Open access

Abstract

Here we present a testing process using a cryogenic 300-mm wafer prober to collect high-volume data on the performance of hundreds of industry-manufactured spin qubit devices at 1.6 K. This provides fast feedback to enable optimization of the CMOS-compatible fabrication process, leading to high yield and low process variation.

Overview

CoChat Summary

TL;DR

The paper demonstrates a 300-mm cryogenic wafer-probing process for industry-manufactured Si/SiGe spin qubit devices, enabling automated high-volume measurements of single-electron transitions across full wafers. It reports high component yield, low process variation, and spin-qubit operation with coherence times and single-qubit fidelities comparable to leading results.

Key findings

  • A 300-mm cryo-prober cools wafers to an electron temperature of 1.6 ± 0.2 K in about 2 h and automates measurements across thousands of devices, including charge-sensing scans of single-electron transitions.
  • On a representative wafer, component yields were: ohmics 100% (1,624/1,624), gates 100% (10,208/10,208), quantum dots 99.8% (3,703/3,712), and full 12QD arrays 96% (223/232).
  • Random matched-pair variation was 59 mV for gate threshold voltage (VT), 61 mV for 1e voltage on a 30-nm SiGe barrier wafer, and 63 mV on a 50-nm barrier wafer; V1e and VT were strongly correlated (ρ > 0.9).
  • Charge-sensing success was 91% across 696 quantum dot sites on 58 die; 98% of 1e voltage data points passed a scan-margin validation. A median of 63% of quantum dots per 12QD device could be set to n = 1e with a common voltage.
  • Qubit measurements gave average coherence times of 5 μs for NatSi and 205 μs for 28Si, with single-qubit Clifford fidelities of 99.90 ± 0.01% and 99.88 ± 0.02% for two 28Si qubits.

Methods

  • Devices were fabricated in Intel’s D1 factory on 300-mm Si wafers with a Si/Si0.7Ge0.3 heterostructure, using optical lithography, extreme ultraviolet lithography for quantum dot gates, and industry deposition, etch, and chemical-mechanical polish processes; gate pitches ranged from 50 to 100 nm.
  • Testing used a Bluefors/AEM Afore cryogenic 300-mm wafer prober with a chuck base temperature of 1.0 K and electron temperature of 1.6 ± 0.2 K; automated alignment and electrical measurements extracted gate-line resistance, ohmic contact resistance, carrier mobility, gate threshold voltage, barrier–barrier scans, and charge-sensing data.
  • Variation was analyzed with a matched-pair voltage-difference method using mirror-symmetric gate pairs; yield was defined from ohmic, gate, quantum dot, and full-device functionality. Qubit measurements were performed in dilution refrigerators using EDSR, randomized benchmarking, Ramsey, and Hahn echo sequences.

Limitations

  • The paper states that the 91% charge-sensing success rate is primarily limited by the measurement algorithm, including automated charge-sensor tuning and sensor instability during scans.
  • The reported 63% voltage-sharing success rate is described as far from the uniformity needed to tune an ensemble of spin qubits with shared voltages; future work at mK temperatures is needed to expand operation of the 12-qubit unit cell.
  • The paper notes that mobility estimates from the cryo-prober overestimate actual mobility by about 30% because of an approximated threshold density, and that 10–30% of Hall bars show greatly reduced mobility due to discrete defect modes such as misfit dislocation pile-ups.
  • The paper states that the mirror-symmetric matched-pair method may need revision for much larger arrays, and that tuning barrier voltages can increase the matched-pair ΔV1e standard deviation by about 20% for devices with notable cross-capacitance, such as the 50-nm SiGe barrier wafer.
  • Evident from the text: wafer-prober measurements are at 1.6 K rather than the mK temperatures used for spin-qubit operation; qubit fidelity and coherence data come from separate dilution-refrigerator measurements on selected devices, not from the full-wafer prober; yield is reported for one representative wafer and 232 12QD devices.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1038/s41586-024-07275-6

