Fitting Characteristics of Eighteen N95 Filtering-Facepiece Respirators
Abstract
Four performance measures were used to evaluate the fitting characteristics of 18 models of N95 filtering-facepiece respirators: (1) the 5th percentile simulated workplace protection factor (SWPF) value, (2) the shift average SWPF value, (3) the h-value, and (4) the assignment error. The effect of fit-testing on the level of protection provided by the respirators was also evaluated. The respirators were tested on a panel of 25 subjects with various face sizes. Simulated workplace protection factor values, determined from six total penetration (face-seal leakage plus filter penetration) tests with re-donning between each test, were used to indicate respirator performance. Five fit-tests were used: Bitrex, saccharin, generated aerosol corrected for filter penetration, PortaCount Plus corrected for filter penetration, and the PortaCount Plus with the N95-Companion accessory. Without fit-testing, the 5th percentile SWPF for all models combined was 2.9 with individual model values ranging from 1.3 to 48.0. Passing a fit-test generally resulted in an increase in protection. In addition, the h-value of each respirator was computed. The h-value has been determined to be the population fraction of individuals who will obtain an adequate level of protection (i.e., SWPF >/=10, which is the expected level of protection for half-facepiece respirators) when a respirator is selected and donned (including a user seal check) in accordance with the manufacturer's instructions without fit-testing. The h-value for all models combined was 0.74 (i.e., 74% of all donnings resulted in an adequate level of protection), with individual model h-values ranging from 0.31 to 0.99. Only three models had h-values above 0.95. Higher SWPF values were achieved by excluding SWPF values determined for test subject/respirator combinations that failed a fit-test. The improvement was greatest for respirator models with lower h-values. Using the concepts of shift average and assignment error to measure respirator performance yielded similar results. The highest level of protection was provided by passing a fit-test with a respirator having good fitting characteristics.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The Quality Factor (QF) is commonly used to compare the performance of materials used to filter particulates under similar experimental conditions [62–81];”
From The protective performance of reusable cloth face masks, disposable procedure masks, KN95 masks and N95 respirators: Filtration and total inward leakage (2021) · cites this paper as [71]
Citations
Cites
NIOSH guide to industrial respiratory protection.
(1987) · doi:10.26616/nioshpub87116 · Cited by 128
Comparison of Five Methods for Fit-Testing N95 Filtering-Facepiece Respirators
Coffey, C. C., Lawrence, R. B., Zhuang, Z., Campbell, D. L., Jensen, P. A., & Myers, W. R. (2002) · Applied Occupational and Environmental Hygiene · doi:10.1080/10473220290107002 · Cited by 70
Self-citationSimulated Workplace Performance of N95 Respirators
Coffey, C. C., Campbell, D. L., & Zhuang, Z. (1999) · American Industrial Hygiene Association Journal · doi:10.1080/00028899908984481 · Cited by 62
Self-citationCoffey, C. C., Campbell, D. L., Myers, W. R., & Zhuang, Z. (1998) · American Industrial Hygiene Association Journal · doi:10.1080/15428119891011036 · Cited by 40
Self-citationASSIGNED PROTECTION FACTORS FOR TWO RESPIRATOR TYPES BASED UPON WORKPLACE PERFORMANCE TESTING
Lenhart, S. W., & Campbell, D. L. (1984) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/28.2.173 · Cited by 39
Self-citationRespiratory Protection as a Function of Respirator Fitting Characteristics and Fit-Test Accuracy
