‘Outwalk’: a protocol for clinical gait analysis based on inertial and magnetic sensors
Open accessThe DOI matches its Crossref record; 30 of the 30 references it cites were found in a registry or an index.
Abstract
A protocol named Outwalk was developed to easily measure the thorax–pelvis and lower-limb 3D kinematics on children with cerebral palsy (CP) and amputees during gait in free-living conditions, by means of an Inertial and Magnetic Measurement System (IMMS). Outwalk defines the anatomical/functional coordinate systems (CS) for each body segment through three steps: (1) positioning the sensing units (SUs) of the IMMS on the subjects’ thorax, pelvis, thighs, shanks and feet, following simple rules; (2) computing the orientation of the mean flexion–extension axis of the knees; (3) measuring the SUs’ orientation while the subject’s body is oriented in a predefined posture, either upright or supine. If the supine posture is chosen, e.g. when spasticity does not allow to maintain the upright posture, hips and knees static flexion angles must be measured through a standard goniometer and input into the equations that define Outwalk anatomical CSs. In order to test for the inter-rater measurement reliability of these angles, a study was carried out involving nine healthy children (7.9 ± 2 years old) and two physical therapists as raters. Results showed RMS error of 1.4° and 1.8° and a negligible worst-case standard error of measurement of 2.0° and 2.5° for hip and knee angles, respectively. Results were thus smaller than those reported for the same measures when performed through an optoelectronic system with the CAST protocol and support the beginning of clinical trials of Outwalk with children with CP.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
QUANTIFYING TEST-RETEST RELIABILITY USING THE INTRACLASS CORRELATION COEFFICIENT AND THE SEM
WEIR, J. P. (2005) · Journal of Strength and Conditioning Research · doi:10.1519/00124278-200502000-00038 · Cited by 4,842
Wu, G., van der Helm, F. C. T., (DirkJan) Veeger, H. E. J., Makhsous, M., Van Roy, P., Anglin, C., Nagels, J., Karduna, A. R., McQuade, K., Wang, X., et al. (2005) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2004.05.042 · Cited by 4,558
Wu, G., Siegler, S., Allard, P., Kirtley, C., Leardini, A., Rosenbaum, D., Whittle, M., D’Lima, D. D., Cristofolini, L., Witte, H., et al. (2002) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00222-6 · Cited by 3,242 · cited 6×
“In fact, the lack of information regarding the position of the sensors implies that the anatomical CSs cannot be defined through the calibration of single anatomical landmarks (as recommended by the ISB [32]).”
BackgroundMethodInfluentialA gait analysis data collection and reduction technique
Davis, R. B., Õunpuu, S., Tyburski, D., & Gage, J. R. (1991) · Human Movement Science · doi:10.1016/0167-9457(91)90046-z · Cited by 3,086
When to use agreement versus reliability measures
de Vet, H. C. W., Terwee, C. B., Knol, D. L., & Bouter, L. M. (2006) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2005.10.015 · Cited by 1,786 · cited 1×
“Following current recommendations about statistical parameters describing reliability [10], we also computed for h and k the standard error of measurement (or ‘typical error’), considering (SEMse) and not considering (SEMnse) the systematic error introduced by T1 and T2.”
MethodModelling and Control of Robot Manipulators
Sciavicco, L., & Siciliano, B. (2000) · Advanced Textbooks in Control and Signal Processing · doi:10.1007/978-1-4471-0449-0 · Cited by 1,701
Cappozzo, A., Catani, F., Della Croce, U., & Leardini, A. (1995) · Clinical Biomechanics · doi:10.1016/0268-0033(95)91394-t · Cited by 1,677 · cited 2×
“The definition of the anatomical CSs was partially based on the external anatomical landmarks described in [6].”
BackgroundMethodAccelerometry: A technique for quantifying movement patterns during walking
Kavanagh, J. J., & Menz, H. B. (2008) · Gait & Posture · doi:10.1016/j.gaitpost.2007.10.010 · Cited by 617 · cited 1×
“Owing to these features, the IMMSs might allow the user to execute and acquire the movement in laboratory-free settings, in a continuous modality, for long periods; therefore, they might allow the user to collect a great number of consecutive gait cycles during spontaneous walking in daily life environments [16].”
