Tear Glucose Analysis for the Noninvasive Detection and Monitoring of Diabetes Mellitus
The DOI matches its Crossref record; 98 of the 101 references it cites were found in a registry or an index.
Abstract
One approach to the noninvasive monitoring of blood glucose concentration is to monitor glucose concentrations in tear fluid. While several methods for sensing glucose in tear fluid have been proposed, controversy remains as to the precise concentrations of tear glucose in normal and diabetic subjects and as to whether tear fluid glucose concentrations correlate with blood glucose concentrations. This review covers the present understanding of the physiology of glucose transport in tears, the regulation of the aqueous tear fraction, and studies of tear glucose concentration over the last 80 years. The various tear collection methods employed greatly influence the measured tear glucose concentrations. Studies that involve mechanical irritation of the conjunctiva during sampling measure the highest tear glucose concentrations, while studies that avoid tear stimulation measure the lowest concentrations. Attempts to monitor tear glucose concentration in situ by using contact lens-based sensing devices are discussed, and new observations are presented of tear glucose concentration obtained by a method designed to avoid tear stimulation. These studies indicate the importance of the sampling method in determining tear glucose concentrations. On the basis of these results, we discuss the future of in vivo tear glucose sensing and outline the studies needed to resolve the remaining questions about the relationship between tear and blood glucose concentrations.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Additionally, crystalline colloidal arrays and holographic sensors have been incorporated into contact lenses to continuously sense glucose using phenylboronic acid derivatives in hydrogels.68”
From Wearables in Medicine (2018) · cites this paper as [68]
In the citing paper
“Studies designed to establish correlations between blood and tear glucose concentrations are inconclusive from both human (243-245) and animal studies (246).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [243]
In the citing paper
“Not only do biomarker molecules in tears diffuse directly from the blood and exhibit close tear–blood concentration correlations, but also tear analysis presents opportunities for the diagnosis of ocular disease.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [95]
In the citing paper
“Not only do biomarker molecules in tears diffuse directly from the blood and exhibit close tear–blood concentration correlations, but also tear analysis presents opportunities for the diagnosis of ocular disease.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [95]
In the citing paper
“In particular, tear glucose concentrations have been demonstrated to exhibit good correlation with blood glucose levels, reflecting the diffusion from the lachrymal artery when the tears are sampled without any eye irritation or stimulation, which can compromise the relationship94, 95, 99-101.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [95]
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Standards of Medical Care in Diabetes
Power, D. (2006) · Diabetes Care · doi:10.2337/diacare.29.02.06.dc05-1593 · Cited by 8,673
Economic Costs of Diabetes in the U.S. in 2002
American Diabetes Association (2003) · Diabetes Care · doi:10.2337/diacare.26.3.917 · Cited by 4,147
Effect of intensive therapy on the microvascular complications of type 1 diabetes mellitus
The Writing Team for the Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications Research Group (2002) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.287.19.2563 · Cited by 941
Functional aspects of the tear film lipid layer
Bron, A. J., Tiffany, J. M., Gouveia, S. M., Yokoi, N., & Voon, L. W. (2004) · Experimental Eye Research · doi:10.1016/j.exer.2003.09.019 · Cited by 782
Standards of Medical Care in Diabetes
American Diabetes Association (2004) · Diabetes Care · doi:10.2337/diacare.27.2007.s15 · Cited by 683
Toward a Glucose Biosensor Based on Surface-Enhanced Raman Scattering
Shafer-Peltier, K. E., Haynes, C. L., Glucksberg, M. R., & Van Duyne, R. P. (2003) · Journal of the American Chemical Society · doi:10.1021/ja028255v · Cited by 653
Determination of tear volume and tear flow.
