Silicone encapsulation of thin-film SiO x , SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study
Open accessThe DOI matches its Crossref record; 67 of the 68 references it cites were found in a registry or an index.
Abstract
Objective. Ensuring the longevity of implantable devices is critical for their clinical usefulness. This is commonly achieved by hermetically sealing the sensitive electronics in a water impermeable housing, however, this method limits miniaturisation. Alternatively, silicone encapsulation has demonstrated long-term protection of implanted thick-film electronic devices. However, much of the current conformal packaging research is focused on more rigid coatings, such as parylene, liquid crystal polymers and novel inorganic layers. Here, we consider the potential of silicone to protect implants using thin-film technology with features 33 times smaller than thick-film counterparts. Approach. Aluminium interdigitated comb structures under plasma-enhanced chemical vapour deposited passivation (SiO x , SiO x N y , SiO x N y + SiC) were encapsulated in medical grade silicones, with a total of six passivation/silicone combinations. Samples were aged in phosphate-buffered saline at 67 ∘C for up to 694 days under a continuous ±5 V biphasic waveform. Periodic electrochemical impedance spectroscopy measurements monitored for leakage currents and degradation of the metal traces. Fourier-transform infrared spectroscopy, x-ray photoelectron spectroscopy, focused-ion-beam and scanning-electron- microscopy were employed to determine any encapsulation material changes. Main results. No silicone delamination, passivation dissolution, or metal corrosion was observed during ageing. Impedances greater than 100 GΩ were maintained between the aluminium tracks for silicone encapsulation over SiO x N y and SiC passivations. For these samples the only observed failure mode was open-circuit wire bonds. In contrast, progressive hydration of the SiO x caused its resistance to decrease by an order of magnitude. Significance. These results demonstrate silicone encapsulation offers excellent protection to thin-film conducting tracks when combined with appropriate inorganic thin films. This conclusion corresponds to previous reliability studies of silicone encapsulation in aqueous environments, but with a larger sample size. Therefore, we believe silicone encapsulation to be a realistic means of providing long-term protection for the circuits of implanted electronic medical devices.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures
![Figure 1. (a) An encapsulated IDC sample soldered to a polyether ether ketone (PEEK) bung with depth electrodes. (b) Diced IDC after fabrication with bond pad electrodes labelled as counter electrode (CE), shield (SH), and working electrode (WE) [27]. (c) IDC adhered and wire bonded to the adapter, with MED-1000 coated over the wire bonds and the entire structure encapsulated in MED-6015. (d) Manufactured IDCs immersed in PBS in borosilicate test tubes (Reproduced from [27]. © IOP Publishing Ltd. CC BY 3.0).](https://figures.cochat.ai/paper-figures/2e5007fab8664296841ebaf6ea82722f/silicone-encapsulation-of-thin-film-sio-sub-x-sub-sio-sub-x-sub-n-sub-y-sub-and-figure-1.jpg)






Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Implant packaging methods to provide a biocompatiblity and corrosion protection through hermetic packages and coatings and non-hermetic encapsulants aim to offer device lifetimes of years to decades [213], [214], [215].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [213]
In the citing paper
“It is not yet clear whether approaches involving softer penetrating electrodes offer a substantially different tradeoff 20,59 .”
From The Layer 7 Cortical Interface: A Scalable and Minimally Invasive Brain–Computer Interface Platform (2022) · cites this paper as [59]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Statsmodels: Econometric and Statistical Modeling with Python
Seabold, S., & Perktold, J. (2010) · Proceedings of the Python in Science Conference · doi:10.25080/majora-92bf1922-011 · Cited by 6,465 · cited 1×
“The OLR was calculated using the Python Statsmodels library [30] on the logarithm (base 10) of the impedance measurements (y= 10mx+ c).”
MethodElectrical conductivity of silica gel in the presence of adsorbed water
Anderson, J. H., & Parks, G. A. (1968) · The Journal of Physical Chemistry · doi:10.1021/j100856a051 · Cited by 461 · cited 1×
“Dissociation of Si–OH and adsorbed water form mobile protons (H3O) that facilitate surface conduction through the porous network of the material [42, 43].”
