Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-Year review of the literature
Abstract
Objective Review of the accuracy and repeatability of Doppler cardiac output (CO) measurements in children.Design Publications in the scientific literature retrieved using a computerized Medline search from 1982-2002 and a manual review of article bibliographies. Studies comparing Doppler flow measurements with thermodilution, Fick, or dye dilution methods in the pediatric critical care setting were identified to assess the bias, precision, and intra- and interobserver repeatability of Doppler CO measurement. Where results were not suitable for comparison and the original measurements available, data were re-analyzed using appropriate statistical methods and presented in comparative tables.Results The precision of pediatric Doppler CO measurements compared to thermodilution, dye dilution, or Fick methods is around 30% and repeatability varies from less than 1% to 22%. Bias is generally less than 10% but varies considerably.Conclusions The bias, precision, and repeatability from study to study indicate that Doppler CO measurements are acceptably reproducible in children, with best results when used to track changes rather than absolute values, and using the transesophageal approach.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Despite its known limitations, including relatively high intra- and interobserver variability (coefficient of variation ranging from 2.1% to 22% and 3.1% to 22%, respectively), echocardiography remains the most widely adopted modality in neonatal and pediatric intensive care.
“Despite its known limitations, including relatively high intra- and interobserver variability (coefficient of variation ranging from 2.1% to 22% and 3.1% to 22%, respectively) [58], echocardiography remains the most widely adopted modality in neonatal and pediatric intensive care.”
From Trends and Challenges in Noninvasive Hemodynamic Monitoring of Neonates Following Cardiac Surgery: A Narrative Review (2025) · cites this paper as [58]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Elevation of Systemic Oxygen Delivery in the Treatment of Critically Ill Patients
Hayes, M., Timmins, A. C., Yau, E. H. S., Palazzo, M. G. A., Hinds, C., & Watson, D. I. (1994) · New England Journal of Medicine · doi:10.1056/nejm199406163302404 · Cited by 1,530 · see citing works
A Trial of Goal-Oriented Hemodynamic Therapy in Critically Ill Patients
Gattinoni, L., Brazzi, L., Pelosi, P. P., Latini, R., Tognoni, G., Pesenti, A. M., & Fumagalli, R. (1995) · New England Journal of Medicine · doi:10.1056/nejm199510193331601 · Cited by 1,486 · see citing works
Critchley, L. A. H., & Critchley, J. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009982611386 · Cited by 1,328 · see citing works
BISHOP, M. H., Shoemaker, W. C., APPEL, P. L., et al. (1995) · doi:10.1097/00005373-199505000-00018 · Cited by 349 · see citing works
Noninvasive pulsed Doppler determination of cardiac output in neonates and children
Alverson, D. C., Eldridge, M. W., Dillon, T., Yabek, S. M., & Berman, W. (1982) · The Journal of Pediatrics · doi:10.1016/s0022-3476(82)80178-9 · Cited by 232 · see citing works
Clinicians’ abilities to estimate cardiac index in ventilated children and infants
Tibby, S. M., Hatherill, M., Marsh, M. J., & Murdoch, I. A. (1997) · Archives of Disease in Childhood · doi:10.1136/adc.77.6.516 · Cited by 224 · see citing works
Darmon, P.-L., Hillel, Z., Mogtader, A. H., et al. (1994) · Anesthesiology · doi:10.1097/00000542-199404000-00011 · Cited by 186 · see citing works
Incidence of Complications in Pediatric Transesophageal Echocardiography: Experience in 1650 Cases
Stevenson, J. G. (1999) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(99)70090-8 · Cited by 150 · see citing works
Shoemaker, W. C. (1990) · Archives of Surgery · doi:10.1001/archsurg.1990.01410220116016 · Cited by 95 · see citing works
Tibby, S. M., Hatherill, M., & Murdoch, I. A. (2000) · Critical Care Medicine · doi:10.1097/00003246-200006000-00061 · Cited by 81 · see citing works
Reproducibility of measurements of cardiac output in newborn infants by Doppler ultrasound.
