The effect of microgeometry, implant thickness and polyurethane chemistry on the foreign body response to subcutaneous implants
The DOI matches its Crossref record; 27 of the 28 references it cites were found in a registry or an index.
Abstract
We addressed the effect of implant thickness, implant porosity, and polyurethane (PU) chemistry on angiogenesis and on the foreign body response in rats. The following materials were implanted subcutaneously for 7 weeks then excised for histologic analysis: a solid PU; a solid polyurethane with silicone and polyethylene oxide (PU-S-PEO); porous expanded polytetrafluoroethylene (ePTFE); and porous polyvinyl alcohol sponge (PVA). Two thicknesses of PU-S-PEO were compared: 300 microns (thin) and 2000 microns (thick). Foreign body capsule (FBC) thickness was much less in PU-S-PEO implants than in PU implants. In addition, FBC were thinner in thin implants than in thick implants. FBC was much more dense in solid implants than in porous implants. As compared with solid implants, porous implants (PVA and ePTFE) led to a marked increase in the number of microvessels that developed adjacent to the implant, as observed both with hematoxylin/eosin staining and with an immunohistochemical anti-endothelial stain. We conclude that the polyethylene oxide and silicone moieties in PU reduce the thickness of the subsequent FBC. In addition, thin implants lead to a thin FBC. Porous implants (PVA and ePTFE) cause more angiogenesis than solid implants. These results may have implications for the measurement of blood-derived analytes by biosensors.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“This suggests that fibers with smaller diameters provided a thicker FBC formation.[ 124 ]”
From Unraveling and Harnessing the Immune Response at the Cell–Biomaterial Interface for Tissue Engineering Purposes (2024) · cites this paper as [124]
In the citing paper
“In macroscale, a similar response was observed for cylindrical polyurethane substrates provoking a larger FBR and thicker FBC in large diameter cylinders of 2 mm compared to 0.3 mm implants.”
From Unraveling and Harnessing the Immune Response at the Cell–Biomaterial Interface for Tissue Engineering Purposes (2024) · cites this paper as [124]
In the citing paper
“They found that increasing the size of implanted materials resulted in larger foreign body reactions, and the formation of a thicker layer of fibrosis around the implant23.”
From Size- and shape-dependent foreign body immune response to materials implanted in rodents and non-human primates (2015) · cites this paper as [23]
In the citing paper
“the physical properties of the implant (e.g., size) impact the FBR, and the large size of the stainless steel cage will influence the overall immune response to the materials.115”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [115]
In the citing paper
“the capsule thickness was ~20% thinner surrounding the smaller implant.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [115]
1 more claim about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Protein adsorption to poly(ethylene oxide) surfaces
Gombotz, W. R., Guanghui, W., Horbett, T. A., & Hoffman, A. S. (1991) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820251211 · Cited by 634
Neovascularization of synthetic membranes directed by membrane microarchitecture
Brauker, J. H., Carr‐Brendel, V. E., Martinson, L. A., Crudele, J., Johnston, W. D., & Johnson, R. C. (1995) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820291208 · Cited by 517 · cited 2×
“We were interested in counting only those vessels that were close to the implant membrane (‘‘close vascular structures’’ [11], so only those vessels that were within the rectangle (within 40 mm of the implant surface) were counted.”
BackgroundMethodPolyurethane Elastomer Biostability
Stokes, K., McVenes, R., & Anderson, J. M. (1995) · Journal of Biomaterials Applications · doi:10.1177/088532829500900402 · Cited by 366 · cited 1×
“Both PUs (PU-S-PEO and PU) used in our study were made with polycarbonate which has been shown to favor biocompatibility in that it limits chronic biodegradation and stress cracking [17,18].”
