Pulse oximetry based on photoplethysmography imaging with red and green light: Calibratability and challenges

Moço, Andreia, Verkruysse, Wim·2021·Journal of Clinical Monitoring and Computing

Abstract

Remotely measuring the arterial blood oxygen saturation (SpO 2) in visible light (Vis) involves different probing depths, which may compromise calibratibility. This paper assesses the feasibility of calibrating camera-based SpO 2 (SpO 2,cam) using red and green light. Camera-based photoplethysmographic (PPG) signals were measured at 46 healthy adults at center wavelengths of 580 nm (green), 675 nm (red), and 840 nm (near-infrared; NIR). Subjects had their faces recorded during normoxia and hypoxia and under gradual cooling. SpO 2,cam estimates in Vis were based on the normalized ratio of camera-based PPG amplitudes in red over green light (RoG). SpO 2,cam in Vis was validated against contact SpO 2 (reference) and compared with SpO 2,cam estimated using red-NIR wavelengths. An RoG-based calibration curve for SpO 2 was determined based on data with a SpO 2 range of 85-100%. We found an [Formula: see text] error of 2.9% (higher than the [Formula: see text] for SpO 2,cam in red-NIR). Additional measurements on normoxic subjects under temperature cooling (from [Formula: see text] to [Formula: see text]) evidenced a significant bias of - 1.7, CI [- 2.7, - 0.7]%. It was also noted that SpO[Formula: see text] estimated at the cheeks was significantly biased (- 3.6, CI [- 5.7, - 1.5]%) with respect to forehead estimations. Under controlled conditions, SpO[Formula: see text] can be calibrated with red and green light but the accuracy is less than that of SpO[Formula: see text] estimated in the usual red-NIR window.

Integrity signals

Registry

Crossref · 10.1007/s10877-019-00449-y

Cites

23 of 23 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Other subjective scales were employed in 13 studies (35.14%).

    “and other subjective scales were employed in 13 studies (35.14%) [30,42,43,44,45,50,52,56,72,78,80,87,88].”
  • A recent study on 46 subjects at center wavelengths of 840 nm, 675 nm, and 580 nm found significant bias at lower temperatures when calibrating a camera-based SpO2 oximeter using red and green light under normoxic and hypoxic conditions.

    “A recent study was performed on 46 subjects at center wavelengths of 840 nm (near-infrared), 675 nm (red), and 580 nm (green) to evaluate the feasibility of calibrating a camera-based SpO2 oximeter using red and green light on data under normoxic and hypoxic conditions found significant bias at lower temperatures [72].”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Moço, A., & Verkruysse, W.
Affiliations
Eindhoven University of Technology (NL)
Corresponding author
Moço, Andreia
Venue
Journal of Clinical Monitoring and Computing
Published
2020-01-01
Volume
35
Issue
1
Pages
123-133
ISSN
  • 1387-1307 (ISSN-L)
  • 1573-2614
OpenAlex
W2998583602
Cited by
55
References
23
Citations by year
2026: 4 · 2025: 6 · 2024: 13 · 2023: 10 · 2022: 11 · 2021: 8 · 2020: 3
FWCI
2.51
Type
journal article
Language
en
Open access
closed
Locations
  • Journal of Clinical Monitoring and Computing (publishedVersion)
  • PubMed (publishedVersion)
  • TU/e Research Portal (submittedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
  • Health Sciences › Medicine › Complementary and alternative medicine › Cardiovascular and exercise physiology
Keywords
pulse oximetry, imaging photoplethysmography, arterial oxygen saturation, visible light, red and green light, calibration, hypoxia, facial measurements
MeSH terms
Adult, Hypoxia, Humans, Light, Oximetry, Oxygen, Photoplethysmography

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review