Advances in MRI around metal
Open accessThe DOI matches its Crossref record; 66 of the 66 references it cites were found in a registry or an index.
Abstract
The prevalence of orthopedic metal implants is continuously rising in the aging society. Particularly the number of joint replacements is increasing. Although satisfying long‐term results are encountered, patients may suffer from complaints or complications during follow‐up, and often undergo magnetic resonance imaging (MRI). Yet metal implants cause severe artifacts on MRI, resulting in signal‐loss, signal‐pileup, geometric distortion, and failure of fat suppression. In order to allow for adequate treatment decisions, metal artifact reduction sequences (MARS) are essential for proper radiological evaluation of postoperative findings in these patients. During recent years, developments of musculoskeletal imaging have addressed this particular technical challenge of postoperative MRI around metal. Besides implant material composition, configuration and location, selection of appropriate MRI hardware, sequences, and parameters influence artifact genesis and reduction. Application of dedicated metal artifact reduction techniques including high bandwidth optimization, view angle tilting (VAT), and the multispectral imaging techniques multiacquisition variable‐resonance image combination (MAVRIC) and slice‐encoding for metal artifact correction (SEMAC) may significantly reduce metal‐induced artifacts, although at the expense of signal‐to‐noise ratio and/or acquisition time. Adding advanced image acquisition techniques such as parallel imaging, partial Fourier transformation, and advanced reconstruction techniques such as compressed sensing further improves MARS imaging in a clinically feasible scan time. This review focuses on current clinically applicable MARS techniques. Understanding of the main principles and techniques including their limitations allows a considerate application of these techniques in clinical practice. Essential orthopedic metal implants and postoperative MR findings around metal are presented and highlighted with clinical examples. Level of Evidence: 4 Technical Efficacy: Stage 3 J. Magn. Reson. Imaging 2017;46:972–991.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Sparse MRI: The application of compressed sensing for rapid MR imaging
Lustig, M., Donoho, D., & Pauly, J. M. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21391 · Cited by 6,952
MR Procedures: Biologic Effects, Safety, and Patient Care
Shellock, F. G., & Crues, J. V. (2004) · Radiology · doi:10.1148/radiol.2323030830 · Cited by 563
Metal-Induced Artifacts in MRI
Hargreaves, B. A., Worters, P. W., Pauly, K. B., Pauly, J. M., Koch, K. M., & Gold, G. E. (2011) · American Journal of Roentgenology · doi:10.2214/ajr.11.7364 · Cited by 559 · cited 10×
“In the algorithmic approach to diagnosis and management of metal-on-metal arthroplasty published in 2012 by the hip society,113 imaging should include ultrasound or metal artifact reduction sequence MRI in addition to plain radiography.113,114,144,145 MARS sequences including VAT and SEMAC have been shown to be useful…”
BackgroundInfluentialLee, M.-J., Kim, S., Lee, S.-A., Song, H.-T., Huh, Y.-M., Kim, D.-H., Han, S. H., & Suh, J.-S. (2007) · RadioGraphics · doi:10.1148/rg.273065087 · Cited by 513
SEMAC: Slice encoding for metal artifact correction in MRI
Lu, W., Pauly, K. B., Gold, G. E., Pauly, J. M., & Hargreaves, B. A. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21967 · Cited by 392 · cited 3×
“In addition to in-plane artifacts, SEMAC reduces throughplane distortions by correcting for signal that is excited in wrong slice positions.(71,72,81) SEMAC is based on a 2D TSE sequence.”
Fat and water magnetic resonance imaging
Bley, T. A., Wieben, O., François, C. J., Brittain, J. H., & Reeder, S. B. (2010) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21895 · Cited by 372 · cited 1×
“water-only and fat-only image reconstructions.(50,60) Images may be reconstructed with and without fat suppression.”