Campbell, D. L., Coffey, C. C., & Lenhart, S. W. (2001) · AIHAJ - American Industrial Hygiene Association · doi:10.1080/15298660108984607 · Cited by 36
Self-citationMcKay, R. T., & Davies, E. (2000) · Applied Occupational and Environmental Hygiene · doi:10.1080/104732200301269 · Cited by 23
1997 American Industrial Hygiene Conference and Exposition
(1997) · Biomedical Safety & Standards · doi:10.1097/00149078-199704010-00016 · Cited by 14
Research works printed but not linked: 10
- National Institute for Occupational Safety and Health (NIOSH), 1995, Federal Register, 60, 30336
- Centers for Disease Control and Prevention, 1998, MMWR, 47, 1045
- American National Standards Institute (ANSI), 1992, American National Standard for Respiratory Protection (ANSI-Z88.2)
- Industrial Safety Equipment Association, 1996, Use and Selection Guide for Non-Powered Air Purifying Particulate Respirators
- Occupational Safety and Health Administration (OSHA), 1998, Final Rule Federal Register, 63, 1278
- Hyatt E. C., 1976, Respirator Protection Factors (LA-6084-MS)
- Hack A., 1974, Selection of respirator test panels representative of U.S. adult facial sizes (LA-5488)
- U.S. Department of Health and Human Services (DHHS), 1999, Policy Statement: Saccharin Use for Respirator Fit Testing
- Coffey C. C., 1998, J. Int. Soc. Respir. Protect., 16, 25
- American National Standards Institute (ANSI), 2001, American National Standard for Respirator Fit Testing Methods (ANSI/AIHA-Z88.10)
Cited by
Sorted by citations
Jensen, P. A., Lambert, L. A., Iademarco, M. F., & Ridzon, R. (2005) · PsycEXTRA Dataset · doi:10.1037/e545842006-001 · Cited by 1,269
Self-citationTransmission of influenza A in human beings
Brankston, G., Gitterman, L., Hirji, Z., Lemieux, C., & Gardam, M. A. (2007) · The Lancet Infectious Diseases · doi:10.1016/s1473-3099(07)70029-4 · Cited by 743
Particle Sizes of Infectious Aerosols: Implications for Infection Control
Fennelly, K. P. (2020) · The Lancet Respiratory Medicine · doi:10.1016/s2213-2600(20)30323-4 · Cited by 621
Natural Ventilation for the Prevention of Airborne Contagion
Escombe, A. R., Oeser, C. C., Gilman, R. H., Navincopa, M., Ticona, E., Pan, W., Martínez, C., Chacaltana, J., Rodríguez, R., Moore, D. A. J., et al. (2007) · PLoS Medicine · doi:10.1371/journal.pmed.0040068 · Cited by 420
Bałazy, A., Toivola, M., Adhikari, A., Sivasubramani, S. K., Reponen, T., & Grinshpun, S. A. (2006) · American Journal of Infection Control · doi:10.1016/j.ajic.2005.08.018 · Cited by 376
Lange, C. G., Abubakar, I., Alffenaar, J. C., Bothamley, G. H., Caminero, J. A., Carvalho, A. C. C., Chang, K. C., Codecasa, L. R., Correia, A., Crudu, V., et al. (2014) · European Respiratory Journal · doi:10.1183/09031936.00188313 · Cited by 308
Health Effects of Wildfire Smoke in Children and Public Health Tools: A Narrative Review
Holm, S. M., Miller, M. D., & Balmes, J. R. (2021) · Journal of Exposure Science & Environmental Epidemiology · doi:10.1038/s41370-020-00267-4 · Cited by 246
Lee, S.-A., Grinshpun, S. A., & Reponen, T. (2008) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/men005 · Cited by 235
Bałazy, A., Toivola, M., Reponen, T. A., Podgórski, A., Zimmer, A. T., & Grinshpun, S. A. (2005) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/mei058 · Cited by 237
Basu, S., Andrews, J. R., Poolman, E. M., Gandhi, N. R., Shah, N., Moll, A. P., Moodley, P., Galvani, A. P., & Friedland, G. H. (2007) · The Lancet · doi:10.1016/s0140-6736(07)61636-5 · Cited by 203
Lee, S.-A., Hwang, D.-C., Li, H.-Y., Tsai, C.-F., Chen, C.-W., & Chen, J.-K. (2016) · Journal of Healthcare Engineering · doi:10.1155/2016/8572493 · Cited by 134