Hudson, I., Houston, A. B., Aitchison, T. C., et al. (1990) · Archives of Disease in Childhood · doi:10.1136/adc.65.1_spec_no.15 · Cited by 79 · see citing works
Mellander, M., Sabel, K., Caidahl, K., et al. (1987) · Pediatric Cardiology · doi:10.1007/bf02427536 · Cited by 71 · see citing works
Endotracheal Cardiac Output Monitor
Wallace, A. W., Salahieh, A., Lawrence, A. T., Spector, K., Owens, C., & Alonso, D. (2000) · Anesthesiology · doi:10.1097/00000542-200001000-00030 · Cited by 44 · see citing works
Rein, A. J. J. T., Hsieh, K., Elixson, M., et al. (1986) · American Heart Journal · doi:10.1016/0002-8703(86)90685-x · Cited by 43 · see citing works
Continuous haemodynamic monitoring in children: use of transoesophageal Doppler
Murdoch, I., Marsh, M., Tibby, S. M., & McLuckie, A. (1995) · Acta Paediatrica · doi:10.1111/j.1651-2227.1995.tb13751.x · Cited by 39 · see citing works
Noninvasive Measurement of Cardiac Output in Critically Ill Children
Mohan, U. R., Britto, J., Habibi, P., et al. (2002) · Pediatric Cardiology · doi:10.1007/s00246-001-0014-2 · Cited by 29 · see citing works
Claflin, K. S., Alverson, D. C., Pathak, D. R., et al. (1988) · Journal of Ultrasound in Medicine · doi:10.7863/jum.1988.7.6.311 · Cited by 27 · see citing works
Notterman, D. A., Castello, F. V., Steinberg, C., Greenwald, B. M., OʼLoughlin, J. E., & Gold, J. P. (1989) · The Journal of Pediatrics · doi:10.1016/s0022-3476(89)80280-x · Cited by 25 · see citing works
Continuous wave doppler cardiac output: Use in pediatric patients receiving inotropic support
Morrow, W. R., Murphy, D. J., Fisher, D. J., et al. (1988) · Pediatric Cardiology · doi:10.1007/bf02080552 · Cited by 25 · see citing works
Wodey, É., Gai, V., Carré, F., & Ecoffey, C. (2001) · Pediatric Anesthesia · doi:10.1046/j.1460-9592.2001.00677.x · Cited by 20 · see citing works
A Comparative Study of Cardiac Output Measurement by Dye Dilution and Pulsed Doppler Ultrasound
Tibballs, J., Osborne, A., & Hockmann, M. (1988) · Anaesthesia and Intensive Care · doi:10.1177/0310057x8801600306 · Cited by 20 · see citing works
Cardiac function in healthy infants and children: Doppler echocardiographic evaluation
Hanséus, K., Bj�rkhem, G., & Lundstr�m, N.-R. (1994) · Pediatric Cardiology · doi:10.1007/bf00795729 · Cited by 19 · see citing works
Pulmonary artery catheterization in children.
Thompson, A. E. (1997) · PubMed · Cited by 15 · see citing works
Sholler, G. F., Whight, C. M., & Gelermajer, J. M. (1986) · The American Journal of Cardiology · doi:10.1016/0002-9149(86)90701-0 · Cited by 14 · see citing works
Non-invasive assessment of cardiac output in children.