BackgroundFast and quiet MRI using a swept radiofrequency
Idiyatullin, D., Corum, C., Park, J.-Y., & Garwood, M. (2006) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2006.05.014 · Cited by 353
A multispectral three‐dimensional acquisition technique for imaging near metal implants
Koch, K. M., Lorbiecki, J. E., Hinks, R. S., & King, K. F. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21856 · Cited by 269
Stradiotti, P., Curti, A., Castellazzi, G., & Zerbi, A. (2009) · European Spine Journal · doi:10.1007/s00586-009-0998-5 · Cited by 241
Imaging near metal with a MAVRIC‐SEMAC hybrid
Koch, K. M., Brau, A. C., Chen, W., Gold, G. E., Hargreaves, B. A., Koff, M., McKinnon, G. C., Potter, H. G., & King, K. F. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22523 · Cited by 230 · cited 5×
“Siemens (Erlangen, Germany) calls its dedicated MARS techniques “WARP” (conventional MARS techniques plus VAT) and “advanced WARP” (WARP plus SEMAC) (in alphabetic order).(25,69,70,72) The efficacy of these advanced metal artifact reduction techniques with respect to artifact reduction as compared to conventional sequ…”
BackgroundMethodInfluentialTotal inhomogeneity correction including chemical shifts and susceptibility by view angle tilting
Cho, Z. H., Kim, D. J., & Kim, Y. K. (1988) · Medical Physics · doi:10.1118/1.596162 · Cited by 217
Minimizing artifacts caused by metallic implants at MR imaging: experimental and clinical studies.
Suh, J. S., Jeong, E. K., Shin, K. H., Cho, J. H., Na, J. B., Kim, D. H., & Han, C. D. (1998) · American Journal of Roentgenology · doi:10.2214/ajr.171.5.9798849 · Cited by 183 · cited 1×
“constant field of view) reduces both in-plane distortions and intravoxel dephasing.(66) Matrix sizes of 512 are recommended.”
BackgroundMagnetic resonance imaging near metal implants
Koch, K. M., Hargreaves, B. A., Pauly, K. B., Chen, W., Gold, G. E., & King, K. F. (2010) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22313 · Cited by 182 · cited 7×
“Siemens (Erlangen, Germany) calls its dedicated MARS techniques “WARP” (conventional MARS techniques plus VAT) and “advanced WARP” (WARP plus SEMAC) (in alphabetic order).(25,69,70,72) The efficacy of these advanced metal artifact reduction techniques with respect to artifact reduction as compared to conventional sequ…”
BackgroundMethodInfluentialMRI Safety Update 2008: Part 2, Screening Patients for MRI
Shellock, F. G., & Spinazzi, A. (2008) · American Journal of Roentgenology · doi:10.2214/ajr.08.1038.2 · Cited by 171
Sutter, R., Ulbrich, E. J., Jellus, V., Nittka, M., & Pfirrmann, C. W. A. (2012) · Radiology · doi:10.1148/radiol.12112408 · Cited by 161 · cited 2×
“It was demonstrated in several studies that SEMAC is superior to standard MR sequences, high bandwidth protocols, and simple VAT with respect to artifact reduction.(1,2,84,85)”
Self-citationMR Imaging with Metal-suppression Sequences for Evaluation of Total Joint Arthroplasty
Talbot, B. S., & Weinberg, E. (2015) · Radiographics · doi:10.1148/rg.2016150075 · Cited by 139 · cited 3×
“Several abnormalities around orthopedic implants may be visualized by using MARS sequences.(49,105) Aseptic loosening of an implant is characterized by a circumferential lucency around the implant or the cement, which is progressively enlarging over time,(44,105) while an osteolysis has a more localized extent (Fig.”
BackgroundMR imaging of articular cartilage of the knee: new methods using ultrashort TEs.
Gold, G. E., Thedens, D. R., Pauly, J. M., Fechner, K. P., Bergman, G. A., Beaulieu, C. F., & Macovski, A. (1998) · American Journal of Roentgenology · doi:10.2214/ajr.170.5.9574589 · Cited by 117
Reduction of blurring in view angle tilting MRI
Butts, K., Pauly, J. M., & Gold, G. E. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20375 · Cited by 108 · cited 2×
“VAT does not increase scan time, but the slice shear induces some blurring.(80) Using thin slices may keep the blurring at an acceptable level.”
BackgroundMethodLamellated Hyperintense Synovitis: Potential MR Imaging Sign of an Infected Knee Arthroplasty
Plodkowski, A. J., Hayter, C. L., Miller, T. T., Nguyen, J., & Potter, H. G. (2012) · Radiology · doi:10.1148/radiol.12120042 · Cited by 108 · cited 1×
“In the presence of a TKA pseudotumors, osteolysis, loosening, periprosthetic fracture, and infection are frequent queries.(157) Other possible complications after TKA are arthrofibrosis, collateral ligament rupture, quadriceps tendon rupture, retropatellar osteoarthritis, and patellar clunk syndrome.”
