Effect of combined movement and heart rate monitor placement on physical activity estimates during treadmill locomotion and free-living
The DOI matches its Crossref record; 37 of the 37 references it cites were found in a registry or an index.
Abstract
A placement effect on activity measures from movement sensors has been reported during treadmill and free-living activity. Positioning of electrodes may impact on movement artifact susceptibility as well as surface ECG waveform amplitudes and thus potentially on the precision by which heart rate (HR) is ascertained from such ECG traces. The purpose of this study was to examine the extent to which placement of the combined HR and movement sensor, Actiheart, influences measurement of HR and movement, and estimates of energy expenditure. A total of 24 participants (20-39 years, 45-109 kg, 1.54-2.05 m, 19-29 kg m(-2)) were recruited. Whilst wearing two monitors, one placed at the level of the third intercostal space (upper position) and one just below the apex of the sternum (lower position), study participants performed level walking, incline walking, and level running on treadmill, and completed at least one day of free-living monitoring. Placement differences in HR data quality, movement counts, and energy expenditure (estimated from combined HR and movement) were analyzed with regression techniques. Quality of HR data was generally higher when monitors were placed in the lower position. This effect was more pronounced in men during both treadmill activity (relative risk, RR [95% CI] of noisy HR data in upper vs. lower position, RR=1.3[0.3; 5.6] in women, RR=174[14; 2,156] in men) and during free-living (RR=1.2[0.4; 3.3] in women, RR=25[9.6; 67] in men). There were minor place…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · cited 2×
“…along with higher ECG amplitudes, both of which may contribute to more reliable HR readings since commonly used mathematical algorithms may easily detect QRS complexes due to higher signal-to-noise ratio (Horan et al. 1980; Pan and Tompkins 1985; Schijvenaars et al. 1997; Searle and Kirkup 2000).”
BackgroundAge-predicted maximal heart rate revisited
Tanaka, H., Monahan, K. D., & Seals, D. R. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01054-8 · Cited by 3,775 · cited 1×
“When participants were unable to complete the protocol, HR had reached at least 90% of age-predicted maximal HR (Tanaka et al. 2001).”
BackgroundThe sources of external work in level walking and running.
Cavagna, G. A., Thys, H., & Zamboni, A. (1976) · The Journal of Physiology · doi:10.1113/jphysiol.1976.sp011613 · Cited by 804 · cited 1×
“This component increases with increasing running intensity, whilst the vertical component remains relatively constant (Cavagna et al. 1976; Chang and Kram 1999).”
BackgroundA direct comparison of wet, dry and insulating bioelectric recording electrodes
Searle, A., & Kirkup, L. (2000) · Physiological Measurement · doi:10.1088/0967-3334/21/2/307 · Cited by 725 · cited 4×
“…which type of electrode to use, since conductive, adhesive, and allergenic properties are different between electrode types (Ask et al. 1979; Burke and Gleeson 2000; Fernandez and Pallas-Areny 2000; Geddes et al. 1962; Hagemann et al. 1985; Patterson 1978; Searle and Kirkup 2000; Xu et al. 1999).”
BackgroundInfluentialReliability and validity of the combined heart rate and movement sensor Actiheart
Brage, S., Brage, N., Franks, P. W., Ekelund, U., & Wareham, N. J. (2005) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1602118 · Cited by 667 · cited 9×
“One limitation of the present study was that golden standard ECG was not available to perform agreement analysis with Actiheart measured HR in the two positions, although the HR algorithm has been successfully validated against ECG in a previous study (Brage et al. 2005).”
BackgroundMethodInfluentialSelf-citationBrage, S., Brage, N., Franks, P. W., Ekelund, U., Wong, M.-Y., Andersen, L. B., Froberg, K., & Wareham, N. J. (2004) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00703.2003 · Cited by 473 · cited 8×
“Data are mean PAI (robust SE) in J min 1 kg 1 (treadmill) or PAEE in kJ day 1 kg 1 (free-living) and corresponding epoch utilization (percentage of total, branch 1/2/3/4) in the branched model (Brage et al. 2004) *P<0.”
