Mechano-acoustic sensing of physiological processes and body motions via a soft wireless device placed at the suprasternal notch
Open accessAbstract
Skin-mounted soft electronics that incorporate high-bandwidth triaxial accelerometers can capture broad classes of physiologically relevant information, including mechano-acoustic signatures of underlying body processes (such as those measured by a stethoscope) and precision kinematics of core-body motions. Here, we describe a wireless device designed to be conformally placed on the suprasternal notch for the continuous measurement of mechano-acoustic signals, from subtle vibrations of the skin at accelerations of around 10-3 m s-2 to large motions of the entire body at about 10 m s-2, and at frequencies up to around 800 Hz. Because the measurements are a complex superposition of signals that arise from locomotion, body orientation, swallowing, respiration, cardiac activity, vocal-fold vibrations and other sources, we exploited frequency-domain analysis and machine learning to obtain-from human subjects during natural daily activities and exercise-real-time recordings of heart rate, respiration rate, energy intensity and other essential vital signs, as well as talking time and cadence, swallow counts and patterns, and other unconventional biomarkers. We also used the device in sleep laboratories and validated the measurements using polysomnography.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
IMUs with high sampling frequency characteristics (3.3 or 6.6 kHz) can capture high spectral features to measure acoustic signals in addition to mechanical signals induced by cardiac activities.
“However, IMUs with high sampling frequency characteristics (3.3 or 6.6 kHz) can capture high spectral features to measure acoustic signals in addition to mechanical signals induced by cardiac activities (26, 27).”
From Miniaturized wireless, skin-integrated sensor networks for quantifying full-body movement behaviors and vital signs in infants (2021) · cites this paper as [26]
The Abstract describes a high-bandwidth accelerometer that captures mechano-acoustic signals, and the Results section on suprasternal-notch measurements reports both cardiac pulses at 20–50 Hz and acoustic components of swallowing at 100–800 Hz. Our copy does not establish the claimed 3.3 or 6.6 kHz sampling frequencies; searches for those values and for sampling rates found no supporting passage. (LLM, conf=0.86)
Experimental — accuracy not yet measuredPrevious reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms.
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [23]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Ror2 signaling regulates Golgi structure and transport through IFT20 for tumor invasiveness
Nishita, M., Park, S.-Y., Nishio, T., Kamizaki, K., Wang, Z., Tamada, K., Takumi, T., Hashimoto, R., Otani, H., Pazour, G. J., et al. (2017) · Scientific Reports · doi:10.1038/s41598-016-0028-x · Cited by 15,947 · see citing works
A Mathematical Introduction to Robotic Manipulation
Murray, R. M., Li, Z., & Sastry, S. (2017) · CRC Press eBooks · doi:10.1201/9781315136370 · Cited by 6,698 · see citing works
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836 · see citing works
Self-citationAASM Scoring Manual Updates for 2017 (Version 2.4)
Berry, R. B., Brooks, R., Gamaldo, C. E., Harding, S. M., Lloyd, R. M., Quan, S. F., Troester, M. M., & Vaughn, B. V. (2017) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.6576 · Cited by 1,605 · see citing works
Clinical measurement of speech and voice
Baken, R. J. (1987) · Cited by 1,454 · see citing works
Fractal design concepts for stretchable electronics
Fan, J. A., Yeo, W.-H., Su, Y., Hattori, Y., Lee, W., Jung, S.-Y., Zhang, Y., Liu, Z., Cheng, H., Falgout, L., et al. (2014) · Nature Communications · doi:10.1038/ncomms4266 · Cited by 1,088 · see citing works
Self-citationA Review of Accelerometry-Based Wearable Motion Detectors for Physical Activity Monitoring
Yang, C.-C., & Hsu, Y.-L. (2010) · Sensors · doi:10.3390/s100807772 · Cited by 1,058 · see citing works
Kim, D., Song, J., Choi, W. M., Kim, H., Kim, R.-H., Liu, Z., Huang, Y. Y., Hwang, K.-C., Zhang, Y., & Rogers, J. A. (2008) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0807476105 · Cited by 740 · see citing works
Self-citationUnobtrusive Sensing and Wearable Devices for Health Informatics