Richardson, J. R., Ferguson, J. G., Hiscox, J., et al. (1998) · Emergency Medicine Journal · doi:10.1136/emj.15.5.304 · Cited by 14 · see citing works
Welch, E. M., Duara, S., Suguihara, C., et al. (2009) · Biology of the Neonate · doi:10.1159/000244101 · Cited by 14 · see citing works
Suprasternal Doppler ultrasound for assessment of stroke distance
Childs, C., Goldring, S., Tann, W., & Hillier, V. F. (1998) · Archives of Disease in Childhood · doi:10.1136/adc.79.3.251 · Cited by 9 · see citing works
Chew, M. S., BS, M., Brandberg, J., Bjærum, S., Baek-Jensen, K., Sloth, E., Ask, P., Hasenkam, J. M., & Janerot-Sjöberg, B. (2000) · Critical Care Medicine · doi:10.1097/00003246-200011000-00022 · Cited by 8 · see citing works
Self-citationWippermann, C.-F., Schranz, D., Huth, R., Zepp, F., Oelert, H., & Jüngst, B.-K. (1992) · Heart · doi:10.1136/hrt.67.2.180 · Cited by 8 · see citing works
Hirsimäki, H., Kero, P., Seppälä, T. T., et al. (1993) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446894 · Cited by 4 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Noori, S., Drabu, B., Soleymani, S., & Seri, I. (2012) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/fetalneonatal-2011-301090 · Cited by 146 · see citing works
Echocardiographic Evaluation of Hemodynamics in Neonates and Children
Singh, Y. (2017) · Frontiers in Pediatrics · doi:10.3389/fped.2017.00201 · Cited by 116 · see citing works
New technologies in pediatric anesthesia
Holtby, H., Skowno, J. J., Kor, D. J., Flick, R. P., & Uezono, S. (2012) · Pediatric Anesthesia · doi:10.1111/pan.12007 · Cited by 35 · see citing works
Advanced Hemodynamic Monitoring in the Neonatal Intensive Care Unit
de Boode, W.-P. (2020) · Clinics in Perinatology · doi:10.1016/j.clp.2020.05.001 · Cited by 34 · see citing works
Nedelcu, C., Ionescu, N. S., Bizubac, A. M., Filip, C., & Cirstoveanu, C. (2025) · Life · doi:10.3390/life15101621 · Cited by 2 · see citing works
“Despite its known limitations, including relatively high intra-and interobserver variability (coefficient of variation ranging from 2.1% to 22% and 3.1% to 22%, respectively) [58], echocardiography remains the most widely adopted modality in neonatal and pediatric intensive care.”
Influential
- Authors
- Chew, M. S., & Poelaert, J.
- Affiliations
- Skåne University Hospital (SE)
- Ghent University Hospital (BE)
- Corresponding author
- Chew, Michelle S.
- Venue
- Intensive Care Medicine
- Published
- 2003-11-01
- Volume
- 29
- Issue
- 11
- Pages
- 1889-1894
- ISSN
- 0342-4642 (ISSN-L)
- 1432-1238
- OpenAlex
- W2074626476
- Cited by
- 151
- References
- 32
- Citations by year
- 2026: 3 · 2025: 5 · 2024: 7 · 2023: 2 · 2022: 6 · 2021: 9 · 2020: 7 · 2019: 11 · 2018: 6 · 2017: 8 · 2016: 4 · 2015: 7 · 2014: 7 · 2013: 10 · 2012: 12
- FWCI
- 3.83
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Intensive Care Medicine (publishedVersion)
- PubMed (publishedVersion)
- Ghent University Academic Bibliography (Ghent University) (submittedVersion)
- Lund University Publications (Lund University) (submittedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Epidemiology › Sepsis Diagnosis and Treatment
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Function and Risk Factors
- Keywords
- measurement accuracy, repeatability, thermodilution, Fick method, dye dilution, esophageal Doppler, pediatric intensive care, measurement bias
- MeSH terms
- Cardiac Output, Child, Critical Care, Data Interpretation, Statistical, Dye Dilution Technique, Humans, Monitoring, Physiologic, Research Design, Sensitivity and Specificity, Thermodilution, Echocardiography, Doppler, Reproducibility of Results, Bias, Echocardiography, Transesophageal
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review