MethodEngineering the tissue which encapsulates subcutaneous implants. I. Diffusion properties
Sharkawy, A. A., Klitzman, B., Truskey, G. A., & Reichert, W. M. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19971205)37:3<401::aid-jbm11>3.0.co;2-e · Cited by 238
den Braber, E. T., de Ruijter, J. E., Ginsel, L. A., von Recum, A. F., & Jansen, J. A. (1996) · Biomaterials · doi:10.1016/0142-9612(96)00032-4 · Cited by 208
den Braber, E. T., de Ruijter, J. E., Smits, H. T. J., Ginsel, L. A., von Recum, A. F., & Jansen, J. A. (1996) · Biomaterials · doi:10.1016/0142-9612(96)85910-2 · Cited by 198 · cited 1×
“, has shown that specific geographic features in substrate microgeometry (for example, channel width) play an important role in the way that fibroblasts orient and attach to substrates [26,27].”
BackgroundQuantitative in-vivo studies on angiogenesis in a rat sponge model.
Andrade, S. P., Fan, T., & Lewis, G. P. (1987) · PubMed Central · Cited by 180
Sharkawy, A. A., Klitzman, B., Truskey, G. A., & Reichert, W. M. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19980615)40:4<586::aid-jbm10>3.0.co;2-e · Cited by 145
A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.
Updike, S. J., Shults, M. C., Gilligan, B. J., & Rhodes, R. K. (2000) · Diabetes Care · doi:10.2337/diacare.23.2.208 · Cited by 135
Tissue Response to Suture Materials Implanted Subcutaneously in a Rabbit Model
Setzen, G., & Williams, E. F. (1997) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-199712000-00023 · Cited by 117 · cited 1×
“Setzen and Williams examined the tissue response to sutures implanted subcutaneously in rabbits (diameter of approximately 200 mm) and found FBC thicknesses of 85–125 mm [22].”
Background
Cited by
Items 101–175 of 175
Characterization and toxicological effects of three-dimensional graphene foams in rats in vivo
Zha, Y., Chai, R., Song, Q., Chen, L., Wang, X., Cheng, G., Tang, M., & Wang, M. (2016) · Journal of Nanoparticle Research · doi:10.1007/s11051-016-3425-y · Cited by 17
Histological biocompatibility of new, non‐absorbable glaucoma deep sclerectomy implant
Kałuzny, J. J., Jóźwicki, W., & Wiśniewska, H. (2006) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.30677 · Cited by 17
Hadavi, E., de Vries, R. H. W., Smink, A. M., de Haan, B., Leijten, J., Schwab, L. W., Karperien, M., de Vos, P., Dijkstra, P. J., & van Apeldoorn, A. A. (2020) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.34686 · Cited by 16
Combination of Nitric Oxide Release and Surface Texture for Mitigating the Foreign Body Response
Malone‐Povolny, M. J., Bradshaw, T. M., Merricks, E. P., Long, C. T., Nichols, T. C., & Schoenfisch, M. H. (2021) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.1c00022 · Cited by 18
Howes, P. D., & Green, M. (2010) · Photochemical & Photobiological Sciences · doi:10.1039/c0pp00106f · Cited by 16
An Implantable Subcutaneous Glucose Sensor Array in Ketosis‐prone Rats: Closed Loop Glycemic Control
Ward, W. K., Wood, M. D., Casey, H. M., Quinn, M. J., & Federiuk, I. F. (2005) · Artificial Organs · doi:10.1111/j.1525-1594.2005.29024.x · Cited by 18
Self-citationIntravital microscopy evaluation of angiogenesis and its effects on glucose sensor performance
Koschwanez, H. E., Reichert, W. M., & Klitzman, B. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32630 · Cited by 15
Janis, A. D., Johnson, C., Ernst, D. M. J., & Brightman, A. O. (2011) · Journal of Biomaterials Applications · doi:10.1177/0885328210391688 · Cited by 15
Blood Flow in the Foreign-Body Capsules Surrounding Surgically Implanted Subcutaneous Devices
Bartoli, C. R., & Godleski, J. J. (2008) · Journal of Surgical Research · doi:10.1016/j.jss.2008.07.045 · Cited by 17
“In relationship to biomaterials and implants of millimeter dimension or smaller, investigators have quantified histological characteristics of the FBC such as thickness [1,7,9–11], collagen content and density [1,5,7,11] fibrinogen content [12], cell composition [7,11,13–15], size and extent of vessel formation [7,10]…”
