Verification, analytical validation, and clinical validation (V3): The foundation of determining fit-for-purpose for Biometric Monitoring Technologies (BioMeTs)

Open access
Goldsack, Jennifer C., Coravos, Andrea, Bakker, Jessie P., Bent, Brinnae, Dowling, Ariel V., Fitzer-Attas, Cheryl, Godfrey, Alan, Godino, Job G. +8 more·2020·npj Digital Medicine

Published version of a preprint

Abstract

Digital medicine is an interdisciplinary field, drawing together stakeholders with expertize in engineering, manufacturing, clinical science, data science, biostatistics, regulatory science, ethics, patient advocacy, and healthcare policy, to name a few. Although this diversity is undoubtedly valuable, it can lead to confusion regarding terminology and best practices. There are many instances, as we detail in this paper, where a single term is used by different groups to mean different things, as well as cases where multiple terms are used to describe essentially the same concept. Our intent is to clarify core terminology and best practices for the evaluation of Biometric Monitoring Technologies (BioMeTs), without unnecessarily introducing new terms. We focus on the evaluation of BioMeTs as fit-for-purpose for use in clinical trials. However, our intent is for this framework to be instructional to all users of digital measurement tools, regardless of setting or intended use. We propose and describe a three-component framework intended to provide a foundational evaluation framework for BioMeTs. This framework includes (1) verification, (2) analytical validation, and (3) clinical validation. We aim for this common vocabulary to enable more effective communication and collaboration, generate a common and meaningful evidence base for BioMeTs, and improve the accessibility of the digital medicine field.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1038/s41746-020-0260-4

Cites

5 of 5 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Strong development or vendor-reported performance does not necessarily translate into reliable clinical utility in practice.

    “These concerns are reinforced by empirical evidence showing that strong development or vendor-reported performance does not necessarily translate into reliable clinical utility in practice [6,7].”

    From Beyond Model Development in Healthcare AI: Post-Development Robustness, Post-Deployment Monitoring, and Lifecycle Governance—A Scoping Review of Reviews (2026) · cites this paper as [7]

    In “Moving from current siloed practices to one universal best practice” (page unavailable), the paper says a BioMeT may perform well under V3 yet be useless if the target population cannot use it appropriately in the intended setting. This supports the distinction between performance and clinical utility, but the cited passage does not provide empirical evidence about strong development or vendor-reported performance failing in practice. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • The development and validation of robust and reliable DHTs require the collaboration of diverse experts and a stepwise process analogous to the development of fluid, imaging, and other conventional laboratory-derived biomarkers.

    “The development and validation of robust and reliable DHTs require the collaboration of diverse experts and a stepwise process analogous to the development of fluid, imaging, and other conventional laboratory‐derived biomarkers. 23 , 24 , 25”
  • The Digital Medicine Society V3 framework outlines and explains the typical verification, analytical validation, and clinical validation processes of biometric monitoring technologies.

    “Another resource is the Digital Medicine Society V3 framework, 24 which outlines and explains the typical verification, analytical validation, and clinical validation processes of biometric monitoring technologies.”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