Mishima, S., Gasset, A. R., Klyce, S. D., & BAUM, J. L. (1966) · PubMed · Cited by 607
Deiss, D., Bolinder, J., Riveline, J.-P., Battelino, T., Bosi, E., Tubiana-Rufi, N., Kerr, D., & Phillip, M. (2006) · Diabetes Care · doi:10.2337/dc06-1134 · Cited by 539
Economic Costs of Diabetes in the U.S. in 2002Preprint
Dall, T. M., Nikolov, P., & Hogan, P. F. (2003) · RePEc: Research Papers in Economics · Cited by 482
Research works printed but not linked: 6
- Michail, 1937, Lacrimal elimination of glucose in diabetic patients, C R Soc Biol Paris, 125, 194
- Fourcroy, 1791, Examen Chimique des Larmes et de l'Humeur des Narines, Annales de Chimie, 10, 113
- Balik, 1965, Method for the biochemical study of tears, Acta Ophthalmol, 43, 190
- Motoji, 1971, The glucose content of tear fluid in normal and diabetic subjects, Jpn J Clin Ophthalmol, 25, 1945
- Michail, 1937, Lacrimal elimination of glucose during alimentary hyperglycemia, CR Soc Biol Paris, 126, 1042
- Michail, 1937, Lachrymal elimination of glucose during adrenaline hyperglucemia, CR Soc Biol Paris, 125, 1095
Cited by
Items 101–187 of 187
Research Progress on Biomimetic Nanomaterials for Electrochemical Glucose Sensors
Chi, L., Zhang, C., Wu, X., Qian, X., Sun, H., He, M., & Guo, C. (2023) · Biomimetics · doi:10.3390/biomimetics8020167 · Cited by 18
Manjunatha, N. K., Raga, S., Gowre, S. C., Miyan, H., & Sharan, P. (2023) · Plasmonics · doi:10.1007/s11468-023-01857-2 · Cited by 17
“The normal cell glucose concentration is 0 to 65 mg/dL, and the affected diabetic cell is 84 mg/dL [37].”
BackgroundContact lens wear and the diabetic corneal epithelium: A happy or disastrous marriage?
Bussan, K. A., & Robertson, D. M. (2018) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2018.09.015 · Cited by 19
del Castillo, J. M. B., Pinazo-Duran, M. D., Sanz-Gonzalez, S. M., Muñoz-Hernández, A. M., García‐Medina, J. J., & Zanón‐Moreno, V. (2020) · Ophthalmic Research · doi:10.1159/000510211 · Cited by 23
Park, S., Nam, D. Y., Jeon, H., Han, J. H., Jang, D., Hwang, J., Park, Y.-S., Han, Y., Choy, Y. B., & Lee, D. Y. (2023) · Biomaterials Research · doi:10.1186/s40824-023-00469-5 · Cited by 18
“5d) [26, 59].”
BackgroundResultInfluentialOrdered Porous Gold Electrodes to Enhance the Sensitivity of Enzyme-Based Glucose Sensors
Paul, M. T. Y., Kinkead, B., & Gates, B. D. (2013) · Journal of The Electrochemical Society · doi:10.1149/2.019402jes · Cited by 16
A Disposable Tear Glucose Biosensor—Part 3: Assessment of Enzymatic Specificity
Lan, K., McAferty, K., Shah, P., Lieberman, E., Patel, D. R., Cook, C. B., & Belle, J. T. L. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500511 · Cited by 18
Assessing and Advancing Technology for the Noninvasive Measurement of Clinical Glucose
Bai, C., Graham, T. L., & Arnold, M. A. (2008) · Analytical Letters · doi:10.1080/00032710802418885 · Cited by 20
“A change in either the color (Baca, Finegold, and Asher, 2007; Ben-Moshe, Alexeev, and Asher 2006; Alexeev et al. 2004) or fluorescence intensity (Badugu, Lakowicz, and Geddes, 2005; March, Lazzaro, and Rastogi 2006; Pickup et al. 2005; Badugu, Lakowicz, and Geddes 2004) of the membrane (lens) is…”
BackgroundInfrared reflectometry of skin: Analysis of backscattered light from different skin layers
Pleitez, M. A., Hertzberg, O., Bauer, A., Lieblein, T., Glasmacher, M., Tholl, H. D., & Mäntele, W. (2017) · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · doi:10.1016/j.saa.2017.04.080 · Cited by 16
A Disposable Tear Glucose Biosensor—Part 4
Belle, J. T. L., Engelschall, E., Lan, K., Shah, P., Saez, N., Maxwell, S., Adamson, T., Abou-Eid, M., McAferty, K., Patel, D. R., et al. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813511741 · Cited by 16