Zheng, Y.-L., Ding, X., Poon, C. C. Y., Lo, B. P. L., Zhang, H., Zhou, X.-L., Yang, G., Zhao, N., & Zhang, Y. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2309951 · Cited by 725 · see citing works
Soft network composite materials with deterministic and bio-inspired designs
Jang, K., Chung, H. U., Xu, S. H., Lee, C. H., Luan, H., Jeong, J., Cheng, H., Kim, G.-T., Han, S. Y., Lee, J., et al. (2015) · Nature Communications · doi:10.1038/ncomms7566 · Cited by 534 · see citing works
Self-citationLiu, Y., Norton, J. J. S., Qazi, R., Zou, Z., Ammann, K. R., Liu, H., Yan, L., Tran, P. L., Jang, K.-I., Lee, J. W., et al. (2016) · Science Advances · doi:10.1126/sciadv.1601185 · Cited by 397 · see citing works
Self-citationUsing Inertial Sensors for Position and Orientation Estimation
Kok, M., Hol, J. D., & Schön, T. B. (2017) · Foundations and Trends® in Signal Processing · doi:10.1561/2000000094 · Cited by 356 · see citing works
Using Inertial Sensors for Position and Orientation Estimation
Kok, M., Hol, J. D., & Schön, T. B. (2017) · now publishers, Inc. eBooks · doi:10.1561/9781680833577 · Cited by 293 · see citing works
Gender recognition from speech. Part II: Fine analysis
Childers, D. G., & Wu, K. (1991) · The Journal of the Acoustical Society of America · doi:10.1121/1.401664 · Cited by 244 · see citing works
Electronic sensor and actuator webs for large-area complex geometry cardiac mapping and therapy
Kim, D., Ghaffari, R., Lu, N., Wang, S., Lee, S. P., Keum, H., D’Angelo, R., Klinker, L. E., Su, Y., Lü, C., et al. (2012) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1205923109 · Cited by 240 · see citing works
Self-citationGyrophone: recognizing speech from gyroscope signals
Michalevsky, Y., Boneh, D., & Nakibly, G. (2014) · Cited by 222 · see citing works
Estimation of Respiratory Rate From Photoplethysmogram Data Using Time–Frequency Spectral Estimation
Chon, K. H., Dash, S., & Ju, K. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2019766 · Cited by 213 · see citing works
Inan, O. T., Baran Pouyan, M., Javaid, A. Q., Dowling, S., Etemadi, M., Dorier, A., Heller, J. A., Bicen, A. O., Roy, S., De Marco, T., et al. (2018) · Circulation: Heart Failure · doi:10.1161/circheartfailure.117.004313 · Cited by 207 · see citing works
Mobile Voice Health Monitoring Using a Wearable Accelerometer Sensor and a Smartphone Platform
Mehta, D. D., Zañartu, M., Feng, S. W., Cheyne, H. A., & Hillman, R. E. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2207896 · Cited by 203 · see citing works
Tadi, M. J., Lehtonen, E., Saraste, A., Tuominen, J., Koskinen, J., Teräs, M., Airaksinen, J. K. E., Pänkäälä, M., & Koivisto, T. (2017) · Scientific Reports · doi:10.1038/s41598-017-07248-y · Cited by 175 · see citing works
Respiratory Rate and Flow Waveform Estimation from Tri-axial Accelerometer Data
Bates, A., Ling, M. J., Mann, J., & Arvind, D. K. (2010) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2010.50 · Cited by 168 · see citing works
Rahman, T., Adams, A. T., Zhang, M., Cherry, E., Zhou, B., Peng, H., & Choudhury, T. (2014) · International Conference on Mobile Systems, Applications, and Services · doi:10.1145/2594368.2594386 · Cited by 159 · see citing works
Sampling frequency affects the processing of Actigraph raw acceleration data to activity counts
Brønd, J. C., & Arvidsson, D. (2015) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00628.2015 · Cited by 148 · see citing works
Upper airway closing pressures in snorers
Issa, F., & Sullivan, C. E. (1984) · Journal of Applied Physiology · doi:10.1152/jappl.1984.57.2.528 · Cited by 143 · see citing works
An Accelerometer-Based Device for Sleep Apnea Screening
Morillo, D., Ojeda, J. L. R., Foix, L. F. C., & Jiménez, A. L. (2009) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2027231 · Cited by 129 · see citing works
Digital stethoscope: technology update
Swarup, S., & Makaryus, A. N. (2018) · Medical Devices Evidence and Research · doi:10.2147/mder.s135882 · Cited by 117 · see citing works
Hemodynamic evaluation of the prone position by transesophageal echocardiography
Toyota, S., & Amaki, Y. (1998) · Journal of Clinical Anesthesia · doi:10.1016/s0952-8180(97)00216-x · Cited by 112 · see citing works
Estimation of Respiration Rate from Three-Dimensional Acceleration Data Based on Body Sensor Network
Liu, G.-Z., Guo, Y.-W., Zhu, Q.-S., Huang, B.-Y., & Wang, L. (2011) · Telemedicine and e-Health · doi:10.1089/tmj.2011.0022 · Cited by 107 · see citing works