BackgroundPiech, D. K., Johnson, B. C., Shen, K., Ghanbari, M. M., Li, K. Y., Neely, R. M., Kay, J. E., Carmena, J. M., Maharbiz, M. M., & Muller, R. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-0518-9 · Cited by 442
Ay, F., & Aydinli, A. (2004) · Optical Materials · doi:10.1016/j.optmat.2003.12.004 · Cited by 271 · cited 1×
“This was attributed to the use of N2O as a gas reactant and N2 as a carrier gas during deposition, and was similarly observed by Ay and Aydinli [35].”
ResultDissolution Behaviors and Applications of Silicon Oxides and Nitrides in Transient Electronics
Kang, S., Hwang, S., Cheng, H., Yu, S., Kim, B. H., Kim, J., Huang, Y., & Rogers, J. A. (2014) · Advanced Functional Materials · doi:10.1002/adfm.201304293 · Cited by 258
BION ™ system for distributed neural prosthetic interfaces
Loeb, G. E., Peck, R. A., Moore, W. H., & Hood, K. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00011-x · Cited by 243
Long-term implants of Parylene-C coated microelectrodes
Schmidt, E. M., Mcintosh, J. S., & Bak, M. J. (1988) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441836 · Cited by 234
Magnetoelectric materials for miniature, wireless neural stimulation at therapeutic frequencies
Singer, A., Dutta, S., Lewis, E., Chen, Z., Chen, J. C., Verma, N., Avants, B., Feldman, A. K., O’Malley, J., Beierlein, M., et al. (2020) · Neuron · doi:10.1016/j.neuron.2020.05.019 · Cited by 202
Plasma‐enhanced chemical vapor deposited silicon carbide as an implantable dielectric coating
Cogan, S. F., Edell, D. J., Guzelian, A. A., Ping Liu, Y., & Edell, R. (2003) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.10152 · Cited by 169 · cited 2×
“The stability of SiC in aqueous environments is illustrated by the ageing studies of Cogan et al [25] and Hsu et al [26].”
BackgroundMethodSelf-citationBarrese, J. C., Aceros, J., & Donoghue, J. P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/026003 · Cited by 161
Corrosion of silicon integrated circuits and lifetime predictions in implantable electronic devices
Vanhoestenberghe, A., & Donaldson, N. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/031002 · Cited by 140 · cited 2×
“Extensive characterisation of microelectronics encapsulation confirms surface cleanliness to be a primary factor affecting encapsulant adhesion and device reliability [36].”
BackgroundSelf-citationPlasma cleaning of metal surfaces
O'Kane, D. F., & Mittal, K. L. (1974) · Journal of Vacuum Science and Technology · doi:10.1116/1.1318069 · Cited by 125 · cited 1×
“Even mild processes (Ar at 40 W for 1 min) have been shown to dramatically reduce carbon contamination on metal surfaces [58].”
BackgroundMinnikanti, S., Diao, G., Pancrazio, J. J., Xie, X., Rieth, L. W., Solzbacher, F., & Peixoto, N. (2013) · Acta Biomaterialia · doi:10.1016/j.actbio.2013.10.031 · Cited by 114
Emerging Encapsulation Technologies for Long-Term Reliability of Microfabricated Implantable Devices
Ahn, S.-H., Jeong, J., & Kim, S. J. (2019) · Micromachines · doi:10.3390/mi10080508 · Cited by 107 · cited 1×
“A useful comparison to our results is given by a recent review by Ahn et al [4] of over 21 encapsulation reliability studies.”
ResultInterlayer Dielectrics for Semiconductor Technologies
(2003) · Elsevier eBooks · doi:10.1016/b978-0-12-511221-5.x5000-4 · Cited by 106
Hsu, J.-M., Tathireddy, P., Rieth, L. W., Normann, A. R., & Solzbacher, F. (2007) · Thin Solid Films · doi:10.1016/j.tsf.2007.04.050 · Cited by 104 · cited 1×
“The stability of SiC in aqueous environments is illustrated by the ageing studies of Cogan et al [25] and Hsu et al [26].”
MethodCaldwell, R., Street, M. G., Sharma, R., Takmakov, P., Baker, B., & Rieth, L. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2019.119731 · Cited by 100
Jeong, J., Laiwalla, F., Lee, J., Ritasalo, R., Pudas, M., Larson, L., Leung, V., & Nurmikko, A. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201806440 · Cited by 96
Plasma‐deposited passivation layers for moisture and water protection
Vogt, M., & Hauptmann, R. (1995) · Surface and Coatings Technology · doi:10.1016/0257-8972(95)08268-9 · Cited by 92
The first 500 sacral anterior root stimulators: implant failures and their repair
Brindley, G. S. (1995) · Spinal Cord · doi:10.1038/sc.1995.3 · Cited by 89 · cited 2×
“As an alternative, silicone encapsulation has previously demonstrated longterm protection of implanted thick-film electronic devices [3].”