BackgroundCorticosteroid Modulation of Tissue Responses to Implanted Sensors
Friedl, K. E. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.898 · Cited by 15
Surface Topography of PLA Implants Defines the Outcome of Foreign Body Reaction: An In Vivo Study
Иванова, Е., Fayzullin, A. L., Минаев, Н. В., Dolganova, I. N., Serejnikova, N. B., Gafarova, E. R., Tokarev, M., Minaeva, E., Aleksandrova, P. V., Reshetov, I. V., et al. (2023) · Polymers · doi:10.3390/polym15204119 · Cited by 14
Imaichi-Kobayashi, S., Kassab, R., Piersigilli, A., Robertson, R., Leonard, C., Long, N., Dean, B., Phaneuf, M., & Ling, V. (2023) · Biomaterials · doi:10.1016/j.biomaterials.2023.122123 · Cited by 13
Gao, X., Yang, J., Gan, X., Lin, Y., Xu, J., Shan, Z., Han, Z., Chen, S., Huang, B., Fan, B., et al. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.152044 · Cited by 13
Overview on Cell-Biomaterial Interactions
Gonçalves, S., Dourado, F. O. Q., & Rodrigues, L. R. (2014) · Springer eBooks · doi:10.1007/978-3-319-12478-0_4 · Cited by 16
Electrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 16
“, decreased encapsulation [348].”
MethodLong-term Evaluation of Porous PEGT/PBT Implants for Soft Tissue Augmentation
Lamme, E. N., Druecke, D., Pieper, J., May, P. S., Kaim, P., Jacobsen, F., Steinau, H., & Steinstraesser, L. (2007) · Journal of Biomaterials Applications · doi:10.1177/0885328206075552 · Cited by 16
Oe, S., Masum, M. A., Ichii, O., Nishimura, T., Nakamura, T., Namba, T., Otani, Y., Nakayama, Y., Elewa, Y. H. A., & Kon, Y. (2020) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.37115 · Cited by 12
“The texture of implant materials affects the formation and 57 histological structure of capsule forms surrounding the implant due to foreign body 58 reactions (2).”
BackgroundInnate and Adaptive Immune Responses in Tissue Engineering
Norton, L. W., & Babensee, J. E. (2009) · Springer eBooks · doi:10.1007/978-3-540-77755-7_50 · Cited by 12
Fabrication of Patterned Integrated Electrochemical Sensors
Mujeeb-U-Rahman, M., Adalian, D., & Scherer, A. (2015) · Journal of Nanotechnology · doi:10.1155/2015/467190 · Cited by 14
“This size reduction also provides major advantages in reducing individual device cost and foreign body response to such devices when used for in vivo applications [8].”
BackgroundScaffold-Assisted Ectopic Transplantation of Internal Organs and Patient-Derived Tumors
Carpenter, R., Oh, H. J., Ham, I., Kim, D., Hur, H., & Lee, J. (2019) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.9b00978 · Cited by 11
In Vivo Delivery of M0, M1, and M2 Macrophage Subtypes via Genipin-Cross-Linked Collagen Biotextile
Isali, I., Mclellan, P., Wong, T. R., Cingireddi, S., Jain, M. K., Anderson, J. M., Hijaz, A., & Akkuş, O. (2022) · Tissue Engineering. Part A · doi:10.1089/ten.tea.2021.0203 · Cited by 10
Systems of conductive skin for power transfer in clinical applications
Kourouklis, A. P., Kaemmel, J., Wu, X., Potapov, E. V., Cesarovic, N., Ferrari, A., Starck, C. T., Falk, V., & Mazza, E. (2021) · European Biophysics Journal · doi:10.1007/s00249-021-01568-8 · Cited by 10
Percutaneous Window Chamber Method for Chronic Intravital Microscopy of Sensor—Tissue Interactions
Koschwanez, H. E., Klitzman, B., & Reichert, W. M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200608 · Cited by 9
Miller, K. J., Cao, W., Ibrahim, M. M., & Levinson, H. (2017) · Advances in Skin & Wound Care · doi:10.1097/01.asw.0000513149.43488.56 · Cited by 10
“Excessive fibrosis around an implant can have severe consequences including device failure.(11,12) Peri-implant fibrosis around a medical device, prosthesis, or biomaterial ultimately results in an end-stage host immune response termed the FBR.”
