Diagnostic accuracy of seismocardiography compared with electrocardiography for the anatomic and physiologic diagnosis of coronary artery disease during exercise testing
Abstract
A multicenter study was performed to compare the diagnostic accuracy of a new technique, seismocardiography, with that of electrocardiography for physiologically and anatomically significant ischemic coronary artery disease (CAD) during exercise stress testing. Five participating centers enrolled 129 patients who had simultaneous seismocardiograms and 12-lead electrocardiograms at the time of their exercise treadmill stress tests. Two different definitions of CAD were used: anatomic and physiologically significant disease. The presence of anatomically significant CAD (> or = 50% diameter stenosis) was documented by coronary angiography. Physiologically significant CAD was defined as present in the same 129 patients when coronary arteriography (> or = 50% diameter stenosis) and thallium-201 scintigraphy (defect on initial postexercise images) were both abnormal. Seismocardiography had a significantly better sensitivity for detecting anatomic CAD than did electrocardiography (73 vs 48%; p < 0.001), without loss of specificity (78 vs 80%; p = NS). Exercise seismocardiography added significant incremental diagnostic information beyond that provided by exercise electrocardiography. Seismocardiography was more sensitive (without less specificity) in women and in patients who did not achieve maximal predicted heart rate. In patients with physiologically significant CAD, the seismocardiogram was also significantly more sensitive (78%) than was the electrocardiogram (55%) (p < 0.02), without loss of specificity (84 vs 74%). Seismocardiography significantly improved sensitivity for the detection of anatomic and physiologic CAD. It is easy to perform and may be a clinically useful adjunct in exercise stress testing.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Visualization of heart vibrations (SCG) provides essential information on mechanical outcomes of myocardial activity, valve motions, and other features absent from ECG data.
“Visualization of heart vibrations, referred to as a seismocardiogram (SCG), is rarely performed in general clinical practice, especially in the NICU/PICU, yet it provides essential information on the mechanical outcomes of myocardial activity, valve motions and other features that are absent from ECG data11.”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [11]
Cites
The meaning and use of the area under a receiver operating characteristic (ROC) curve.
Hanley, J. A., & McNeil, B. J. (1982) · Radiology · doi:10.1148/radiology.143.1.7063747 · Cited by 21,856 · see citing works
Heart Disease: A Textbook of Cardiovascular Medicine.
(1984) · Annals of Internal Medicine · doi:10.7326/0003-4819-101-3-415_3 · Cited by 4,119 · see citing works
Analysis of probability as an aid in the clinical diagnosis of coronary artery disease.
Diamond, G. A., & Forrester, J. S. (1979) · New England Journal of Medicine · doi:10.1056/nejm197906143002402 · Cited by 2,664 · see citing works
Exercise-induced ST depression in the diagnosis of coronary artery disease. A meta-analysis.
Gianrossi, R., Detrano, R., Mulvihill, D., Lehmann, K. G., Dubach, P., Colombo, A., McArthur, D. L., & Froelicher, V. F. (1989) · Circulation · doi:10.1161/01.cir.80.1.87 · Cited by 840 · see citing works
Weiner, D. A., Ryan, T. J., McCabe, C. H., Kennedy, J. W., Schloss, M. G., Tristani, F. E., CHAITMAN, B. R., & Fisher, L. D. (1979) · New England Journal of Medicine · doi:10.1056/nejm197908023010502 · Cited by 604 · see citing works
Behavioral Health: A Handbook of Health Enhancement and Disease Prevention
Herd, A. J., & Jd, M. (1984) · Medical Entomology and Zoology · Cited by 511 · see citing works
Sketch, M. H., Mohiuddin, S. M., Lynch, J. D., Zencka, A. E., & Runco, V. (1975) · The American Journal of Cardiology · doi:10.1016/0002-9149(75)90521-4 · Cited by 264 · see citing works
Jones, R. H., McEwan, P., Newman, G. E., Port, S. C., Rerych, S. K., Scholz, P., Upton, M. T., Peter, C. A., Austin, E. H., Leong, K.-H., et al. (1981) · Circulation · doi:10.1161/01.cir.64.3.586 · Cited by 219 · see citing works
Ritchie, J. L., Zaret, B. L., Strausś, H. W., Pitt, B., Berman, D. S., Schelbert, H. R., Ashburn, W. L., Berger, H. J., & Hamilton, G. W. (1978) · The American Journal of Cardiology · doi:10.1016/0002-9149(78)90926-8 · Cited by 208 · see citing works
Fleming, R. M., Kirkeeide, R. L., Smalling, R. W., Gould, K. L., & Stuart, Y. (1991) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(91)90752-u · Cited by 177 · see citing works
Heart rate adjustment of ST segment depression for improved detection of coronary artery disease.