Impact of multiple consecutive donnings on filtering facepiece respirator fit
Bergman, M. S., Viscusi, D. J., Zhuang, Z., Palmiero, A. J., Powell, J. B., & Shaffer, R. E. (2011) · American Journal of Infection Control · doi:10.1016/j.ajic.2011.05.003 · Cited by 127
Self-citationComparison of Performance of Three Different Types of Respiratory Protection Devices
Lawrence, R. B., Duling, M. G., Calvert, C. A., & Coffey, C. C. (2006) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620600829211 · Cited by 127
Self-citationThe Effect of Subject Characteristics and Respirator Features on Respirator Fit
Zhuang, Z., Coffey, C. C., & Ann, R. B. (2005) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620500391668 · Cited by 93
Self-citationYassi, A., Moore, D. M., FitzGerald, J. M., Bigelow, P., Hon, C., & Bryce, E. (2005) · Journal of Occupational and Environmental Medicine · doi:10.1097/01.jom.0000150207.18085.41 · Cited by 93
Transmission and Institutional Infection Control of TuberculosisPreprint
Nardell, E. A. (2015) · Cold Spring Harbor Perspectives in Medicine · doi:10.1101/cshperspect.a018192 · Cited by 93
Masks and respirators for prevention of respiratory infections: a state of the science review
Greenhalgh, T., MacIntyre, C. R., Baker, M. G., Bhattacharjee, S., Chughtai, A. A., Fisman, D. N., Kunasekaran, M. P., Kvalsvig, A., Lupton, D. A., Oliver, M., et al. (2024) · Clinical Microbiology Reviews · doi:10.1128/cmr.00124-23 · Cited by 110
Wein, L. M. (2009) · Operations Research · doi:10.1287/opre.1090.0695 · Cited by 80
McMahon, E., Wada, K., & Dufresne, A. (2008) · American Journal of Infection Control · doi:10.1016/j.ajic.2007.10.014 · Cited by 75
Duncan, S., Bodurtha, P., & Naqvi, S. (2021) · PLOS ONE · doi:10.1371/journal.pone.0258191 · Cited by 75
Wilkinson, I. J., Pisaniello, D., Ahmad, J., & Edwards, S. (2010) · Infection Control and Hospital Epidemiology · doi:10.1086/655460 · Cited by 68
Histoplasmosis: Protecting Workers at Risk
Lenhart, S. W., Schafer, M. P., Singal, M., & Hajjeh, R. A. (2004) · Cited by 61
Simulated workplace protection factors for half-facepiece respiratory protective devices
Duling, M. G., Lawrence, R. B., Slaven, J. E., & Coffey, C. C. (2007) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620701346925 · Cited by 52
Self-citationDevelopment of an Advanced Respirator Fit-Test Headform
Bergman, M. S., Zhuang, Z., Hanson, D. F., Heimbuch, B. K., McDonald, M. J., Palmiero, A. J., Shaffer, R. E., Harnish, D. A., Husband, M., & Wander, J. D. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2013.816434 · Cited by 52
Self-citationHon, C., Danyluk, Q., Bryce, E., Janssen, B., Neudorf, M., Yassi, A., Shen, H., & Astrakianakis, G. (2016) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2016.1237030 · Cited by 51
Quantifying exposure risk: Surgical masks and respirators
Diaz, K. T., & Smaldone, G. C. (2010) · American Journal of Infection Control · doi:10.1016/j.ajic.2010.06.002 · Cited by 49
Assessing Infection Control Measures for Pandemic Influenza
Wein, L. M., & Atkinson, M. P. (2009) · Risk Analysis · doi:10.1111/j.1539-6924.2009.01232.x · Cited by 46
Lee, J. Y. (2016) · Tuberculosis & respiratory diseases · doi:10.4046/trd.2016.79.4.234 · Cited by 48
Fitting Characteristics of N95 Filtering-Facepiece Respirators Used Widely in China
Yu, Y.-Y., Jiang, L.-M., Zhuang, Z., Liu, Y., Wang, X., Liu, J., Ming-na, Y., & Chen, W. (2014) · PLoS ONE · doi:10.1371/journal.pone.0085299 · Cited by 42
Self-citationLee, M. C., Takaya, S., Long, R., & Joffe, A. M. (2008) · Infection Control and Hospital Epidemiology · doi:10.1086/591860 · Cited by 41
Does Wearing a Surgical Facemask or N95-Respirator Impair Radio Communication?