BackgroundMethodResultInfluentialSelf-citationA comparative evaluation of three accelerometry-based physical activity monitors
WELK, G. J., BLAIR, S. N., WOOD, K., JONES, S., & THOMPSON, R. W. (2000) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200009001-00008 · Cited by 343 · cited 4×
“To our knowledge, there are no other studies which have assessed the effect of placement on accelerometers attached to the two positions used in this study but Welk and colleagues reported a placement effect for the uni-axial CSA accelerometer (also known as the MTI Actigraph) but not for the tri-axial Tritrac and the…”
BackgroundInfluentialEndurance training does not enhance total energy expenditure in healthy elderly persons
Goran, M. I., & Poehlman, E. T. (1992) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1992.263.5.e950 · Cited by 275 · cited 2×
“…comparisons between different cultural settings and time eras, and finally in lifestyle intervention studies to ascertain the extent to which the intervention was delivered or indeed accompanied by compensatory behavior in other activity domains (Goran and Poehlman 1992; Wareham and Rennie 1998).”
BackgroundWareham, N. J., & Rennie, K. L. (1998) · PubMed · Cited by 256
Self-citationRehrig, P. W., Park, S.-E., Trolier-McKinstry, S., Messing, G. L., Jones, B., & Shrout, T. R. (1999) · Journal of Applied Physics · doi:10.1063/1.370943 · Cited by 231
Effect of Monitor Placement and of Activity Setting on the MTI Accelerometer Output
YNGVE, A., NILSSON, A., SJ??STR??M, M., & EKELUND, U. (2003) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000048829.75758.a0 · Cited by 192 · cited 3×
“2000) and lower back during treadmill activity but not during free-living (Yngve et al. 2003).”
BackgroundSelf-citationA micropower dry-electrode ECG preamplifier
Burke, M. J., & Gleeson, D. T. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.821734 · Cited by 164 · cited 2×
“…is which type of electrode to use, since conductive, adhesive, and allergenic properties are different between electrode types (Ask et al. 1979; Burke and Gleeson 2000; Fernandez and Pallas-Areny 2000; Geddes et al. 1962; Hagemann et al. 1985; Patterson 1978; Searle and Kirkup 2000; Xu et al.…”
BackgroundMetabolic cost of generating horizontal forces during human running
Chang, Y.-H., & Kram, R. (1999) · Journal of Applied Physiology · doi:10.1152/jappl.1999.86.5.1657 · Cited by 164 · cited 1×
“This component increases with increasing running intensity, whilst the vertical component remains relatively constant (Cavagna et al. 1976; Chang and Kram 1999).”
BackgroundSimultaneous measurement of heart rate and body motion to quantitate physical activity
HASKELL, W. L., YEE, M. C., EVANS, A., & IRBY, P. J. (1993) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199301000-00015 · Cited by 155 · cited 2×
“…Fax: +44-1223-330316 N. Brage Æ K. Froberg Institute of Sports Science and Clinical Biomechanics, University of Southern Denmark, Odense, Denmark Haskell et al. 1993; Moon and Butte 1996; Rennie et al. 2000; Strath et al. 2001 but the advantage of adding a measure of HR to movement…”
BackgroundSimultaneous heart rate-motion sensor technique to estimate energy expenditure
STRATH, S. J., BASSETT, D. R., SWARTZ, A. M., & THOMPSON, D. L. (2001) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200112000-00022 · Cited by 134 · cited 4×
“N. Brage AE K. Froberg Institute of Sports Science and Clinical Biomechanics, University of Southern Denmark, Odense, Denmark Eur J Appl Physiol (2006) 96: 517–524 DOI 10.1007/s00421-005-0112-6 Haskell et al. 1993; Moon and Butte 1996; Rennie et al. 2000; Strath et al. 2001 but the advantage of adding a measure of HR…”
BackgroundInfluentialValidity of the simultaneous heart rate-motion sensor technique for measuring energy expenditure
STRATH, S. J., BASSETT, D. R., THOMPSON, D. L., & SWARTZ, A. M. (2002) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200205000-00025 · Cited by 112 · cited 2×
“This is somewhat concerning, since most analytical models of combined HR and movement depend more on using the HR data during higher intensities of physical activity, that is when there is most likely also a lot of movement (Brage et al. 2004; Rennie et al. 2000; Strath et al. 2001, 2002 ).”