BackgroundSensor Biocompatibility and Biofouling in Real‐Time Monitoring
Voskerician, G., & Anderson, J. M. (2006) · Wiley eBooks · doi:10.1002/9780471740360.ebs1370 · Cited by 10
Lužajić-Božinovski, T., Todorović, V., Milošević, I., Prokić, B., Gajdov, V., Nešović, K., Mišković-Stanković, V., & Marković, D. (2021) · Journal of Biomaterials Applications · doi:10.1177/08853282211046119 · Cited by 7
Silicone-containing thermoresponsive membranes to form an optical glucose biosensor
Dong, P. X., Singh, K. A., Soltes, A. M., Ko, B. S., Gaharwar, A. K., McShane, M. J., & Grunlan, M. A. (2022) · Journal of Materials Chemistry B · doi:10.1039/d2tb01192a · Cited by 7
Barsch, F., Mamilos, A., Schmitt, V. H., Babel, M., Winter, L., Wagner, W., Winther, H., Ottomann, C., Niedermair, T., Schreml, S., et al. (2022) · Cells · doi:10.3390/cells11182834 · Cited by 6
“They found thicker fibrotic capsules and higher capsule densities in the solid material group [88].”
BackgroundDynamic and living devices for overcoming fibrosis of implanted biomaterials
Jackson, J. M., Testu, L., & Abramson, A. (2025) · Bioengineering & Translational Medicine · doi:10.1002/btm2.70048 · Cited by 9
“30 Another group developed poly (lactic-co-glycolic) acid (PLGA)/polyvinyl alcohol (PVA) hydrogel-coated biosensors of three different sizes (0.3 (cid:1) 0.3 (cid:1) 3 mm 3 ; 0.5 (cid:1) 0.5 (cid:1) 5 mm 3 ; 0.75 (cid:1) 0.75 (cid:1) 9 mm 3 ) and similarly found capsular thickness increased with…”
InfluentialSafety, biodegradability, and biocompatibility considerations of long-acting drug delivery systems
Mishra, D., Glover, K., Gade, S., Sonawane, R., & Singh, T. R. R. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-821749-8.00008-2 · Cited by 7
Cen, C., Zhang, Y., Cao, Y., Hu, C., Tang, L., Liu, C., Wang, T., & Peng, W. (2025) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.4c01849 · Cited by 6
“62 Roughened grafts have been found to furnish better traction, reduce tissue motion, and generate a thinner membrane.”
