Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study

Open access
Lamont, C, Grego, T, Nanbakhsh, K, Shah Idil, A, Giagka, V, Vanhoestenberghe, A, Cogan, S, Donaldson, N · 2021 · Journal of Neural Engineering

Abstract

Objective. Ensuring the longevity of implantable devices is critical for their clinical usefulness. This is commonly achieved by hermetically sealing the sensitive electronics in a water impermeable housing, however, this method limits miniaturisation. Alternatively, silicone encapsulation has demonstrated long-term protection of implanted thick-film electronic devices. However, much of the current conformal packaging research is focused on more rigid coatings, such as parylene, liquid crystal polymers and novel inorganic layers. Here, we consider the potential of silicone to protect implants using thin-film technology with features 33 times smaller than thick-film counterparts. Approach. Aluminium interdigitated comb structures under plasma-enhanced chemical vapour deposited passivation (SiO x , SiO x N y , SiO x N y + SiC) were encapsulated in medical grade silicones, with a total of six passivation/silicone combinations. Samples were aged in phosphate-buffered saline at 67 ∘C for up to 694 days under a continuous ±5 V biphasic waveform. Periodic electrochemical impedance spectroscopy measurements monitored for leakage currents and degradation of the metal traces. Fourier-transform infrared spectroscopy, x-ray photoelectron spectroscopy, focused-ion-beam and scanning-electron- microscopy were employed to determine any encapsulation material changes. Main results. No silicone delamination, passivation dissolution, or metal corrosion was observed during ageing. Impedances greater than 100 GΩ were maintained between the aluminium tracks for silicone encapsulation over SiO x N y and SiC passivations. For these samples the only observed failure mode was open-circuit wire bonds. In contrast, progressive hydration of the SiO x caused its resistance to decrease by an order of magnitude. Significance. These results demonstrate silicone encapsulation offers excellent protection to thin-film conducting tracks when combined with appropriate inorganic thin films. This conclusion corresponds to previous reliability studies of silicone encapsulation in aqueous environments, but with a larger sample size. Therefore, we believe silicone encapsulation to be a realistic means of providing long-term protection for the circuits of implanted electronic medical devices.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/1741-2552/abf0d6