BackgroundRautaharju, P. M., Park, L., Rautaharju, F., & Crow, R. S. (1998) · Journal of Electrocardiology · doi:10.1016/s0022-0736(98)90003-6 · Cited by 104 · cited 1×
“Moreover, breast protuberance has been shown to have only small effects on ECG amplitudes (Rautaharju et al. 1998).”
BackgroundEthnic differences in ECG amplitudes in North American white, black, and Hispanic men and women
Rautaharju, P. M., Zhou, S., & Calhoun, H. P. (1994) · Journal of Electrocardiology · doi:10.1016/s0022-0736(94)80040-5 · Cited by 102 · cited 3×
“For example, R wave amplitude from a 12-lead ECG is normally much higher in lead V5 than for example in lead aVL (Rautaharju et al. 1994), and in a study comparing nine electrode positions, covering the left chest area from the second rib to 20 cm below”
BackgroundInfluentialA combined heart rate and movement sensor: proof of concept and preliminary testing study
Rennie, K., Rowsell, T., Jebb, S., Holburn, D., & Wareham, N. (2000) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1600973 · Cited by 83 · cited 3×
“…Institute of Sports Science and Clinical Biomechanics, University of Southern Denmark, Odense, Denmark Haskell et al. 1993; Moon and Butte 1996; Rennie et al. 2000; Strath et al. 2001 but the advantage of adding a measure of HR to movement registration would logically also depend on the…”
BackgroundSelf-citationMoon, J. K., & Butte, N. F. (1996) · Journal of Applied Physiology · doi:10.1152/jappl.1996.81.4.1754 · Cited by 68 · cited 1×
“…Brage Æ K. Froberg Institute of Sports Science and Clinical Biomechanics, University of Southern Denmark, Odense, Denmark Haskell et al. 1993; Moon and Butte 1996; Rennie et al. 2000; Strath et al. 2001 but the advantage of adding a measure of HR to movement registration would logically…”
BackgroundAg-AgCl electrode noise in high-resolution ECG measurements.
Fernández, M., & Pallas-Areny, R. (2000) · PubMed · Cited by 64
Avons, P., Garthwaite, P. H., Davies, H. L., Murgatroyd, P. R., & James, W. P. T. (1988) · PubMed · Cited by 42
Effect of electrode positioning on ECG interpretation by computer
Schijvenaars, B. J. A., Kors, J. A., van Herpen, G., Kornreich, F., & van Bemmel, J. H. (1997) · Journal of Electrocardiology · doi:10.1016/s0022-0736(97)80010-6 · Cited by 42 · cited 2×
“…along with higher ECG amplitudes, both of which may contribute to more reliable HR readings since commonly used mathematical algorithms may easily detect QRS complexes due to higher signal-to-noise ratio (Horan et al. 1980; Pan and Tompkins 1985; Schijvenaars et al. 1997; Searle and Kirkup 2000).”
BackgroundChallenges Facing Validation of Noninvasive Electrical Imaging of the Heart
Nash, M. P., & Pullan, A. J. (2005) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2005.00608.x · Cited by 41 · cited 1×
“These matrix factors influence the ECG amplitude, which combined with noise elements affects the nature of the final observable signal between any two points on the surface (Horan et al. 1980; Li et al. 2003; Nash and Pullan 2005; Schijvenaars et al. 1995).”
BackgroundTime and frequency dependence of disposable ECG electrode-skin impedance.
Olson, W. H., Schmincke, D. R., & Henley, B. L. (1979) · PubMed · Cited by 30
Interpolation of body surface potential maps
Schijvenaars, B. J. A., Kors, J. A., van Herpen, G., Kornreich, F., & van Bemmel, J. H. (1995) · Journal of Electrocardiology · doi:10.1016/s0022-0736(95)80034-4 · Cited by 25 · cited 1×
“These matrix factors influence the ECG amplitude, which combined with noise elements affects the nature of the final observable signal between any two points on the surface (Horan et al. 1980; Li et al. 2003; Nash and Pullan 2005; Schijvenaars et al. 1995).”