Lopez‐Silva, T. L., Anderson, C. F., & Schneider, J. P. (2025) · ACS Central Science · doi:10.1021/acscentsci.4c02198 · Cited by 5
New opportunities for bioscaffold‐enabled spinal cord injury repair
Qi, X., Fu, Y., Su, Z., Li, L., Chakrabarti, S., Li, P., Wu, Y., Liu, F., Gao, T., Dong, Z., et al. (2025) · BMEMat · doi:10.1002/bmm2.70025 · Cited by 7
Impact of biomaterials’ physical properties on cellular and molecular responses
He, J., & Burgess, D. J. (2020) · Elsevier eBooks · doi:10.1016/b978-0-08-102967-1.00005-0 · Cited by 6
Subdermal implants of poly(L-lactic acid) with plasticizer: an ultrastructural study in rats
Silva, D. R. M., Joazeiro, P. P., Duek, E. A. D. R., & Alberto-Rincon, M. C. (2005) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/156856206774879018 · Cited by 5
Esophagus: A Tissue Engineering Challenge
Ritchie, A. C., Chian, K. S., Ratner, B. D., & Beckstead, B. L. (2006) · The electrical engineering handbook/Electrical engineering handbook series/Electrical engineering handbook · doi:10.1201/9781420003871.ch54 · Cited by 5
Development of 3D-printed biodegradable composite scaffolds for tissue engineering applications
Serra, T. (2014) · UPCommons institutional repository (Universitat Politècnica de Catalunya) · doi:10.5821/dissertation-2117-95344 · Cited by 15
“6 Synergistically coupling the state of the art in scaffolds fabrication techniques [9] with the advances in the field of developmental biology [10] lead to new strategies in tissue engineering such as the attainment of faster and cheaper high-throughput analysis systems and even the reduction of animal testing [10].”
BackgroundBar-Shai, N., Mann, D., Shalom, A., Nyska, A., & Golberg, A. (2025) · Scientific Reports · doi:10.1038/s41598-025-19632-0 · Cited by 5
“Porous structures promote cell infiltration, connective tissue development, and angiogenesis, whereas dense, non-porous scaffolds can hinder infiltration and diffusion, limit tissue integration 38,56,57,62,63 .”
Immunomodulatory Delivery Materials for Tissue Repair
Est-Witte, S. E., Jain, M., Ben‐Akiva, E., Yang, J., Warren, T., Tzeng, S. Y., & Green, J. J. (2025) · Advanced Healthcare Materials · doi:10.1002/adhm.202501400 · Cited by 4
Emerging Trends in Immunomodulatory Nanomaterials Toward Cancer Therapy
Mukherjee, A., Madamsetty, V. S., & Mukherjee, S. (2021) · Synthesis lectures on biomedical engineering · doi:10.2200/s01083ed1v01y202103bme061 · Cited by 4
Lamination for subdermal implant fixation
Hori, B. D., Petrell, R. J., Trites, A. W., & Godbey, T. (2009) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.31369 · Cited by 5
Biocompatibility of Supercritical CO 2 -Treated Titanium Implants in a Rat Model
Hill, C. M., Kang, Q. K., Wahl, C., Jimenez, A., LaBerge, M., Drews, M., Matthews, M., & An, Y. H. (2006) · The International Journal of Artificial Organs · doi:10.1177/039139880602900413 · Cited by 5
“Under l ight microscopy each soft t issue/implant interface was evaluated using a previously described method (11).”
BackgroundMethodThe Role of Piezoelectric Materials for Antibacterial Therapy
Dey, A., Bag, N., Mondal, D., Roy, J., Ghosh, S., Roy, S., Guo, B., & Das, S. (2025) · Wiley eBooks · doi:10.1002/9783527846979.ch16 · Cited by 3
Beta‐adrenoceptor blockade delays granulation tissue formation in polyurethane sponge implants
Romana‐Souza, B., dos Santos, J. S., Desmoulière, A., & Monte‐Alto‐Costa, A. (2009) · Journal of Cutaneous Pathology · doi:10.1111/j.1600-0560.2008.01110.x · Cited by 3