“Glucose has long been known to be present in tear fluid (TF), making TF a particularly attractive alternative sampling source for glucose detection.(5) While there are still many challenges to overcome, self-monitoring of tear glucose (SMTG) has been heavily studied because of its potential painless and noninvasive ap…”
BackgroundChen, J., & Furukawa, H. (2023) · Infrared Physics & Technology · doi:10.1016/j.infrared.2023.104717 · Cited by 15
A Disposable Tear Glucose Biosensor—Part 2: System Integration and Model Validation
Belle, J. T. L., Bishop, D. K., Vossler, S. R., Patel, D. R., & Cook, C. B. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400210 · Cited by 14
Hansen, J. S., Petersen, J. F., Høeg-Jensen, T., & Christensen, J. B. (2012) · Tetrahedron Letters · doi:10.1016/j.tetlet.2012.08.066 · Cited by 15
Ratiometric glucose sensing based on fluorescent oxygen films and glucose oxidase
Su, F., Zhang, L., Kong, X., Lee, F., Tian, Y., & Meldrum, D. R. (2017) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2017.03.004 · Cited by 16
López-Peña, G., Ortiz-Mansilla, E., Arranz, A., Bogdan, N., Manso‐Silván, M., & Rodríguez, E. M. (2024) · Colloids and Surfaces B Biointerfaces · doi:10.1016/j.colsurfb.2024.113934 · Cited by 12
Park, S. H., & Pak, J. J. (2024) · Sensors · doi:10.3390/s24216945 · Cited by 14
Continuous glucose monitoring: Minimally and non-invasive technologies
Lee, J. E., Sridharan, B., Kim, D., Sung, Y., Park, J. H., & Lim, H. G. (2025) · Clinica Chimica Acta · doi:10.1016/j.cca.2025.120358 · Cited by 14
Donora, M., González‐Fernández, E., Quintero, A. V., Smet, H. D., & Underwood, I. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.126671 · Cited by 14
Data Reliability and Quality in Body Area Networks for Diabetes Monitoring
Huzooree, G., Khedo, K. K., & Joonas, N. (2018) · EAI/Springer Innovations in Communication and Computing · doi:10.1007/978-3-030-00865-9_4 · Cited by 15
Design and Development of an In Vitro Tear Replenishment System
Mohammadi, S., Postnikoff, C., Wright, A. M., & Gorbet, M. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1045-1 · Cited by 12
Bio-chemical markers of chronic, non-infectious disease in the human tear film
Alotaibi, S., Markoulli, M. P., Ozkan, J., & PAPAS, E. B. (2021) · Clinical and Experimental Optometry · doi:10.1080/08164622.2021.1974282 · Cited by 14
Stress biomarkers in the tear film
Banbury, L. K. (2009) · Cross Connect (Southern Cross University) · Cited by 20
Devi, L. S., Paily, R. P., & Dasmahapatra, A. K. (2024) · Polymer · doi:10.1016/j.polymer.2024.127947 · Cited by 12
Fundamental Characteristics of a Glucose Transistor with a Chemically Functional Interface
Uematsu, Y., Kajisa, T., & Sakata, T. (2017) · ChemElectroChem · doi:10.1002/celc.201700419 · Cited by 11
Kim, H., Lee, S., Lee, K. W., Kim, E. S., Kim, H., Im, H., Yoon, H. C., Ko, J., & Kim, E. (2023) · Dyes and Pigments · doi:10.1016/j.dyepig.2023.111287 · Cited by 14
Dang, D. K., Xuyen, V. T. K., Ta, Q. T. H., & Nhiem, L. T. (2025) · RSC Advances · doi:10.1039/d5ra05656j · Cited by 13
Toward clinical translation of biosensors for chronic kidney disease
Rath, R. J., Talebian, S., Mayol, B., Ding, S., Brasier, N., Jeerapan, I., Sempionatto, J. R., Naficy, S., & Silva, S. M. (2025) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2025.118328 · Cited by 12
Intra-Body Wireless Power and Data Links for Biomedical Microsystems
Chen, W., Gao, C., Zhu, Y., Lin, L., & Li, J. (2023) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas58349.2023.10388442 · Cited by 10
“It has also been integrated with a contact lens for detection of glucose levels in the aqueous humor [25] [26], which challenges the wireless connectivity, considering the temperature sensitiveness of the cornea (maximum at 35.41 ± 0.7 ℃ ) [27].”