Kligfield, P., Ameisen, O., & Okin, P. M. (1989) · Circulation · doi:10.1161/01.cir.79.2.245 · Cited by 175 · see citing works
Fortuin, N. J., & Weiss, J. L. (1977) · Circulation · doi:10.1161/01.cir.56.5.699 · Cited by 167 · see citing works
Presti, C. F., Armstrong, W. F., & FEIGENBAUM, H. B. (1988) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(88)80093-2 · Cited by 125 · see citing works
Salerno, D. M., & Zanetti, J. M. (1990) · Cited by 123 · see citing works
Lachterman, B., Lehmann, K. G., Detrano, R., Neutel, J. M., & Froelicher, V. F. (1990) · Circulation · doi:10.1161/01.cir.82.1.44 · Cited by 100 · see citing works
SCARBOROUGH, W. R., Mason, R. E., Davis, F. W., SINGEWALD, M. L., Baker, B. M., & Lore, S. A. (1952) · American Heart Journal · doi:10.1016/0002-8703(52)90095-1 · Cited by 77 · see citing works
Salerno, D. M., Zanetti, J. M., Green, L. A., Mooney, M. R., Madison, J. D., & Tassel, R. A. V. (1991) · The American Journal of Cardiology · doi:10.1016/0002-9149(91)90744-6 · Cited by 68 · see citing works
Noninvasive diagnosis of coronary artery disease: the cardiokymographic stress test.
Silverberg, R. A., Diamond, G. A., Vas, R., Tzivoni, D., Swan, H. J. C., & Forrester, J. S. (1980) · Circulation · doi:10.1161/01.cir.61.3.579 · Cited by 57 · see citing works
Diamond, G. A. (1987) · Medical Decision Making · doi:10.1177/0272989x8700700407 · Cited by 52 · see citing works
Exercise Seismocardiography for Detection of Coronary Artery Disease
Salerno, D. M., Zanetti, J. M., Poliac, L. C., Crow, R. S., Hannan, P. J., Wang, K., Goldenberg, I. F., & Tassel, R. A. V. (2017) · American Journal of Noninvasive Cardiology · doi:10.1159/000470383 · Cited by 35 · see citing works
THE BALLISTOCARDIOGRAM IN CORONARY ARTERY DISEASE
Taymor, R. C. (1952) · JAMA · doi:10.1001/jama.1952.02930060001001 · Cited by 31 · see citing works
Accuracy of cardiokymography during exercise testing: Results of a multicenter study
Weiner, D. A. (1985) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(85)80105-4 · Cited by 30 · see citing works
Marcus, M. L., White, C. W., & Kirchner, P. T. (1986) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(86)80378-3 · Cited by 27 · see citing works
Goldschlager, N., & Sox, H. C. (1988) · American Heart Journal · doi:10.1016/0002-8703(88)90628-x · Cited by 27 · see citing works
A lesson from the controversy about heart rate adjustment of ST segment depression.
Bobbio, M., & Detrano, R. (1991) · Circulation · doi:10.1161/01.cir.84.3.1410 · Cited by 22 · see citing works
Richard, D., & Birrer, R. B. (1988) · PubMed · Cited by 21 · see citing works
Silvestre, A., Sandhu, G., Desser, K. B., & Benchimol, A. (1978) · The American Journal of Cardiology · doi:10.1016/0002-9149(78)90931-1 · Cited by 17 · see citing works
The academic life of a noninvasive test.
