A modified procedure for gas‐source isotope ratio mass spectrometry: the long‐integration dual‐inlet (LIDI) methodology and implications for clumped isotope measurements
The DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
RATIONALE: High‐precision stable isotope measurements in gas‐source isotope ratio mass spectrometry are generally carried out by repeated comparison of the composition of an unknown sample with that of a working gas (WG) through a dual‐inlet (DI). Due to the established DI protocols, however, most of the sample gas is wasted rather than measured, which is a major problem when sample size is limited. Here we propose a new methodology allowing the measurement of a much larger portion of the available sample. METHODS: We tested a new measurement protocol, the long‐integration dual‐inlet (LIDI) method, which consists of a single measurement of the sample for 200 to 600 seconds followed by a single measurement of the WG. The isotope ratios of the sample are calculated by comparison of the beam ratios of the WG and sample at equivalent intensities of the major ion beam. RESULTS: Three isotopically very different CO 2 samples were analyzed. The LIDI measurements of large samples (50 to 100 µmol of CO 2 ) measured at quasi‐constant beam sizes, and of small samples (1.5 to 2 µmol of CO 2 ) measured in micro‐volume mode, generated results that are indistinguishable from the standard DI measurements for carbon, oxygen and clumped isotope compositions. The external precision of Δ 47 using the LIDI protocol (~±0.007 ‰) is similar to that of the state of the art DI measurements. CONCLUSIONS: For traditional and clumped isotope measurements of CO 2 , the LIDI protocol allows the measurement of a much larger portion of the sample gas rather than only ~20 % of it. In addition, the sample can be measured at higher signal intensity and for longer time, allowing the measurement of smaller samples while preserving precision. We suggest that other gases commonly used for stable isotope measurements with gas‐source mass spectrometry would also benefit from this new protocol. Copyright © 2014 John Wiley & Sons, Ltd.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Coplen, T. B. (2011) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.5129 · Cited by 1,831
13C–18O bonds in carbonate minerals: a new kind of paleothermometer
Ghosh, P., Adkins, J., Affek, H., Balta, B., Guo, W., Schauble, E. A., Schrag, D., & Eiler, J. M. (2006) · Geochimica et Cosmochimica Acta · doi:10.1016/j.gca.2005.11.014 · Cited by 914
“Clumped-isotope” geochemistry—The study of naturally-occurring, multiply-substituted isotopologues
Eiler, J. M. (2007) · Earth and Planetary Science Letters · doi:10.1016/j.epsl.2007.08.020 · Cited by 774
Defining an absolute reference frame for ‘clumped’ isotope studies of CO2
Dennis, K. J., Affek, H. P., Passey, B. H., Schrag, D. P., & Eiler, J. M. (2011) · Geochimica et Cosmochimica Acta · doi:10.1016/j.gca.2011.09.025 · Cited by 627
Passey, B. H., Levin, N. E., Cerling, T. E., Brown, F. H., & Eiler, J. M. (2010) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1001824107 · Cited by 449
Paleoclimate reconstruction using carbonate clumped isotope thermometry
Eiler, J. M. (2011) · Quaternary Science Reviews · doi:10.1016/j.quascirev.2011.09.001 · Cited by 436
Huntington, K. W., Eiler, J. M., Affek, H. P., Guo, W., Bonifacie, M., Yeung, L. Y., Thiagarajan, N., Passey, B., Tripati, A., Daëron, M., et al. (2009) · Journal of Mass Spectrometry · doi:10.1002/jms.1614 · Cited by 436
18O13C16O in Earth’s atmosphere
Eiler, J. M., & Schauble, E. (2004) · Geochimica et Cosmochimica Acta · doi:10.1016/j.gca.2004.05.035 · Cited by 362
Bernasconi, S. M., Hu, B., Wacker, U., Fiebig, J., Breitenbach, S. F. M., & Rutz, T. (2013) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.6490 · Cited by 166
Self-citationBreitenbach, S. F. M., & Bernasconi, S. M. (2011) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.5052 · Cited by 164
Self-citation
Research works printed but not linked: 16
- 10.1063/1.1740961 · DOI · Paper page
- 10.1063/1.1745698 · DOI · Paper page
- 10.1002/(SICI)1096-9888(199603)31:3<225::AID-JMS319>3.0.CO;2-L · DOI · Paper page
- 10.1016/0146-6380(90)90147-R · DOI · Paper page
- 10.1016/B978-044451114-0/50040-5 · DOI
- 10.1029/95PA03773 · DOI
- 10.1002/rcm.6400 · DOI · Paper page
- 10.1016/j.epsl.2012.05.013 · DOI
- 10.1073/pnas.1109444108 · DOI · Paper page
- 10.1016/j.ijms.2012.10.014 · DOI
- 10.1029/2012JD017992 · DOI · Paper page
- 10.1016/j.gca.2005.08.021 · DOI · Paper page
- 10.1002/rcm.905 · DOI
- 10.1002/rcm.906 · DOI · Paper page
- 10.1063/1.1745483 · DOI
- 10.1002/rcm.6436 · DOI · Paper page
Cited by
Bernasconi, S. M., Daëron, M., Bergmann, K. D., Bonifacie, M., Meckler, A. N., Affek, H. P., Anderson, N., Bajnai, D., Barkan, E., Beverly, E., et al. (2021) · Geochemistry Geophysics Geosystems · doi:10.1029/2020gc009588 · Cited by 340
Self-citationBernasconi, S. M., Müller, I. A., Bergmann, K. D., Breitenbach, S. F. M., Fernandez, A., Hodell, D. A., Jaggi, M., Meckler, A. N., Millan, I., & Ziegler, M. (2018) · Geochemistry, Geophysics, Geosystems · doi:10.1029/2017gc007385 · Cited by 301
Self-citationJohn, C. M., & Bowen, D. (2016) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.7720 · Cited by 247
Meckler, A. N., Ziegler, M., Millán, M. I., Breitenbach, S. F. M., & Bernasconi, S. M. (2014) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.6949 · Cited by 239
“Further reductions in sample amounts required to achieve this precision are anticipated with a recently proposed different measurement protocol, the long-integration dualinlet (LIDI) method.([23]) These methodological improvements will further expand the applicability of clumped isotopes geochemistry to materials that…”
MethodSelf-citationSchauer, A. J., Kelson, J., Saenger, C., & Huntington, K. W. (2016) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.7743 · Cited by 157
Metadata
- Authors
- Hu, B., Radke, J., Schlüter, H., Heine, F. T., Zhou, L., & Bernasconi, S. M.
- Venue
- Rapid Communications in Mass Spectrometry
- Publisher
- Wiley
- Published
- 2014-05-22
- Volume
- 28
- Issue
- 13
- Pages
- 1413-1425
- ISSN
- 0951-4198 (ISSN-L)
- 1097-0231
- OpenAlex
- W2147109978
- Cited by
- 91
- References
- 15
- FWCI
- 1.90
- Type
- journal article
- Language
- en
- Open access
- closed
- Funders
- National Natural Science Foundation of China (41073008)
- Sino Swiss Science and Technology Cooperation (EG 03-032010)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A modified procedure for gas‐source isotope ratio mass spectrometry: the long‐integration dual‐inlet (LIDI) methodology and implications for clumped isotope measurements. https://app.cochat.ai/papers/10.1002/rcm.6909
- The paper
- Hu, B., Radke, J., Schlüter, H., Heine, F. T., Zhou, L., & Bernasconi, S. M. (2014). A modified procedure for gas‐source isotope ratio mass spectrometry: the long‐integration dual‐inlet (LIDI) methodology and implications for clumped isotope measurements. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.6909
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.