Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study
Open accessAbstract
Objective. Ensuring the longevity of implantable devices is critical for their clinical usefulness. This is commonly achieved by hermetically sealing the sensitive electronics in a water impermeable housing, however, this method limits miniaturisation. Alternatively, silicone encapsulation has demonstrated long-term protection of implanted thick-film electronic devices. However, much of the current conformal packaging research is focused on more rigid coatings, such as parylene, liquid crystal polymers and novel inorganic layers. Here, we consider the potential of silicone to protect implants using thin-film technology with features 33 times smaller than thick-film counterparts. Approach. Aluminium interdigitated comb structures under plasma-enhanced chemical vapour deposited passivation (SiO x , SiO x N y , SiO x N y + SiC) were encapsulated in medical grade silicones, with a total of six passivation/silicone combinations. Samples were aged in phosphate-buffered saline at 67 ∘C for up to 694 days under a continuous ±5 V biphasic waveform. Periodic electrochemical impedance spectroscopy measurements monitored for leakage currents and degradation of the metal traces. Fourier-transform infrared spectroscopy, x-ray photoelectron spectroscopy, focused-ion-beam and scanning-electron- microscopy were employed to determine any encapsulation material changes. Main results. No silicone delamination, passivation dissolution, or metal corrosion was observed during ageing. Impedances greater than 100 GΩ were maintained between the aluminium tracks for silicone encapsulation over SiO x N y and SiC passivations. For these samples the only observed failure mode was open-circuit wire bonds. In contrast, progressive hydration of the SiO x caused its resistance to decrease by an order of magnitude. Significance. These results demonstrate silicone encapsulation offers excellent protection to thin-film conducting tracks when combined with appropriate inorganic thin films. This conclusion corresponds to previous reliability studies of silicone encapsulation in aqueous environments, but with a larger sample size. Therefore, we believe silicone encapsulation to be a realistic means of providing long-term protection for the circuits of implanted electronic medical devices.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures
![Figure 1. (a) An encapsulated IDC sample soldered to a polyether ether ketone (PEEK) bung with depth electrodes. (b) Diced IDC after fabrication with bond pad electrodes labelled as counter electrode (CE), shield (SH), and working electrode (WE) [27]. (c) IDC adhered and wire bonded to the adapter, with MED-1000 coated over the wire bonds and the entire structure encapsulated in MED-6015. (d) Manufactured IDCs immersed in PBS in borosilicate test tubes (Reproduced from [27]. © IOP Publishing Ltd. CC BY 3.0).](/paper-figures/2e5007fab8664296841ebaf6ea82722f/silicone-encapsulation-of-thin-film-sio-sub-x-sub-sio-sub-x-sub-n-sub-y-sub-and-figure-1.jpg)






Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Implant packaging methods to provide a biocompatiblity and corrosion protection through hermetic packages and coatings and non-hermetic encapsulants aim to offer device lifetimes of years to decades [213], [214], [215].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [213]
In the citing paper
“It is not yet clear whether approaches involving softer penetrating electrodes offer a substantially different tradeoff 20,59 .”
From The Layer 7 Cortical Interface: A Scalable and Minimally Invasive Brain–Computer Interface Platform (2022) · cites this paper as [59]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Statsmodels: Econometric and Statistical Modeling with Python
Seabold, S., & Perktold, J. (2010) · Proceedings of the Python in Science Conference · doi:10.25080/majora-92bf1922-011 · Cited by 6,465
Electrical conductivity of silica gel in the presence of adsorbed water
Anderson, J. H., & Parks, G. A. (1968) · The Journal of Physical Chemistry · doi:10.1021/j100856a051 · Cited by 461
Piech, D. K., Johnson, B. C., Shen, K., Ghanbari, M. M., Li, K. Y., Neely, R. M., Kay, J. E., Carmena, J. M., Maharbiz, M. M., & Muller, R. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-0518-9 · Cited by 442
Ay, F., & Aydinli, A. (2004) · Optical Materials · doi:10.1016/j.optmat.2003.12.004 · Cited by 271
Dissolution Behaviors and Applications of Silicon Oxides and Nitrides in Transient Electronics