Carbon fiber intramedullary nails reduce artifact in postoperative advanced imaging
Zimel, M. N., Hwang, S., Riedel, E., & Healey, J. H. (2015) · Skeletal Radiology · doi:10.1007/s00256-015-2158-9 · Cited by 105
MR Imaging of Knee Arthroplasty Implants
Fritz, J. M., Lurie, B., & Potter, H. G. (2015) · Radiographics · doi:10.1148/rg.2015140216 · Cited by 100 · cited 1×
“Knee After the hip, the knee is the second most frequently replaced joint.(2,44,155,156) MARS techniques work successfully at the knee despite the minor surrounding soft tissue and the more complex metal anatomy (Fig.”
Fritz, J. M., Ahlawat, S., Demehri, S., Thawait, G. K., Raithel, E., Gilson, W. D., & Nittka, M. (2016) · Investigative Radiology · doi:10.1097/rli.0000000000000317 · Cited by 91
Fritz, J. M., Lurie, B., & Miller, T. T. (2013) · Seminars in Musculoskeletal Radiology · doi:10.1055/s-0033-1348098 · Cited by 90
Metal Artifact Reduction With MAVRIC SL at 3-T MRI in Patients With Hip Arthroplasty
Choi, S.-J., Koch, K. M., Hargreaves, B. A., Stevens, K. J., & Gold, G. E. (2014) · American Journal of Roentgenology · doi:10.2214/ajr.13.11785 · Cited by 80 · cited 4×
“MAVRIC-SL excites singles slices like SEMAC but uses the pulse profiles from MAVRIC.69,70 Not only standard TSE sequences and STIR, but also ultrashort echo time sequences (UTE)86 are compatible with MAVRIC.”
MethodInfluentialRamos‐Cabrer, P., van Duynhoven, J., van der Toorn, A., & Nicolay, K. (2004) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2004.01.061 · Cited by 73
Fritz, J. M., Fritz, B., Thawait, G. K., Raithel, E., Gilson, W. D., Nittka, M., & Mont, M. A. (2016) · Skeletal Radiology · doi:10.1007/s00256-016-2437-0 · Cited by 70 · cited 3×
“Using CS, the phase-encoding steps may be reduced, resulting in an 8-fold acceleration of k-space-encoding and in a dramatic reduction of the acquisition time.(99,102) Image quality with CS-SEMAC is comparable to SEMAC sequences, but with a scan time of 4–5 minutes instead of 10– 12 minutes, which drastically increase…”
MethodOptimization of metal artefact reduction (MAR) sequences for MRI of total hip prostheses
Toms, A. P., Smith-Bateman, C., Malcolm, P. N., et al. (2010) · Clinical Radiology · doi:10.1016/j.crad.2009.12.014 · Cited by 64 · cited 3×
“Doubling or tripling the bandwidth to 500–800 Hz/pixel is recommended, although this substantially reduces the SNR.(64) Consequently, the scan averages may need to be increased to receive enough signal, which increases the scan time.”
BackgroundCompressed‐Sensing multispectral imaging of the postoperative spine
Worters, P. W., Sung, K., Stevens, K. J., Koch, K. M., & Hargreaves, B. A. (2012) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.23750 · Cited by 64
Quantifying image distortion of orthopedic materials in magnetic resonance imaging
Koff, M. F., Shah, P. H., Koch, K. M., & Potter, H. G. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.23991 · Cited by 64 · cited 2×
“MAVRIC is a spatially nonselective 3D acquisition technique to reduce metal artifacts.(4,30,70) The 3D phase-encoding does not suffer from through-plane distortions, as is the case for any gradient based slice-selection.”
BackgroundFilli, L., Luechinger, R., Frauenfelder, T., Beck, S., Guggenberger, R., Farshad‐Amacker, N. A., & Andreisek, G. (2014) · Skeletal Radiology · doi:10.1007/s00256-014-2057-5 · Cited by 62 · cited 8×
“In the algorithmic approach to diagnosis and management of metal-on-metal arthroplasty published in 2012 by the hip society,113 imaging should include ultrasound or metal artifact reduction sequence MRI in addition to plain radiography.113,114,144,145 MARS sequences including VAT and SEMAC have been shown to be useful…”
BackgroundInfluentialSutter, R., Hodek, R., Fucentese, S. F., Nittka, M., & Pfirrmann, C. W. A. (2013) · American Journal of Roentgenology · doi:10.2214/ajr.13.10531 · Cited by 60 · cited 2×
“It was demonstrated in several studies that SEMAC is superior to standard MR sequences, high bandwidth protocols, and simple VAT with respect to artifact reduction.(1,2,84,85)”
BackgroundSelf-citationAssessment of MRI Issues at 7 T for 28 Implants and Other Objects
Dula, A. N., Virostko, J. M., & Shellock, F. G. (2014) · American Journal of Roentgenology · doi:10.2214/ajr.13.10777 · Cited by 57 · cited 1×
“Certain implants may be acceptable, whereas others may not be safe at 7T.(41,42) Among these potentially unsafe implants are endovascular grafts and several orthopedic implants.”
