Raman spectroscopy as a process analytical technology for pharmaceutical manufacturing and bioprocessing

Open access
Esmonde-White, Karen A., Cuellar, Maryann, Uerpmann, Carsten, Lenain, Bruno, Lewis, Ian R. · 2017 · Analytical and Bioanalytical Chemistry

The DOI matches its Crossref record; 102 of the 103 references it cites were found in a registry or an index.

Abstract

Adoption of Quality by Design (QbD) principles, regulatory support of QbD, process analytical technology (PAT), and continuous manufacturing are major factors effecting new approaches to pharmaceutical manufacturing and bioprocessing. In this review, we highlight new technology developments, data analysis models, and applications of Raman spectroscopy, which have expanded the scope of Raman spectroscopy as a process analytical technology. Emerging technologies such as transmission and enhanced reflection Raman, and new approaches to using available technologies, expand the scope of Raman spectroscopy in pharmaceutical manufacturing, and now Raman spectroscopy is successfully integrated into real-time release testing, continuous manufacturing, and statistical process control. Since the last major review of Raman as a pharmaceutical PAT in 2010, many new Raman applications in bioprocessing have emerged. Exciting reports of in situ Raman spectroscopy in bioprocesses complement a growing scientific field of biological and biomedical Raman spectroscopy. Raman spectroscopy has made a positive impact as a process analytical and control tool for pharmaceutical manufacturing and bioprocessing, with demonstrated scientific and financial benefits throughout a product's lifecycle.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1007/s00216-016-9824-1

Cites

102 of 102 references verified0 citation errors · 0 not checkable

Statistics

Figures

Fig. 1 Schematic comparing variants of sample excitation (solid line) and signal collection [dashed line(s)] used in Raman spectroscopy in measuring turbid media. (a) Backscattered Raman is a commonly used geometry that uses a single site of excitation with collection of signal close (<1 mm) to the excitation. As applied to Raman microscopy, this approach is called epi-illumination as a single microscope objective is used to excite the sample and collect Raman signal. The sampling volume in backscattered Raman is generally small, both in the lateral and axial dimensions. Thus, backscattered Raman is a good approach for measuring a surface. (b) Overlapping a defocused or wide laser beam with multiple collection fibers in a backscattering-like geometry is called wide area Raman (also known as large volumetric or global illumination Raman). (c) Separating collection fibers from the sample excitation by a small distance (Δd = 1–6 mm) enables collection of diffusely scattered Raman photons, known as a spatially-offset Raman spectroscopy (SORS). (d) Transmission Raman collects Raman photons diffusely scattered through a sample. Wide area, spatially-offset Raman spectroscopy (SORS) and transmission Raman provides representative sampling in turbid media and enables collection of Raman signal from buried layers in a layered sample
Fig. 1 Schematic comparing variants of sample excitation (solid line) and signal collection [dashed line(s)] used in Raman spectroscopy in measuring turbid media. (a) Backscattered Raman is a commonly used geometry that uses a single site of excitation with collection of signal close (<1 mm) to the excitation. As applied to Raman microscopy, this approach is called epi-illumination as a single microscope objective is used to excite the sample and collect Raman signal. The sampling volume in backscattered Raman is generally small, both in the lateral and axial dimensions. Thus, backscattered Raman is a good approach for measuring a surface. (b) Overlapping a defocused or wide laser beam with multiple collection fibers in a backscattering-like geometry is called wide area Raman (also known as large volumetric or global illumination Raman). (c) Separating collection fibers from the sample excitation by a small distance (Δd = 1–6 mm) enables collection of diffusely scattered Raman photons, known as a spatially-offset Raman spectroscopy (SORS). (d) Transmission Raman collects Raman photons diffusely scattered through a sample. Wide area, spatially-offset Raman spectroscopy (SORS) and transmission Raman provides representative sampling in turbid media and enables collection of Raman signal from buried layers in a layered sampleCC BY · PubMed Central
Fig. 2 A time course of predicted Raman concentrations of phenol and ether for a second pilot plant batch, with off-line HPLC and theoretical limit of ether. In situ Raman spectroscopy was used to control a heterogeneous etherification reaction, with Raman measurements of ether %w/w used to predict end of reaction. Reagent was added from 0 to ~180 min followed by a line wash, which accounts for the profile disturbance at ~ 240 min. In the process description and batch sheet, the reaction would have been completed at ~ 1250 min and at that time a sample would be collected for offline HPLC analysis (square in figure). In situ Raman data showed reaction completion nearly 600 min before stipulated time. The data suggest that batch cycle time could be reduced by several hours when moving up to the commercial manufacturing scale, improving process efficiency. Reprinted with permission from Hart, Richard J., Nicholas I. Pedge, Alan R. Steven, and Kevin Sutcliffe. “In Situ Monitoring of a Heterogeneous Etherification Reaction Using Quantitative Raman Spectroscopy.” Organic Process Research & Development 19, no. 1 (January 16, 2015): 196–202. doi:10.1021/op500027w. Copyright 2016 American Chemical Society
Fig. 2 A time course of predicted Raman concentrations of phenol and ether for a second pilot plant batch, with off-line HPLC and theoretical limit of ether. In situ Raman spectroscopy was used to control a heterogeneous etherification reaction, with Raman measurements of ether %w/w used to predict end of reaction. Reagent was added from 0 to ~180 min followed by a line wash, which accounts for the profile disturbance at ~ 240 min. In the process description and batch sheet, the reaction would have been completed at ~ 1250 min and at that time a sample would be collected for offline HPLC analysis (square in figure). In situ Raman data showed reaction completion nearly 600 min before stipulated time. The data suggest that batch cycle time could be reduced by several hours when moving up to the commercial manufacturing scale, improving process efficiency. Reprinted with permission from Hart, Richard J., Nicholas I. Pedge, Alan R. Steven, and Kevin Sutcliffe. “In Situ Monitoring of a Heterogeneous Etherification Reaction Using Quantitative Raman Spectroscopy.” Organic Process Research & Development 19, no. 1 (January 16, 2015): 196–202. doi:10.1021/op500027w. Copyright 2016 American Chemical SocietyCC BY · PubMed Central
Fig. 3 (a) Raman spectra after fluorescence corrections of reference components: ethanol, glucose, water, glycerol, and lactic acid, and (b) standard plot of baseline-corrected spectra for a simulated fermentation of 100 g/L glucose up to 50 g/L. Ethanol measured at 877 cm−1 (diamond), y = 21.406x + 12.553, R2 = 0.9959; 1046 cm−1 (multiplication sign), y = 3.0716x + 138.07, R2 = 0.9569 at 1455 cm − 1 (triangle), y = 6.4871x + 138.52, R2 = 0.984. Glucose measured at 514 cm−1 (circle), y = 2.8793x + 11.362, R2 = 0.9973, and 1,123 cm−1 (square), y = 4.2116x + 24.177, R2 = 0.994. Reprinted with permission from Iversen, Jens A., Rolf W. Berg, and Birgitte K. Ahring. “Quantitative Monitoring of Yeast Fermentation Using Raman Spectroscopy.” Analytical and Bioanalytical Chemistry 406, no. 20 (2014): 4911–4019. doi:10.1007/s00216-014-7897-2. Copyright 2014 Springer.
Fig. 3 (a) Raman spectra after fluorescence corrections of reference components: ethanol, glucose, water, glycerol, and lactic acid, and (b) standard plot of baseline-corrected spectra for a simulated fermentation of 100 g/L glucose up to 50 g/L. Ethanol measured at 877 cm−1 (diamond), y = 21.406x + 12.553, R2 = 0.9959; 1046 cm−1 (multiplication sign), y = 3.0716x + 138.07, R2 = 0.9569 at 1455 cm − 1 (triangle), y = 6.4871x + 138.52, R2 = 0.984. Glucose measured at 514 cm−1 (circle), y = 2.8793x + 11.362, R2 = 0.9973, and 1,123 cm−1 (square), y = 4.2116x + 24.177, R2 = 0.994. Reprinted with permission from Iversen, Jens A., Rolf W. Berg, and Birgitte K. Ahring. “Quantitative Monitoring of Yeast Fermentation Using Raman Spectroscopy.” Analytical and Bioanalytical Chemistry 406, no. 20 (2014): 4911–4019. doi:10.1007/s00216-014-7897-2. Copyright 2014 Springer.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 102 references