Kligfield, P., & Okin, P. M. (1990) · Circulation · doi:10.1161/01.cir.82.6.2284 · Cited by 11 · see citing works
Grandi, A., Barzizza, F., Bernardi, L., Venco, A., & Finardi, G. (1984) · PubMed · Cited by 7 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Ballistocardiography and Seismocardiography: A Review of Recent Advances
Inan, O. T., Migeotte, P.-F., Park, K.-S., Etemadi, M., Tavakolian, K., Casanella, R., Zanetti, J., Tank, J., Funtova, I., Prisk, G. K., et al. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2361732 · Cited by 775 · see citing works
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-0792-9 · Cited by 482 · see citing works
Contactless seismocardiography via deep learning radars
Ha, U., Assana, S., & Adib, F. (2020) · ACM Annual International Conference on Mobile Computing and Networking (MobiCom) · doi:10.1145/3372224.3419982 · Cited by 203 · see citing works
Di Rienzo, M., Vaini, E., Castiglioni, P., Merati, G., Meriggi, P., Parati, G., Faini, A., & Rizzo, F. (2013) · Autonomic Neuroscience · doi:10.1016/j.autneu.2013.04.005 · Cited by 163 · see citing works
Seismocardiography: Past, present and future
Zanetti, J. M., & Tavakolian, K. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6611170 · Cited by 152 · see citing works
Gupta, P., Moghimi, M. J., Jeong, Y., Gupta, D., Inan, O. T., & Ayazi, F. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0225-7 · Cited by 125 · see citing works
Definition of Fiducial Points in the Normal Seismocardiogram
Sørensen, K., Schmidt, S. E., Jensen, A. S., Søgaard, P., & Struijk, J. J. (2018) · Scientific Reports · doi:10.1038/s41598-018-33675-6 · Cited by 111 · see citing works
Heart monitoring systems—A review
Jain, P. K., & Tiwari, A. K. (2014) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2014.08.014 · Cited by 101 · see citing works
Etemadi, M., Inan, O. T., Heller, J. A., Hersek, S., Klein, L., & Roy, S. (2016) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2015.2405480 · Cited by 95 · see citing works
“Furthermore, the results of a large multicenter study demonstrated that combining SCG and ECG signals improved the predictive accuracy of detecting CAD compared to using ECG alone [18].”
Mant, J., McManus, R. J., Oakes, R., Delaney, B. C., Barton, P. M., Deeks, J. J., Hammersley, L., Davies, R., Davies, M., & Hobbs, R. F. (2004) · Health Technology Assessment · doi:10.3310/hta8020 · Cited by 85 · see citing works
Santucci, F., Presti, D. L., Massaroni, C., Schena, E., & Setola, R. (2022) · Sensors · doi:10.3390/s22155805 · Cited by 47 · see citing works
Motion Artifact Cancellation of Seismocardiographic Recording From Moving Subjects
Yang, C., & Tavassolian, N. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2573269 · Cited by 42 · see citing works
Korzeniowska‐Kubacka, I., Bilińska, M., & Piotrowicz, R. (2005) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2005.00547.x · Cited by 42 · see citing works
Marcus, G. M., Sorrell, V. L., Zanetti, J. M., Bosnos, M., Baweja, G., PERLICK, D., Ott, P., Indik, J. H., He, D. S., & Gear, K. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00894.x · Cited by 41 · see citing works
Jia, L., Shi, L., Liu, C., Xu, J., Gao, Y., Sun, C., Liu, S., & Wu, G. (2022) · IEEE Internet of Things Journal · doi:10.1109/jiot.2022.3145863 · Cited by 38 · see citing works
Simplified detection of myocardial ischemia by seismocardiography
Becker, M., Roehl, A. B., Siekmann, U., Koch, A., de la Fuente, M., Roissant, R., Radermacher, K., Marx, N., & Hein, M. (2013) · Herz · doi:10.1007/s00059-013-3851-x · Cited by 35 · see citing works