BackgroundNonhermeticity of Polymeric Lid Sealants
Traeger, R. (1977) · IEEE Transactions on Parts, Hybrids, and Packaging · doi:10.1109/tphp.1977.1135193 · Cited by 89 · cited 1×
“It is not obvious that a material as waterpermeable as silicone would be useful but, for the years in which medical implants must remain functional in the body, all polymers are permeable towater vapour [5].”
BackgroundInfluence of absorbed water components on SiOCH low-k reliability
Li, Y., Ciofi, I., Carbonell, L., Heylen, N., Aelst, J. V., Baklanov, M. R., Groeseneken, G., Maex, K., & Tökei, Z. (2008) · Journal of Applied Physics · doi:10.1063/1.2966578 · Cited by 85
Xie, X., Rieth, L., Caldwell, R., Diwekar, M., Tathireddy, P., Sharma, R., & Solzbacher, F. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2266542 · Cited by 81 · cited 1×
“The most promising results presented were from multi-layer coatings that incorporated atomic layer deposited (ALD) aluminium oxide (Al2O3) or hafnium oxide (HfO2), neither of which demonstrated a change in insulating properties when aged at temperatures above 80 ◦C in PBS for 6 months [46, 47].”
BackgroundEncapsulation of integrated circuits
White, M. L. (1969) · Proceedings of the IEEE · doi:10.1109/proc.1969.7344 · Cited by 78 · cited 1×
“What is critical is that the encapsulant remains adhesively-bonded to the surfaces between the conductors so that there is no void into which the vapour can condense to form a conductive liquid film [6].”
BackgroundDaves, W., Krauß, A., Behnel, N., Häublein, V., Bauer, A. J., & Frey, L. (2011) · Thin Solid Films · doi:10.1016/j.tsf.2011.02.089 · Cited by 75 · cited 1×
“0 [29] to correct the ATR measurements.”
An in situ XPS study of oxygen plasma cleaning of aluminum surfaces
Li, H., Belkind, A., Jansen, F., & Orban, Z. (1997) · Surface and Coatings Technology · doi:10.1016/s0257-8972(97)00079-0 · Cited by 75
Aspects of silicone rubber as an encapsulant for neurological prostheses: Part 1 Osmosis
Donaldson, P. E. K. (1991) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446293 · Cited by 59
Biodurability of Retrieved Silicone Gel Breast Implants
Brandon, H. J., Jerina, K. L., Wolf, C. J., & Young, V. L. (2003) · Plastic & Reconstructive Surgery · doi:10.1097/01.prs.0000060795.16982.1c · Cited by 49
Compensation effect during water desorption from siloxane-based spin-on dielectric thin films
Proost, J., Baklanov, M. R., Maex, K., & Delaey, L. (2000) · Journal of Vacuum Science & Technology B Microelectronics and Nanometer Structures Processing Measurement and Phenomena · doi:10.1116/1.591189 · Cited by 46 · cited 1×
“Temperatures between 190 ◦C– 355 ◦C are required to remove surface bound water [44, 45].”
BackgroundWater-induced room-temperature oxidation of Si–H and –Si–Si– bonds in silicon oxide
Liao, W., & Lee, S. (1996) · Journal of Applied Physics · doi:10.1063/1.362915 · Cited by 46
The Limits of Hermeticity Test Methods for Micropackages
Vanhoestenberghe, A., & Donaldson, N. (2011) · Artificial Organs · doi:10.1111/j.1525-1594.2011.01222.x · Cited by 39
Self-citationBalde, J. W. (1991) · IEEE Transactions on Components Hybrids and Manufacturing Technology · doi:10.1109/33.87316 · Cited by 39
Packaging of Implantable Microsystems
Najafi, K. (2007) · 2007 IEEE Sensors · doi:10.1109/icsens.2007.4388335 · Cited by 38 · cited 1×
“Very small package sizes are beyond the sensitivity of spectrometers to validate hermeticity and increasing the number of conductors and feedthroughs in the package is non-trivial [1, 2].”
BackgroundWater‐Induced Corrosion of Materials Used for Semiconductor Passivation
Osenbach, J. W. (1993) · Journal of The Electrochemical Society · doi:10.1149/1.2221147 · Cited by 37 · cited 2×
“of tracks in buried passivations which has also been observed during reliability studies of microelectronics [20].”
