Evaluation of Commercial Glucometer Test Strips for Potential Measurement of Glucose in Tears
The DOI matches its Crossref record; 36 of the 37 references it cites were found in a registry or an index.
Abstract
Tear glucose measurements have been suggested as a potential alternative to blood glucose monitoring for diabetic patients. While previous work has reported that there is a correlation between blood and tear glucose levels in humans, this link has not been thoroughly established and additional clinical studies are needed. Herein, we evaluate the potential of using commercial blood glucose test strips to measure glucose in tears. Of several blood glucose strips evaluated, only one brand exhibits the low detection limit required for quantitating glucose in tears. Calibration of these strips in the range of 0-100 μM glucose with an applied potential of 150 mV to the working electrode yields a sensitivity of 0.127 nA/μM and a limit of quantitation (LOQ) of 9 μM. The strips also exhibit ≤13% error (n = 3) for 25, 50, and 75 μM glucose in the presence of 10 μM acetaminophen, 100 μM ascorbic acid, and 100 μM uric acid. Measurements of glucose in tears from nine normal (nondiabetic) fasting human subjects using strips yielded glucose values within the range of 5-148 μM (mean = 47 μM, median = 43 μM), similar to those for human tears reported by others with more complex LC-MS methods. The glucometer strip method could facilitate more clinical studies to determine whether tear glucose and blood glucose levels sufficiently correlate for application to routine measurements in tears to supplement blood glucose testing. This would be especially helpful for children, adolescents, other Type 1 diabetics, and also for Type 2 diabetics who require treatment with insulin and cannot tolerate multiple finger sticks per day.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The biosensor has a detection limit of 0.18 mg/dL and shows good linearity over the typical range of glucose concentrations in a tear film (0.18-10.5 mg/dL).”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [165]
In the citing paper
“To be analytically useful, sensing techniques require a very low detection limit (since glucose is present in tear fluid at levels 50-100 times lower than blood), high analytical sensitivity and selectivity, and the ability to quantitatively measure a very small sample in a short period of time (Figure 4E) 165.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [165]
In the citing paper
“Yao et al. have integrated functional contact lenses consisting of a differential glucose biosensor module, metal interconnections, a readout circuit, an antenna, and a telecommunication circuit to monitor tear glucose levels wirelessly, continuously and non-invasively 165.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [165]
In the citing paper
“Studies designed to establish correlations between blood and tear glucose concentrations are inconclusive from both human (243-245) and animal studies (246).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [245]
Citations
Cites
Danaei, G., Finucane, M. M., Lu, Y., Singh, G. M., Cowan, M. J., Paciorek, C. J., Lin, J. K., Farzadfar, F., Khang, Y.-H., Stevens, G. A., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(11)60679-x · Cited by 3,568
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Raman spectroscopy for noninvasive glucose measurements
Enejder, A. M. K., Scecina, T. G., Oh, J., Hunter, M., Shih, W.-C., Sasic, S., Horowitz, G. L., & Feld, M. S. (2005) · Journal of Biomedical Optics · doi:10.1117/1.1920212 · Cited by 271
Current development in non-invasive glucose monitoring
Ferrante do Amaral, C. E., & Wolf, B. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.06.003 · Cited by 269
Glucose monitoring by reverse iontophoresis
Potts, R. O., A. Tamada, J., & J. Tierney, M. (2002) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.210 · Cited by 243
Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus
Baca, J. T., Finegold, D. N., & Asher, S. A. (2007) · The Ocular Surface · doi:10.1016/s1542-0124(12)70094-0 · Cited by 187
Measurement of tear glucose levels with amperometric glucose biosensor/capillary tube configuration
Yan, Q., Peng, B., Su, G., Cohan, B. E., Major, T. C., & Meyerhoff, M. E. (2011) · Analytical Chemistry · doi:10.1021/ac201700c · Cited by 185
Self-citationNovel FAD-Dependent Glucose Dehydrogenase for a Dioxygen-Insensitive Glucose Biosensor
TSUJIMURA, S., KOJIMA, S., KANO, K., IKEDA, T., SATO, M., SANADA, H., & OMURA, H. (2006) · Bioscience, Biotechnology, and Biochemistry · doi:10.1271/bbb.70.654 · Cited by 175
Tear Glucose Dynamics in Diabetes Mellitus
Lane, J. D., Krumholz, D. M., Sack, R. A., & Morris, C. (2006) · Current Eye Research · doi:10.1080/02713680600976552 · Cited by 161
The Technology Behind Glucose Meters: Test Strips
Hönes, J., Müller, P., & Surridge, N. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0005 · Cited by 156
Research works printed but not linked: 2
- Choy C. K. M., 2000, Invest. Ophthalmol. Vis. Sci., 41, 3293
- Kalisz H. M, 2000, Biochem. J., 349, 553
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Electrochemical Glucose Sensing: Is There Still Room for Improvement?
Witkowska Nery, E., Kundys, M., Jeleń, P. S., & Jönsson-Niedziółka, M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03151 · Cited by 292
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210
Gabriel, E. F. M., Garcia, P. T., Cardoso, T. M. G., Lopes, F. M., Martins, F. T., & Coltro, W. K. T. (2016) · The Analyst · doi:10.1039/c6an00430j · Cited by 204
Wen, D., Liu, W., Haubold, D., Zhu, C., Oschatz, M., Holzschuh, M., Wolf, A., Simon, F., Kaskel, S., & Eychmüller, A. (2016) · ACS Nano · doi:10.1021/acsnano.5b07505 · Cited by 207
Metadata
- Authors
- Cha, K. H., Jensen, G. C., Balijepalli, A. S., Cohan, B. E., & Meyerhoff, M. E.
- Affiliations
- University of Michigan (US)
- EyeLab Group LLC, 2350 Washtenaw Avenue No. 4, Ann Arbor, Michigan, 48104-4525, United States
- Corresponding author
- Meyerhoff, Mark E.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2014-01-24
- Volume
- 86
- Issue
- 3
- Pages
- 1902-1908
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2094203185
- Semantic Scholar
- fab7c610aa10576f9c465e749ce19ca2dc3d222f
- Cited by
- 67
- References
- 37
- Citations by year
- 2026: 5 · 2025: 1 · 2024: 6 · 2023: 6 · 2022: 5 · 2021: 8 · 2020: 7 · 2019: 5 · 2018: 4 · 2017: 2 · 2016: 10 · 2015: 6 · 2014: 2
- FWCI
- 3.54
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- tear glucose, blood glucose monitoring, diabetes mellitus, non-invasive glucose measurement, limit of quantification, acetaminophen interference, ascorbic acid interference, uric acid interference
- MeSH terms
- Electrochemistry, Glucose, Humans, Reagent Strips, Tears, Blood Glucose Self-Monitoring, Limit of Detection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of Commercial Glucometer Test Strips for Potential Measurement of Glucose in Tears. https://app.cochat.ai/papers/10.1021/ac4040168
- The paper
- Cha, K. H., Jensen, G. C., Balijepalli, A. S., Cohan, B. E., & Meyerhoff, M. E. (2014). Evaluation of Commercial Glucometer Test Strips for Potential Measurement of Glucose in Tears. Analytical Chemistry. https://doi.org/10.1021/ac4040168
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.