Automated blood pressure measuring devices: how are they clinically validated for accuracy?
The DOI matches its Crossref record; 29 of the 30 references it cites were found in a registry or an index.
Abstract
Accurate blood pressure (BP) assessment is essential for the optimal diagnosis and management of hypertension. Contemporary clinical practice guidelines strongly endorse use of automated cuff blood pressure measuring devices (BPMD) as the preferred means of measuring and monitoring BP in the office, at home and with ambulatory blood pressure monitoring. To ensure that they are accurate, automated BPMDs should undergo clinical validation testing, performed using an established clinical validation standard. Unfortunately, most BPMDs sold on the global market have not been clinically validated. Furthermore, in the last thirty years, several different clinical validation protocols have been published, with major differences apparent between these standards, causing controversy with respect to which standard is considered acceptable for clinical validation. Complexly worded standards, multiple revisions, and firewalled access also contribute to a lack of understanding and use of clinical validation standards and the number of expert centers performing clinical validations is small. Recently, joint society collaborations have led to creation of the AAMI/ESH/ISO universal standard for the clinical validation of automated cuff BPMDs. Not only is this unified standard a necessary step, but oversight from regulators and influential stakeholders to ensure that only clinically validated BPMDs can be marketed is additionally needed.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Automated BPMDs that have passed accuracy testing versus a reference auscultatory sphygmomanometer using a scientifically accepted validation protocol are recommended for clinical use globally.”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [14]
In the citing paper
“Since each unique automated BPMD can have different processes, algorithms and fitting functions to derive BPs (none of which are publicly disclosed), the accuracy of each BPMD needs to be individually determined by comparison to a BP reference standard using a scientifically accepted validation protocol [14, 57, 58].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [14]
Citations
Cites
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Whelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Hypertension · doi:10.1161/hyp.0000000000000065 · Cited by 7,391
Stanaway, J. D., Afshin, A., Gakidou, E., Lim, S. S., Abate, D., Abate, K. H., Abbafati, C., Abbasi, N., Abbastabar, H., Abd-Allah, F., et al. (2018) · The Lancet · doi:10.1016/s0140-6736(18)32225-6 · Cited by 4,943
2020 International Society of Hypertension Global Hypertension Practice Guidelines
Unger, T., Borghi, C., Charchar, F., Khan, N. A., Poulter, N. R., Prabhakaran, D., Ramirez, A., Schlaich, M., Stergiou, G. S., Tomaszewski, M., et al. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.15026 · Cited by 4,132
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association
Muntner, P., Shimbo, D., Carey, R. M., Charleston, J. B., Gaillard, T., Misra, S., Myers, M. G., Ogedegbe, G., Schwartz, J. E., Townsend, R. R., et al. (2019) · Hypertension · doi:10.1161/hyp.0000000000000087 · Cited by 1,512
2020 International Society of Hypertension global hypertension practice guidelines
Unger, T., Borghi, C., Charchar, F., Khan, N. A., Poulter, N. R., Prabhakaran, D., Ramirez, A., Schlaich, M., Stergiou, G. S., Tomaszewski, M., et al. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002453 · Cited by 1,283
Stergiou, G. S., Palatini, P., Parati, G., O’Brien, E., Januszewicz, A., Lurbe, E., Persu, A., Mancia, G., Kreutz, R., & on behalf of the European Society of Hypertension Council and the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002843 · Cited by 859
O??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772
OʼBrien, E., Pickering, T., Asmar, R., Myers, M., Parati, G., Staessen, J., Mengden, T., Imai, Y., Waeber, B., Palatini, P., et al. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200202000-00002 · Cited by 744
Rabi, D. M., McBrien, K. A., Sapir-Pichhadze, R., Nakhla, M., Ahmed, S. B., Dumanski, S. M., Butalia, S., Leung, A. A., Harris, K. C., Cloutier, L., et al. (2020) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2020.02.086 · Cited by 583 · cited 1×
“National and international organizations have called for clinically validated BPMDs to be used in clinical practice and for home monitoring [8, 10, 11, 31, 32].”
Background
Research works printed but not linked: 1
- Kaplan NM, Commentary on the sixth report of the Joint National Committee (JNC-6). Am J Hypertens. 1998;11:134–6.
Cited by
Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00693-x · Cited by 84
Self-citationSharman, J. E., Ordunez, P., Brady, T., Parati, G., Stergiou, G., Whelton, P. K., Padwal, R., Olsen, M. H., Delles, C., Schutte, A. E., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00747-0 · Cited by 31
Self-citationHow to find and use validated blood pressure measuring devices
Picone, D. S., Padwal, R. S., Stergiou, G. S., Cohen, J. B., McManus, R. J., Eckert, S., Asayama, K., Atkins, N., Rakotz, M., Lombardi, C., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00718-5 · Cited by 23
Self-citationLombardi, C., Picone, D. S., Sharman, J. E., Campbell, N. R. C., Farias, R. F., Guerre, S., Gittens, A., Paccot, M., Villacres, N., González, Y. V., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00706-9 · Cited by 12
Lee, J., Kim, Y. M., Yoo, S.-W., Kim, D.-Y., Shin, S., Na, J. O., Shin, J., Park, K., & Lee, E. M. (2025) · Hypertension Research · doi:10.1038/s41440-025-02364-z · Cited by 3
Metadata
- Authors
- Ringrose, J. S., & Padwal, R.
- Affiliations
- University of Alberta (CA)
- Corresponding author
- Ringrose, Jennifer S.
- Venue
- Journal of Human Hypertension
- Published
- 2022-10-11
- Volume
- 37
- Issue
- 2
- Pages
- 101-107
- ISSN
- 0950-9240 (ISSN-L)
- 1476-5527
- OpenAlex
- W4304190118
- Semantic Scholar
- 3c2639262a64d90b38d12b3f3ee096ba5777c58e
- Cited by
- 9
- References
- 30
- Citations by year
- 2025: 3 · 2024: 1 · 2023: 1 · 2022: 4
- FWCI
- 0.46
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springernature.com/gp/researchers/text-and-data-mining (tdm)
- www.springernature.com/gp/researchers/text-and-data-mining (vor)
- Locations
- Journal of Human Hypertension (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Clinical practice guidelines implementation
- Keywords
- clinical validation, validation standards, hypertension diagnosis, ambulatory blood pressure monitoring, home blood pressure monitoring, regulatory oversight, measurement accuracy
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Societies, Medical, Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Automated blood pressure measuring devices: how are they clinically validated for accuracy? https://app.cochat.ai/papers/10.1038/s41371-022-00761-2
- The paper
- Ringrose, J. S., & Padwal, R. (2022). Automated blood pressure measuring devices: how are they clinically validated for accuracy? Journal of Human Hypertension. https://doi.org/10.1038/s41371-022-00761-2
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.