BackgroundInfluentialQuantification of tear glucose levels and their correlation with blood glucose levels in dogs
Lee, E., Kang, S., Shim, J., Jeong, D., Jeong, Y., Ahn, J., & Seo, K. (2022) · Veterinary Medicine and Science · doi:10.1002/vms3.788 · Cited by 8
“Owing to these two reasons, including the origin of tears and barriers in the epitheliums of the cornea and conjunctiva, the concentration of TG is closely related to the methods of tear collection (Baca et al., 2007; van Haeringen & Glasius, 1977).”
BackgroundMethodInfluentialFully inkjet-printed glucose assay fabricated on highly porous pigment coating
Jutila, E., Koivunen, R., Bollström, R., & Gane, P. A. C. (2020) · Microfluidics and Nanofluidics · doi:10.1007/s10404-020-02344-4 · Cited by 8
Yetisen, A. K. (2014) · Springer theses · doi:10.1007/978-3-319-13584-7_5 · Cited by 9
Wen, X., Ng, T. K., Zhang, G., Chen, H., Wu, Z., Liu, Q., & Zhang, M. (2025) · Molecular Biomedicine · doi:10.1186/s43556-025-00297-0 · Cited by 10
Designing Minimally Invasive Preocular Contact Tips for Potential Application in Tear Collection
Lee, S. H., Cho, Y. C., Nam, D. Y., Huh, B. K., Kim, S.-N., Lee, D. H., Park, C. G., & Choy, Y. B. (2018) · Cornea · doi:10.1097/ico.0000000000001603 · Cited by 7
Changes in tear glucose and insulin concentrations following an oral glucose tolerance test
Chao, C., Lema, C., Redfern, R. L., & Richdale, K. (2022) · Clinical and Experimental Optometry · doi:10.1080/08164622.2022.2111204 · Cited by 6
“88 mg/dl, respectively).(7,10,11,30) Consistent with blood glucose, tear glucose levels in this study increased by 1.”
Innovative advancement of contact lenses for noninvasive diagnosis and therapy: A mini review
Wen, X., Liu, Q., & Zhang, M. (2023) · MedComm – Biomaterials and Applications · doi:10.1002/mba2.40 · Cited by 7
NanoLC-timsTOF-Assisted Analysis of Glycated Albumin in Diabetes-Affected Plasma and Tears
Tan, Y., Toba, E. A. D. L., Rubakhin, S. S., Labriola, L. T., Canfield, C., Pan, D., & Sweedler, J. V. (2023) · Journal of the American Society for Mass Spectrometry · doi:10.1021/jasms.3c00331 · Cited by 6
Tear Cocktail: Composition of Tears
Prashar, A. (2019) · Springer eBooks · doi:10.1007/978-981-13-7169-1_4 · Cited by 8
From Vision Correction to Drug Delivery: Unraveling the Potential of Therapeutic Contact Lens
Saini, A., Sharma, M., Singh, I., & Swami, R. (2024) · Current Drug Delivery · doi:10.2174/0115672018270396231213074746 · Cited by 7
Steinmetz, A., Jörn, U., Oechtering, G. U., & Heilmann, R. M. (2018) · Veterinary Ophthalmology · doi:10.1111/vop.12615 · Cited by 6
Energy harvesting system integrated on wearable contact lens
Chen, L., Shaker, G. S. A., & Safavi‐Naeini, S. (2015) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/aps.2015.7304565 · Cited by 5
“The conventional practice of pricking, where a glucometer draws blood through a small piercing of the skin, is currently being challenged by the strong correlation of glucose biomarker in basal tear fluid to that in blood: glucose in tear is enough to deduce its overall level in the body [1].”