Cites

71 of 71 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 Cryo-prober measurement flow. a, The cryo-prober cools 300-mm wafers (upper image) to an electron temperature of 1.6 K in around 2 h. Lower image shows a cross-sectional transmission electron micrograph of a Si/SiGe quantum dot qubit device. Gates are false-coloured. Scale bar, 100 nm. b, When the wafer is cold, device pads are aligned to the probe pins using wafer stage controls and machine vision feedback. The stage lifts the device pads into contact with the probe pins to connect devices to measurement electronics at room temperature. Device pads are 100 × 100 μm2 with 150-μm pitch. c, With the device in contact, a wide variety of measurements can be performed to extract device data. d, After repeating this process on many devices across the wafer, device data can be used for statistical analysis of wafer-scale trends.
Fig. 1 Cryo-prober measurement flow. a, The cryo-prober cools 300-mm wafers (upper image) to an electron temperature of 1.6 K in around 2 h. Lower image shows a cross-sectional transmission electron micrograph of a Si/SiGe quantum dot qubit device. Gates are false-coloured. Scale bar, 100 nm. b, When the wafer is cold, device pads are aligned to the probe pins using wafer stage controls and machine vision feedback. The stage lifts the device pads into contact with the probe pins to connect devices to measurement electronics at room temperature. Device pads are 100 × 100 μm2 with 150-μm pitch. c, With the device in contact, a wide variety of measurements can be performed to extract device data. d, After repeating this process on many devices across the wafer, device data can be used for statistical analysis of wafer-scale trends.CC BY · PubMed Central
Fig. 2 Process optimization aided by cryo-prober feedback. a, Schematic of the gate structure in an optimized spin qubit array. Gates designed to accumulate charge are coloured yellow, blue and green, whereas gates designed to deplete charge (screening gates) are coloured red. Scale in the vertical direction is approximate. b,c, Spin qubit device variation and electrostatics performance are improved through optimization of the gate stack. Three versions of the device fabrication are highlighted. Only the third version includes the lower screening layer shown in a. b, Gate VT variation both within and between wafers is improved after process optimization. Box plots show the median and interquartile range of each distribution. Whiskers mark the maximum and minimum values excluding outliers, which are defined as points removed from the median by more than 1.5 times the interquartile range. c, Representative quantum dot transport measurements are shown for each of the three versions, with improvements made to disorder and stability.
Fig. 2 Process optimization aided by cryo-prober feedback. a, Schematic of the gate structure in an optimized spin qubit array. Gates designed to accumulate charge are coloured yellow, blue and green, whereas gates designed to deplete charge (screening gates) are coloured red. Scale in the vertical direction is approximate. b,c, Spin qubit device variation and electrostatics performance are improved through optimization of the gate stack. Three versions of the device fabrication are highlighted. Only the third version includes the lower screening layer shown in a. b, Gate VT variation both within and between wafers is improved after process optimization. Box plots show the median and interquartile range of each distribution. Whiskers mark the maximum and minimum values excluding outliers, which are defined as points removed from the median by more than 1.5 times the interquartile range. c, Representative quantum dot transport measurements are shown for each of the three versions, with improvements made to disorder and stability.CC BY · PubMed Central
Fig. 3 Threshold voltage statistics from 12QD arrays. a, Tiled array of I–V curves taken on 12QD devices across a wafer. I–V curves from a single device are shown in the inset, including 27 gates from the linear quantum dot array. b, Schematic of the measurement configuration overlaid on an inline scanning electron microscopy image of a representative 12QD device. Quantum dot locations are indicated by dashed circles. Gates are false-coloured by function: yellow for plunger gates, blue for barrier gates, green for reservoir gates and red for centre screening gate. Scale bar, 100 nm. c, Histograms of gate VT values across the 12QD array. Data are taken from 232 12QD devices across a wafer. d, Histogram of ΔVT calculated between matched gate pairs using the VT dataset shown in c.