BackgroundRunge, V. M. (2013) · Investigative Radiology · doi:10.1097/01.rli.0000434380.71793.d3 · Cited by 56 · cited 8×
“In the algorithmic approach to diagnosis and management of metal-on-metal arthroplasty published in 2012 by the hip society,113 imaging should include ultrasound or metal artifact reduction sequence MRI in addition to plain radiography.113,114,144,145 MARS sequences including VAT and SEMAC have been shown to be useful…”
BackgroundInfluentialMR Imaging Near Metallic Implants Using MAVRIC SL
Gutierrez, L. B., Do, B. H., Gold, G. E., Hargreaves, B. A., Koch, K. M., Worters, P. W., & Stevens, K. J. (2014) · Academic Radiology · doi:10.1016/j.acra.2014.09.010 · Cited by 55
MR imaging near metal with undersampled 3D radial UTE‐MAVRIC sequences
Carl, M., Koch, K. M., & Du, J. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24219 · Cited by 54 · cited 1×
“Using an undersampled 3D radial UTEMAVRIC sequence, imaging of tissues with short T2 such as tendons, ligaments, and cortical bone has been described as feasible adjacent to metallic implants.(87)”
MethodUlbrich, E. J., Sutter, R., Aguiar, R. F., Nittka, M., & Pfirrmann, C. W. A. (2012) · American Journal of Roentgenology · doi:10.2214/ajr.11.8233 · Cited by 50 · cited 3×
“A very efficient measure for reducing in-plane artifacts in conventional MR sequences is increasing the receiver bandwidth.(52,62,64) If stronger imaging gradients are applied, each voxel is encoded by a wider frequency range.”
BackgroundSelf-citationPostoperative MRI of the shoulder
Beltrán, L., Bencardino, J. T., & Steinbach, L. S. (2014) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24570 · Cited by 48
Skare, S. T., & Andersson, J. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20528 · Cited by 47
Evaluation of 39 medical implants at 7.0 T
Feng, D. X., McCauley, J. P., Morgan-Curtis, F., Salam, R. A., Pennell, D., Loveless, M. E., & Dula, A. N. (2015) · British Journal of Radiology · doi:10.1259/bjr.20150633 · Cited by 45 · cited 2×
“Among these potentially unsafe implants are endovascular grafts and several orthopedic implants.(41)”
BackgroundRadzi, S., Cowin, G., Robinson, M., Pratap, J., Volp, A., Schuetz, M., & Schmutz, B. (2014) · PubMed · doi:10.3978/j.issn.2223-4292.2014.03.06 · Cited by 44 · cited 1×
“5T and 3T MRI can be used for evaluation of the articular surface of the ankle, if titanium screws are at least 3 mm distant.(158) MRI with Dixon sequences and spoiled gradient-echo (SPGR) sequences enabled significantly improved visualization of articular cartilage in patients with metal implants of the ankle, with r…”
Metal artifacts caused by gradient switching
Graf, H., Steidle, G., Martirosian, P., Lauer, U. A., & Schick, F. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20524 · Cited by 42
Kumar, N. M., Netto, C. D. C., Schon, L. C., & Fritz, J. M. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000350 · Cited by 42 · cited 3×
“For many clinical situations, using increased readout bandwidth is considered to be the most effective parameter for artifact reduction.(62)”
BackgroundMethodResultJungmann, P. M., Ganter, C., Schaeffeler, C., Bauer, J. S., Baum, T. H., Meier, R., Nittka, M., Pohlig, F., Rechl, H., von Eisenhart‐Rothe, R., et al. (2015) · PLoS ONE · doi:10.1371/journal.pone.0124922 · Cited by 41 · cited 6×
“Exemplarily, while for some implants seven SEMAC steps may be sufficient since the through-plane distortion only involves the adjacent three slices, for stainless-steel or complex implants, 13 or even more SEMAC steps may not be sufficient to account for the entire through-plane distortion.(7,32) The amount of SEMAC s…”
BackgroundMethodInfluentialSelf-citationLiebl, H., Heilmeier, U., Lee, S., Nardo, L., Patsch, J., Schuppert, C., Han, M., Rondak, I., Banerjee, S., Koch, K. M., et al. (2014) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24668 · Cited by 40
