Clinical extracorporeal hybrid liver support – phase I study with primary porcine liver cells
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
The objective of this study was to evaluate the feasibility and safety of a hybrid liver support system with extracorporeal plasma separation and bioreactor perfusion in patients with acute liver failure (ALF) who had already fulfilled the criteria for high urgency liver transplantation (LTx). Eight patients (one male, seven female) were treated in terms of bridging to transplantation. The mean age was 36.5 yr (range 20 to 58). Etiology of liver failure was drug‐related in two patients, hepatitis B infection in three patients, and unknown for three patients. The bioreactors were charged with primary liver cells from specific pathogen‐free pigs. Cell viability varied between 91 and 98%. Continuous liver support treatment over a period of 8 to 46 h (mean 27.3 h) was safely performed and well‐tolerated by all patients. No complications associated with the therapy were observed during the follow‐up period. Thrombocytopenia was considered to be an effect of the plasma separation. Subsequently, all patients were transplanted successfully and were observed over at least 3 yr with an organ and patient survival rate of 100%. Screening of patient's sera for antibodies specific for porcine endogenous retroviruses (PERVs) showed no reactivity – either prior to application of the system, or after extracorporeal treatment. The results encourage us to continue the development of the technology, and further studies appear to be justified. The bioreactor technology has been integrated into a modular extracorporeal liver support (MELS) system, combining biologic liver support with artificial detoxification technology.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Infection of human cells by an endogenous retrovirus of pigs
Patience, C., Takeuchi, Y., & Weiss, R. A. (1997) · Nature Medicine · doi:10.1038/nm0397-282 · Cited by 953 · cited 1×
“The reported identification of PERV that may infect human cells in vitro [16, 17] and may be pathogenic for the transplant recipient [18] had implications for the use of xenogeneic cells and tissue in general.”
BackgroundPrinciples of Tissue Engineering
Pegg, D. E. (1999) · Cryobiology · doi:10.1006/cryo.1999.2214 · Cited by 564
Watanabe, F. D., Mullon, C. J.-P., Hewitt, W. R., Arkadopoulos, N., Kahaku, E., Eguchi, S., Khalili, T., Arnaout, W., Shackleton, C. R., Rozga, J., et al. (1997) · Annals of Surgery · doi:10.1097/00000658-199705000-00005 · Cited by 453 · cited 2×
“Sussman’s Extracorporeal Liver Assist Device (ELAD) [10] and Demetriou’s Bioartificial Liver (HepatAssistTM, Circe Biomedical, Lexington, MA, USA) [11, 12] use modified concepts of hemodialysis cartridges hosting the liver cells.”
BackgroundThe Hepatix Extracorporeal Liver Assist Device: Initial Clinical Experience
Sussman, N. L., Gislason, G. T., Conlin, C. A., & Kelly, J. H. (1994) · Artificial Organs · doi:10.1111/j.1525-1594.1994.tb02221.x · Cited by 277 · cited 3×
“M"ller1, TH. Steinm"ller1, J. Denner2, P. Neuhaus1 and J. C. Gerlach1,3 1General, Visceral and Transplantation Surgery, Charit – Campus Virchow Klinikum, HumboldtUniversity Berlin, Berlin, Germany, 2Robert Koch Institute, Berlin, Germany, 3Department of Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA,…”
BackgroundBIOREACTOR FOR A LARGER SCALE HEPATOCYTE IN VITRO PERFUSION
Gerlach, J. C., Encke, J., Hole, O., Müller, C., Ryan, C. J., & Neuhaus, P. (1994) · Transplantation · doi:10.1097/00007890-199411150-00002 · Cited by 184 · cited 1×
“These hollow fibers represent the artificial equivalent of the liver lobuli’s vascular system [3].”
BackgroundSelf-citationProductive Infection of Human Primary Cells and Cell Lines with Porcine Endogenous Retroviruses
Specke, V., Rubant, S., & Denner, J. (2001) · Virology · doi:10.1006/viro.2001.0934 · Cited by 150 · cited 1×
“The reported identification of PERV that may infect human cells in vitro [16, 17] and may be pathogenic for the transplant recipient [18] had implications for the use of xenogeneic cells and tissue in general.”
BackgroundSelf-citationGerlach, J. C., Schnoy, N., Encke, J., Smith, M. D., Müller, C., & Neuhaus, P. (1995) · Hepatology · doi:10.1002/hep.1840220226 · Cited by 106 · cited 1×
“The liver cells (400 to 600 g of hepatocytes in coculture with non-parenchymal cells) aggregate and immobilize on the outer surface of and between the capillaries and spontaneously form new liver sinusoid-like structures [4].”