Fig. 3 Threshold voltage statistics from 12QD arrays. a, Tiled array of I–V curves taken on 12QD devices across a wafer. I–V curves from a single device are shown in the inset, including 27 gates from the linear quantum dot array. b, Schematic of the measurement configuration overlaid on an inline scanning electron microscopy image of a representative 12QD device. Quantum dot locations are indicated by dashed circles. Gates are false-coloured by function: yellow for plunger gates, blue for barrier gates, green for reservoir gates and red for centre screening gate. Scale bar, 100 nm. c, Histograms of gate VT values across the 12QD array. Data are taken from 232 12QD devices across a wafer. d, Histogram of ΔVT calculated between matched gate pairs using the VT dataset shown in c.CC BY · PubMed Central
Fig. 4 Single-electron voltage statistics from 12QD arrays. a, Example charge-sensing measurement. Bright lines represent electron number transitions in the quantum dot. The red star marks where the single-electron (1e) voltages are extracted for the plunger and barrier gates. The red triangle indicates where the voltage difference between the plunger and barrier gates is extracted. Dashed lines bound regions of extreme tunnel rate (Γ) (see Methods). b, Histograms of 1e plunger and barrier voltages across the 12QD array. Data are taken from a wafer with a 50-nm SiGe barrier. c,d, Histograms of 1e plunger voltage differences calculated between matched gate pairs on a wafer with 30-nm (c) and 50-nm (d) SiGe barriers. e,f, Histograms of 1e–2e addition voltage taken on a wafer with 30-nm (e) and 50-nm (f) SiGe barriers. The uncertainties shown are standard deviations. g,h, Histograms of voltage difference between plunger and barrier gates at the 1e transition taken on a wafer with 30-nm (g) and 50-nm (h) barriers. a.u., arbitrary units.
Fig. 4 Single-electron voltage statistics from 12QD arrays. a, Example charge-sensing measurement. Bright lines represent electron number transitions in the quantum dot. The red star marks where the single-electron (1e) voltages are extracted for the plunger and barrier gates. The red triangle indicates where the voltage difference between the plunger and barrier gates is extracted. Dashed lines bound regions of extreme tunnel rate (Γ) (see Methods). b, Histograms of 1e plunger and barrier voltages across the 12QD array. Data are taken from a wafer with a 50-nm SiGe barrier. c,d, Histograms of 1e plunger voltage differences calculated between matched gate pairs on a wafer with 30-nm (c) and 50-nm (d) SiGe barriers. e,f, Histograms of 1e–2e addition voltage taken on a wafer with 30-nm (e) and 50-nm (f) SiGe barriers. The uncertainties shown are standard deviations. g,h, Histograms of voltage difference between plunger and barrier gates at the 1e transition taken on a wafer with 30-nm (g) and 50-nm (h) barriers. a.u., arbitrary units.CC BY · PubMed Central
Extended Data Fig. 1 From single electrons to spin qubits. a, Scanning electron microscopy image of the cobalt micromagnets fabricated on an Intel 12Q device to enable electric dipole spin resonance for single-qubit control. The white arrow indicates the direction of magnetization M. The dashed line shows where the linear array of quantum dots is formed with respect to the micromagnets. b, Rabi oscillations between the spin-up |1⟩ and the spin-down state |0⟩ driven by EDSR. c, Randomized benchmarking of single-qubit Clifford gates for two qubits, Q1 and Q2, from a 28Si device (dev12). The difference in the measured spin-up probability is plotted for two different starting states, |0⟩ or |1⟩, as a function of sequence length m. From exponential fits (solid lines) of the data, we estimate average Clifford-gate fidelities of 99.90 ± 0.01% and 99.88 ± 0.02% for Q1 and Q2, respectively. d, A Ramsey sequence performed on Q3 from dev12. By fitting the decay (solid line), we extract T2^(*)=15.6μs. e, Cumulative T2^(*) as a function of measurement time for a subset of devices described in g. The dephasing time saturates at long measurements to the limit T2^(*)(∞). f, A Hahn echo sequence performed on Q3 from dev12. By fitting the decay (solid line), we extract T2^Echo=225μs. g, T2^(*) (stars), T2^(*)(∞) (diamonds) and T2^Echo (circles) data points measured from 39 qubits formed in 14 devices (dev1–dev14) from five different wafers (w1–w5). Two device types are featured: a linear array of three qubits (3Q) or 12 qubits (12Q). The colour of each point corresponds to the position of the qubit in the array, which is labelled Q1–Q3 for the 3Q samples and Q1–Q12 for the 12Q samples. Error bars represent uncertainty of the fit.