BackgroundTiangco, C. E., Andar, A. U., Quarterman, J. C., Ge, X., Sevilla, F. B., III, Rao, G., Stinchcomb, A. L., Bunge, A. L., & Tolosa, L. M. (2017) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-017-0289-7 · Cited by 7
A Disposable Tear Glucose Biosensor—Part 5: Improvements in Reagents and Tear Sampling Component
Lin, C., Ito, Y., Deng, A., Johns, J., Matloff, D. S., Cook, C. B., Sode, K., & Belle, J. T. L. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818769944 · Cited by 7
Lemp, M. A., & Beuerman, R. W. (2011) · Elsevier eBooks · doi:10.1016/b978-0-323-06387-6.00010-6 · Cited by 4
Tear Fluid Photonic Crystal Contact Lens Noninvasive Glucose Sensors
Baca, J. T., & Myllylä, R. A. (2008) · Series in medical physics and biomedical engineering · doi:10.1201/9781584889755.ch13 · Cited by 7
Self-citationSensor Ecosystems for Health Self‐Monitoring: Focus on Diabetes Management
Ganguly, A., Churcher, N. K. M., Upasham, S., Dhamu, V. N., & Prasad, S. (2023) · Wiley eBooks · doi:10.1002/9783527834266.ch44 · Cited by 4
Tear glucose is associated with the presence and severity of diabetic retinopathy
Cao, N., Feng, L., Xu, W., He, F., Xiao, C., Liu, Y., Xu, W., Cui, J., Guo, Y., Yao, L., et al. (2025) · International Journal of Retina and Vitreous · doi:10.1186/s40942-025-00636-x · Cited by 4
“However, this method is invasive [9] and inconvenient, for it causes pain, loss of finger sensitivity and scar formation [10], which compromises the compliance of patients.”
Tang, L., Zheng, J., Li, Z., Liu, F., Wang, L., & Yu, W. W. (2026) · Science China Materials · doi:10.1007/s40843-025-3780-3 · Cited by 3
Smart contact lens–trained digital twin for device-free personalized uric acid prediction
Song, H., Hong, Y., Kim, D., Seo, H., Park, W., Paek, J., Lee, D., Cho, S. K., Ahn, S. S., Kim, J., et al. (2026) · Science Advances · doi:10.1126/sciadv.aee8506 · Cited by 3
“This transient spike is attributed to UA leakage into the interstitial space or epithelial cells, which can distort the SUA-TUA correlation ( 32 , 57 , 58 ).”
Wang, H., He, Y., Wang, Q., Zhang, X., Liu, Y., & Zhang, J. (2025) · Langmuir · doi:10.1021/acs.langmuir.5c02962 · Cited by 4
Noninvasive and Ultrasensitive Glucose Detection via Electrolyte-Gated IGZO Thin-Film Transistors
Hwang, C., Park, S., & Kim, D. Y. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3661008 · Cited by 2
“Tears, in particular, have garnered significant interest due to their ease of collection, minimal discomfort during sampling, relatively low infection risk, and rapid glucose responsiveness attributed to the “plasma leakage” phenomenon [18], [19], [20].”
Teutsch, T., Mesch, M., Giessen, H. W., & Tarín, C. (2017) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2017.08.913 · Cited by 2
Development of a semi-quantitative tear film based method for public screening of diabetes mellitus
Takkar, B., Mukherjee, S., Chauhan, R. C., & Venkatesh, P. (2019) · Medical Hypotheses · doi:10.1016/j.mehy.2019.02.043 · Cited by 2
“The tear film has lower glucose concentrations than blood [14,15], and therefore the ability of the test to function at low glucose concentrations is a concern with regards to our hypothesis.”
BackgroundPhysical Sensors: Holographic Sensors
Lowe, C. R., & Moghaddam, G. K. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-822548-6.10037-8 · Cited by 2
Integrating voltammetry and substrate-enhanced luminescence for noninvasive glucose sensing
López-Peña, G., Vilar, A. P., Jaafar, A., Simón‐Fuente, S., Arranz, A., Ribagorda, M., Manso‐Silván, M., & Rodríguez, E. M. (2025) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2025.100591 · Cited by 1
Recent advances in smart contact lenses
Adibag, Z., Ghanbarzadeh, M., Salati, M. A., Rad, M. E., Ansari, R., Ta, C. N., Fadel, D., Mofidfar, M., & Abbasi, F. (2025) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2025.100734 · Cited by 1
Zhang, A., Jia, W., Wei, W., Qu, Z., Su, M., Bu, Y., Du, B., & Xiong, L. (2026) · Microchemical Journal · doi:10.1016/j.microc.2026.117235 · Cited by 1
Saraf, S. (2016) · Journal of International Crisis and Risk Communication Research · Cited by 2
Prashar, A. (2019) · Springer eBooks · doi:10.1007/978-981-13-7169-1_6 · Cited by 1
Turja, M. S., Kwon, T.-H., Kim, H., & Kim, K. (2024) · International Conference on Consumer Electronics · doi:10.1109/icce59016.2024.10444367 · Cited by 1
“Novel noninvasive methodologies employ various body tissues and fluids, such as skin [7], [8], [10], saliva [12], and tears [13].”
