Gamage, P. T., & Taebi, A. (2024) · Journal of Medical Devices · doi:10.1115/1.4066368 · Cited by 8
Exploring the Impact of Sensor Location on Seismocardiography-Derived Cardiac Time Intervals
Mann, A. J., Gamage, P. T., Kakavand, B., & Taebi, A. (2023) · Journal of Engineering and Science in Medical Diagnostics and Therapy · doi:10.1115/1.4063203 · Cited by 7
Copper‐Leaf‐Based Process for Imperceptible Computational Electronics
de Mulatier, S., Coulon, D., Delattre, R., Blayac, S., & Ramuz, M. P. (2019) · Advanced Electronic Materials · doi:10.1002/aelm.201900787 · Cited by 9
Development and Characterization of a Self-Tightening Tourniquet System
Vega, S. J., Torres, S. I. H., Berard, D., Boice, E. N., & Snider, E. J. (2022) · Sensors · doi:10.3390/s22031122 · Cited by 6
Recent trend in stretchable composite sensors for wearable robot applications
Zaheer, M. U., & Chang, S. (2023) · Advanced Composite Materials · doi:10.1080/09243046.2023.2270379 · Cited by 6
Park, B., Ok, J., Park, S., Yang, Y. J., Jeon, H., & Kim, T. (2026) · Advanced Materials · doi:10.1002/adma.202517030 · Cited by 6
Artificial Intelligence-Driven Wearable Sensors for Cardiovascular Health Monitoring
Deng, H. T., Ma, B., Chen, G., Chi, J., Tong, J., Yu, F., Lü, B., Chen, Y., Zhao, C., Zhang, C., et al. (2026) · ACS Sensors · doi:10.1021/acssensors.5c04840 · Cited by 8
Piezoelectric biosensors for healthcare applications
Mutegoa, E., & Sahini, M. G. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-21658-9.02001-5 · Cited by 10
Han, L., Liang, W., Liu, Y., Xue, W., Wang, J., Yang, Z., Zhou, Q., Dong, Y., & Wang, X. (2024) · Applied Materials Today · doi:10.1016/j.apmt.2024.102077 · Cited by 7
Schmitter‐Edgecombe, M., Luna, C. M., & Cook, D. J. (2022) · Springer eBooks · doi:10.1007/978-3-031-11389-5_4 · Cited by 10
Multimodal Hydrogel Bioelectronics for Closed‐Loop Diagnosis and Treatment of Cardiac Diseases
Wang, F., Xue, Y., Wang, Y., Wang, Q., Wen, P., Chen, X., Wang, C., Shan, L., Wu, K., Lu, B., et al. (2026) · Advanced Functional Materials · doi:10.1002/adfm.202527184 · Cited by 10
Liu, R., Swanson, S., Huang, E., Krejsa, M. R., MacKenzie, J. D., Kleweno, C. P., Taroc, A., Nelson, L. Y., & Seibel, E. J. (2025) · Journal of Medical Devices · doi:10.1115/1.4068595 · Cited by 4
An analytical review for cloud computing in healthcare system based biosensors
Awsaj, M. K., Mashhadany, Y. I. M. A., & Fourati, L. C. (2022) · International Journal of Health Sciences · doi:10.53730/ijhs.v6ns9.12297 · Cited by 6
“[38]outlined a smartwear device dependant on a sensor of PPG located on the finger, betweenfinger glove and wrist strap.”
BackgroundGao, Z., Wang, Y., Yu, K., Dai, Z., Song, T., Zhang, J., Huang, C., Zhang, H., & Yang, H. (2024) · Sensors · doi:10.3390/s24072235 · Cited by 4
“…of multi-frequency vibration also requires high sensor sensitivity and PVDF sensors without a substrate have the best sensitivity because they allow sufficient epidermis deformation [11]. sensors without a substrate have the best sensitivity because they allow su ffi cient epidermis deformation [11].”
BackgroundMethodInfluentialDesign and Fabrication of Integrated Hollow Concave Cilium Mems Heart Sound Sensor
Wang, B., Li, Y., Yang, Y., Cui, J., Zhang, G., Wang, R., Zhang, W., He, C., Shi, P., & Wang, S. (2022) · Micromachines · doi:10.3390/mi13122174 · Cited by 6
“presented an E-tattoo sensor capable of recording ECG, SCG, and heart sounds in 2019 [14].”
Self-citationWireless Fully Passive Package-Embedded Seismocardiogram With RF Backscattering
Sayeed, S. Y. B., Navaz, H. V., Volakis, J. L., & Raj, P. M. (2023) · IEEE Transactions on Components Packaging and Manufacturing Technology · doi:10.1109/tcpmt.2023.3298915 · Cited by 3
Rigid and soft microelectrodes for electrophysiology measurement
Gong, H., Weng, W., Zhang, S., Gao, Z., & Hu, N. (2025) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2025.100591 · Cited by 4
Muto, V., Andreozzi, E., Cappelli, C., Centracchio, J., Meo, G. D., Esposito, D., Bifulco, P., & De, D. (2023) · Electronics · doi:10.3390/electronics12204359 · Cited by 3
Wireless wearable devices for continuous monitoring of body sounds and motions
Kim, S. H., Jeanne, E., Shalish, W., Yoo, J., & Rogers, J. A. (2024) · Clinical and Translational Medicine · doi:10.1002/ctm2.1593 · Cited by 4
“…soft electronic devices designed for sensing subtle body vibrations and sounds, with additional capabilities in tracking bulk motions of the body and noises 5,7–10 in the ambient environment through the use of multi-sensor designs and sound separation data analytics, respectively (Figure 1).”