Effects of body position on autonomic regulation of cardiovascular function in young, healthy adults
Watanabe, N., Reece, J., & Polus, B. (2007) · Chiropractic & Manual Therapies · doi:10.1186/1746-1340-15-19 · Cited by 106 · see citing works
Curone, D., Bertolotti, G. M., Cristiani, A., Secco, E. L., & Magenes, G. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2010.2050696 · Cited by 105 · see citing works
Effects of supine, prone, and lateral positions on cardiovascular and renal variables in humans
Pump, B., Talleruphuus, U., Christensen, N. J., Warberg, J., & Norsk, P. (2002) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00619.2001 · Cited by 104 · see citing works
Estimation of respiratory waveform using an accelerometer
Hung, P. D., Bonnet, S., Guillemaud, R., Castelli, E., & Yen, P. T. N. (2008) · IEEE International Symposium on Biomedical Imaging (ISBI) · doi:10.1109/isbi.2008.4541291 · Cited by 99 · see citing works
Javaid, A. Q., Ashouri, H., Dorier, A., Etemadi, M., Heller, J. A., Roy, S., & Inan, O. T. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2600945 · Cited by 88 · see citing works
Physiological Acoustic Sensing Based on Accelerometers: A Survey for Mobile Healthcare
Hu, Y., Kim, E. G., Cao, G., Liu, S., & Xu, Y. P. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1111-8 · Cited by 86 · see citing works
Estimation of respiratory waveform and heart rate using an accelerometer
Phan, D. H., Bonnet, S., Guillemaud, R., Castelli, E., & Thi, N. Y. P. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4650316 · Cited by 86 · see citing works
Dudik, J. M., Coyle, J. L., & Sejdić, E. (2015) · IEEE Transactions on Human-Machine Systems · doi:10.1109/thms.2015.2408615 · Cited by 85 · see citing works
Dash, S., Shelley, K. H., Silverman, D. G., & Chon, K. H. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2038226 · Cited by 85 · see citing works
Kumari, S. K., & Mathana, J. M. (2018) · Journal of Medical Systems · doi:10.1007/s10916-018-1115-2 · Cited by 81 · see citing works
Jin, A., Yin, B., Morren, G., Duric, H., & Aarts, R. M. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5333116 · Cited by 75 · see citing works
Shandhi, M. M. H., Semiz, B., Hersek, S., Goller, N., Ayazi, F., & Inan, O. T. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2895775 · Cited by 74 · see citing works
Time and time–frequency characterization of dual-axis swallowing accelerometry signals
Lee, J., Steele, C. M., & Chau, T. T. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/9/008 · Cited by 73 · see citing works
Self-citationA comparative analysis of swallowing accelerometry and sounds during saliva swallows
Dudik, J. M., Jestrović, I., Luan, B., Coyle, J. L., & Sejdić, E. (2015) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-14-3 · Cited by 72 · see citing works
Posatskiy, A. O., & Chau, T. T. (2011) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2011.09.004 · Cited by 68 · see citing works
A wearable system for the seismocardiogram assessment in daily life conditions
Rienzo, M. D., Meriggi, P., Rizzo, F., Vaini, E., Faini, A., Merati, G., Parati, G., & Castiglioni, P. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091058 · Cited by 61 · see citing works
SNORING: LINEAR STABILITY ANALYSIS ANDIN-VITROEXPERIMENTS
Aurégan, Y., & Dépollier, C. (1995) · Journal of Sound and Vibration · doi:10.1006/jsvi.1995.0577 · Cited by 59 · see citing works
An ear-worn continuous ballistocardiogram (BCG) sensor for cardiovascular monitoring
He, D. D., Winokur, E. S., & Sodini, C. G. (2012) · doi:10.1109/embc.2012.6347123 · Cited by 59 · see citing works
Cumulative Irritation in Older and Younger Skin: a Comparison
Schwindt, D. A. (1998) · Acta Dermato Venereologica · doi:10.1080/000155598441864 · Cited by 58 · see citing works
Garment-based detection of falls and activities of daily living using 3-axis MEMS accelerometer
Nyan, M. N., Tay, F. E. H., Manimaran, M., & Seah, K. H. W. (2006) · Journal of Physics Conference Series · doi:10.1088/1742-6596/34/1/175 · Cited by 56 · see citing works
Respiratory rate assessments using a dual-accelerometer device
Lapi, S., Lavorini, F., Borgioli, G., Calzolai, M., Masotti, L. F., Pistolesi, M., & Fontana, G. A. (2013) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2013.11.003 · Cited by 55 · see citing works
Fajdiga, I. (2005) · CHEST Journal · doi:10.1378/chest.128.2.896 · Cited by 45 · see citing works
Accelerometer based system for continuous respiratory rate monitoring