Thomas, F., Allen, C., Butts, W. J., Rhoades, C., Brandon, C., & Handrahan, D. L. (2011) · Air Medical Journal · doi:10.1016/j.amj.2010.12.007 · Cited by 41
Winter, S., Thomas, J. H., Stephens, D. P., & Davis, J. S. (2010) · Critical Care and Resuscitation · doi:10.1016/s1441-2772(23)01352-2 · Cited by 39
Avoiding COVID-19: Aerosol GuidelinesPreprint
Evans, M. (2020) · medRxiv · doi:10.1101/2020.05.21.20108894 · Cited by 41
Chopra, J., Abiakam, N., Kim, H., Metcalf, C., Worsley, P. R., & Cheong, Y. C. (2021) · BMJ Global Health · doi:10.1136/bmjgh-2021-005537 · Cited by 37
Guha, S., McCaffrey, B., Hariharan, P., & Myers, M. R. (2016) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2016.1237029 · Cited by 31
Levine, C., Grady, C., Block, T. R., Hurley, H. J., Russo, R., Peixoto, B., Frees, A., Ruiz, A., & Alland, D. (2020) · Journal of Hospital Infection · doi:10.1016/j.jhin.2020.10.007 · Cited by 32
Schmitt, J., & Wang, J. (2022) · Indoor Air · doi:10.1111/ina.13127 · Cited by 35
Respirator Donning in Post-Hurricane New Orleans
Cummings, K. J., Cox‐Ganser, J. M., Riggs, M. A., Edwards, N. T., & Kreiss, K. (2007) · Emerging infectious diseases · doi:10.3201/eid1305.061490 · Cited by 32
Or, P. P.-L., Chung, J. W. Y., & Wong, T. K. S. (2018) · Journal of Clinical Nursing · doi:10.1111/jocn.14268 · Cited by 30
Errors Associated with Three Methods of Assessing Respirator Fit
Coffey, C. C., Lawrence, R. B., Zhuang, Z., Duling, M. G., & Campbell, D. L. (2005) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620500455398 · Cited by 30
Self-citationRespiratory Issues in Beef and Pork Production: Recommendations From an Expert Panel
Essen, S. V., Moore, G., Gibbs, S. G., & Larson, K. R. L. (2010) · Journal of Agromedicine · doi:10.1080/1059924x.2010.486283 · Cited by 29
Online National Health Agency Mask Guidance for the Public in Light of COVID-19: Content Analysis
Laestadius, L. I., Wang, Y., Taleb, Z. B., Kalan, M. E., Cho, Y. I., & Manganello, J. A. (2020) · JMIR Public Health and Surveillance · doi:10.2196/19501 · Cited by 28
Personal protective equipment and evidence‐based advice for surgical departments during COVID ‐19
Tan, L., Kovoor, J. G., Williamson, P., Tivey, D. R., Babidge, W., Collinson, T. G., Hewett, P. J., Hugh, T. J., Padbury, R. T. A., Langley, S. J., et al. (2020) · ANZ Journal of Surgery · doi:10.1111/ans.16194 · Cited by 27
Shaffer, R. E., & Janssen, L. L. (2014) · American Journal of Infection Control · doi:10.1016/j.ajic.2014.10.021 · Cited by 27
Fit Testing Respirators for Public Health Medical Emergencies
Brosseau, L. M. (2010) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2010.514782 · Cited by 26
Bergman, M. S., Zhuang, Z., Brochu, E., & Palmiero, A. J. (2015) · Journal of the International Society for Respiratory Protection · doi:10.66828/001c.166885 · Cited by 25
Self-citationDo Various Respirator Models Fit the Workers in the Norwegian Smelting Industry?
