The Layer 7 Cortical Interface: A Scalable and Minimally Invasive Brain–Computer Interface Platform

Open access
Hettick, Mark, Ho, Elton, Poole, Adam J., Monge, Manuel, Papageorgiou, Demetrios, Takahashi, Kazutaka, LaMarca, Morgan, Trietsch, Daniel +13 more·2022·bioRxiv (Cold Spring Harbor Laboratory)

Abstract

Progress toward the development of brain–computer interfaces has signaled the potential to restore, replace, and augment lost or impaired neurological function in a variety of disease states. Existing approaches to developing high-bandwidth brain–computer interfaces rely on invasive surgical procedures or brain-penetrating electrodes, which limit addressable applications of the technology and the number of eligible patients. Here we describe a novel approach to constructing a neural interface, comprising conformable thin-film electrode arrays and a minimally invasive surgical delivery system that together facilitate bidirectional communication with large portions of the cortical surface (enabling both recording and stimulation). We demonstrate the feasibility and safety of delivering reversible implants containing over 2,000 microelectrodes to multiple functional regions in both hemispheres of the brain simultaneously, without requiring a craniotomy or damaging the cortical surface, at an effective insertion rate faster than 40 ms per channel. We further evaluate the performance of this system immediately following implantation for high-density neural recording and visualizing cortical surface activity at spatial and temporal resolutions and extents not previously possible in multiple preclinical large animal studies as well as in a five-patient pilot clinical study involving both anesthetized and awake neurosurgical patients. We characterize the spatial scales at which sensorimotor activity and speech are represented at the cortical surface, demonstrate accurate neural decoding of somatosensory, visual, and volitional walking activity, and achieve precise neuromodulation through cortical stimulation at sub-millimeter scales. The resulting system generates 90 Gb/h of electrophysiologic data, and demonstrates the highly scalable nature of micro-electrocorticography and its utility for next-generation brain-computer interfaces that may expand the patient population that could benefit from neural interface technology.

Integrity signals

Registry

Crossref · 10.1101/2022.01.02.474656

Cites

105 of 105 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 20 cited sources checked · 0 supported · 0 flagged

  • Findingown data

    The authors showed that implants with more than 2,000 microelectrodes could be delivered safely and feasibly to several functional areas in both hemispheres of the Göttingen minipig brain at the same time.

    “We demonstrate the safety and feasibility of rapidly delivering reversible implants containing over 2,000 microelectrodes to multiple functional regions in both hemispheres of the Göttingen minipig brain simultaneously, without requiring a craniotomy, at an effective insertion rate faster than 40 ms per channel, without damaging the cortical surface.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Safety and feasibility in Göttingen minipigs predict safety and feasibility in humans.
  • Findingown data

    After optimization, electrode connectivity yield exceeded 93%, as confirmed by impedance testing of 4,232 electrodes across eight arrays.

    “Process yield was evaluated with respect to electrode connectivity and was confirmed to be >93% after optimization, by impedance characterization of 4,232 electrodes from eight arrays.”
    Quote in the paperSubstantial claim
  • Findingown data

    The pia stayed intact after the array was placed and removed.

    “The pia remained intact following array placement and removal, and we found no macroscopic evidence of tissue disruption at the cortical surface corresponding to array placement.”
    Quote in the paperSubstantial claim
  • Findingown data

    Measurements of platinum electrode impedance showed that impedance magnitude corresponded directly to electrode area.

    “Impedance spectra of the platinum electrodes (shown in the photographs and photomicrographs of Figures 2a and 2b, respectively) verified direct correspondence between electrode area and impedance magnitude (Figure 2b).”
    Quote in the paperNot a substantial claim
  • Findingown data

    Each 529-channel module has an effective area of 0.48 cm2.

    “The effective area of the microelectrode arrays described here is 0.48 cm 2 per 529-channel module, and we have demonstrated that these arrays can be connected in a modular fashion and inserted simultaneously, at an effective rate exceeding 40 s per cm 2 of cortical surface.”
    Quote in the paperSubstantial claim
  • May combine two claims. 35
  • Classified as probably not a substantial claim. 16
  • May combine two claims.

    Findingown data

    For each electrode size, EIS showed the usual rise in cutoff frequency as size decreased, and 1 kHz impedance increased as expected with smaller electrodes.