Human Life Signs Detection Using High-Sensitivity Pulsed Laser Vibrometer
Wang, C.-C., TRIVEDI, S. B., Jin, F., Stepanov, S. I., Chen, Z., Khurgin, J. B., Rodriguez, P., & Prasad, N. S. (2007) · IEEE Sensors Journal · doi:10.1109/jsen.2007.905041 · Cited by 34 · see citing works
Jerosch‐Herold, M., Zanetti, J. M., Merkle, H., Poliac, L. C., Huang, H., Mansoor, A., Zhao, F., & Wilke, N. M. (1999) · International journal of cardiac imaging · doi:10.1023/a:1006364518204 · Cited by 33 · see citing works
A Clinical Method for Estimation of VO2max Using Seismocardiography
Sørensen, K., Poulsen, M. K., Karbing, D. S., Søgaard, P., Struijk, J. J., & Schmidt, S. E. (2020) · International Journal of Sports Medicine · doi:10.1055/a-1144-3369 · Cited by 31 · see citing works
Multichannel seismocardiography: an imaging modality for investigating heart vibrations
Munck, K., Sørensen, K., Struijk, J. J., & Schmidt, S. E. (2020) · Physiological Measurement · doi:10.1088/1361-6579/abc0b7 · Cited by 27 · see citing works
Estimation of heart rate and respiratory rate from the seismocardiogram under resting state
Lin, Y., & Jhou, Y.-F. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101779 · Cited by 26 · see citing works
Performance analysis of seismocardiography for heart sound signal recording in noisy scenarios
Jain, P. K., Tiwari, A. K., & Chourasia, V. S. (2016) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2016.1139203 · Cited by 25 · see citing works
Yao, J., Tridandapani, S., Auffermann, W. F., Wick, C. A., & Bhatti, P. T. (2018) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2018.2869141 · Cited by 24 · see citing works
Assantachai, P., Chatthanawaree, W., Thamlikitkul, V., Praditsuwan, R., & Pisalsarakij, D. (2002) · PubMed · Cited by 24 · see citing works
Artifact Noise Removal Techniques on Seismocardiogram Using Two Tri-Axial Accelerometers
Luu, L., & Dinh, A. (2018) · Sensors · doi:10.3390/s18041067 · Cited by 22 · see citing works
Isilay, Z. M., Racca, V., Pezzano, A., Tavanelli, M., & Rienzo, M. D. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.825918 · Cited by 22 · see citing works
Systolic and diastolic cardiac function time intervals and exercise capacity in women
Libonati, J. R., Colby, A. M., Caldwell, T. M., KASPARIAN, R., & GLASSBERG, H. L. (1999) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199902000-00009 · Cited by 22 · see citing works
Correlation between seismocardiogram and systolic blood pressure
Imtiaz, M. S., Shrestha, R., Dhillon, T., Yousuf, K. A., Saeed, B., Dinh, A., & Wahid, K. A. (2013) · Conference proceedings - Canadian Conference on Electrical and Computer Engineering · doi:10.1109/ccece.2013.6567773 · Cited by 22 · see citing works
Heart Activity Monitoring on Smartphone
Dinh, A. (2011) · Cited by 18 · see citing works
Exercise training improves left ventricular isovolumic relaxation
Libonati, J. R. (2000) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200008000-00007 · Cited by 18 · see citing works
Standalone Systolic Profile Detection of Non-Contact SCG Signal With LSTM Network
Li, Y., Xia, Z., & Zhang, Y. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2957382 · Cited by 17 · see citing works
Zhang, G., Cottrell, A. C., Henry, I., & McCombie, D. B. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591454 · Cited by 15 · see citing works
Clustering of SCG Events Using Unsupervised Machine Learning