BackgroundResultSuaning, G. J., Schuettler, M., Ordonez, J. S., & Lovell, N. H. (2007) · 2007 3rd International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/cne.2007.369597 · Cited by 34 · cited 1×
“Similar to the methods described in [28], we diluted the MED3-4013 pre-polymer in n-Heptane (99%, Fisher Scientific) to improve its viscosity and allow for dip-coating.”
MethodCharacterization of functional relationship between temperature and microelectronic reliability
Lall, P., PECHT, M. G., & Hakim, E. B. (1995) · Microelectronics Reliability · doi:10.1016/0026-2714(95)93067-k · Cited by 33 · cited 1×
“Importantly, both ionic contamination and charge injection of MOSFET gate oxides exhibit an inverse temperature acceleration factor [67].”
BackgroundTesting for MOS IC Failure Modes
Edwards, D. G. (1982) · IEEE Transactions on Reliability · doi:10.1109/tr.1982.5221212 · Cited by 32
Sim, S. P., & Lawson, R. W. (1979) · IEEE International Reliability Physics Symposium proceedings · doi:10.1109/irps.1979.362878 · Cited by 32
Silicone rubber encapsulation for an endoscopically implantable gastrostimulator
Lonys, L., Vanhoestenberghe, A., Julémont, N., Godet, S., Delplancke, M., Mathys, P., & Nonclercq, A. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1236-9 · Cited by 31
Self-citationBehavior of a ‐ SiN : H and a ‐ SiON : H Films in Condensed Water
Osenbach, J., & Knolle, W. R. (1992) · Journal of The Electrochemical Society · doi:10.1149/1.2069076 · Cited by 28 · cited 16×
“Desorption of nitrogen then proceeded by the expected reaction between water and Si–N, releasing ammonia and forming SiO2 [16, 40, 41].”
BackgroundResultInfluentialIannuzzi, M. (1983) · IEEE Transactions on Components Hybrids and Manufacturing Technology · doi:10.1109/tchmt.1983.1136166 · Cited by 28
Donaldson, P. E. K., & Aylett, B. J. (1995) · PubMed · doi:10.1007/bf02510794 · Cited by 24
Aspects of silicone rubber as encapsulant for neurological prostheses Part 4: Two-part rubbers
Donaldson, P. E. K. (1997) · Medical & Biological Engineering & Computing · doi:10.1007/bf02530051 · Cited by 23
The permeability of silicone rubber to metal compounds: Relevance to implanted devices
Donaldson, N., Baviskar, P., Cunningham, J., & Wilson, D. (2012) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.33257 · Cited by 22
Self-citationIannuzzi, M. (1983) · IEEE Transactions on Components Hybrids and Manufacturing Technology · doi:10.1109/tchmt.1983.1136165 · Cited by 22
Apparatus to investigate the insulation impedance and accelerated life-testing of neural interfaces
Donaldson, N., Lamont, C. A., Idil, A. S., Mentink, M., & Perkins, T. A. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aadeac · Cited by 22 · cited 6×
“(b) Diced IDC after fabrication with bond pad electrodes labelled as counter electrode (CE), shield (SH), and working electrode (WE) [27].”
BackgroundMethodSelf-citationDevelopment of an active high-density transverse intrafascicular micro-electrode probe
Verplancke, R., Cauwe, M., Schaubroeck, D., Cuypers, D., Vandecasteele, B., Mader, L., Vanhaverbeke, C., Ballini, M., O’Callaghan, J., Goikoetxea, E., et al. (2020) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ab5df2 · Cited by 21
Lee, H.-I., Park, J.-B., Xianyu, W., Kim, K., Chung, J. G., Kyoung, Y. K., Byun, S., Yang, W. Y., Park, Y. Y., Kim, S. M., et al. (2017) · Scientific Reports · doi:10.1038/s41598-017-14291-2 · Cited by 21 · cited 13×
“Desorption of nitrogen then proceeded by the expected reaction between water and Si–N, releasing ammonia and forming SiO2 [16, 40, 41].”