Føreland, S., Robertsen, O., & Hegseth, M. N. (2019) · Safety and Health at Work · doi:10.1016/j.shaw.2019.06.004 · Cited by 24
Infection Prevention Precautions for Routine Anesthesia Care During the SARS-CoV-2 Pandemic
Bowdle, A. T., Jelačić, S., Shishido, S., & Munoz-Price, L. S. (2020) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000005169 · Cited by 28
Protecting health care workers from tuberculosis: A 10-year experience
Welbel, S. F., French, A. L., Bush, P., DeGuzman, D., & Weinstein, R. A. (2009) · American Journal of Infection Control · doi:10.1016/j.ajic.2009.01.004 · Cited by 23
Davidson, C. S., Green, C. F., Gibbs, S. G., Schmid, K. K., Panlilio, A. L., Jensen, P. A., & Scarpino, P. V. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2013.818243 · Cited by 23
Self-citationClaudio, C. V., Ribeiro, R. P., Martins, J. T., Marziale, M. H. P., Solci, M. C., & Dalmas, J. C. (2017) · Revista Latino-Americana de Enfermagem · doi:10.1590/1518-8345.1561.2853 · Cited by 23
Respiratory Source Control Versus Receiver Protection: Impact of Facemask Fit
Mansour, M. M., & Smaldone, G. C. (2013) · Journal of Aerosol Medicine and Pulmonary Drug Delivery · doi:10.1089/jamp.2012.0998 · Cited by 22
(2013) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/met054 · Cited by 22
Evaluation of the Benefit of the User Seal Check on N95 Filtering Facepiece Respirator Fit
Viscusi, D. J., Bergman, M. S., Zhuang, Z., & Shaffer, R. E. (2012) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2012.683757 · Cited by 21
Self-citationKim, H., Jung-Eun, B. A. E. K., Seo, H., Lee, J.-E., Myong, J., Lee, S., & Lee, J. (2015) · Industrial Health · doi:10.2486/indhealth.2014-0259 · Cited by 19
Radonovich, L. J., Perl, T. M., Davey, V. J., & Cohen, H. J. (2009) · Disaster Medicine and Public Health Preparedness · doi:10.1097/dmp.0b013e3181be830c · Cited by 20
Hariharan, P., Sharma, N., Guha, S., Banerjee, R. K., D’Souza, G. A., & Myers, M. R. (2021) · Scientific Reports · doi:10.1038/s41598-021-89604-7 · Cited by 19
Brown, C. K. (2019) · Health Security · doi:10.1089/hs.2018.0056 · Cited by 15
Myong, J., Byun, J., Cho, Y., Seo, H., Jung-Eun, B. A. E. K., Koo, J., & Kim, H. (2015) · Industrial Health · doi:10.2486/indhealth.2015-0072 · Cited by 15
Au, S. S. W., Gomersall, C. D., Leung, P., & Li, P. T. Y. (2010) · Journal of Hospital Infection · doi:10.1016/j.jhin.2010.01.017 · Cited by 15
Protecting Health Care Workers From Tuberculosis, 2013
Behrman, A. J., Buchta, W. G., Budnick, L. D., Hodgson, M. J., Raymond, L. W., Russi, M. B., Spillmann, S. J., & Swift, M. D. (2013) · Journal of Occupational and Environmental Medicine · doi:10.1097/jom.0b013e3182a0d7cd · Cited by 15
Development of a Dust Respirator by Improving the Half Mask Frame Design
Bazaluk, O., Чеберячко, С. І., Cheberiachko, Y., Deryugin, O., Lozynskyi, V. G., Книш, І. М., Saik, P. B., & Naumov, M. (2021) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph18105482 · Cited by 13
Apisarnthanarak, A., Warren, D. K., & Fraser, V. J. (2007) · Clinical Infectious Diseases · doi:10.1086/522538 · Cited by 13
Myers, M. R., Hariharan, P., Guha, S., & Yan, J. (2016) · Mathematical Medicine and Biology A Journal of the IMA · doi:10.1093/imammb/dqw018 · Cited by 12
Wilson, A. M., Jones, R. M., Lerma, V. L., Abney, S. E., King, M., Weir, M. H., Sexton, J. D., Noakes, C. J., & Reynolds, K. A. (2021) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2021.1926468 · Cited by 12
The characteristics, treatment and prevention of laboratory animal allergy
Corradi, M., Ferdenzi, E., & Mutti, A. (2012) · Lab Animal · doi:10.1038/laban.163 · Cited by 12