Extended Data Fig. 1 From single electrons to spin qubits. a, Scanning electron microscopy image of the cobalt micromagnets fabricated on an Intel 12Q device to enable electric dipole spin resonance for single-qubit control. The white arrow indicates the direction of magnetization M. The dashed line shows where the linear array of quantum dots is formed with respect to the micromagnets. b, Rabi oscillations between the spin-up |1⟩ and the spin-down state |0⟩ driven by EDSR. c, Randomized benchmarking of single-qubit Clifford gates for two qubits, Q1 and Q2, from a 28Si device (dev12). The difference in the measured spin-up probability is plotted for two different starting states, |0⟩ or |1⟩, as a function of sequence length m. From exponential fits (solid lines) of the data, we estimate average Clifford-gate fidelities of 99.90 ± 0.01% and 99.88 ± 0.02% for Q1 and Q2, respectively. d, A Ramsey sequence performed on Q3 from dev12. By fitting the decay (solid line), we extract T2^(*)=15.6μs. e, Cumulative T2^(*) as a function of measurement time for a subset of devices described in g. The dephasing time saturates at long measurements to the limit T2^(*)(∞). f, A Hahn echo sequence performed on Q3 from dev12. By fitting the decay (solid line), we extract T2^Echo=225μs. g, T2^(*) (stars), T2^(*)(∞) (diamonds) and T2^Echo (circles) data points measured from 39 qubits formed in 14 devices (dev1–dev14) from five different wafers (w1–w5). Two device types are featured: a linear array of three qubits (3Q) or 12 qubits (12Q). The colour of each point corresponds to the position of the qubit in the array, which is labelled Q1–Q3 for the 3Q samples and Q1–Q12 for the 12Q samples. Error bars represent uncertainty of the fit.CC BY · PubMed Central
Extended Data Fig. 2 Electron temperature measurement in the cryo-prober. a, Charge-stability diagram showing the configuration in which electron temperature is extracted. b, 1D measurement across the transition indicated by the dashed red line in a with theoretical fit overlaid.
Extended Data Fig. 2 Electron temperature measurement in the cryo-prober. a, Charge-stability diagram showing the configuration in which electron temperature is extracted. b, 1D measurement across the transition indicated by the dashed red line in a with theoretical fit overlaid.CC BY · PubMed Central
Extended Data Fig. 3 Process optimization aided by cryo-prober feedback. a,b, Optical images of two test structures with the same pad layout: gate-line resistance test structure (a) and Hall bar (b). In each case, the active probe pads are highlighted and a schematic of the measurement is shown. c,d, Improvements in device metrics from process optimization. Box plots show the median and interquartile range of each distribution. Whiskers mark the maximum and minimum values excluding outliers, which are defined as points removed from the median by more than 1.5 times the interquartile range. c, Gate-line resistance is reduced through optimization of the gate process and introduction of superconducting materials. d, Estimated carrier mobility is increased through improvements in epitaxy and increase in quantum well depth. e, Hall measurements taken in a conventional cryostat show mobility as a function of carrier density for two samples with the QW B process and different SiGe barrier thickness.
Extended Data Fig. 3 Process optimization aided by cryo-prober feedback. a,b, Optical images of two test structures with the same pad layout: gate-line resistance test structure (a) and Hall bar (b). In each case, the active probe pads are highlighted and a schematic of the measurement is shown. c,d, Improvements in device metrics from process optimization. Box plots show the median and interquartile range of each distribution. Whiskers mark the maximum and minimum values excluding outliers, which are defined as points removed from the median by more than 1.5 times the interquartile range. c, Gate-line resistance is reduced through optimization of the gate process and introduction of superconducting materials. d, Estimated carrier mobility is increased through improvements in epitaxy and increase in quantum well depth. e, Hall measurements taken in a conventional cryostat show mobility as a function of carrier density for two samples with the QW B process and different SiGe barrier thickness.CC BY · PubMed Central
Extended Data Fig. 4 Charge-sensing scan analysis. a–e, Sequence to extract transition curve data. a, Raw lock-in data are taken in by the analysis algorithm. b, A first-order Gaussian filter is applied to remove background charge sensor features from the data. c, A maximum filter is applied to locate points of high signal. d, Local maxima are filtered and binned into ‘curve segments’. e, Curve segments are merged into a set of continuous transition curves. The coordinates of these transition curves are collected and used to analyse 1e and 2e transition voltage statistics. f,g, Validation of 1e transition data. f, Histogram of 1e–2e addition voltage statistics from a wafer with a 50-nm SiGe barrier. The vertical dashed line indicates two times the median addition voltage (0.158 V). g, Histogram of the 1e scan margin, or distance between the purported 1e transition and the left edge of the scan window, extracted from charge-sensing scans on the same wafer as f. 98% of scans have a margin more than twice the median addition voltage.