    “EIS measurements for each electrode size display typical positive shifts in cutoff frequencies with decreases in electrode size, and impedance magnitude at 1 kHz also follows the expected upward trend with decreasing electrode size (Figure 2b).”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    The platform supported dense neural recording, localized cortical stimulation, and accurate decoding of neural signals.

    “We further demonstrate the performance of this system for high-density neural recording, focal cortical stimulation, and accurate neural decoding.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The strongest activation formed a ring about 1 mm in radius around the stimulating electrode, sparing the area immediately next to it.

    “The most highly activated region formed an annulus of radius approximately 1 mm around, but not immediately surrounding the stimulating electrode.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingreasoning

    Each custom amplifier PCB contains eight RHD2164 64-channel amplifier chips and one RHS2116 16-channel stimulator/amplifier chip, enabling simultaneous recording from up to 528 channels and stimulation from up to 16 channels.

    “The custom amplifier printed circuit boards (PCBs) used to interface with the implanted electrode arrays each contained eight of the RHD2164 64-channel amplifier chips and one of the RHS2116 16-channel stimulator/amplifier chips, allowing for simultaneous recording from up to 528 channels and stimulation from up to 16 channels.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The authors defined a safe stimulation regime for thin-film surface microelectrodes, reached visual cortex minimally invasively, confirmed placement over working visual cortex, stimulated it, and recorded evoked activity.

    “Here we characterize a safe operating regime for cortical stimulation from thin-film cortical surface microelectrodes, demonstrate the ability to access visual cortex in a minimally invasive fashion, electrophysiologically confirm array placement over functioning visual cortex, stimulate the cortical surface, and monitor stimulus-evoked cortical activity.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingreasoning

    The platform offers many channels, dense electrode coverage, and two-way function.

    “The system provides high channel counts and high spatial densities of electrode coverage, as well as bidirectional functionality (both neural recording and neural stimulation).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    CV measurement placed the water window at about 1.85 V, with cathodic and anodic limits of 0.65 V and 1.2 V.

    “Briefly, the water window was determined by CV measurement to be approximately 1.85 V (with a 0.65 V cathodic and 1.2 V anodic limit), and biphasic pulses were applied at 50 Hz to a test electrode within the range of anticipated stimulation parameters to confirm safe stimulation (Figure 7a).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Impedance mapping in vivo indicated high electrode yield and uniform impedance.

    “In vivo electrode impedance mapping results confirm high electrode yield and impedance uniformity (Figure 2b, inset).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The authors demonstrated three core functions: recording, decoding, and cortical stimulation.

    “However, certain functions are commonly required of braincomputer interfaces, and we demonstrate three key functions here: neural recording, neural decoding, and cortical stimulation.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Decoding with many electrodes across the whole array outperformed any single electrode, reaching 73% accuracy for the left and 67% for the right in offline analysis.

    “Multichannel decoding using the large number of electrodes available from the entire array was more robust than decoding using any single electrode, with accuracies reaching 73% on the left and 67% on the right in off-line decoding.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The implanted arrays recorded neural activity across many channels, including spontaneous cortical activity, somatosensory evoked responses, and visual evoked responses.

    “Implanted arrays were used for multichannel neural recording using several paradigms, capturing both spontaneous cortical electrographic activity (Figure 5b) as well as somatosensory evoked potentials (Figure 5e) and visual evoked potentials (Figure 5f).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The arrays can be linked modularly and inserted together at an effective rate above 40 s per cm2 of cortex.

    “The effective area of the microelectrode arrays described here is 0.48 cm 2 per 529-channel module, and we have demonstrated that these arrays can be connected in a modular fashion and inserted simultaneously, at an effective rate exceeding 40 s per cm 2 of cortical surface.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The team inserted double modules in vivo (1,058 channels over 0.96 cm2) and quadruple modules in vitro (2,116 channels each over 1.92 cm2).

    “We performed in vivo insertions of doubly connected modules (1,058 channels over 0.96 cm 2 of cortical surface area), and in vitro insertions of quadruply connected modules (2,116 channels each covering 1.92 cm 2 of surface area).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    A double module halved the effective average insertion rate, and in vivo speeds reached under 20 ms per electrode.

    “We note that use of a doubly connected module halves the effective average insertion rate, and in this manner have achieved effective speeds as fast as <20 ms per electrode in vivo.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The authors showed that high-density microelectrode arrays could be inserted safely and feasibly with a new minimally invasive cranial micro-slit method.