Gamage, P. T., Azad, M. K., Taebi, A., Sandler, R. H., & Mansy, H. A. (2020) · Springer eBooks · doi:10.1007/978-3-030-36844-9_7 · Cited by 14 · see citing works
Seismocardiography: Practical implementation and feasibility
Paukkunen, M. (2014) · Aaltodoc (Aalto University) · Cited by 14 · see citing works
Korzeniowska‐Kubacka, I., Kuśmierczyk‐Droszcz, B., Bilińska, M., Dobraszkiewicz-Wasilewska, B., Mazurek, K., & Piotrowicz, R. (2006) · Via Medica Journals · Cited by 12 · see citing works
Zia, J. S., Kimball, J. P., Rozell, C. J., & Inan, O. T. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3014040 · Cited by 10 · see citing works
Timing detection and seismocardiography waveform extraction
Nguyen, H., Zhang, J., & Nam, Y. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346733 · Cited by 9 · see citing works
The impact of selection bias on measurement of noninvasive test accuracy
Salerno, D. M., Wang, K., Goldenberg, I. F., & Tassel, R. A. V. (1993) · The American Journal of Cardiology · doi:10.1016/0002-9149(93)90164-8 · Cited by 9 · see citing works
Anwar, T., Zubair, M., Rahman, M. D., Sakib, N., Kabir, M. A.-U., Faruk, T., & Islam, M. K. (2019) · International Conference on Advances in Science, Engineering and Robotics Technology (ICASERT) · doi:10.1109/icasert.2019.8934619 · Cited by 8 · see citing works
Intelligent Recognition of Electrocardiograms Using Selective Neuron Networks and Deep Learning
Mazurov, M. E. (2017) · Advances in intelligent systems and computing · doi:10.1007/978-3-319-67349-3_17 · Cited by 7 · see citing works
Modeling Consistent Dynamics of Cardiogenic Vibrations in Low-Dimensional Subspace
Zia, J. S., Kimball, J. P., Hersek, S., & Inan, O. T. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.2980979 · Cited by 7 · see citing works
Noninvasive myocardial contractility monitoring with seismocardiography during simulated dives.
Koch, A., McCormack, P., Schwanecke, A., Schnoor, P. W., Buslaps, C., Tetzlaff, K., & Rieckert, H. (2003) · PubMed · Cited by 7 · see citing works
A new seismocardiography segmentation algorithm for diastolic timed vibrations
Khosrow-Khavar, F., Tavakolian, K., Soleimani-Nouri, M., Kamińska, B., & Menon, C. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6611238 · Cited by 6 · see citing works
Vibration Isolation for Seismocardiogram Measurement in the OpenMRI-Guided Operating Theater
Naemura, K., & Iseki, H. (2003) · JSME International Journal Series C · doi:10.1299/jsmec.46.1426 · Cited by 6 · see citing works
Zhang, H., Wang, Z., Dong, K., Huat, N. S., & Lin, Z. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8036933 · Cited by 6 · see citing works
Casanella, R., Khosrow-Khavar, F., Schmidt, S. E., Zanetti, J. M., & Tavakolian, K. (2019) · Series in bioengineering · doi:10.1007/978-981-10-5092-3_3 · Cited by 6 · see citing works
Johnson, E., Etemadi, M., Malaisrie, S. C., McCarthy, P. M., Markl, M., & Barker, A. J. (2019) · Journal of Cardiac Surgery · doi:10.1111/jocs.14289 · Cited by 5 · see citing works
Cardiac cycle and heart rate calculation based on seismocardiogram
Imtiaz, M. S., Shrestha, R., Dhillon, T., Yousuf, K. A., Saeed, B., Dinh, A., & Wahid, K. A. (2013) · Conference proceedings - Canadian Conference on Electrical and Computer Engineering · doi:10.1109/ccece.2013.6567772 · Cited by 5 · see citing works
Agam, A., Agam, A., Korsgaard, E., Haugstrup, T. Y., Kristensen, C. B., Møgelvang, R., Kragholm, K. H., Emerek, K. J. G., Søgaard, P., & Schmidt, S. E. (2025) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztaf037 · Cited by 4 · see citing works