“This response may be explained by the microgeometry of the material: implants (such as PU sponges) with a large number of pores and channels may result in a prolonged inflammatory response and delayed deposition and reorganization of the collagen matrix.(22) Development of normal granulation tissue requires a precisel…”
BackgroundDendritic cell force-migration coupling on aligned fiber networks
Hernandez-Padilla, C., Joosten, B., Franco, A. T., Cambi, A., van den Dries, K., & Nain, A. S. (2024) · Biophysical Journal · doi:10.1016/j.bpj.2024.07.011 · Cited by 3
The Macrophage in Wound Healing Surrounding Implanted Devices
Godek, M. L., & Grainger, D. W. (2009) · Wiley eBooks · doi:10.1002/9780470567319.ch2 · Cited by 3
Strategies to Overcome Biological Barriers to Biosensing
Ward, W. K., & Duman, H. M. (2009) · Wiley eBooks · doi:10.1002/9780470567319.ch3 · Cited by 3
Self-citationA Review of the Management of Implanted Medical Devices for Diabetes: Trends and Directions
Edman, C. F., & Drinan, D. D. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200609 · Cited by 3
5.4 Biomaterials and the Microvasculature ☆
Joddar, B., Sieminski, A. L., Hogrebe, N. J., Tennant, C. J., & Gooch, K. J. (2016) · Elsevier eBooks · doi:10.1016/b978-0-12-803581-8.09820-9 · Cited by 4
Agrawal, C. M., Ong, J. L., Appleford, M. R., & Mani, G. (2013) · Cambridge University Press eBooks · doi:10.1017/cbo9781139035545.012 · Cited by 1
Optimization Techniques for Miniaturized Integrated Electrochemical Sensors
Mujeeb-U-Rahman, M., & Scherer, A. (2015) · CaltechAUTHORS (California Institute of Technology) · Cited by 1
Immune-instructive materials and surfaces for medical applications
Fisher, L. E., & Ghaemmaghami, A. M. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-821440-4.00008-6 · Cited by 1
Transdermal wires for improved integration in vivo
Kourouklis, A. P., Kaemmel, J., Wu, X., Baños, M., Chanfon, A., de Brot, S., Ferrari, A., Cesarovic, N., Falk, V., & Mazza, E. (2023) · Biomaterials Advances · doi:10.1016/j.bioadv.2023.213568 · Cited by 1
Koschwanez, H. E. (2009) · DukeSpace (Duke University) · Cited by 2
Modulation of macrophage polarization with surface immobilized bioactive molecules
Chen, A. (2016) · ResearchWorks at the University of Washington (University of Washington) · Cited by 1
Gad, S. C. (2014) · Elsevier eBooks · doi:10.1016/b978-0-12-386454-3.00846-0 · Cited by 1
Host responses to microgel-based biomaterial interfaces
Bridges, A. W. (2008) · SMARTech Repository (Georgia Institute of Technology) · Cited by 1
Schwartz, M. (2005) · UA Campus Repository (The University of Arizona)
Soft Tissue Structure and Functionality
Voskerician, G. (2012) · Wiley eBooks · doi:10.1002/9781118129906.ch2
Cao, H. (2011) · DR-NTU (Nanyang Technological University) · doi:10.32657/10356/46428
Reconstruction of Orbital Wall Fracture An experimental and clinical study
Kontio, R. K. (2005) · Työväentutkimus Vuosikirja · Cited by 2
Babasola, I. O. (2012) · QSpace (Queen's University Library)
Sylvester, M. L. (2013) · ResearchWorks at the University of Washington (University of Washington)
Zhang, L. (2012) · ResearchWorks at the University of Washington (University of Washington)
In Vivo Studies of the Foreign Body Reaction to Biomedical Polymers
Yang, J. H. (2013) · OhioLink ETD Center (Ohio Library and Information Network)
Biomaterials and the Microvasculature
Joddar, B., Sieminski, A. L., Tennant, C. J., & Gooch, K. J. (2011) · Elsevier eBooks · doi:10.1016/b978-0-08-055294-1.00157-4
Esophagus: A Tissue Engineering Challenge
Ritchie, A. C., Chian, K. S., Beckstead, B. L., & Ratner, B. D. (2007) · CRC Press eBooks · doi:10.1201/9781420008333.ch25