    “We have demonstrated the safety and feasibility of inserting our high-density microelectrode arrays using a novel, minimally invasive “cranial micro-slit” technique (Figure 3).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Delivering 50,000 pulses at 10 nC per phase and 100,000 at 20 nC per phase (50 and 100 μA, 200 μs width) left test electrode impedance unchanged and kept polarization potentials inside the water window.

    “A total of 50,000 pulses of 10 nC per phase and 100,000 pulses of 20 nC per phase (50 and 100 µA respectively with a 200 µs pulse width) were found to cause no change in impedance on the test electrode, as well as polarization potentials within the water window, so stimulation currents in this range were determined to be acceptable for the 30 pulses per electrode used in vivo.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingown data

    Intraoperative endoscopic view immediately following subdural placement of one thin-film array shows the electrode tail and traces entering the subdural space through a cranial slit.

    “(b) Intraoperative endoscopic view immediately following subdural placement of one thin-film array, after shim removal, showing the electrode tail and traces entering the subdural space through a cranial slit”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingown data

    Intraoperative fluoroscopic image shows four 529-electrode arrays in place, two over each hemisphere.

    “(d) Intraoperative fluoroscopic image (dorsoventral projection) showing four 529-electrode arrays in place, two over each hemisphere.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Somatosensory evoked potentials correspond to tactile stimulation of the left hindlimb, with traces spanning 1 s and maximum peak-to-peak signal approximately 80 μV.

    “(e) Somatosensory evoked potentials corresponding to tactile stimulation of the left hindlimb.Traces span 1 s and the maximum peak-to-peak signal is approximately 80 µV.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    The Layer 7 Cortical Interface acquires neural activity and delivers stimulation.

    “This schematic illustration shows how neural activity is acquired from and stimulated by the Layer 7 Cortical Interface.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Findingnone given

    The sampling rate for recording is set at 20 kHz per channel, generating data at a rate over 2.5 GB per minute for each set of 1,024 channels.

    “The sampling rate for recording is set at 20 kHz per channel, generating data at a rate over 2.5 GB per minute for each set of 1,024 channels.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Findingnone given

    The recording computers interface with either controller via a custom configuration of Intan RHX Data Acquisition Software, allowing real-time event-triggered averaging.

    “The recording computers interface with either controller via a custom configuration of the Intan Technologies RHX Data Acquisition Software, which allows for real-time event-triggered averaging in addition to base functionality.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingnone given

    Each amplifier board allows a hardware reference from one of 16 sites distributed across the array.

    “In addition, each board allows for a hardware reference from one of 16 sites distributed across the array.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    More distant regions exhibit less evoked activity while still being responsive to stimulation.

    “While still being responsive to stimulation, more distant regions exhibit less evoked activity.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingown data

    Visual evoked potentials are obtained from photostimulation of the left eye.

    “(f) Visual evoked potentials from photostimulation of the left eye.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisecitation

    Electrocorticography has facilitated capture of high-quality signals for effective use in brain-computer interfaces in several applications, including motor and speech neural prostheses.

    “In practice, electrocorticography (ECoG) has already facilitated capture of high quality signals for effective use in brain-computer interfaces in several applications, including motor and speech neural prostheses 7,22,24,56,[64][65][66][67][68][69][70] .”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites 7,22,24,56,[64][65][66][67][68][69][70]

      Not checked in the free run.

  • May combine two claims.

    Premisereasoning

    Non-penetrating cortical surface microelectrodes are a potentially attractive alternative and form the basis of the described system.

    “For this reason, non-penetrating cortical surface microelectrodes represent a potentially attractive alternative and form the basis of the system described here 53,63 .”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites [53] Wellman, S. M. et al. A Materials Roadmap to Functional Neural Interface Design. Adv. Funct. Mater. 28, 1701269 (2018). source

      Not checked in the free run.

    • cites [63] 2017, Epidemiology of worldwide spinal cord injury: a literature review, JN source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    High-density microelectrode arrays have been replacing macroelectrode arrays, offering smaller features and improved spatial resolution.

    “As a result, high-density microelectrode arrays have been replacing more traditional macroelectrode arrays, offering smaller features and improved spatial resolution [53][54][55][56][57][58] .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [53][54][55][56][57][58]

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Existing brain-computer interfaces depend on invasive surgery or electrodes that penetrate the brain, which restricts the applications they can address and the number of patients eligible.