BackgroundImproved ?active? electrodes for recording bioelectric signals in work physiology
Hagemann, B., Luhede, G., & Luczak, H. (1985) · European Journal of Applied Physiology · doi:10.1007/bf00426306 · Cited by 25 · cited 2×
“…which type of electrode to use, since conductive, adhesive, and allergenic properties are different between electrode types (Ask et al. 1979; Burke and Gleeson 2000; Fernandez and Pallas-Areny 2000; Geddes et al. 1962; Hagemann et al. 1985; Patterson 1978; Searle and Kirkup 2000; Xu et al. 1999).”
BackgroundECG Electrodes: A Study of Electrical and Mechanical Long‐term Properties
Ask, P., ÖDerg, P. Å., Ödman, S., Tenland, T., & Skogh, M. (1979) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1979.tb01440.x · Cited by 23 · cited 2×
“Related to this issue is which type of electrode to use, since conductive, adhesive, and allergenic properties are different between electrode types (Ask et al. 1979; Burke and Gleeson 2000; Fernandez and Pallas-Areny 2000; Geddes et al. 1962; Hagemann et al. 1985; Patterson 1978; Searle and Kirkup…”
BackgroundHoran, L. G., Hand, R. C., Flowers, N. C., Johnson, J., & Sridharan, M. R. (1980) · Journal of Electrocardiology · doi:10.1016/s0022-0736(80)80081-1 · Cited by 23 · cited 3×
“These matrix factors influence the ECG amplitude, which combined with noise elements affects the nature of the final observable signal between any two points on the surface (Horan et al. 1980; Li et al. 2003; Nash and Pullan 2005; Schijvenaars et al. 1995).”
BackgroundRecording respiration and the electrocardiogram with common electrodes.
Geddes, L. A., HOFF, H. E., Dm, H., Hinds, M., & Baker, L. A. (1962) · PubMed · Cited by 21
SIMULTANEOUS HEART RATE-MOTION SENSOR TECHNIQUE
Strath, S. J., Bassett, D. R., Swartz, A. M., & Thompson, D. L. (2001) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200105001-01423 · Cited by 16 · cited 2×
“…Science and Clinical Biomechanics, University of Southern Denmark, Odense, Denmark Haskell et al. 1993; Moon and Butte 1996; Rennie et al. 2000; Strath et al. 2001 but the advantage of adding a measure of HR to movement registration would logically also depend on the precision by which HR…”
BackgroundThe electrical characteristics of some commercial ECG electrodes
PATTERSON, R. P. (1978) · Journal of Electrocardiology · doi:10.1016/s0022-0736(78)80025-9 · Cited by 16 · cited 4×
“It is very likely that heavier skin preparation would result in less noise on the HR recordings (Fernandez and Pallas-Areny 2000; Klingler et al. 1979; Olson et al. 1979; Patterson 1978), but researchers would need to balance this against the acceptability of such procedures in the population, for which the method is…”
BackgroundInfluentialA two stage mass spectrometer for nuclear physics applications
Wilson, H., Munro, R., Hardy, R. W. D., & Daly, N. R. (1961) · Nuclear Instruments and Methods · doi:10.1016/0029-554x(61)90215-4 · Cited by 14
Body Surface Laplacian Electrocardiogram of Ventricular Depolarization in Normal Human Subjects
Li, G., Lian, J., SALLA, P., Cheng, J. Z., RAMACHANDRA, I., SHAH, P., Avitall, B., & He, B. (2003) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2003.02199.x · Cited by 12 · cited 1×
“These matrix factors influence the ECG amplitude, which combined with noise elements affects the nature of the final observable signal between any two points on the surface (Horan et al. 1980; Li et al. 2003; Nash and Pullan 2005; Schijvenaars et al. 1995).”
BackgroundAnalysis of electrocardiograms for subcutaneous monitors
Burke, M. C., Song, Z., Jenkins, J. M., Alberts, M., Priore, J. D., & ARZBAECHER, R. C. (2003) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2003.09.064 · Cited by 8 · cited 2×
“…for example in lead aVL (Rautaharju et al. 1994), and in a study comparing nine electrode positions, covering the left chest area from the second rib to 20 cm below (sternum to the mid-clavicular line), highest amplitudes were most frequently found in the two lower quadrants (Burke et al. 2003).”
BackgroundA comparison of gel-to-gel and skin measurements of electrode impedance.