Dillenseger, J., Molière, S., Choquet, P., Goetz, C., Ehlinger, M., & Bierry, G. (2016) · Skeletal Radiology · doi:10.1007/s00256-016-2338-2 · Cited by 39 · cited 10×
“In the algorithmic approach to diagnosis and management of metal-on-metal arthroplasty published in 2012 by the hip society,113 imaging should include ultrasound or metal artifact reduction sequence MRI in addition to plain radiography.113,114,144,145 MARS sequences including VAT and SEMAC have been shown to be useful…”
BackgroundMethodInfluentialZou, Y.-F., Chu, B., Wang, C.-B., & Hu, Z. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2014.12.001 · Cited by 37
Shellock, F. G., & Karacozoff, A. M. (2013) · American Journal of Roentgenology · doi:10.2214/ajr.12.10332 · Cited by 33
Optimized Imaging of the Postoperative Spine
McLellan, A. M., Daniel, S. C., Corcuera‐Solano, I., Joshi, V., & Tanenbaum, L. N. (2014) · Neuroimaging Clinics of North America · doi:10.1016/j.nic.2014.01.005 · Cited by 29 · cited 1×
“Spine For metal artifact reduced imaging of the spine, the frequency-encoding direction should be positioned along the long axis of the pedicle screws in sagittal and transverse images (anterior–posterior).(121) The utility of SEMAC for artifact reduction in spine imaging has been demonstrated in several studies.”
BackgroundSparse‐SEMAC: rapid and improved SEMAC metal implant imaging using SPARSE‐SENSE acceleration
Otazo, R., Nittka, M., Bruno, M. T., Raithel, E., Geppert, C., Gyftopoulos, S., Recht, M. P., & Rybak, L. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26342 · Cited by 29 · cited 1×
“Using CS, the phase-encoding steps may be reduced, resulting in an 8-fold acceleration of k-space-encoding and in a dramatic reduction of the acquisition time.(99,102) Image quality with CS-SEMAC is comparable to SEMAC sequences, but with a scan time of 4–5 minutes instead of 10– 12 minutes, which drastically increase…”
MethodOff‐resonance suppression for multispectral MR imaging near metallic implants
den Harder, J. M., van Yperen, G. H., Blume, U. A., & Bos, C. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25126 · Cited by 26 · cited 3×
“Disadvantages of ORS include signal voids and loss of homogeneity.(88)”
BackgroundMethodInfluentialRipple artifact reduction using slice overlap in slice encoding for metal artifact correction
den Harder, J. M., van Yperen, G. H., Blume, U. A., & Bos, C. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25127 · Cited by 22 · cited 1×
“1).(83) It was demonstrated in several studies that SEMAC is superior to standard MR sequences, high bandwidth protocols, and simple VAT with respect to artifact reduction.”
BackgroundOlivieri, L., Cross, R. J., O’Brien, K., Ratnayaka, K., & Hansen, M. S. (2015) · Pediatric Radiology · doi:10.1007/s00247-015-3366-0 · Cited by 22
Saeedi, M., Thomas, A., & Shellock, F. G. (2015) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2015.01.012 · Cited by 22
Imaging the Knee in the Setting of Metal Hardware
Chang, E. Y., Bae, W. C., & Chung, C. B. (2014) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2014.07.009 · Cited by 21 · cited 14×
“In the algorithmic approach to diagnosis and management of metal-on-metal arthroplasty published in 2012 by the hip society,113 imaging should include ultrasound or metal artifact reduction sequence MRI in addition to plain radiography.113,114,144,145 MARS sequences including VAT and SEMAC have been shown to be useful…”
BackgroundInfluentialKnee implant imaging at 3 Tesla using high‐bandwidth radiofrequency pulses
Bachschmidt, T. J., Sutter, R., Jakob, P. M., Pfirrmann, C. W. A., & Nittka, M. (2014) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24729 · Cited by 20 · cited 2×
“Using high-bandwidth RF pulses has also been shown to substantially reduce through-plane distortion artifacts, at the cost of an increased blurring.(61) In contrast to previous assumptions, new studies have shown that longer echo trains do not reduce metal artifacts.”
