Accuracy of a blood glucose monitoring system that recognizes insufficient sample blood volume and allows application of more blood to the same test strip
The DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
Background: Self-monitoring of blood glucose (SMBG) can play a key role in diabetes management. The CONTOUR®NEXT ONE (Ascensia Diabetes Care, Parsippany, NJ, USA) blood glucose monitoring system (BGMS) has been developed for use with CONTOUR®NEXT test strips. The meter connects to the CONTOUR™ Diabetes application on a mobile device, and the Second-Chance® sampling feature recognizes insufficient sample volume, prompting the patient to apply more blood to the same strip if needed (sample re-application).Research design and methods: Two studies evaluated CONTOUR®NEXT ONE BGMS accuracy with sample re-application: a laboratory study (performance under various controlled conditions) and a clinical study (performance in the hands of 52 people with diabetes). A formal International Organization for Standardization (ISO) evaluation has been reported previously; here, accuracy criteria in the laboratory and clinical studies were aligned with ISO 15197:2013 sections 6.3.3 and 8.2, respectively.Results: All results under all conditions in the laboratory study, and with glucose concentrations <100 and ≥100 mg/dL in the clinical study, were within ±15 mg/dL or ±15% of reference measurements. In both studies, all results were within Zone A of the Parkes-Consensus Error Grid.Conclusions: These data demonstrate that the accuracy of the CONTOUR®NEXT ONE BGMS was maintained with sample re-application.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Certain meter system characteristics, such as the use of lancing devices that are less painful (21) and the ability to reapply blood to a strip with an insufficient initial sample, or meters with integrated speech that can read aloud glucose levels for visually impaired individuals (22), may also be beneficial to people with diabetes (23) and may make BGM less burdensome to perform.”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [23]
Citations
Cites
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620 · cited 1×
“In agreement with the laboratory testing, the Parkes-Consensus Error Grid stipulated a minimal level of error, indicative of no effect on clinical action [17].”
Hou, C., Carter, B., Hewitt, J., Francisa, T., & Mayor, S. (2016) · Diabetes Care · doi:10.2337/dc16-0346 · Cited by 411 · cited 1×
“HbA1c compared with controls, suggesting some apps confer a glycemic control benefit to patients [7].”
BackgroundGlucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406 · cited 1×
“que) can affect the accuracy of BGMS measurements [9], and it has been estimated that over 90% of BGMS errors are operator-dependent [10].”
BackgroundContent Analysis of Mobile Health Applications on Diabetes Mellitus
Izahar, S., Lean, Q. Y., Hameed, M. A., Murugiah, M. K., Patel, R. P., Al-Worafi, Y. M., Wong, T. W., & Ming, L. C. (2017) · Frontiers in Endocrinology · doi:10.3389/fendo.2017.00318 · Cited by 370 · cited 1×
“education/advice) have been developed to facilitate diabetes self-management [6].”
BackgroundFactors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348 · cited 1×
“que) can affect the accuracy of BGMS measurements [9], and it has been estimated that over 90% of BGMS errors are operator-dependent [10].”
BackgroundDURÁN, A., MARTÍN, P., RUNKLE, I., PÉREZ, N., ABAD, R., FERNÁNDEZ, M., DEL VALLE, L., SANZ, M. F., & CALLE‐PASCUAL, A. L. (2010) · Journal of Diabetes · doi:10.1111/j.1753-0407.2010.00081.x · Cited by 156 · cited 1×
“outcomes in insulin-naïve patients [2,3,20,21].”
Mannucci, E., Antenore, A., Giorgino, F., & Scavini, M. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817719290 · Cited by 90 · cited 3×
“Structured SMBG in patients with type 2 diabetes (T2D) can provide useful information on the impact of behavior (i.e. physical activity and diet) and treatment, and has been associated with lower HbA1c [8] and improved quality of life [1] versus the use of unstructured glucose testing.”
BackgroundPolonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Axel-Schweitzer, M., Petersen, B., & Wagner, R. S. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0073 · Cited by 85 · cited 1×
“outcomes in insulin-naïve patients [2,3,20,21].”
Structured self‐monitoring of blood glucose reduces glycated hemoglobin in insulin‐treated diabetes
Kato, N., Cui, J., & Kato, M. (2013) · Journal of Diabetes Investigation · doi:10.1111/jdi.12072 · Cited by 36 · cited 1×
“outcomes in insulin-naïve patients [2,3,20,21].”