Cites

67 of 67 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1. (a) An encapsulated IDC sample soldered to a polyether ether ketone (PEEK) bung with depth electrodes. (b) Diced IDC after fabrication with bond pad electrodes labelled as counter electrode (CE), shield (SH), and working electrode (WE) [27]. (c) IDC adhered and wire bonded to the adapter, with MED-1000 coated over the wire bonds and the entire structure encapsulated in MED-6015. (d) Manufactured IDCs immersed in PBS in borosilicate test tubes (Reproduced from [27]. © IOP Publishing Ltd. CC BY 3.0).
Figure 1. (a) An encapsulated IDC sample soldered to a polyether ether ketone (PEEK) bung with depth electrodes. (b) Diced IDC after fabrication with bond pad electrodes labelled as counter electrode (CE), shield (SH), and working electrode (WE) [27]. (c) IDC adhered and wire bonded to the adapter, with MED-1000 coated over the wire bonds and the entire structure encapsulated in MED-6015. (d) Manufactured IDCs immersed in PBS in borosilicate test tubes (Reproduced from [27]. © IOP Publishing Ltd. CC BY 3.0).CC BY · PubMed Central
Figure 2. Impedance magnitude scatter plots and regressions for Bung (n = 3) and Adapter (n = 4) control samples. The different colours represent the four different control samples. The 95% confidence interval and 90% prediction interval is given by the dotted line and grey area, respectively.
Figure 2. Impedance magnitude scatter plots and regressions for Bung (n = 3) and Adapter (n = 4) control samples. The different colours represent the four different control samples. The 95% confidence interval and 90% prediction interval is given by the dotted line and grey area, respectively.CC BY · PubMed Central
Figure 3. Average Bode (a) impedance magnitude and (b) phase plots for each batch’s impedance measurements over the duration of the experiment (± standard deviation shaded in grey). (c) 10 mHz impedance measurements and regressions for samples without WB failures. The different colours represent different IDC samples. Dashed lines and open faced markers represent corrected early measurements (detailed in S6). The 95% confidence interval and 90% prediction interval is given by the dotted line and grey area, respectively. Regressions with a significant change in impedance (p<0.05) are noted.
Figure 3. Average Bode (a) impedance magnitude and (b) phase plots for each batch’s impedance measurements over the duration of the experiment (± standard deviation shaded in grey). (c) 10 mHz impedance measurements and regressions for samples without WB failures. The different colours represent different IDC samples. Dashed lines and open faced markers represent corrected early measurements (detailed in S6). The 95% confidence interval and 90% prediction interval is given by the dotted line and grey area, respectively. Regressions with a significant change in impedance (p<0.05) are noted.CC BY · PubMed Central
Figure 4. (a) High and (b) low wavenumber FTIR measurements of representative IDCs from batches C, D and F, stripped of silicone after ageing. Peaks are normalised with respect to their maximum to show the relative changes to the passivation materials.
Figure 4. (a) High and (b) low wavenumber FTIR measurements of representative IDCs from batches C, D and F, stripped of silicone after ageing. Peaks are normalised with respect to their maximum to show the relative changes to the passivation materials.CC BY · PubMed Central
Figure 5. XPS depth profile of (a) unaged and (b) aged batch C sample (aged 570 days) after decapsulation. SEM+FIB cross sections of SiO x samples at 30 000× magnification (left) and labelled illustration (right) for (c) unaged and (d) aged (570 days) IDCs (batch C, SiO x passivation).
Figure 5. XPS depth profile of (a) unaged and (b) aged batch C sample (aged 570 days) after decapsulation. SEM+FIB cross sections of SiO x samples at 30 000× magnification (left) and labelled illustration (right) for (c) unaged and (d) aged (570 days) IDCs (batch C, SiO x passivation).CC BY · PubMed Central
Figure 6. (a) Cross section with impedances pathways of IDC with hydrated passivation. (b) Equivalent circuit with PBS grounded (Bridged-T model). (c) Fits of the C || (R+C) and Bridged-T model to average batch C measurements (values given in table 2). Models were fitted to the measurements between 10 mHz and 100 Hz.
Figure 6. (a) Cross section with impedances pathways of IDC with hydrated passivation. (b) Equivalent circuit with PBS grounded (Bridged-T model). (c) Fits of the C || (R+C) and Bridged-T model to average batch C measurements (values given in table 2). Models were fitted to the measurements between 10 mHz and 100 Hz.CC BY · PubMed Central
Figure 7. Diagram showing the progressive hydration of the SiOx layer. The capacitance between the metal track and the conductive layer increases as the unhydrated layer gets thinner.
Figure 7. Diagram showing the progressive hydration of the SiOx layer. The capacitance between the metal track and the conductive layer increases as the unhydrated layer gets thinner.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 68 references

Cited by

See all 28

Metadata

Authors
Lamont, C., Grego, T., Nanbakhsh, K., Shah Idil, A., Giagka, V, Vanhoestenberghe, A., Cogan, S., & Donaldson, N.
Affiliations
  • University College London (GB)
  • Delft University of Technology (NL)
  • Fraunhofer Institute for Reliability and Microintegration (DE)
  • Royal National Orthopaedic Hospital (GB)
  • The University of Texas at Dallas (US)
Corresponding author
Lamont, C
Venue
Journal of Neural Engineering
Published
2021-04-06
Volume
18
Issue
5
Pages
055003
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W3139338037
Cited by
28
References
68
Citations by year
2026: 2 · 2025: 10 · 2024: 4 · 2023: 3 · 2022: 5 · 2021: 4
FWCI
1.84
Type
journal article
Language
en
Open access
Licence
Locations
  • Journal of Neural Engineering (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion, other-oa)
  • Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
  • Research Repository (Delft University of Technology) (submittedVersion)
  • Research Repository (Delft University of Technology) (submittedVersion, cc-by)
  • Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
Funders
  • Wellcome Trust (102037)
  • Electronic Components and Systems for European Leadership (Ecsel- 783132-Position-II-2017-IA)
  • Engineering and Physical Sciences Research Council (NS/A000026/1)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Electronic Packaging and Soldering Technologies
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Semiconductor materials and devices
Keywords
silicone encapsulation, thin-film electronics, biomedical implants, plasma-enhanced chemical vapor deposition, long-term aging, phosphate buffered saline, leakage current, metal corrosion
MeSH terms
Electronics, Silicones, Reproducibility of Results, Prostheses and Implants, Coated Materials, Biocompatible

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study. https://app.cochat.ai/papers/10.1088/1741-2552/abf0d6
The paper
Lamont, C., Grego, T., Nanbakhsh, K., Shah Idil, A., Giagka, V, Vanhoestenberghe, A., Cogan, S., & Donaldson, N. (2021). Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/abf0d6

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.