BackgroundInfluentialNanbakhsh, K., Ritasalo, R., Serdijn, W. A., & Giagka, V. (2020) · IEEE Electronic Components and Technology Conference (ECTC) · doi:10.1109/ectc32862.2020.00081 · Cited by 21
Self-citationEdell, D. J. (2004) · Series on bioengineering and biomedical engineering · doi:10.1142/9789812561763_0016 · Cited by 20
Aspects of silicone rubber as encapsulant for neurological prostheses
Donaldson, P. E. K., & Aylett, B. J. (1995) · Medical & Biological Engineering & Computing · doi:10.1007/bf02510501 · Cited by 19
Improvement of water-related hot-carrier reliability by using ECR plasma-SiO/sub 2/
Machida, K., Shimoyama, N., Takahashi, J., Takahashi, Y., Yabumoto, N., & Arai, E. (1994) · IEEE Transactions on Electron Devices · doi:10.1109/16.285021 · Cited by 17
Haque, M. S., Naseem, H. A., & Brown, W. D. (1997) · Journal of The Electrochemical Society · doi:10.1149/1.1837995 · Cited by 16
Plasma Processing for Silicon Oxynitride Films
Cavallari, C., & Gualandris, F. (1987) · Journal of The Electrochemical Society · doi:10.1149/1.2100655 · Cited by 15 · cited 1×
“SiOxNy can maintain a similarly high resistivity as SiOx and provides a better barrier to water and ions [24].”
BackgroundNanbakhsh, K., Kluba, M., Pahl, B., Bourgeois, F., Dekker, R. H., Serdijn, W. A., & Giagka, V. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8857702 · Cited by 15 · cited 1×
“Amore severe test of the insulation is a continuous or ‘DC’ bias [31].”
BackgroundSelf-citationA Chip Integrity Monitor for Evaluating Moisture/Ion Ingress in mm-Sized Single-Chip Implants
Akgun, Ö. C., Nanbakhsh, K., Giagka, V., & Serdijn, W. A. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.3007484 · Cited by 11 · cited 1×
“The authors have recently introduced a new platform fully integrated on a CMOS IC, which can be used to detect potential ingress of moisture/ions through the ICs metal layers [68].”
BackgroundSelf-citationIannuzzi, M. (1981) · IEEE Transactions on Components Hybrids and Manufacturing Technology · doi:10.1109/tchmt.1981.1135825 · Cited by 10 · cited 3×
“The use of air plasma, as opposed to pure O2 makes it difficult to validate the current process from the results of Iannuzzi [55], however we think the ionic bombardment is a key factor in the plasma cleaning process, which can be achieved with Ar, O2, N2 or air plasma [57].”
BackgroundMethodKuliasha, C. A., & Judy, J. W. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8856692 · Cited by 9 · cited 1×
“Indeed, the diffusion of water through silicone encapsulants has been shown to cause failure of anisotropic conductive adhesives for back-end packages during reactive ageing studies [37].”
BackgroundIon Instabilities in MOS Structures
Kriegler, R. J. (1974) · IEEE International Reliability Physics Symposium proceedings · doi:10.1109/irps.1974.362654 · Cited by 8
Accelerated-stress reliability evaluation for an encapsulated wireless cortical stimulator
Suh, S., Troyk, P. R., & Hu, Z. (2014) · doi:10.1109/embc.2014.6943623 · Cited by 8 · cited 1×
“Despite these concerns, a more recent ageing study of the Intracortical Visual Prosthesis from Illinois Institute of Technology showed no failures of four IC devices encapsulated in Dow 96-083 silicone aged at 121 ◦C for over 91 days [22].”
BackgroundEvaluation of moisture induced hot-carrier degradation of N- and P-MOSFETs using steam stress
Lee, Y.-H., Hui, S., & Sabi, B. (2002) · Technical digest - International Electron Devices Meeting/Technical digest · doi:10.1109/iedm.1993.347296 · Cited by 7
Moisture resistance of insulating films for compound semiconductor devices
Oku, T., Totsuka, M., Shiga, T., Takemi, M., & Okumura, M. (2014) · Cited by 7
The Direct-current Conductivity of Silica Gel
Muroya, M.-A., & Kondo, S. (1974) · Bulletin of the Chemical Society of Japan · doi:10.1246/bcsj.47.1533 · Cited by 5 · cited 1×
“Dissociation of Si–OH and adsorbed water form mobile protons (H3O) that facilitate surface conduction through the porous network of the material [42, 43].”