Effectiveness of facemasks for pediatric populations against submicron-sized aerosols
Guha, S., Mejía-Alfaro, A., Hariharan, P., & Myers, M. R. (2015) · American Journal of Infection Control · doi:10.1016/j.ajic.2015.03.032 · Cited by 13
Reducing Respirator Fit Test Errors: A Multi-Donning Approach
Campbell, D. L., Coffey, C. C., Jensen, P. A., & Zhuang, Z. (2005) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620500182174 · Cited by 12
Self-citationLofgren, D. J. (2012) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2012.733578 · Cited by 11
Qualitative fitting characteristics of filtering face-piece respirators on Iranian people
Fakherpour, A., Jahangiri, M., & Seif, M. (2020) · Journal of Environmental Health Science and Engineering · doi:10.1007/s40201-020-00484-x · Cited by 11
Enhancement of peripheral seal of medical face masks using a 3-dimensional–printed custom frame
Ahmed, A. S. I., Zhong, Z., Suprono, M. S., Savignano, R., Riter, H., Oyoyo, U. E., Wilson, A., Reece, R., Kim, J., Cho, E.-H., et al. (2021) · The Journal of the American Dental Association · doi:10.1016/j.adaj.2021.03.011 · Cited by 12
Results of Inspections in Health Hazard Industries in a Region of the State of Washington
Lofgren, D. J. (2008) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459620802066133 · Cited by 11
Rengasamy, S., Walbert, G. F., Newcomb, W. E., Coffey, C. C., Wassell, J. T., & Szalajda, J. V. (2014) · The Annals of Occupational Hygiene · doi:10.1093/annhyg/meu095 · Cited by 10
Self-citationBowdle, A. T., Jelačić, S., Munoz-Price, L. S., Cohen, M., Sai Krishna, M., & Brosseau, L. M. (2021) · Anesthesiology · doi:10.1097/aln.0000000000004005 · Cited by 9
Extreme weather—Wildfires & pregnancy
Basilio, E., & Zlatnik, M. G. (2023) · Seminars in Perinatology · doi:10.1016/j.semperi.2023.151839 · Cited by 9
Fakherpour, A., Jahangiri, M., Yousefinejad, S., & Seif, M. (2019) · MethodsX · doi:10.1016/j.mex.2019.05.034 · Cited by 9
Sekoguchi, S., Shirasaka, T., Ando, H., Ikegami, K., & Ogami, A. (2020) · Industrial Health · doi:10.2486/indhealth.2020-0056 · Cited by 8
Fakherpour, A., Jahangiri, M., Yousefinejad, S., Seif, M., & Banaee, S. (2019) · Industrial Health · doi:10.2486/indhealth.2109-0019 · Cited by 8
Investigation of the Filtration Properties of Medical Masks
Akalın, M., Usta, İ., Koçak, D., & Ozen, M. S. (2010) · Elsevier eBooks · doi:10.1533/9780857090348.93 · Cited by 8
Rengasamy, S., & Eimer, B. C. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2013.818224 · Cited by 8
Manikin-based size-resolved penetrations of CE-marked filtering facepiece respirators
Serfozo, N., Ondráček, J., Otahal, P., Lazaridis, M., & Ždı́mal, V. (2017) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2017.1358816 · Cited by 8
Qualitative Fit Testing of Medium - Size N95/FFP2 Respirators on Iranian Health Care Workers
Honarbakhsh, M., Jahangiri, M., Ghaem, H., Ghorbani, M., Omidvari, F., khorasani, M. A., & Shabani, F. (2018) · Health Scope · doi:10.5812/jhealthscope.62884 · Cited by 8
The effect of N95 designs on respirator fit and its associations with gender and facial dimensions
Hasni, N. A. K., Ismail, R., Robat, R. M., Mohamad, N., Suib, F. A., Pahrol, M. A., Mahmud, H., Osman, B., Lim, Y. C., Seman, Z., et al. (2023) · PLoS ONE · doi:10.1371/journal.pone.0288105 · Cited by 7
Zhuang, Z., Bergman, M. S., Eimer, B. C., & Shaffer, R. E. (2013) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2013.818237 · Cited by 7
Self-citationAssessment of Indoor Air Quality for Group-Housed Macaques (Macaca spp.)