    “Existing brain-computer interfaces rely on invasive surgical procedures or brain-penetrating electrodes, which limit addressable applications of the technology and the number of eligible patients.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites 20,[59][60][61][62]

      Not checked in the free run.

  • Classified as probably not a substantial claim.

    Premisecitation

    It is not yet clear whether softer penetrating electrodes offer a substantially different tradeoff.

    “It is not yet clear whether approaches involving softer penetrating electrodes offer a substantially different tradeoff 20,59 .”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites [20] source

      Not checked in the free run.

    • cites [59] Lamont, C. et al. Silicone encapsulation of thin-film SiO x, SiO x N y and SiC for modern electronic medical implants: a comparative long-term ageing study. J. Neural Eng. 18, 055003 (2021). source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Penetrating microelectrode arrays enable high-spatial-resolution recordings for brain-computer interfaces but cause invasiveness and tissue damage that increase with the number of implanted electrodes.

    “Microelectrode arrays that penetrate the brain have facilitated high-spatial-resolution recordings for brain-computer interfaces, but at the cost of invasiveness and tissue damage that scale with the number of implanted electrodes 20,[59][60][61][62] .”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites 20,[59][60][61][62]

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Surface micro-electrocorticography arrays offer an appealing balance between invasiveness and signal quality and are attractive alternatives or complements to penetrating arrays.

    “For these reasons surface microarrays (micro-electrocorticography or µECoG arrays) embody an appealing tradeoff between invasiveness and signal quality, and are attractive as an alternative or complement to arrays of electrodes that penetrate the cortical surface.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Penetrating microelectrode arrays enable single-cell recording but are more invasive and vary in tissue damage and long-term stability.

    “Microelectrode arrays that penetrate the brain facilitate single-cell recordings, at the cost of greater invasiveness, with varying degrees of brain tissue damage and long-term stability [92][93][94] .”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [92][93][94]

      Not checked in the free run.

  • May combine two claims.

    Interpretationown data

    The arrays are intended for subdural placement through 400 μm skull cuts made with a precision oscillating blade, with real-time imaging and fiberoptic endoscopy, at over 1,000 electrodes per minute.

    “The microelectrode arrays are designed to be inserted into the subdural space through 400-micron-wide skull incisions made using a precision oscillating blade, guided by real-time imaging and fiberoptic endoscopy, at a rate of more than 1,000 electrodes per minute, and to provide 1,000-fold higher electrode density than standard cortical grids.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Recent studies have shown that high-density electrocorticography can decode speech representations, allowing words and short sentences to be articulated.

    “More recent work has demonstrated the utility of high-density ECoG in decoding neural representations of speech, enabling the articulation of words and short sentences 6,69,99 .”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites [6] source

      Not checked in the free run.

    • cites [69] Vomero, M. et al. Conformable polyimide-based μECoGs: Bringing the electrodes closer to the signal source. Biomaterials 255, 120178 (2020). source

      Not checked in the free run.

    • cites [99] source

      Not checked in the free run.

  • May combine two claims.

    Interpretationown data

    Activity near kilohertz frequencies can be relatively uncorrelated even over short distances, which fits nearby but weakly coupled neurons.

    “High-frequency activity approaching the kilohertz range can exhibit relatively uncorrelated activity even at short range, consistent with the coexistence of nearby but relatively uncoupled neurons.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    The platform may speed progress in decoding and encoding neural signals and may broaden the group of patients who could benefit from neural interfaces.

    “Such a system promises to accelerate efforts to better decode and encode neural signals, and to expand the patient population that could benefit from neural interface technology.”
    Quote in the paperNot a substantial claim
    • Rests on an assumption: Safety and feasibility in Göttingen minipigs predict safety and feasibility in humans.
  • May combine two claims.

    Interpretationreasoning

    The Layer 7 Cortical Interface is a modular, scalable, and minimally invasive brain-computer interface platform.

    “The system we describe here, the Layer 7 Cortical Interface, is a modular, scalable, minimally invasive brain-computer interface platform.”
    Quote in the paperSubstantial claim
  • Classified as probably not a substantial claim.

    Premisecitation

    Electrocorticography and micro-electrocorticography have enabled high-resolution temporal and spatial recordings useful in vivo, including for brain-computer interfaces.