Development of paper-gate transistor toward direct detection from microbiological fluids
Kajisa, T., & Sakata, T. (2017) · Japanese Journal of Applied Physics · doi:10.7567/jjap.56.04cm07 · Cited by 1
Zhan, D., Chen, L., Tse, W. H., Chen, L., & Zhang, J. (2017) · Proceedings of the International Conference of Theoretical and Applied Nanoscience and Nanotechnology · doi:10.11159/tann17.122 · Cited by 1
“It has been found that tear glucose level is around 5 times higher in diabetes patients than in normal people and the tear glucose has been identified to have a positive correlation with the blood glucose based on previous research[13-15].”
ResultCrying on the Spot: Collection and Storage
Prashar, A. (2019) · Springer eBooks · doi:10.1007/978-981-13-7169-1_2 · Cited by 1
Biochemical studies of tears in patients with diabetes mellitus and dry eye syndrome
Махмутовна, Н. А., & Хамитова, Г. Х. (2013) · Kazan medical journal · doi:10.17816/kmj1931 · Cited by 1
Determination of loss tangent of human tear film at 9.8 GHz
Bansal, N., Dhaliwal, A. S., & Mann, K. S. (2015) · AIP conference proceedings · doi:10.1063/1.4929297
Building An Electrochemical Contact Lens Biosensor
Jordan, C. T. (2016) · CommonKnowledge Research Repository (Pacific University Oregon) · Cited by 1
Simple Methods of Detecting Disease Factors in Tears and Bloods
Jeong, S., Kim, K., & Noh, H. (2016) · Journal of Life Sciences and Technologies · doi:10.18178/jolst.4.1.26-29
“[13][16] For instance, wireless contact-lens for tear glucose”
BackgroundMicro-channel Embedded Soft Contact Lens for Tear Biomolecule Sampling
Kim, O., Park, H., Song, Y.-W., & Kim, J. (2016) · Journal of Biomedical Engineering Research · doi:10.9718/jber.2016.37.5.178
“그 중에서 눈물과의 연관성이 검증된 전신질환으로는 유방암과 다발성 경화증, 전립선 암, 그리고 당뇨병이 있다[1-4].”
InfluentialNon-invasive Blood Glucose Estimation Methodology Using Predictive Glucose Homeostasis Models
Panwar, M., Acharyya, A., & Shafik, R. (2018) · International Symposium on Embedded Computing and System Design (ISED) · doi:10.1109/ised.2018.8703991
Fink, J. K. (2021) · Wiley eBooks · doi:10.1002/9781119851547.ch6
Fink, J. K. (2022) · Wiley eBooks · doi:10.1002/9781119858072.ch1
Valero, M., Pola, P., Falaiye, O., Zhao, L., Shahriar, H., & Ahamed, S. I. (2022) · doi:10.2196/preprints.38664
Cunningham, D. D., & Stenken, J. A. (2009) · Wiley eBooks · doi:10.1002/9780470567319.fmatter
Prashar, A. (2019) · Springer eBooks · doi:10.1007/978-981-13-7169-1_1
Polyvinyl alcohol/starch modified cotton thread for glucose distance-based colorimetric detection
Punnoy, P. (2019) · Chula Digiverse · doi:10.58837/chula.the.2019.394
The Clinical Accuracy of Non-Invasive Glucose Monitoring for ex vivo Artificial Pancreas
Levy, M. M., & Levy, Y. M. (2016) · Journal of Emerging Investigators · doi:10.59720/16-008
“(2014) found that the glucose level in tears is somewhere between 1 to 10 mg/dL for a non-diabetic and 2 to 35 mg/dL for a diabetic patient (17).”
