Innovative measurement of routine physiological variables (heart rate, respiratory rate and oxygen saturation) using a remote photoplethysmography imaging system: A prospective comparative trial protocol
Open accessAbstract
Introduction Physiological signals are essential for assessing human health. The absence of a medical device to carry out these measurements remotely is one of the main limitations of telemedicine. Remote photoplethysmography imaging (rPPG) makes it possible to use a camera video to measure some of the most valuable physiological variables: heart rate (HR), respiratory rate (RR) and oxygen saturation (SpO 2 ). Our objective was to evaluate the value of such remote measurements compared with existing contact point measurements techniques in real-life clinical settings. Methods and analysis Prospective hospital-based study that will recruit 1045 patients who require a pulmonary function test. For each patient, measurements of HR, RR and SpO 2 , using a standard acquisition system, will be carried out concomitantly with the measurements made by the rPPG system. 30, 60 and 120 s time frames will be used to take measurements. Age, gender and skin phototype will also be collected. The intraclass coefficient correlation will be performed to determine the accuracy and precision of the rPPG algorithm readings. Ethics and dissemination The study protocol has been approved by the French Agency for the Safety of Health Products (ANSM registration no. ID RCB 2020-A02428-31) and by a French ethics committee (CPP OUEST I-TOURS-2020T1-30 DM at 30 October 2020). Results will be published in peer-reviewed journals, at scientific conferences and through press releases. Trial registration number NCT04660318 .
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Among the 11/35 low risk studies, 3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation.
“3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation [46,48,70]”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [70]
Objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [70]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339 · see citing works
Virtually perfect? Telemedicine for COVID-19
Hollander, J. E., & Carr, B. G. (2020) · New England Journal of Medicine · doi:10.1056/nejmp2003539 · Cited by 2,970 · see citing works
Remote plethysmographic imaging using ambient light
Verkruysse, W., Svaasand, L. O., & Nelson, J. S. (2008) · Optics Express · doi:10.1364/oe.16.021434 · Cited by 1,728 · see citing works
Non-contact, automated cardiac pulse measurements using video imaging and blind source separation
Poh, M., McDuff, D., & Picard, R. W. (2010) · Optics Express · doi:10.1364/oe.18.010762 · Cited by 1,652 · see citing works
Covid-19 and Health Care’s Digital Revolution
Keesara, S., Jonas, A., & Schulman, K. (2020) · New England Journal of Medicine · doi:10.1056/nejmp2005835 · Cited by 1,126 · see citing works
Dry-Contact and Noncontact Biopotential Electrodes: Methodological Review
Chi, Y. M., Jung, T., & Cauwenberghs, G. (2010) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2010.2084078 · Cited by 1,069 · see citing works
Physiological parameter monitoring from optical recordings with a mobile phone
Scully, C. G., Lee, J., Meyer, J., Gorbach, A. M., Granquist-Fraser, D., Mendelson, Y., & Chon, K. H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2163157 · Cited by 452 · see citing works
Contact-Free Measurement of Cardiac Pulse Based on the Analysis of Thermal Imagery
Garbey, M., Sun, N., Merla, A., & Pavlidis, I. T. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.891930 · Cited by 402 · see citing works
Heart rate measurement based on a time-lapse image.
Takano, C., & Ohta, Y. (2007) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2006.09.006 · Cited by 345 · see citing works
Nakajima, K., TAMURA, T., & Miike, H. (1996) · Medical Engineering & Physics · doi:10.1016/1350-4533(95)00066-6 · Cited by 244 · see citing works
ECG measurement on a chair without conductive contact
Lim, Y. G., Kim, K. K., & Park, S. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.872823 · Cited by 233 · see citing works
Humphreys, K., Ward, T. E., & Markham, C. (2007) · Review of Scientific Instruments · doi:10.1063/1.2724789 · Cited by 214 · see citing works
Noncontact Respiration Rate Measurement System Using an Ultrasonic Proximity Sensor
Min, S. D., Kim, J. K., Shin, H., Yun, Y. H., Lee, C. K., & Lee, M. (2010) · IEEE Sensors Journal · doi:10.1109/jsen.2010.2044239 · Cited by 168 · see citing works
Monitoring of electrocardiograms in bed without utilizing body surface electrodes
Ishijima, M. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237680 · Cited by 161 · see citing works
Liu, H., Wang, Y., & Wang, L. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346371 · Cited by 60 · see citing works
Moço, A., & Verkruysse, W. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00449-y · Cited by 55 · see citing works
Photoplethysmography: Part 1 Comparison with laser Doppler flowmetry
Lindberg, L. G., Tamura, T., & Öberg, P. Å. (1991) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446294 · Cited by 46 · see citing works
Optical remote sensing of heartbeats
Parra, J. E., & Costa, G. D. (2001) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.438115 · Cited by 8 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
- Authors
- Allado, E., Poussel, M., Moussu, A., Saunier, V., Bernard, Y., Albuisson, E., & Chenuel, B.
- Affiliations
- Centre Hospitalier Régional et Universitaire de Nancy (FR)
- OMEOS, Nancy, Lorraine, France
- Corresponding author
- Allado, Edem
- Venue
- BMJ Open
- Publisher
- BMJ
- Published
- 2021-08
- Volume
- 11
- Issue
- 8
- Pages
- e047896
- ISSN
- 2044-6055 (ISSN-L)
- PMC
- PMC8365801
- OpenAlex
- W3192398007
- Cited by
- 22
- References
- 18
- Citations by year
- 2026: 3 · 2025: 5 · 2024: 3 · 2023: 3 · 2022: 7 · 2021: 1
- FWCI
- 1.06
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- BMJ Open (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Agence Nationale de Sécurité du Médicament et des Produits de Santé
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- remote photoplethysmography, heart rate measurement, respiratory rate measurement, oxygen saturation, non-contact physiological monitoring, telemedicine, pulmonary function tests, precision and accuracy
- MeSH terms
- Heart Rate, Humans, Oxygen, Prospective Studies, Photoplethysmography, Respiratory Rate
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review