1 retracted

Flagged first, then by citations

Research works printed but not linked: 30

  1. [2]Perakslis E Coravos A Is health-care data the new blood? Lancet Digital Health 2019 1 e8 e9 10.1016/S2589-7500(19)30001-9 33323242 · DOI
  2. [3]Coravos A Digital medicine: a primer on measurement Digit Biomark. 2019 3 31 71 10.1159/000500413 32095767 PMC7015383 · DOI
  3. [5]FDA-NIH Biomarker Working Group. BEST (Biomarkers, EndpointS, and other Tools) Resource. (Food and Drug Administration (US), 2016). 27010052
  4. [6]Versi E ‘Gold standard’ is an appropriate term BMJ 1992 305 187 10.1136/bmj.305.6846.187-b PMC1883235 1515860 · DOI
  5. [9]IEEE Computer Society. IEEE Standard for System, Software, and Hardware Verification and Validation. IEEE Std 1012-2016 (Revision of IEEE Std 1012-2012/ Incorporates IEEE Std 1012-2016/Cor1-2017) 1–260 (2017). 10.1109/IEEESTD.2017.8055462. · DOI
  6. [20]Bignardi GE Validation and verification of automated urine particle analysers J. Clin. Pathol. 2017 70 94 101 10.1136/jclinpath-2016-203958 27802413 · DOI
  7. [24]Witt DR Kellogg RA Snyder MP Dunn J Windows into human health through wearables data analytics Curr. Opin. Biomed. Eng. 2019 9 28 46 10.1016/j.cobme.2019.01.001 31832566 PMC6907085 · DOI
  8. [25]McCamley J Donati M Grimpampi E Mazzà C An enhanced estimate of initial contact and final contact instants of time using lower trunk inertial sensor data Gait Posture 2012 36 316 318 10.1016/j.gaitpost.2012.02.019 22465705 · DOI
  9. [26]Trojaniello D Cereatti A Della Croce U Accuracy, sensitivity and robustness of five different methods for the estimation of gait temporal parameters using a single inertial sensor mounted on the lower trunk Gait Posture 2014 40 487 492 10.1016/j.gaitpost.2014.07.007 25085660 · DOI
  10. [27]Hutchinson, L. et al. Operator bias errors are reduced using standing marker alignment device for repeated visit studies. J. Biomech. Eng.140, 041001 (2018). 29114765
  11. [28]Godfrey A Del Din S Barry G Mathers JC Rochester L Instrumenting gait with an accelerometer: a system and algorithm examination Med. Eng. Phys. 2015 37 400 407 10.1016/j.medengphy.2015.02.003 25749552 PMC4381862 · DOI
  12. [35]Huynh, Q. T., Nguyen, U. D., Irazabal, L. B., Ghassemian, N. & Tran, B. Q. Optimization of an accelerometer and gyroscope-based fall detection algorithm. J. Sens. (2015). 10.1155/2015/452078. · DOI
  13. [36]Pham MH Validation of a lower back “wearable”-based sit-to-stand and stand-to-sit algorithm for patients with parkinson’s disease and older adults in a home-like environment Front. Neurol. 2018 9 652 10.3389/fneur.2018.00652 30158894 PMC6104484 · DOI
  14. [37]Kovalchick C Can composite digital monitoring biomarkers come of age? A framework for utilization J. Clin. Transl. Sci. 2017 1 373 380 10.1017/cts.2018.4 29707260 PMC5916505 · DOI
  15. [39]Wang R Accuracy of wrist-worn heart rate monitors JAMA Cardiol. 2017 2 104 10.1001/jamacardio.2016.3340 27732703 · DOI · Paper page
  16. [40]Braybrook M An ambulatory tremor Score for parkinson’s disease J. Parkinsons Dis. 2016 6 723 731 10.3233/JPD-160898 27589540 · DOI
  17. [41]Panescu, D. Medical device development. in 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society 5591–5594 (2009). 10.1109/IEMBS.2009.5333490. 19964134 · DOI
  18. [43]IEEE Standard for System, Software, and Hardware Verification and Validation. IEEE Std 1012-2016 (Revision of IEEE Std 1012-2012/ Incorporates IEEE Std 1012-2016/Cor1-2017) 1–260 (2017). 10.1109/IEEESTD.2017.8055462. · DOI
  19. [44]National Academies of Sciences, Engineering, and Medicine. An Evidence Framework for Genetic Testing. (The National Academies Press, 2017). 10.17226/24632. 28418631 · DOI
  20. [45]Giles D Draper N Neil W Validity of the Polar V800 heart rate monitor to measure RR intervals at rest Eur. J. Appl. Physiol. 2016 116 563 571 10.1007/s00421-015-3303-9 26708360 PMC4751190 · DOI
  21. [47]Hernando D Garatachea N Almeida R Casajús JA Bailón R Validation of heart rate monitor polar rs800 for heart rate variability analysis during exercise J. Strength Cond. Res. 2018 32 716 10.1519/JSC.0000000000001662 27749728 · DOI
  22. [48]Frasch, M. G. et al. Can a heart rate variability biomarker identify the presence of autism spectrum disorder in eight year old children? arXiv:1808.08306 [q-bio] (2018). · arXiv
  23. [49]Karpman C LeBrasseur NK DePew ZS Novotny PJ Benzo RP Measuring gait speed in the out-patient clinic: methodology and feasibility Respir. Care 2014 59 531 537 10.4187/respcare.02688 23983271 · DOI
  24. [50]Fortune E Lugade V Morrow M Kaufman K Validity of using tri-axial accelerometers to measure human movement – Part II: step counts at a wide range of gait velocities Med. Eng. Phys. 2014 36 659 669 10.1016/j.medengphy.2014.02.006 24656871 PMC4030415 · DOI
  25. [51]König A Objective measurement of gait parameters in healthy and cognitively impaired elderly using the dual-task paradigm Aging Clin. Exp. Res 2017 29 1181 1189 10.1007/s40520-016-0703-6 28130713 PMC5674109 · DOI
  26. [54]Parvinian B Scully C Wiyor H Kumar A Weininger S Regulatory considerations for physiological closed-loop controlled medical devices used for automated critical care: food and drug administration workshop discussion topics Anesth. Analg. 2018 126 1916 1925 10.1213/ANE.0000000000002329 28763355 PMC6233305 · DOI
  27. [55]Allen, N. & Gupta, A. Current Diabetes Technology: Striving for the Artificial Pancreas. Diagnostics (Basel)9, 31 (2019). PMC6468523 30875898
  28. [57]Watanabe N Development and Validation of a Novel Cuff-Less Blood Pressure Monitoring Device. J Am Coll Cardiol Basic Trans JACC: Basic to Translational Science 2017 2 631 642 30062178 PMC6058997
  29. [58]IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices. IEEE Std 1708-2014 1–38 (2014). 10.1109/IEEESTD.2014.6882122. · DOI
  30. [60]IEEE standard for software verification and validation. (Institute of Electrical and Electronics Engineers, 1998).

Cited by

Sorted by citations

Authors
Goldsack, J. C., Coravos, A., Bakker, J. P., Bent, B., Dowling, A. V., Fitzer-Attas, C., Godfrey, A., Godino, J. G., Gujar, N., Izmailova, E., Manta, C., Peterson, B., Vandendriessche, B., Wood, W. A., Wang, K. W., & Dunn, J.
Affiliations
  • Harvard University (US)
  • Duke University (US)
  • Northumbria University (GB)
  • University of California San Diego (US)
  • Curis (United States) (US)
  • Case Western Reserve University (US)
  • University of North Carolina at Chapel Hill (US)
Corresponding author
Dunn, Jessilyn
Venue
npj Digital Medicine
Published
2020-04-14
Volume
3
Issue
1
Pages
55-55
ISSN
2398-6352 (ISSN-L)
OpenAlex
W3016443966
Cited by
557
References
5
Citations by year
2026: 138 · 2025: 95 · 2024: 108 · 2023: 82 · 2022: 52 · 2021: 54 · 2020: 27 · 2013: 1
FWCI
12.18
Type
journal article
Language
en
Open access
Locations
  • npj Digital Medicine (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Northumbria Research Link (Northumbria University) (acceptedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
Duke FORGE Fellowship
Topics
  • Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
  • Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
Keywords
verification, analytical method validation, clinical validation, fit for purpose, digital medicine, terminology standardization, clinical trials, evaluation framework

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review