BackgroundElectrochemical analysis of biological fluids
Brown, M. S., Kelly, M., & Koh, A. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-99147-6.00003-x
López-Peña, G., Vilar, A. P., Jaafar, A., Simón‐Fuente, S., Arranz, A., Ribagorda, M., Manso‐Silván, M., & Rodríguez, E. M. (2024) · SSRN Electronic Journal · doi:10.2139/ssrn.4861112
Non‐Invasive Blood Sugar Detection
Kumar, A., & Gupta, V. (2025) · Wiley eBooks · doi:10.1002/9781394302895.ch18
Future applications of contact lenses
Phan, C., Walsh, K., & Jones, L. (2025) · Elsevier eBooks · doi:10.1016/b978-0-323-87929-3.00034-7
Non-Fullerene Organic Semiconductor ITIC as a Redox Mediator in Electrochemical Glucose Biosensors
Papi, M. A. P., Scheidweiler, V. G., Cassemiro, S. M., Akcelrud, L., Bergamini, M. F., & Junior, L. H. M. (2025) · Sensors · doi:10.3390/s25247535
“These features indicate that chitosan not only stabilizes the biolayer but also increases surface roughness, favoring enzyme retention [15].”
Park, S., Nam, D. Y., 전희재, Han, J. H., Jang, D., Hwang, J., Park, Y.-S., Han, Y.-G., Choy, Y. B., & Lee, D. Y. (2023) · Figshare · doi:10.6084/m9.figshare.24864169
Bokhary, K. A., Alzaid, S., Alqahtani, R. A., & Jamous, H. K. (2026) · Research Square · doi:10.21203/rs.3.rs-8525049/v1
Functionalized nanothin films for noninvasive biosensing
Jose, J., Muhammed Ali, V. P., Uthara, K., Mk, S., Sandeep, B. S., Anagha, C., Vijay, A., Radhakrishnan, E. K., & Pereira, M. L. G. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-18929-6.00018-9
Migliavacca, M., Zhang, X., Li, J., He, S., Zhang, M., & Pompa, P. P. (2026) · Journal of Nanobiotechnology · doi:10.1186/s12951-026-04374-6 · Cited by 1
“Over the past few decades, research works indicate that tear glucose (TG) mirrors blood glucose (BG) dynamics but at lower concentrations and with matrix-specific confounders (e.g., reflex tearing, sampling artifacts) [66].”
Machine Learning-Based Information Processing for Smart Contact Lens Tear Glucose Sensing
Li, Z. (2026) · World Conference on Computing and Communication Technologies (WCCCT) · doi:10.1109/wccct69960.2026.11549629
“The tested glucose range covered both typical (0.16 mM to 0.5 mM) [18] and high level (0.9 mM).”
InfluentialJi, J., Wan, H., An, S., Duan, J., Tang, X., & Zeng, D. (2026) · Electrochimica Acta · doi:10.1016/j.electacta.2026.149700
Smart contact lens for personalized glucose level sensing and forecasting
Paek, J., Park, W., Seo, H., Song, H., Kim, D., Chung, W. G., An, S., Hong, Y., Kim, J., Lee, Y.-H., et al. (2026) · Device · doi:10.1016/j.device.2026.101330
Metadata
- Authors
- Baca, J. T., Finegold, D. N., & Asher, S. A.
- Affiliations
- University of Pittsburgh (US)
- Corresponding author
- Asher, Sanford A.
- Venue
- The Ocular Surface
- Publisher
- Elsevier BV
- Published
- 2007-10
- Volume
- 5
- Issue
- 4
- Pages
- 280-293
- ISSN
- 1542-0124 (ISSN-L)
- OpenAlex
- W2062561734
- Semantic Scholar
- f72af5dfc6f5b8d76792108fc0e9d5ed522c11a5
- Cited by
- 187
- References
- 101
- Citations by year
- 2026: 10 · 2025: 13 · 2024: 13 · 2023: 20 · 2022: 14 · 2021: 10 · 2020: 10 · 2019: 14 · 2018: 21 · 2017: 10 · 2016: 8 · 2015: 7 · 2014: 10 · 2013: 6 · 2012: 3
- FWCI
- 3.06
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Ocular Surface (publishedVersion)
- PubMed (publishedVersion)
- D-Scholarship@Pitt (University of Pittsburgh) (acceptedVersion)
- Funders
- National Institutes of Health (1f31eb004181-01a1)
- National Institute of Biomedical Imaging and Bioengineering (1 F31 EB004181-01A1, R01EB004132)
- Topics
- Health Sciences › Medicine › Ophthalmology › Ocular Surface and Contact Lens
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Keywords
- tear glucose, diabetes mellitus, non-invasive glucose monitoring
- MeSH terms
- Animals, Aqueous Humor, Diabetes Mellitus, Glucose, Humans, Monitoring, Physiologic, Tears
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.