BackgroundSelf-citationImmunosuppression by Retroviruses: Implications for Xenotransplantation
Denner, J. (1998) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1998.tb09119.x · Cited by 97 · cited 1×
“The reported identification of PERV that may infect human cells in vitro [16, 17] and may be pathogenic for the transplant recipient [18] had implications for the use of xenogeneic cells and tissue in general.”
BackgroundSelf-citationWhich hepatocyte will it be? Hepatocyte choice for bioartificial liver support systems
Tsiaoussis, J., Newsome, P. N., Nelson, L. J., Hayes, P. C., & Plevris, J. N. (2001) · Liver Transplantation · doi:10.1053/jlts.2001.20845 · Cited by 96 · cited 1×
“support, which would correspond to 150 to 300 g of vital liver cells for a 70-kg patient [7, 8].”
BackgroundTacke, S. J., Bodusch, K., Berg, A., & Denner, J. (2001) · Xenotransplantation · doi:10.1034/j.1399-3089.2001.00080.x-i1 · Cited by 92
Self-citation
Cited by
Acute liver failure: Summary of a workshop†
Lee, W. M., Squires, R. H., Nyberg, S. L., Doo, E., & Hoofnagle, J. H. (2008) · Hepatology · doi:10.1002/hep.22177 · Cited by 757
Lee, W. M., Stravitz, T. R., & Larson, A. M. (2012) · Hepatology · doi:10.1002/hep.25551 · Cited by 550
Artificial and bioartificial liver devices: present and future
Carpentier, B., Gautier, A., & Legallais, C. (2009) · Gut · doi:10.1136/gut.2008.175380 · Cited by 213
Infection Barriers to Successful Xenotransplantation Focusing on Porcine Endogenous Retroviruses
Denner, J., & Tönjes, R. R. (2012) · Clinical Microbiology Reviews · doi:10.1128/cmr.05011-11 · Cited by 208
Self-citationThe Billion Cell Construct: Will Three-Dimensional Printing Get Us There?
Miller, J. S. (2014) · PLoS Biology · doi:10.1371/journal.pbio.1001882 · Cited by 199
“In extracorporeal liver-assist devices, live cells can be loaded into bioreactor chambers outside the body and then connected in a closed loop with host blood circulation so that the biochemical benefit from cells in the device will positively affect the patient [4,5].”
BackgroundNadim, M. K., Kellum, J. A., Davenport, A., Wong, F., Davis, C., Pannu, N., Tolwani, A., Bellomo, R., & Genyk, Y. S. (2012) · Critical Care · doi:10.1186/cc11188 · Cited by 195
“Rationale: Extracorporeal support systems in liver disease can be divided into two broad categories: non-cell [96-131] and cell-based systems (Table 7) [132-134].”
BackgroundLiver support strategies: cutting-edge technologies
Struecker, B., Raschzok, N., & Sauer, I. M. (2014) · Nature Reviews Gastroenterology & Hepatology · doi:10.1038/nrgastro.2013.204 · Cited by 165
Self-citation3D organotypic HepaRG cultures as in vitro model for acute and repeated dose toxicity studies
Mueller, D., Krämer, L., Hoffmann, E., Klein, S., & Noor, F. (2014) · Toxicology in Vitro · doi:10.1016/j.tiv.2013.06.024 · Cited by 159
Tsuruga, Y., Kiyono, T., Matsushita, M., Takahashi, T., Kasai, H., & Todo, S. (2008) · Cell Transplantation · doi:10.3727/096368908786991542 · Cited by 66
Diekmann, S., Bader, A., & Schmitmeier, S. (2006) · Cytotechnology · doi:10.1007/s10616-006-6336-4 · Cited by 32
Metadata
- Authors
- Sauer, I. M., Kardassis, D., Zeillinger, K., Pascher, A., Gruenwald, A., Pless, G., Irgang, M., Kraemer, M., Puhl, G., Frank, J., Müller, A. R., Steinmüller, T., Denner, J., Neuhaus, P., & Gerlach, J. C.
- Venue
- Xenotransplantation
- Publisher
- Wiley
- Published
- 2003-08-27
- Volume
- 10
- Issue
- 5
- Pages
- 460-469
- ISSN
- 0908-665X (ISSN-L)
- 1399-3089
- OpenAlex
- W2081437656
- Cited by
- 181
- References
- 16
- FWCI
- 11.60
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.