BackgroundGait patterns in spastic hemiplegia in children and young adults
(1987) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-198769030-00016 · Cited by 524 · cited 1×
“[30] and diplegic belonging to forms II–IV of”
BackgroundA new method for estimating joint parameters from motion data
Schwartz, M. H., & Rozumalski, A. (2005) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2004.03.009 · Cited by 512 · cited 3×
“It has been widely demonstrated that the knee rotations other than the FE are strongly affected by the soft-tissue artefact problem when measured through skin-mounted markers and are therefore unreliable [20, 25, 27].”
Background
Cited by
IMU-Based Joint Angle Measurement for Gait Analysis
Seel, T., Raisch, J., & Schauer, T. (2014) · Sensors · doi:10.3390/s140406891 · Cited by 867
Survey of Motion Tracking Methods Based on Inertial Sensors: A Focus on Upper Limb Human Motion
Filippeschi, A., Schmitz, N., Miezal, M., Bleser, G., Ruffaldi, E., & Stricker, D. (2017) · Sensors · doi:10.3390/s17061257 · Cited by 361
Robert-Lachaine, X., Mecheri, H., Larue, C., & Plamondon, A. (2017) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1537-2 · Cited by 329
“[12] may indicate that the Outwalk model [6] is more closely related to a model based on anatomical landmarks than the MVN model.”
MethodConcurrent validation of Xsens MVN measurement of lower limb joint angular kinematics
Zhang, J.-T., Novak, A. C., Brouwer, B., & Li, Q. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/8/n63 · Cited by 314
“Similarity of the flexion/extension waveforms is confirmed by the high CMC values of 0.96 or higher.”
BackgroundMethodResultInfluentialFong, D. T.-P., & Chan, Y.-Y. (2010) · Sensors · doi:10.3390/s101211556 · Cited by 286
Wearable inertial sensors for human movement analysis
Iosa, M., Picerno, P., Paolucci, S., & Morone, G. (2016) · Expert Review of Medical Devices · doi:10.1080/17434440.2016.1198694 · Cited by 281
Picerno, P. (2017) · Gait & Posture · doi:10.1016/j.gaitpost.2016.11.008 · Cited by 237
Ferrari, A., Cutti, A. G., Garofalo, P., Raggi, M., Heijboer, M., Cappello, A., & Davalli, A. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-009-0544-y · Cited by 181
“In [ 4 ] a protocol, named ‘‘Outwalk’’, was proposed to measure the kinematics of the pelvis relative to the thorax (TP joint), and of the lower limbs with an IMMS.”
MethodInfluentialSelf-citationTeufl, W., Miezal, M., Taetz, B., Fröhlich, M., & Bleser, G. (2019) · PLOS ONE · doi:10.1371/journal.pone.0213064 · Cited by 129
Cereatti, A., Gurchiek, R., Mündermann, A., Fantozzi, S., Horak, F., Delp, S., & Aminian, K. (2024) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2024.112225 · Cited by 79
Lebel, K., Boissy, P., Nguyen, H., & Duval, C. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-017-0347-6 · Cited by 65
“however, be aware that the accuracy of any technological equipment used for mobility assessment does not only depend upon the technological capability to measure accurately motion in the context of use, but also on soft tissue artifacts due to the positioning and fixation of the modules as well as errors due to anatom…”
Reliability and precision of 3D wireless measurement of scapular kinematics
van den Noort, J. C., Wiertsema, S. H., Hekman, K. M. C., Schönhuth, C. P., Dekker, J., & Harlaar, J. (2014) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1186-2 · Cited by 60
Robust Sensor Fusion and Biomimetic Control of a Lower-Limb Exoskeleton With Multimodal Sensors
Arceo, J. C., Yu, L., & Bai, S. (2025) · IEEE Transactions on Automation Science and Engineering · doi:10.1109/tase.2024.3421318 · Cited by 19
“For example, in [6] the Outwalk protocol for clinical gait analysis using magnetic and inertial sensors was introduced.”
Ding, Z., Yang, C., Ma, J., Wei, J., & Jiang, F. (2020) · Multimedia Tools and Applications · doi:10.1007/s11042-019-07911-8 · Cited by 4
Metadata
- Authors
- Cutti, A. G., Ferrari, A., Garofalo, P., Raggi, M., Cappello, A., & Ferrari, A.
- Venue
- Medical & Biological Engineering & Computing
- Published
- 2009-11-13
- Volume
- 48
- Issue
- 1
- Pages
- 17-25
- ISSN
- 0140-0118 (ISSN-L)
- 1741-0444
- OpenAlex
- W2037819518
- Cited by
- 218
- References
- 30
- FWCI
- 4.43
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.