Extended Data Fig. 4 Charge-sensing scan analysis. a–e, Sequence to extract transition curve data. a, Raw lock-in data are taken in by the analysis algorithm. b, A first-order Gaussian filter is applied to remove background charge sensor features from the data. c, A maximum filter is applied to locate points of high signal. d, Local maxima are filtered and binned into ‘curve segments’. e, Curve segments are merged into a set of continuous transition curves. The coordinates of these transition curves are collected and used to analyse 1e and 2e transition voltage statistics. f,g, Validation of 1e transition data. f, Histogram of 1e–2e addition voltage statistics from a wafer with a 50-nm SiGe barrier. The vertical dashed line indicates two times the median addition voltage (0.158 V). g, Histogram of the 1e scan margin, or distance between the purported 1e transition and the left edge of the scan window, extracted from charge-sensing scans on the same wafer as f. 98% of scans have a margin more than twice the median addition voltage.CC BY · PubMed Central
Extended Data Fig. 5 Voltage-sharing analysis. a, Histogram of 1e and 2e electron voltages taken across a full wafer. Each voltage is plotted relative to the common voltage (Vcommon) assigned to the 12QD device from which the data point is taken. 1e voltages above Vcommon (green) represent dots tuned to n = 0e. 2e voltages below Vcommon (orange) represent dots tuned to n ≥ 2e. All other data points (blue and grey) represent dots tuned to n = 1e. b, Scatter plot indicating the fraction of quantum dots tuned to various electron configurations at Vcommon for 12QD devices across a wafer. The fraction of dots in n = 1e represents the success rate, giving a median success rate of 63%. Box plots show the median and interquartile range of each distribution. Whiskers mark the maximum and minimum values excluding outliers, which are defined as points removed from the median by more than 1.5 times the interquartile range. c,d, Simulated success rate for tuning all quantum dots in an array to n = 1e with a common voltage, plotted as a function of array size in number of quantum dots. Vcommon is chosen with two methods: one method in which Vcommon is optimized to maximize the n = 1e fraction and one method in which Vcommon is set based on the mean of V1e and V2e data. The dashed horizontal line indicates the expected success rate for normally distributed V1e and V2e data in the limit of large array size. Stars indicate the success rate extracted from the measured 12QD data using each method. Simulations are performed with experimentally observed V1e variation (c) and with four times lower V1e variation (d).
Extended Data Fig. 5 Voltage-sharing analysis. a, Histogram of 1e and 2e electron voltages taken across a full wafer. Each voltage is plotted relative to the common voltage (Vcommon) assigned to the 12QD device from which the data point is taken. 1e voltages above Vcommon (green) represent dots tuned to n = 0e. 2e voltages below Vcommon (orange) represent dots tuned to n ≥ 2e. All other data points (blue and grey) represent dots tuned to n = 1e. b, Scatter plot indicating the fraction of quantum dots tuned to various electron configurations at Vcommon for 12QD devices across a wafer. The fraction of dots in n = 1e represents the success rate, giving a median success rate of 63%. Box plots show the median and interquartile range of each distribution. Whiskers mark the maximum and minimum values excluding outliers, which are defined as points removed from the median by more than 1.5 times the interquartile range. c,d, Simulated success rate for tuning all quantum dots in an array to n = 1e with a common voltage, plotted as a function of array size in number of quantum dots. Vcommon is chosen with two methods: one method in which Vcommon is optimized to maximize the n = 1e fraction and one method in which Vcommon is set based on the mean of V1e and V2e data. The dashed horizontal line indicates the expected success rate for normally distributed V1e and V2e data in the limit of large array size. Stars indicate the success rate extracted from the measured 12QD data using each method. Simulations are performed with experimentally observed V1e variation (c) and with four times lower V1e variation (d).CC BY · PubMed Central
Extended Data Fig. 6 Further analysis of 1e voltage data. a, Threshold voltage plotted as a function of 1e voltage for the set of plunger gates used in the matched-pair analysis of Fig. 4c,d. Blue (red) data points are taken from a wafer with a 30-nm (50-nm) SiGe barrier. Linear fits to each dataset are shown as solid lines. The correlation coefficient (ρ) is also shown, indicating a high level of correlation for both datasets: 0.92 (0.93) for the wafer with a 30-nm (50-nm) SiGe barrier. b,c, Median of matched-pair 1e voltage difference distributions from the wafer with a 50-nm SiGe barrier, plotted as a function of gate pair for sets of mirror-symmetric pairs (b) and nearest-neighbour pairs (c). d,e, Histograms of matched-pair 1e voltage differences from combined distributions of gate pairs for sets of mirror-symmetric pairs (d) and nearest-neighbour pairs (e).