Preejith, S. P., Jeelani, A., Maniyar, P., Joseph, J., & Sivaprakasam, M. (2017) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2017.7985870 · Cited by 45 · see citing works
An Online Swallow Detection Algorithm Based on the Quadratic Variation of Dual-Axis Accelerometry
Damouras, S., Sejdic, E., Steele, C. M., & Chau, T. (2010) · IEEE Transactions on Signal Processing · doi:10.1109/tsp.2010.2043972 · Cited by 43 · see citing works
Wearable Motion-Based Heart Rate at Rest: A Workplace Evaluation
Hernandez, J., McDuff, D., Quigley, K. S., Maes, P., & Picard, R. W. (2018) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2018.2877484 · Cited by 22 · see citing works
A daily living activity remote monitoring system for solitary elderly people
Maki, H., Ogawa, H., Matsuoka, S., Yonezawa, Y., & Caldwell, W. M. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091357 · Cited by 21 · see citing works
Measuring Respiration and Heart Rate using Two Acceleration Sensors on a Fully Embedded Platform
Vertens, J., Fischer, F., Heyde, C., et al. (2015) · International Congress on Sport Sciences Research and Technology Support (icSPORTS) · doi:10.5220/0005604000150023 · Cited by 21 · see citing works
Biomedical Signals and Sensors II
Kaniušas, E. (2015) · Biological and medical physics, biomedical engineering · doi:10.1007/978-3-662-45106-9 · Cited by 17 · see citing works
A Pilot Study on BSN-Based Ubiquitous Energy Expenditure Monitoring
Lin, S., Wang, L., Huang, B., Zhang, Y., Wu, X., & Zhao, J. (2009) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.56 · Cited by 8 · see citing works
A Miniaturized MEMS Motion Processing System for Nuclear Medicine Imaging Applications
Tadi, M. J., Lehtonen, E., Teuho, J., et al. (2016) · Computing in cardiology · doi:10.22489/cinc.2016.042-452 · Cited by 8 · see citing works
Sensing by Acoustic Biosignals
Kaniušas, E. (2015) · Biological and medical physics, biomedical engineering · doi:10.1007/978-3-662-45106-9_4 · Cited by 6 · see citing works
Makarenkova, A., Poreva, A., & Slozko, M. (2017) · IEEE Ukraine Conference on Electrical and Computer Engineering (UKRCON) · doi:10.1109/ukrcon.2017.8100499 · Cited by 5 · see citing works
Application of Acceleration Sensors in Physiological Experiments
Vavrinský, E., Moskal'vá, D., Darříček, M., et al. (2014) · Journal of Electrical Engineering · doi:10.2478/jee-2014-0049 · Cited by 2 · see citing works
A MEMS accelerometers based system for the measurement of lung sound delays
Pompilio, P. P., Sgura, A., Pedotti, A., et al. (2010) · Cairo International Biomedical Engineering Conference · doi:10.1109/cibec.2010.5716042 · Cited by 1 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- [45]Titze IR & Martin DW Principles of Voice Production The Journal of the Acoustical Society of America 104, (Prentice Hall, 1998).
Cited by
Sorted by citations
Continuous monitoring of deep-tissue haemodynamics with stretchable ultrasonic phased arrays
Wang, C., Qi, B., Lin, M., Zhang, Z., Makihata, M., Liu, B., Zhou, S., Huang, Y.-H., Hu, H., Gu, Y., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00763-4 · Cited by 274 · see citing works
Self-citationShin, J. H., Kwon, J., Kim, J. U., Ryu, H., Ok, J., Joon Kwon, S., Park, H., & Kim, T.-I. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00164-w · Cited by 123 · see citing works
Self-citationJeong, H., Lee, J. Y., Lee, K., Kang, Y. J., Kim, J.-T., Avila, R., Tzavelis, A., Kim, J., Ryu, H., Kwak, S. S., et al. (2021) · Science Advances · doi:10.1126/sciadv.abg3092 · Cited by 121 · see citing works
Self-citationRyu, D., Kim, D. H., Price, J. T., Lee, J. Y., Chung, H. U., Allen, E., Walter, J. R., Jeong, H., Cao, J., Kulikova, E., et al. (2021) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2100466118 · Cited by 105 · see citing works
Self-citationPillai, S., Upadhyay, A., Sayson, D., Nguyen, B. H., & Tran, S. D. (2021) · Molecules · doi:10.3390/molecules27010165 · Cited by 95 · see citing works
Jeong, H., Kwak, S. S., Sohn, S., Lee, J. Y., Lee, Y. J., O’Brien, M. K., Park, Y., Avila, R., Kim, J.-T., Yoo, J.-Y., et al. (2021) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2104925118 · Cited by 59 · see citing works
“6B summarizes the heart and respiratory rate (26, 27) determined from such data averaged over a recording time of 60 min at each of these three ages in the form of bar graphs for LR (gray) and ER (red) subjects.”