BackgroundFunctional materials-enabled flexible electronic skin for flow field decoding
Hu, X., Sheng, G., Chen, Y., & Zhao, F. (2025) · Frontiers in Electronics · doi:10.3389/felec.2025.1528802 · Cited by 2
Low-cost and convenient screening of disease using analysis of physical measurements and recordings
Chandra, J., Lin, R., Kancherla, D., Scott, S. C., Sul, D., Andrade, D., Marzouk, S., Iyer, J. M., Wasswa, W., Villanueva, C., et al. (2024) · PLOS Digital Health · doi:10.1371/journal.pdig.0000574 · Cited by 2
Mazumdar, H., Khondakar, K. R., Das, S., & Kaushik, A. K. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202501304 · Cited by 3
Wu, Y., Liu, J., Hong, H., et al. (2026) · Advanced Materials Technologies · doi:10.1002/admt.202502277 · Cited by 1
An Electron Devices Perspective on the Technology Road Map for Implanted Electronics: A Review
Mahmud, F., Bello, A., Vijayrania, P., Kumar, A., & Budhraja, V. (2026) · IEEE Electron Devices Magazine · doi:10.1109/med.2026.3706522 · Cited by 1
Lee, J., Hwang, Y. M., & Park, S. (2023) · Applied Sciences · doi:10.3390/app132413101 · Cited by 1
“…material and device integration technology have created a new era for wearable and flexible healthcare sensors that include cardiopulmonary monitoring devices [10,12,23–31], more practical and simplified devices with evidence regarding their use in the real-world intensive care unit (ICU) are needed.”
BackgroundTang, C., Wang, H., Dou, Y., & Lai, P. (2024) · ACS Omega · doi:10.1021/acsomega.4c05316 · Cited by 1
Kim, J., Kim, H., & Yoon, H. (2025) · Korean Journal of Chemical Engineering · doi:10.1007/s11814-025-00436-x · Cited by 1
“3B) [48].”
Wearable Technologies in Lifestyle Medicine
Mechanick, J. I., & Zhao, S. (2020) · Springer eBooks · doi:10.1007/978-3-030-48088-2_13 · Cited by 2
Ag/polypyrrole composites for polarizable dry microsensor used in bioimpedance measurement
Telipan, G., Pîslaru‐Dănescu, L., Ion, I., et al. (2020) · CAS proceedings · doi:10.1109/cas50358.2020.9267982 · Cited by 1
“As a result, wearable devices are being developed to perform continuous cardiovascular monitoring for telemedicine and outpatients [1].”
BackgroundSmart Seismocardiography: A Machine Learning Approach for Automatic Data Processing
López-Rico, O. Y., & Ramírez-Chavarría, R. G. (2021) · International Electronic Conference on Sensors and Applications (ECSA) · doi:10.3390/ecsa-8-11325 · Cited by 1
Rhythmusmonitoring: „New tools on the horizon“
von Falkenhausen, A. S., Fichtner, S., Kääb, S., & Sinner, M. F. (2020) · Herzschrittmachertherapie + Elektrophysiologie · doi:10.1007/s00399-020-00706-7 · Cited by 1
Soft, wearable devices to monitor electrophysiological signals and gaseous biomarkers
Herbert, R., Popoola, M. O., Zhang, W., & Cheng, H. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-99147-6.00006-5 · Cited by 1
Metadata
- Authors
- Ha, T., Tran, J., Liu, S., Jang, H., Jeong, H., Mitbander, R., Huh, H., Qiu, Y., Duong, J., Wang, R. L., Wang, P., Tandon, A., Sirohi, J., & Lu, N.
- Affiliations
- The University of Texas at Austin (US)
- Texas Materials Institute (US)
- Children's Medical Center (US)
- Corresponding author
- Lu, Nanshu
- ORCID
- Venue
- Advanced Science
- Publisher
- Wiley
- Published
- 2019-05-21
- Volume
- 6
- Issue
- 14
- Pages
- 1900290-1900290
- ISSN
- 2198-3844 (ISSN-L)
- PMC
- PMC6662084
- OpenAlex
- W2945869729
- Cited by
- 220
- References
- 63
- Citations by year
- 2026: 15 · 2025: 28 · 2024: 33 · 2023: 34 · 2022: 44 · 2021: 34 · 2020: 26 · 2019: 6
- FWCI
- 9.68
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Advanced Science (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Funders
- Office of Naval Research (N00014‐16‐1‐2044)
- Air Force Office of Scientific Research (FA9550‐15‐1‐0112)
- National Science Foundation (CNS‐1738293, UL1TR001105)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- seismocardiography, stretchable sensors, polyvinylidene fluoride, piezoelectric polymers, electrocardiogram, chest vibration, digital image correlation, motion artifact reduction, systolic time intervals, non-invasive blood pressure estimation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.