Maaskant, A., Janssen, I., Wouters, I., van Eerdenburg, F. V., Remarque, E., Langermans, J., & Bakker, J. (2022) · Animals · doi:10.3390/ani12141750 · Cited by 6
Davidson, C. S., Green, C. F., Panlilio, A. L., Jensen, P. A., Stover, B. H., Roselle, G. A., Gibbs, S. G., & Scarpino, P. V. (2010) · Indoor and Built Environment · doi:10.1177/1420326x10378805 · Cited by 7
Self-citationNIOSH-NPPTL, 6. C. M. R., & Zhuang, Z. (2015) · Journal of Nanotechnology and Materials Science · doi:10.15436/2377-1372.15.015 · Cited by 7
Self-citationPreventing Transmission of Mycobacterium Tuberculosis—A Refocused Approach
Nardell, E. A. (2019) · Clinics in Chest Medicine · doi:10.1016/j.ccm.2019.07.010 · Cited by 7
Occupation-Related Respiratory Infections Revisited
Ling, D. I., & Menzies, D. (2010) · Infectious Disease Clinics of North America · doi:10.1016/j.idc.2010.04.013 · Cited by 6
Protecting Health Care Workers from Tuberculosis
Medical Center Occupational Health Section and Occupational and Environmental Lung Disorders Committee (2008) · Journal of Occupational and Environmental Medicine · doi:10.1097/jom.0b013e318176c523 · Cited by 6
Yan, R., Chillrud, S. N., Magadini, D. L., & Yan, B. (2020) · Journal of the International Society for Respiratory Protection · doi:10.66828/001c.166423 · Cited by 5
Consistency and Reproducibility of Bioaerosol Delivery for Infectivity Studies on Mice
Stone, B. R. (2010) · Cited by 4
Qualitative fit test outcomes for N95 filtering facepiece respirators in health care workers
Wardhan, R., Patel, V., Neff, A. B., Cook, R., White, J. D., & Vasilopoulos, T. (2022) · American Journal of Infection Control · doi:10.1016/j.ajic.2022.05.022 · Cited by 4
Personal respiratory protection and prevention of occupational tuberculosis.
Fennelly, K. P. (2005) · PubMed · Cited by 4
Danyluk, Q., Hon, C., Astrakianakis, G., Bryce, E., Janssen, B., & Yassi, A. (2023) · doi:10.32920/23935236.v1 · Cited by 4
Assessment of Two Personal Breathing Recording Devices in a Simulated Healthcare Environment
Zhu, J., He, X., Guffey, S. E., Bergman, M. S., Lee, E. G., & Zhuang, Z. (2018) · Journal of the International Society for Respiratory Protection · doi:10.66828/001c.166805 · Cited by 3
Self-citationSimulated workplace protection factor study of a quarter-facepiece elastomeric respirator
Sietsema, M., Hamza, H., & Brosseau, L. M. (2022) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2022.2145014 · Cited by 3
P2/N95 fit testing and the risk of COVID-19 in Healthcare Workers
Collis, B. J., TACEY, M. A., McGrath, C., Madigan, V., Kainer, M. A., Tramontana, A. R., & Aboltins, C. A. (2023) · Infection Disease & Health · doi:10.1016/j.idh.2023.07.005 · Cited by 3
Evaluation of Rigidity of Surgical N95 Respirators Using a Manikin-System: A Pilot Study.
Rengasamy, S., & Niezgoda, G. (2019) · PubMed · Cited by 3
Coffey, C. C., Miller, C., & Szalajda, J. V. (2021) · Journal of Occupational and Environmental Hygiene · doi:10.1080/15459624.2021.1976411 · Cited by 3
Self-citation
Metadata
- Authors
- Coffey, C. C., Lawrence, R. B., Campbell, D. L., Zhuang, Z., Calvert, C. A., & Jensen, P. A.
- Affiliations
- National Institute for Occupational Safety and Health (US)
- Venue
- Journal of Occupational and Environmental Hygiene
- Publisher
- Informa UK Limited
- Published
- 2004-04
- Volume
- 1
- Issue
- 4
- Pages
- 262-271
- ISSN
- 1545-9624 (ISSN-L)
- 1545-9632
- OpenAlex
- W2023424240
- Cited by
- 119
- References
- 8
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 1 · 2023: 5 · 2022: 6 · 2021: 11 · 2020: 11 · 2019: 8 · 2018: 4 · 2017: 4 · 2016: 5 · 2015: 6 · 2014: 5 · 2013: 8 · 2012: 4
- FWCI
- 4.98
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Occupational and Environmental Hygiene (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Infection Control and Ventilation
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Occupational Health and Safety in Workplaces
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Occupational Health and Performance
- Keywords
- N95 respirators, filtering facepiece respirators, respirator fit testing, workplace protection factor, H-value, respirator selection, filter penetration, face size
- MeSH terms
- Adult, Equipment Design, Face, Female, Filtration, Humans, Male, Materials Testing, Respiratory Protective Devices, Sensitivity and Specificity, Occupational Exposure, Workplace
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review