BackgroundResultSelf-citationLee, J. B., Cha, J. G., Lee, M. H., Lee, Y. K., Lee, E. H., & Jeon, C. H. (2012) · Skeletal Radiology · doi:10.1007/s00256-012-1449-7 · Cited by 17
Kaushik, S. S., Karr, R., Runquist, M. L., Marszalkowski, C. S., Sharma, A., Rand, S. D., Maiman, D. J., & Koch, K. M. (2016) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25321 · Cited by 17
MRI Findings Associated with Recalled Modular Femoral Neck Rejuvenate and ABG Implants
Walsh, C. P., Hubbard, J. C., Nessler, J. P., & Markel, D. C. (2015) · The Journal of Arthroplasty · doi:10.1016/j.arth.2015.05.021 · Cited by 16
Trammell, T. R., Flint, K., & Ramsey, C. J. (2012) · Journal of Bone and Joint Surgery · doi:10.2106/jbjs.k.01470 · Cited by 15 · cited 1×
“However, even implants that cause more severe artifacts on MRI do not necessarily preclude proper radiological evaluation of the anatomic region, when imaging parameters are adjusted accordingly.(34,40)”
Deligianni, X., Bieri, O., Elke, R., Wischer, T. K., & Egelhof, T. (2015) · RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren · doi:10.1055/s-0041-104893 · Cited by 14 · cited 1×
“The larger the area of field inhomogeneities, the more slice-encoding steps need to be selected.(82) Reducing these steps reduces scan time.”
BackgroundSchmidt, M. A., Panek, R., Colgan, R. M., Hughes, J. C., Sohaib, A., Saran, F. H., Murray, J. R., Bernard, J., Revell, P., Nittka, M., et al. (2016) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2016.05.004 · Cited by 13
Qi, S., Wu, Z., Mu, Y.-F., Gao, L.-L., Yang, J., Zuo, P., Nittka, M., Liu, Y., Wang, H., & Yin, H. J. (2016) · Medicine · doi:10.1097/md.0000000000003184 · Cited by 12
Jin, Z. W., Guan, Y., Yu, G., & Sun, Y. (2016) · BioMed Research International · doi:10.1155/2016/1429892 · Cited by 11
Technical Update on Magnetic Resonance Imaging of the Shoulder
Vieira, R. L. R., Rybak, L., & Recht, M. P. (2012) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2012.01.005 · Cited by 10 · cited 2×
“Additionally, spherical components such as the humeral head replacement cause typical cloverleaf artifacts.(132,134) In cases with nonmetallic implants or suture anchors, standard sequences may be used”
Grierson, J., Lamb, C. R., & David, F. H. (2012) · Veterinary and Comparative Orthopaedics and Traumatology · doi:10.3415/vcot-12-03-0031 · Cited by 9 · cited 1×
“The frequency-encoding gradient polarity has also been described to have an influence on the appearance of metal artifacts.(26) The phase-encoding direction is spared by the artifacts, since the relative phase of each phase-encoding step is not affected by a B0 field offset.”
BackgroundSimulated field maps for susceptibility artefact correction in interventional MRI
Kochan, M., Daga, P., Burgos, N., White, M. J., Cardoso, M. J., Mancini, L., Winston, G. P., McEvoy, A. W., Thornton, J. S., Yousry, T. A., et al. (2015) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-015-1253-7 · Cited by 5 · cited 1×
“4).(53,54) This technique avoids additional artifacts caused by fat suppression techniques, but one needs to be aware of the fact that it does not eliminate geometric distortion artifacts occurring on the original images.”
Background
Cited by
MRI in traumatic spinal cord injury: from clinical assessment to neuroimaging biomarkers
Freund, P., Seif, M., Weiskopf, N., Friston, K., Fehlings, M. G., Thompson, A. J., & Curt, A. (2019) · The Lancet Neurology · doi:10.1016/s1474-4422(19)30138-3 · Cited by 222
Metadata
- Authors
- Jungmann, P. M., Agten, C. A., Pfirrmann, C. W., & Sutter, R.
- Venue
- Journal of Magnetic Resonance Imaging
- Publisher
- Wiley
- Published
- 2017-03-25
- Volume
- 46
- Issue
- 4
- Pages
- 972-991
- ISSN
- 1053-1807 (ISSN-L)
- 1522-2586
- OpenAlex
- W2603054283
- Cited by
- 227
- References
- 66
- FWCI
- 10.03
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.