    “In practice, however, electrocorticography (ECoG) and microelectrocorticography (µECoG) have facilitated high temporal-and spatial-resolution recordings for effective use in in vivo applications 56,64,65 , including brain-computer interfaces.”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites [56] Deku, F. et al. Amorphous silicon carbide ultramicroelectrode arrays for neural stimulation and recording. J. Neural Eng. 15, 016007 (2018). source

      Not checked in the free run.

    • cites [64] 2021, A Systematic Review of the Incidence, Prevalence, Costs, and Activity and Work Limitations of Amputation, Osteoarthritis, Rheumatoid Arthritis, Back Pain, Multiple Sclerosis, Spinal Cord Injury, Stroke, and Traumatic Brain Injury in the United States: A 2019 Update. Archives of Physical Medicine and Rehabilitation

      Not checked in the free run.

    • cites [65] source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given

    The platform is intended to bring high-density, high-channel-count neural interfaces to the many millions of neurologic patients who could benefit.

    “The system is designed to deliver the benefits of high-density, high-channel-count neural interfaces to the millions of patients with neurologic disorders who stand to benefit from this technology.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given

    The arrays are intended to give 1,000 times the electrode density of standard cortical grids.

    “The microelectrode arrays are designed to be inserted into the subdural space through 400-micron-wide skull incisions made using a precision oscillating blade, guided by real-time imaging and fiberoptic endoscopy, at a rate of more than 1,000 electrodes per minute, and to provide 1,000-fold higher electrode density than standard cortical grids.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Brain-computer interface decoding can work with multiunit activity, local field potentials, or micro-electrocorticography, and does not require single-unit spikes.

    “Effective neural decoding for brain-computer interfaces can be achieved using multiunit activity, local field potentials, or micro-electrocorticography, without relying on single-unit action potentials 22,[84][85][86][87][88][89][90][91] .”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites 22,[84][85][86][87][88][89][90][91]

      Not checked in the free run.

  • May combine two claims.

    Interpretationown data

    The platform can be deployed quickly and minimally invasively over large cortical areas, reversibly and without trauma.

    “Here we have described a modular and highly scalable brain-computer interface platform capable of rapid, minimally invasive surgical deployment over multiple large areas of the cortical surface in a reversible and atraumatic manner.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    New thin-film micro-electrocorticography arrays have safely conformed to the brain for months and reliably recorded multi-unit and single-unit activity in some locations.

    “Recently developed thin-film µECoG arrays have demonstrated the ability to conform to the brain surface safely for several months while reliably recording multi-unit and single-unit activity in certain anatomic settings 53,57 .”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites [53] Wellman, S. M. et al. A Materials Roadmap to Functional Neural Interface Design. Adv. Funct. Mater. 28, 1701269 (2018). source

      Not checked in the free run.

    • cites [57] Li et al. Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO2/Al2O3/HfO2 Triple-Interlayer for Biomedical Implants. Coatings 9, 579 (2019). source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Multiple groups have used surface arrays to link neural activity to motor function for controlling prostheses in patients with physical impairment.

    “Several groups have used surface array techniques to correlate neural activity with motor function for the control of neural prostheses in physically impaired patients 7,67,68,[95][96][97][98] .”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites 7,67,68,[95][96][97][98]

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Surface-conforming arrays are much less invasive, easier to insert and remove without tissue damage, and possibly more stable long-term, though being farther from neurons may limit single-cell communication.

    “Microelectrode arrays that conform to the brain surface are considerably less invasive, more easily inserted and removed without damage to brain tissue, and potentially more stable in the long-term, but their greater distance from target neurons may limit single-cell communication.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Study design and conductreasoning

    General anesthesia was maintained with isoflurane at levels sufficient to produce analgesia without suppressing electrocorticographic activity.

    “General anesthesia was maintained with isoflurane at levels sufficient to produce analgesia without suppressing electrocorticographic activity, a balance that was facilitated by the minimally invasive nature of the procedure.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductown data

    A stimulation waveform was used for in vitro confirmation of safe polarization potential.

    “(a) Stimulation waveform used for in vitro confirmation of safe polarization potential, with the 100 µA current overlaid on the waveform for reference.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given

    The microelectrode arrays provide multichannel data used in electrophysiologic paradigms for recording spontaneous and stimulus-evoked neural activity, decoding, and focal stimulation across functional brain regions.