Pfützner, A., Schipper, C., Ramljak, S., Flacke, F., Sieber, J., Forst, T. M., & Musholt, P. B. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700612 · Cited by 32 · cited 2×
“able BGMs from 2013 confirmed application of insufficient sample volume may cause BGMs to generate incorrect blood glucose results, and in some cases, without triggering an error message on the meter to alert the individual that insufficient sample volume had been applied [11].”
BackgroundChristiansen, M. P., Greene, C., Pardo, S. A., Warchal-Windham, M. E., Harrison, B., Morin, R., & Bailey, T. S. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817691301 · Cited by 31 · cited 2×
“Previous data demonstrate the CONTOUR®NEXT ONE BGMS meets ISO 15197:2013 requirements for BGMS for self-testing in managing diabetes mellitus [13].”
MethodSelf-citationFreckmann, G., Jendrike, N. R. M., Baumstark, A., Pleus, S., Liebing, C., & Haug, C. (2018) · Diabetes Therapy · doi:10.1007/s13300-018-0392-6 · Cited by 25
Nishimura, A., Harashima, S.-I., Fujita, Y., Tanaka, D., Wang, Y., Liu, Y., & Inagaki, N. (2017) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2016.08.019 · Cited by 18 · cited 1×
“outcomes in insulin-naïve patients [2,3,20,21].”
Gao, L., Ji, L., Su, Q., Feng, B., Shan, Z., Hu, R., Xing, X., & Xue, Y. (2015) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2015.08.004 · Cited by 17 · cited 1×
“diet) and treatment, and has been associated with lower HbA1c [8] and improved quality of life [1] versus the use of unstructured glucose testing.”
BackgroundUsing Radar Plots to Demonstrate the Accuracy and Precision of 6 Blood Glucose Monitoring Systems
Pardo, S. A., Dunne, N., & Simmons, D. A. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817713026 · Cited by 14 · cited 1×
“A radar plot (a relatively new method for the communication of BGMS results that intuitively resembles a target, with proximity to the center indicative of level of BGMS accuracy, and degree of dispersal indicative of level of precision [18,19]) was generated using the mean absolute difference and the mean absolute re…”
MethodIn This Issue of Diabetes Care
(2018) · Diabetes Care · doi:10.2337/dc19-ti01 · Cited by 12
Use of Radar Plots to Compare the Analytical Performance of Five Blood Glucose Monitoring Systems
Pardo, S. A., Dunne, N., & Simmons, D. A. (2019) · Therapeutic Delivery · doi:10.4155/tde-2019-0046 · Cited by 1 · cited 1×
“A radar plot (a relatively new method for the communication of BGMS results that intuitively resembles a target, with proximity to the center indicative of level of BGMS accuracy, and degree of dispersal indicative of level of precision [18,19]) was generated using the mean absolute difference and the mean absolute re…”
Method
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
7. Diabetes Technology: Standards of Care in Diabetes—2026
Bajaj, M. S., McCoy, R. G., Balapattabi, K., Bannuru, R. R., Bellini, N. J., Bennett, A. K., Beverly, E. A., Early, K. B., ChallaSivaKanaka, S., Echouffo-Tcheugui, J. B., et al. (2025) · Diabetes Care · doi:10.2337/dc26-s007 · Cited by 293
Metadata
- Authors
- Harrison, B., & Brown, D.
- Affiliations
- Sensor Technology, Ascensia Diabetes Care, Valhalla, NY, USA
- Corresponding author
- Brown, Daniel
- Venue
- Expert Review of Medical Devices
- Publisher
- Informa UK Limited
- Published
- 2020-01-02
- Volume
- 17
- Issue
- 1
- Pages
- 75-82
- ISSN
- 1743-4440 (ISSN-L)
- 1745-2422
- OpenAlex
- W2994924405
- Cited by
- 12
- References
- 18
- Citations by year
- 2025: 1 · 2024: 2 · 2023: 2 · 2022: 2 · 2021: 2 · 2020: 3
- FWCI
- 1.42
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Expert Review of Medical Devices (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- self-monitoring of blood glucose, blood glucose meter accuracy, diabetes management, ISO 15197
- MeSH terms
- Blood Glucose, Blood Specimen Collection, Blood Volume, Diabetes Mellitus, Humans, Regression Analysis, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.