Extended Data Fig. 6 Further analysis of 1e voltage data. a, Threshold voltage plotted as a function of 1e voltage for the set of plunger gates used in the matched-pair analysis of Fig. 4c,d. Blue (red) data points are taken from a wafer with a 30-nm (50-nm) SiGe barrier. Linear fits to each dataset are shown as solid lines. The correlation coefficient (ρ) is also shown, indicating a high level of correlation for both datasets: 0.92 (0.93) for the wafer with a 30-nm (50-nm) SiGe barrier. b,c, Median of matched-pair 1e voltage difference distributions from the wafer with a 50-nm SiGe barrier, plotted as a function of gate pair for sets of mirror-symmetric pairs (b) and nearest-neighbour pairs (c). d,e, Histograms of matched-pair 1e voltage differences from combined distributions of gate pairs for sets of mirror-symmetric pairs (d) and nearest-neighbour pairs (e).CC BY · PubMed Central
Extended Data Fig. 7 Barrier–barrier scans representing three versions of wafer fabrication. Wafer-scale maps of barrier–barrier scans are shown to represent the three versions of wafer fabrication highlighted in Fig. 2b,c: high-κ stack A with high-temperature spacer (a), high-κ stack B with high-temperature spacer (b) and high-κ stack B with low-temperature spacer and an integrated screening gate layer (c). Each set of scans shows a measurement from one quantum dot per device and represents the complete set from which the individual examples in Fig. 2c are taken. Scans are arranged by device location on the wafer. For the first two sets of measurements, only half of die are measured by sampling in a chequerboard pattern across the wafer. Further missing scans are because of non-yielding quantum dots on the earlier versions of fabrication.
Extended Data Fig. 7 Barrier–barrier scans representing three versions of wafer fabrication. Wafer-scale maps of barrier–barrier scans are shown to represent the three versions of wafer fabrication highlighted in Fig. 2b,c: high-κ stack A with high-temperature spacer (a), high-κ stack B with high-temperature spacer (b) and high-κ stack B with low-temperature spacer and an integrated screening gate layer (c). Each set of scans shows a measurement from one quantum dot per device and represents the complete set from which the individual examples in Fig. 2c are taken. Scans are arranged by device location on the wafer. For the first two sets of measurements, only half of die are measured by sampling in a chequerboard pattern across the wafer. Further missing scans are because of non-yielding quantum dots on the earlier versions of fabrication.CC BY · PubMed Central
Extended Data Fig. 8 Charge-sensing data from wafer with a 30-nm SiGe barrier. Charge-sensing scans are grouped by 12QD device and arranged by wafer location. Scans with unresolved transitions and/or fitting errors are removed. 1e and 2e transition curves identified by the analysis algorithm are plotted in blue and orange, respectively.
Extended Data Fig. 8 Charge-sensing data from wafer with a 30-nm SiGe barrier. Charge-sensing scans are grouped by 12QD device and arranged by wafer location. Scans with unresolved transitions and/or fitting errors are removed. 1e and 2e transition curves identified by the analysis algorithm are plotted in blue and orange, respectively.CC BY · PubMed Central
Extended Data Fig. 9 Charge-sensing data from wafer with a 50-nm SiGe barrier. Charge-sensing scans are grouped by 12QD device and arranged by wafer location. Scans with unresolved transitions and/or fitting errors are removed. 1e and 2e transition curves identified by the analysis algorithm are plotted in blue and orange, respectively.
Extended Data Fig. 9 Charge-sensing data from wafer with a 50-nm SiGe barrier. Charge-sensing scans are grouped by 12QD device and arranged by wafer location. Scans with unresolved transitions and/or fitting errors are removed. 1e and 2e transition curves identified by the analysis algorithm are plotted in blue and orange, respectively.CC BY · PubMed Central
Extended Data Fig. 10 Charge sensing of double quantum dots across a wafer. Charge-sensing scans are taken on eight double quantum dots per 12QD device (two pairs of quantum dots for each charge sensor) and arranged by wafer location.
Extended Data Fig. 10 Charge sensing of double quantum dots across a wafer. Charge-sensing scans are taken on eight double quantum dots per 12QD device (two pairs of quantum dots for each charge sensor) and arranged by wafer location.CC BY · PubMed Central