BackgroundMethodSelf-citationO’Brien, M. K., Botonis, O. K., Larkin, E., Carpenter, J., Martin-Harris, B., Maronati, R., Lee, K., Cherney, L. R., Hutchison, B., Xu, S., et al. (2021) · Digital Biomarkers · doi:10.1159/000517144 · Cited by 37 · see citing works
Self-citationDavies, C., Lee, J. Y., Walter, J., Kim, D., Yu, L., Park, J., Blake, S., Kalluri, L., Cziraky, M., Stanek, E., et al. (2022) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.10194 · Cited by 23 · see citing works
Self-citationWalter, J. R., Lee, J. Y., Snoll, B., Park, J. B., Kim, D. H., Xu, S., & Barnhart, K. (2022) · Sleep Medicine · doi:10.1016/j.sleep.2022.09.026 · Cited by 15 · see citing works
Self-citation
- Authors
- Lee, K., Ni, X., Lee, J. Y., Arafa, H., Pe, D. J., Xu, S., Avila, R., Irie, M., Lee, J. H., Easterlin, R. L., Kim, D. H., Chung, H. U., Olabisi, O. O., Getaneh, S., Chung, E., Hill, M., Bell, J., Jang, H., Liu, C., Park, J. B., Kim, J., Kim, S. B., Mehta, S., Pharr, M., Tzavelis, A., Reeder, J. T., Huang, I., Deng, Y., Xie, Z., Davies, C. R., Huang, Y., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- University of Illinois Urbana-Champaign (US)
- Texas A&M University (US)
- Carle Foundation Hospital (US)
- Corresponding author
- Xie, Zhaoqian
- Davies, Charles R.
- Huang, Yonggang
- Rogers, John A.
- ORCID
- Lee, KunHyuck
- Ni, Xiaoyue
- Lee, Jong Yoon
- Arafa, Hany
- Xu, Shuai
- Avila, Raudel
- Lee, Joo Hee
- Kim, Dong Hyun
- Chung, Ha Uk
- Hill, Marc
- Jang, Hokyung
- Liu, Claire
- Park, Jun Bin
- Kim, Jungwoo
- Kim, Sung Bong
- Mehta, Sunita
- Pharr, Matt
- Tzavelis, Andreas
- Reeder, Jonathan T.
- Huang, Ivy
- Deng, Yujun
- Xie, Zhaoqian
- Davies, Charles R.
- Huang, Yonggang
- Rogers, John A.
- Venue
- Nature Biomedical Engineering
- Published
- 2019-11-25
- Volume
- 4
- Issue
- 2
- Pages
- 148-158
- ISSN
- 2157-846X (ISSN-L)
- PMC
- PMC7035153
- OpenAlex
- W2990691150
- Cited by
- 349
- References
- 62
- Citations by year
- 2026: 34 · 2025: 43 · 2024: 64 · 2023: 53 · 2022: 65 · 2021: 63 · 2020: 27
- FWCI
- 13.66
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Nature Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- U.S. Department of Health & Human Services | NIH | National Institute on Aging (R41AG062023, R43AG060812)
- National Science Foundation Graduate Research Fellowship
- Indo-US Science and Technology Forum
- National Natural Science Foundation of China (11402134)
- National Science Foundation (CMMI1635443)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- suprasternal notch, triaxial accelerometer, vital signs monitoring, heart rate, respiration rate, energy intensity, swallowing, cadence, sleep patterns, polysomnography validation, frequency domain analysis, machine learning
- MeSH terms
- Clavicle, Equipment Design, Humans, Monitoring, Physiologic, Physiological Phenomena, Signal Processing, Computer-Assisted, Skin Physiological Phenomena, Sleep, Vibration, Biosensing Techniques, Exercise, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review