BackgroundLeung, C., & Ong, E. (2003) · Elsevier eBooks · doi:10.1016/b978-012511221-5/50007-6 · Cited by 4
A Bayesian Demonstration of Reliability for Encapsulated Implanted Electronics
Lamont, C. A., Mazza, F., & Donaldson, N. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717034 · Cited by 4 · cited 1×
“This is further evidenced by previous work from our group, which demonstrated only a single encapsulation failure for 36 foundry-processed IDC structures coated in MED6015 silicone, aged under a constant 5 V DC bias for over 302 days at between 47 ◦Cand 87 ◦C [48].”
ResultSelf-citationRole of Adhesion Phenomenon in the Reliability of Electronic Packaging
Viswanadham, P. (2014) · doi:10.1002/9781118831373.ch6 · Cited by 3 · cited 1×
“The use of air plasma, as opposed to pure O2 makes it difficult to validate the current process from the results of Iannuzzi [55], however we think the ionic bombardment is a key factor in the plasma cleaning process, which can be achieved with Ar, O2, N2 or air plasma [57].”
BackgroundLamont, C. A. (2020) · UCL Discovery (University College London) · Cited by 2
Self-citation
Cited by
Ultra‐Thin Flexible Encapsulating Materials for Soft Bio‐Integrated Electronics
Sang, M., Kim, K., Shin, J., & Yu, K. J. (2022) · Advanced Science · doi:10.1002/advs.202202980 · Cited by 161
Closed-loop optogenetic control of the dynamics of neural activity in non-human primates
Zaaimi, B., Turnbull, M., Hazra, A., Wang, Y., Gándara, C., McLeod, F., McDermott, E. E., Escobedo‐Cousin, E., Idil, A. S., Bailey, R. G., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00945-8 · Cited by 72
Self-citationThanjavur, N., Laxmi, B., & Kim, Y. (2025) · Biosensors · doi:10.3390/bios15010038 · Cited by 63
“Examples of materials used include poly(l-lactide-co-ε -caprolactone) (PLCL), polytetrafluoroethylene (PTFE), and silicon dioxide (SiO 2 ) nanoparticles [105–107].”
Nanbakhsh, K., Idil, A. S., Lamont, C. A., Dücső, C., Akgun, Ö. C., Horváth, D., Tóth, K., Meszéna, D., Ulbert, I., Mazza, F., et al. (2025) · Nature Communications · doi:10.1038/s41467-024-55298-4 · Cited by 49
“Full details of the sample preparation can be found elsewhere 71 .”
Self-citationThe Layer 7 Cortical Interface: A Scalable and Minimally Invasive Brain–Computer Interface Platform
Hettick, M., Ho, E., Poole, A. J., Monge, M., Papageorgiou, D., Takahashi, K., LaMarca, M., Trietsch, D., Reed, K., Murphy, M., et al. (2022) · doi:10.1101/2022.01.02.474656 · Cited by 31
Metadata
- Authors
- Lamont, C., Grego, T., Nanbakhsh, K., Shah Idil, A., Giagka, V, Vanhoestenberghe, A., Cogan, S., & Donaldson, N.
- Affiliations
- University College London (GB)
- Delft University of Technology (NL)
- Fraunhofer Institute for Reliability and Microintegration (DE)
- Royal National Orthopaedic Hospital (GB)
- The University of Texas at Dallas (US)
- Corresponding author
- Lamont, C
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2021-04-06
- Volume
- 18
- Issue
- 5
- Pages
- 055003
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC8208634
- OpenAlex
- W3139338037
- Cited by
- 28
- References
- 68
- Citations by year
- 2026: 2 · 2025: 10 · 2024: 4 · 2023: 3 · 2022: 5 · 2021: 4
- FWCI
- 1.84
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (vor)
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- Locations
- Journal of Neural Engineering (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion, other-oa)
- Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
- Research Repository (Delft University of Technology) (submittedVersion)
- Research Repository (Delft University of Technology) (submittedVersion, cc-by)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Funders
- Wellcome Trust (102037)
- Electronic Components and Systems for European Leadership (Ecsel- 783132-Position-II-2017-IA)
- Engineering and Physical Sciences Research Council (NS/A000026/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electronic Packaging and Soldering Technologies
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Semiconductor materials and devices
- Keywords
- silicone encapsulation, thin-film electronics, biomedical implants, plasma-enhanced chemical vapor deposition, long-term aging, phosphate buffered saline, leakage current, metal corrosion
- MeSH terms
- Electronics, Silicones, Reproducibility of Results, Prostheses and Implants, Coated Materials, Biocompatible
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.