Kang, S., Hwang, S., Cheng, H., Yu, S., Kim, B. H., Kim, J., Huang, Y., & Rogers, J. A. (2014) · Advanced Functional Materials · doi:10.1002/adfm.201304293 · Cited by 258
BION ™ system for distributed neural prosthetic interfaces
Loeb, G. E., Peck, R. A., Moore, W. H., & Hood, K. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00011-x · Cited by 243
Long-term implants of Parylene-C coated microelectrodes
Schmidt, E. M., Mcintosh, J. S., & Bak, M. J. (1988) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441836 · Cited by 234
Magnetoelectric materials for miniature, wireless neural stimulation at therapeutic frequencies
Singer, A., Dutta, S., Lewis, E., Chen, Z., Chen, J. C., Verma, N., Avants, B., Feldman, A. K., O’Malley, J., Beierlein, M., et al. (2020) · Neuron · doi:10.1016/j.neuron.2020.05.019 · Cited by 202
Plasma‐enhanced chemical vapor deposited silicon carbide as an implantable dielectric coating
Cogan, S. F., Edell, D. J., Guzelian, A. A., Ping Liu, Y., & Edell, R. (2003) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.10152 · Cited by 169
Self-citationBarrese, J. C., Aceros, J., & Donoghue, J. P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/026003 · Cited by 161
Cited by
Ultra‐Thin Flexible Encapsulating Materials for Soft Bio‐Integrated Electronics
Sang, M., Kim, K., Shin, J., & Yu, K. J. (2022) · Advanced Science · doi:10.1002/advs.202202980 · Cited by 161
Closed-loop optogenetic control of the dynamics of neural activity in non-human primates
Zaaimi, B., Turnbull, M., Hazra, A., Wang, Y., Gándara, C., McLeod, F., McDermott, E. E., Escobedo‐Cousin, E., Idil, A. S., Bailey, R. G., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00945-8 · Cited by 72
Self-citationThanjavur, N., Laxmi, B., & Kim, Y. (2025) · Biosensors · doi:10.3390/bios15010038 · Cited by 63
Nanbakhsh, K., Idil, A. S., Lamont, C. A., Dücső, C., Akgun, Ö. C., Horváth, D., Tóth, K., Meszéna, D., Ulbert, I., Mazza, F., et al. (2025) · Nature Communications · doi:10.1038/s41467-024-55298-4 · Cited by 46
Self-citationThe Layer 7 Cortical Interface: A Scalable and Minimally Invasive Brain–Computer Interface Platform
Hettick, M., Ho, E., Poole, A. J., Monge, M., Papageorgiou, D., Takahashi, K., LaMarca, M., Trietsch, D., Reed, K., Murphy, M., et al. (2022) · doi:10.1101/2022.01.02.474656 · Cited by 31
Metadata
- Authors
- Lamont, C., Grego, T., Nanbakhsh, K., Shah Idil, A., Giagka, V, Vanhoestenberghe, A., Cogan, S., & Donaldson, N.
- Affiliations
- University College London (GB)
- Delft University of Technology (NL)
- Fraunhofer Institute for Reliability and Microintegration (DE)
- Royal National Orthopaedic Hospital (GB)
- The University of Texas at Dallas (US)
- Corresponding author
- Lamont, C
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2021-04-06
- Volume
- 18
- Issue
- 5
- Pages
- 055003
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC8208634
- OpenAlex
- W3139338037
- Cited by
- 28
- References
- 68
- Citations by year
- 2026: 2 · 2025: 10 · 2024: 4 · 2023: 3 · 2022: 5 · 2021: 4
- FWCI
- 1.84
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (vor)
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- Locations
- Journal of Neural Engineering (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion, other-oa)
- Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
- Research Repository (Delft University of Technology) (submittedVersion)
- Research Repository (Delft University of Technology) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Funders
- Wellcome Trust (102037)
- Electronic Components and Systems for European Leadership (Ecsel- 783132-Position-II-2017-IA)
- Engineering and Physical Sciences Research Council (NS/A000026/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electronic Packaging and Soldering Technologies
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Semiconductor materials and devices
- Keywords
- silicone encapsulation, thin-film electronics, biomedical implants, plasma-enhanced chemical vapor deposition, long-term aging, phosphate buffered saline, leakage current, metal corrosion
- MeSH terms
- Electronics, Silicones, Reproducibility of Results, Prostheses and Implants, Coated Materials, Biocompatible
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study. https://app.cochat.ai/papers/10.1088/1741-2552/abf0d6
- The paper
- Lamont, C., Grego, T., Nanbakhsh, K., Shah Idil, A., Giagka, V, Vanhoestenberghe, A., Cogan, S., & Donaldson, N. (2021). Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/abf0d6
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.