Research works printed but not linked: 4

  1. Slater JB, Tedesco TM, Fairchild RC, Lewis IR. Raman spectrometry and its adaptation to the industrial environment. In: Lewis IR, Edwards HGM, editors. Handbook of Raman spectroscopy: from the research laboratory to the process line. New York: Marcel-Dekker; 2001.
  2. Kourti T, Davis B. The business benefits of quality by design (QbD). Pharm Eng. 2012;32(4):1–10.
  3. Moore CMV. Regulatory perspective on real time release testing (RTRT). Washington DC: Proceedings of the AAPS Annual Meeting; 2011.
  4. Gervasio GJ, Pelletier MJ. On-line Raman analysis of PCl3 reactor material. J Process Anal Chem. 1997;3(1/2):7–11.

Cited by

See all 351

Metadata

Authors
Esmonde-White, K. A., Cuellar, M., Uerpmann, C., Lenain, B., & Lewis, I. R.
Venue
Analytical and Bioanalytical Chemistry
Published
2016-08-04
Volume
409
Issue
3
Pages
637-649
ISSN
  • 1618-2642 (ISSN-L)
  • 1618-2650
OpenAlex
W2497039129
Cited by
332
References
103
Citations by year
2026: 43 · 2025: 36 · 2024: 42 · 2023: 31 · 2022: 35 · 2021: 36 · 2020: 50 · 2019: 26 · 2018: 23 · 2017: 8 · 2016: 1
FWCI
2.34
Type
journal article
Language
en
Open access
Locations
  • Analytical and Bioanalytical Chemistry (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
Funders
National Institute of Arthritis and Musculoskeletal and Skin Diseases (T32AR007080)
Topics
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
  • Physical Sciences › Chemistry › Analytical Chemistry › Spectroscopy and Chemometric Analyses
  • Physical Sciences › Chemistry › Analytical Chemistry › Analytical Methods in Pharmaceuticals
Keywords
Raman spectroscopy, process analytical technology, pharmaceutical manufacturing, bioprocessing, quality by design, continuous production, real-time release testing, statistical process control, in situ monitoring
MeSH terms
Biotechnology, Spectrum Analysis, Raman, Technology, Pharmaceutical

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Raman spectroscopy as a process analytical technology for pharmaceutical manufacturing and bioprocessing. https://app.cochat.ai/papers/10.1007/s00216-016-9824-1
The paper
Esmonde-White, K. A., Cuellar, M., Uerpmann, C., Lenain, B., & Lewis, I. R. (2017). Raman spectroscopy as a process analytical technology for pharmaceutical manufacturing and bioprocessing. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-016-9824-1

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.