    “The microelectrode arrays provide multichannel data that is used in a variety of electrophysiologic paradigms to perform neural recording of both spontaneous and stimulus-evoked neural activity as well as decoding and focal stimulation of neural activity across a variety of functional brain regions.”
    Quote in the paperSubstantial claim
49 of 49
  • The authors state this as an assumption.

    For substantial clinical benefit, implantation must be minimally invasive, reversible, and free of neural tissue damage.

    “In order to achieve meaningful clinical impact on the large populations of patients who stand to benefit from brain-computer interface technologies, surgical procedures involved in implanting neural interfaces should be minimally invasive and reversible, and avoid damaging neural tissue.”
  • The authors state this as an assumption.

    Damage to tissue and overall procedure duration must not scale with the number of channels.

    “Tissue damage and total procedural time cannot be allowed to increase in proportion to channel number, as contemporary channel counts, already reaching the many thousands, will likely increase by orders of magnitude with further progress in this field.”
  • The paper relies on this without support.

    Safety and feasibility in Göttingen minipigs predict safety and feasibility in humans.

    “We demonstrate the safety and feasibility of rapidly delivering reversible implants containing over 2,000 microelectrodes to multiple functional regions in both hemispheres of the Göttingen minipig brain simultaneously, without requiring a craniotomy, at an effective insertion rate faster than 40 ms per channel, without damaging the cortical surface.”
  • The authors state this as an assumption.

    Implants can be expected to cause some tissue encapsulation.

    “Any implant may be expected to induce a certain degree of tissue encapsulation, but the anticipated tissue damage associated with removing multiple penetrating microelectrodes is greater than that associated with removing and replacing cortical surface arrays.”
  • The paper relies on this without support.

    Taking out several penetrating microelectrodes is expected to damage more tissue than removing and replacing surface arrays.

    “Any implant may be expected to induce a certain degree of tissue encapsulation, but the anticipated tissue damage associated with removing multiple penetrating microelectrodes is greater than that associated with removing and replacing cortical surface arrays.”
  • The paper relies on this without support.

    Göttingen minipigs have skull thickness comparable to that of adult humans.

    “selected for well characterized functional neuroanatomy as well as skull thickness comparable to that of adult humans”
6 of 6

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Approaches to increase density and lower invasiveness of ECoG are currently being developed for humans and animal models.

    “Different approaches to increase the density and lower the invasiveness of ECoG are currently being developed for use in humans and animal models176–182.”

    From The speech neuroprosthesis (2024) · cites this paper as [182]

Cites

Flagged first, then by citations

Research works printed but not linked: 4

  1. 2008, Neural Stimulation and Recording Electrodes. Annu. Rev. Biomed. Eng, 10, 275
  2. 2021, A Systematic Review of the Incidence, Prevalence, Costs, and Activity and Work Limitations of Amputation, Osteoarthritis, Rheumatoid Arthritis, Back Pain, Multiple Sclerosis, Spinal Cord Injury, Stroke, and Traumatic Brain Injury in the United States: A 2019 Update. Archives of Physical Medicine and Rehabilitation, 102, 115
  3. 2017, National and State Estimates of the Numbers of Adults and Children with Active Epilepsy — United States, 2015. MMWR Morb. Mortal. Wkly. Rep, 66, 821
  4. 2011, Highly Conformable Conducting Polymer Electrodes for In Vivo Recordings. Adv. Mater, 23, H268

Cited by

Sorted by citations

Authors
Hettick, M., Ho, E., Poole, A. J., Monge, M., Papageorgiou, D., Takahashi, K., LaMarca, M., Trietsch, D., Reed, K., Murphy, M., Rider, S., Gelman, K. R., Byun, Y. W., Hanson, T., Tolosa, V., Lee, S.-H., Bhatia, S., Konrad, P. E., Mager, M., Mermel, C. H., & Rapoport, B. I.
Corresponding author
Rapoport, Benjamin I.
Venue
bioRxiv (Cold Spring Harbor Laboratory)
Publisher
openRxiv
Published
2022-01-02
OpenAlex
W4205266536
Cited by
31
References
104
Citations by year
2026: 6 · 2025: 14 · 2024: 7 · 2023: 3 · 2022: 1
Type
posted content
Language
en
Open access
Locations
bioRxiv (Cold Spring Harbor Laboratory) (acceptedVersion)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Neural dynamics and brain function
Keywords
brain-computer interface, cortical surface mapping, neural decoding, neuromodulation, sensorimotor cortex, speech decoding

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-05.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-05.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review