Verification

Claims the paper makes

8 Valid · 7 Not valid · 9 Not checked · 146 Own results and interpretations

89 more claims

Create a free account to see all 170 claims and every source verdict.

See all claims — free account
  • Probably not a substantial claim. 26
  • It may state two separate claims as one. 57
  • The quote only partly states this claim. 1

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

2 of 2

Citations

Cites

Show all 90 references

Research works printed but not linked: 4

  1. Pillarisetty, R. et al. in 2019 IEEE International Electron Devices Meeting (IEDM) 31.5.1–31.5.4 (IEEE, 2019).
  2. Li, R. et al. in 2020 IEEE International Electron Devices Meeting (IEDM) 38.3.1–38.3.4 (IEEE, 2020).
  3. Kotlyar, R. et al. in 2022 International Electron Devices Meeting (IEDM) 8.4.1–8.4.4 (IEEE, 2022).
  4. Contamin, L.C. et al. in 2022 International Electron Devices Meeting (IEDM) 22.1.1–22.1.4 (IEEE, 2022).

Cited by

See all 178

Metadata

Authors
Neyens, S., Zietz, O. K., Watson, T. F., Luthi, F., Nethwewala, A., George, H. C., Henry, E., Islam, M., Wagner, A. J., Borjans, F., Connors, E. J., Corrigan, J., Curry, M. J., Keith, D., Kotlyar, R., Lampert, L. F., Mądzik, M. T., Millard, K., Mohiyaddin, F. A., Pellerano, S., Pillarisetty, R., Ramsey, M., Savytskyy, R., Schaal, S., Zheng, G., Ziegler, J., Bishop, N. C., Bojarski, S., Roberts, J., & Clarke, J. S.
Affiliations
Intel (United States) (US)
Corresponding author
  • Neyens, Samuel
  • Clarke, James S.
Venue
Nature
Published
2024-05-01
Volume
629
Issue
8010
Pages
80-85
ISSN
  • 0028-0836 (ISSN-L)
  • 1476-4687
OpenAlex
W4396544305
Cited by
178
References
90
Citations by year
2026: 68 · 2025: 77 · 2024: 32
FWCI
20.85
Type
journal article
Language
en
Open access
Licence
Locations
  • Nature (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • arXiv (Cornell University) (submittedVersion, cc-by-nc-nd)
  • RePEc: Research Papers in Economics (submittedVersion)
  • arXiv (Cornell University)
Topics
  • Physical Sciences › Physics and Astronomy › Condensed Matter Physics › Quantum and electron transport phenomena
  • Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Semiconductor Quantum Structures and Devices
  • Physical Sciences › Materials Science › Materials Chemistry › Diamond and Carbon-based Materials Research
Keywords
spin qubits, process variation, CMOS-compatible fabrication

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (qwen3.8-omni-flash and gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Probing single electrons across 300-mm spin qubit wafers. https://app.cochat.ai/papers/10.1038/s41586-024-07275-6
The paper
Neyens, S., Zietz, O. K., Watson, T. F., Luthi, F., Nethwewala, A., George, H. C., Henry, E., Islam, M., Wagner, A. J., Borjans, F., Connors, E. J., Corrigan, J., Curry, M. J., Keith, D., Kotlyar, R., Lampert, L. F., Mądzik, M. T., Millard, K., Mohiyaddin, F. A., . . . Clarke, J. S. (2024). Probing single electrons across 300-mm spin qubit wafers. Nature. https://doi.org/10.1038/s41586-024-07275-6

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.