Application of laser doppler vibrocardiography for human heart auscultation
Koegelenberg, S. (2014) · SUNScholar (Stellenbosch University) · Cited by 4 · see citing works
Zienkiewicz, A., Korhonen, V. O., Kiviniemi, V. V., & Myllylä, T. (2024) · Biosensors · doi:10.3390/bios14120621 · Cited by 3 · see citing works
An accelerometer based system to measure myocardial performance index during stress testing
Dinh, A., Bui, F. M., & Nguyen, T. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591820 · Cited by 3 · see citing works
Carleton, R. N., Teckchandani, T. A., Neary, J. P., Samayoa, J., Khoury, J. M. B., Maguire, K. Q., Krätzig, G. P., & Asmundson, G. J. G. (2025) · Journal of Psychiatric Research · doi:10.1016/j.jpsychires.2025.10.044 · Cited by 2 · see citing works
Sejsmokardiografia — nieinwazyjna metoda diagnostyki niedokrwienia w chorobie niedokrwiennej serca
Korzeniowska‐Kubacka, I. (2001) · Via Medica Journals · Cited by 2 · see citing works
A Low Cost Sensing Device to Detect Cardiac Timing and Function
Dinh, A., Ko, S., Neary, J. P., & Quarrie, D. M. (2012) · IEEE International Mixed-Signals, Sensors and Systems Test Workshop (IMS3TW) · doi:10.1109/ims3tw.2012.29 · Cited by 1 · see citing works
Naemura, K., & Iseki, H. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1403962 · Cited by 1 · see citing works
Przydatność sejsmokardiografii wysiłkowej w rozpoznawaniu choroby niedokrwiennej serca
Korzeniowska‐Kubacka, I., Rausińska-Nocny, L., Mazurek, K., Wasilewska, B., & Bątkowska, E. R. I. K. (2001) · Via Medica Journals · Cited by 1 · see citing works
Paukkunen, M., & Linnavuo, M. (2014) · Advances in systems analysis, software engineering, and high performance computing book series · doi:10.4018/978-1-4666-6252-0.ch011 · Cited by 1 · see citing works
Seismocardiography: Interpretation and Clinical Application
Sørensen, K. (2021) · doi:10.54337/aau451017905 · Cited by 1 · see citing works
Luu, L. (2017) · HARVEST (University of Saskatchewan)
Detection and prediction of cardiac quiescence for computed tomography coronary angiography
Wick, C. A. (2014) · SMARTech Repository (Georgia Institute of Technology)
- Authors
- Wilson, R. A., Bamrah, V. S., Lindsay, J., Schwaiger, M., & Morganroth, J.
- Affiliations
- Oregon Health & Science University (US)
- Corresponding author
- Wilson, Richard A.
- ORCID
- Schwaiger, Markus
- Venue
- The American Journal of Cardiology
- Publisher
- Elsevier BV
- Published
- 1993-03
- Volume
- 71
- Issue
- 7
- Pages
- 536-545
- ISSN
- 0002-9149 (ISSN-L)
- OpenAlex
- W1978300338
- Cited by
- 61
- References
- 29
- Citations by year
- 2025: 2 · 2024: 1 · 2022: 3 · 2021: 1 · 2020: 8 · 2019: 5 · 2018: 3 · 2017: 3 · 2016: 4 · 2015: 1 · 2014: 6 · 2013: 6 · 2012: 2
- FWCI
- 0.39
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- The American Journal of Cardiology (publishedVersion)
- PubMed (publishedVersion)
- Deep Blue (University of Michigan) (acceptedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- seismocardiography, electrocardiogram, coronary artery disease, exercise stress testing, diagnostic accuracy, myocardial ischemia, thallium-201 scintigraphy, coronary angiography, sensitivity and specificity, women, submaximal exercise
- MeSH terms
- Coronary Disease, Electrocardiography, Exercise Test, Female, Heart Function Tests, Humans, Male, Middle Aged, Predictive Value of Tests, ROC Curve, Sensitivity and Specificity, Thallium Radioisotopes, Tomography, Emission-Computed, Single-Photon, Logistic Models, Coronary Angiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review