Klingler, D. R., Schoenberg, A. A., Worth, N. P., Egleston, C. F., & Burkart, J. A. (1979) · PubMed · Cited by 6
Preparation and characterization of new electrocardiogram electrodes
Xu, P., Carotenuto, G., Nicolais, L., ZHENG, Z., & Kuang, X. (1999) · Journal of Materials Science Materials in Medicine · doi:10.1023/a:1008965718192 · Cited by 3 · cited 2×
“…which type of electrode to use, since conductive, adhesive, and allergenic properties are different between electrode types (Ask et al. 1979; Burke and Gleeson 2000; Fernandez and Pallas-Areny 2000; Geddes et al. 1962; Hagemann et al. 1985; Patterson 1978; Searle and Kirkup 2000; Xu et al. 1999).”
Background
Cited by
Strath, S. J., Kaminsky, L. A., Ainsworth, B. E., Ekelund, U., Freedson, P. S., Gary, R. A., Richardson, C. R., Smith, D. T., & Swartz, A. M. (2013) · Circulation · doi:10.1161/01.cir.0000435708.67487.da · Cited by 1,029
Self-citationWarren, J. M., Ekelund, U., Besson, H., Mezzani, A., Geladas, N., & Vanhees, L. (2010) · European Journal of Cardiovascular Prevention & Rehabilitation · doi:10.1097/hjr.0b013e32832ed875 · Cited by 587
Self-citationAssessment of physical activity: a critical appraisal
Westerterp, K. R. (2009) · European Journal of Applied Physiology · doi:10.1007/s00421-009-1000-2 · Cited by 459
“However, the general shortcoming of heart rate monitoring over longer time under free-living conditions still is the data quality (Brage et al. 2006).”
BackgroundDowd, K. P., Szeklicki, R., Minetto, M. A., Murphy, M. H., Polito, A., Ghigo, E., van der Ploeg, H., Ekelund, U., Maciaszek, J., Stemplewski, R., et al. (2018) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-017-0636-2 · Cited by 385
“Study methodologies included the wearing of devices over the left and right hip [337, 370, 385, 387, 406, 413, 415, 417], over the hip and lower back [409], the wearing of devices side by side at the same location on the hip [301, 344, 411, 414, 416–418], devices worn at 3 intercostal space and just below the apex of…”
MethodSelf-citationPhysical Activity and Pregnancy: Cardiovascular Adaptations, Recommendations and Pregnancy Outcomes
Melzer, K., Schutz, Y., Boulvain, M., & Kayser, B. (2010) · Sports Medicine · doi:10.2165/11532290-000000000-00000 · Cited by 314
Metadata
- Authors
- Brage, S., Brage, N., Ekelund, U., Luan, J., Franks, P. W., Froberg, K., & Wareham, N. J.
- Affiliations
- MRC Elsie Widdowson Laboratory (GB)
- University of Southern Denmark (DK)
- Corresponding author
- Brage, Søren
- Venue
- European Journal of Applied Physiology
- Published
- 2005-12-13
- Volume
- 96
- Issue
- 5
- Pages
- 517-524
- ISSN
- 0020-9376 (ISSN-L)
- 1439-6319
- 1439-6327
- OpenAlex
- W2062313928
- Semantic Scholar
- a6f62112feb4574123692230b80f1ac1b231293a
- Cited by
- 127
- References
- 37
- Citations by year
- 2025: 2 · 2024: 4 · 2023: 3 · 2022: 1 · 2021: 3 · 2020: 6 · 2019: 5 · 2018: 10 · 2017: 5 · 2016: 17 · 2015: 6 · 2014: 11 · 2013: 9 · 2012: 9
- FWCI
- 1.40
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- European Journal of Applied Physiology (publishedVersion)
- PubMed (publishedVersion)
- University of Southern Denmark Research Portal (University of Southern Denmark) (submittedVersion)
- Funders
- Wellcome Trust
- Topics
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Physical Activity and Health
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Keywords
- Actiheart, wearable activity trackers, energy expenditure estimation, treadmill locomotion, free-living physical activity, ECG signal quality, sex differences, incline walking, running
- MeSH terms
- Adult, Electrocardiography, Electrodes, Energy Metabolism, Exercise Test, Female, Heart Rate, Humans, Locomotion, Male, Movement, Ribs, Sex Factors, Sternum, Reproducibility of Results, Artifacts
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.