Gad, S. C. (2005) · Elsevier eBooks · doi:10.1016/b0-12-369400-0/00430-0
Immunomodulation and Various Strategies Effecting Immune Response
Mukherjee, A., Madamsetty, V. S., & Mukherjee, S. (2021) · Synthesis lectures on biomedical engineering · doi:10.1007/978-3-031-01669-1_3
Gad, S. C. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-824315-2.00789-2
Bioactive Composite Material Based on Polyurethane with Isoniazid
Кулєш, Д. В., Galatenko, N. А., Rozhnova, R. A., & Nechaeva, L. Y. (2023) · American Journal of Polymer Science and Technology · doi:10.11648/j.ajpst.20230901.11
Sannikov, A. B., & Shaydakov, E. V. (2023) · Ambulatornaya khirurgiya = Ambulatory Surgery (Russia) · doi:10.21518/akh2023-004
Selecting Biocompatible Biomaterials for Stem Cell-Derived β-Cell Transplantation
de Vries, R. H. W., & van Apeldoorn, A. A. (2023) · Springer eBooks · doi:10.1007/978-3-031-41943-0_6
Amara, H. B., Farjam, P., Lutz, T., Omar, O., Palmquist, A., Lieleg, O., Browne, M., Taylor, A., Verkerke, G. J., Rouwkema, J., et al. (2024) · Acta Biomaterialia · doi:10.1016/j.actbio.2024.05.042
“The biocompatibility of polyurethanes in general, and polycarbonate urethane in particular, has been traditionally inferred from long-term morphological studies in animal models ( 25, 26 ) or humans (27, 28) , and from cytocompatibility assays in vitro (29) .”
Biomimetic scaffolds based on stem cells for neural tissue engineering: a review
Kakehbaraei, S., Jalili, C., Bahreini, A., Nasta, T. Z., Kakebaraei, S., & Arab‐Zozani, M. (2026) · Stem Cell Research & Therapy · doi:10.1186/s13287-026-05182-3
“This reaction may result in the secretion of several cytotoxic agents, such as reactive oxygen species and proteolytic enzymes, which can compromise the integrity of the scaffold and nearby healthy tissue [241, 242].”
Metadata
- Authors
- Ward, W. K., Slobodzian, E. P., Tiekotter, K. L., & Wood, M. D.
- Affiliations
- Legacy Health (US)
- Corresponding author
- Ward, W.Kenneth
- ORCID
- Wood, Michael D
- Venue
- Biomaterials
- Publisher
- Elsevier BV
- Published
- 2002-11
- Volume
- 23
- Issue
- 21
- Pages
- 4185-4192
- ISSN
- 0142-9612 (ISSN-L)
- OpenAlex
- W2051278851
- Semantic Scholar
- a4b440b4403876866d6fbb47c3906a6f4d696195
- Cited by
- 176
- References
- 28
- Citations by year
- 2026: 2 · 2025: 10 · 2024: 8 · 2023: 11 · 2022: 9 · 2021: 15 · 2020: 11 · 2019: 7 · 2018: 7 · 2017: 7 · 2016: 7 · 2015: 10 · 2014: 4 · 2013: 10 · 2012: 7
- FWCI
- 4.49
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- Locations
- Biomaterials (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5r01dk055297-03)
- National Institute of Diabetes and Digestive and Kidney Diseases (R01 DK55297)
- Topics
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Physical Sciences › Materials Science › Surfaces, Coatings and Films › Polymer Surface Interaction Studies
- Physical Sciences › Materials Science › Biomaterials › Hydrogels: synthesis, properties, applications
- Keywords
- foreign body reaction, subcutaneous implants, polyurethane, implant thickness, angiogenesis, poly(ethylene oxide), silicone, porous implants, expanded polytetrafluoroethylene, polyvinyl alcohol sponge, microgeometry, biosensors
- MeSH terms
- Animals, Biocompatible Materials, Foreign-Body Reaction, Male, Microscopy, Electron, Scanning, Polyethylene Glycols, Polytetrafluoroethylene, Polyurethanes, Polyvinyl Alcohol, Porosity, Rats, Sprague-Dawley, Neovascularization, Physiologic, Prostheses and Implants, Subcutaneous Tissue, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.