Interphase gap as a means to reduce electrical stimulation thresholds for epiretinal prostheses
Open accessThe DOI matches its Crossref record; 13 of the 13 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: Epiretinal prostheses are designed to restore functional vision to the blind by electrically stimulating surviving retinal neurons. These devices have classically employed symmetric biphasic current pulses in order to maintain a balance of charge. Prior electrophysiological and psychophysical studies in peripheral nerve show that adding an interphase gap (IPG) between the two phases makes stimulation more efficient than pulses with no gap. This led us to investigate the effect of IPG duration on retinal stimulation thresholds. APPROACH: We measured retinal ganglion cell (RGC) electrical thresholds in salamander retina and phosphene perceptual thresholds in epiretinal prosthesis patients during stimulation with different IPG lengths. We also built Hodgkin-Huxley-type models of RGCs to further study how IPG affects thresholds. MAIN RESULTS: In general, there was a negative exponential correlation between threshold and IPG duration. Durations greater than or equal to ~0.5 ms reduced salamander RGC thresholds by 20-25%. Psychophysical testing in five retinal prosthesis patients indicated that stimulating with IPGs can decrease perceptual thresholds by 10-15%. Results from computational models of RGCs corroborated the observed behavior. SIGNIFICANCE: Incorporating interphase gaps can reduce the power consumption of epiretinal prostheses and increase the available dynamic range of phosphene size and brightness.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Hartong, D. T., Berson, E. L., & Dryja, T. P. (2006) · The Lancet · doi:10.1016/s0140-6736(06)69740-7 · Cited by 2,924
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 1×
“This causes a blocking effect, in which the anodic phase of one pulse hyperpolarizes the cell membrane, thus raising the threshold of action potential initiation for the next pulse [13, 26, 27].”
BackgroundWhich elements are excited in electrical stimulation of mammalian central nervous system: A review
Ranck, J. B. (1975) · Brain Research · doi:10.1016/0006-8993(75)90364-9 · Cited by 2,024
Visual perception in a blind subject with a chronic microelectronic retinal prosthesis
Humayun, M. S., Weiland, J. D., Fujii, G. Y., Greenberg, R., Williamson, R., Little, J., Mech, B., Cimmarusti, V., Van Boemel, G., Dagnelie, G., et al. (2003) · Vision Research · doi:10.1016/s0042-6989(03)00457-7 · Cited by 856 · cited 2×
“Each of these factors has been shown to affect perceptual threshold [4, 5, 32, 33].”
BackgroundSelf-citationInterim results from the international trial of Second Sight's visual prosthesis
Humayun, M. S., Dorn, J. D., da Cruz, L., Dagnelie, G., Sahel, J.-A., Stanga, P. E., Cideciyan, A. V., Duncan, J. L., Eliott, D., Filley, E., et al. (2012) · Ophthalmology · doi:10.1016/j.ophtha.2011.09.028 · Cited by 700
Self-citationThe Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve Stimulation
Gorman, P. H., & Mortimer, J. T. (1983) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1983.325041 · Cited by 313 · cited 2×
“Indeed, other studies have found the effects of IPG to vary with pulse duration [9, 10, 13, 28].”
BackgroundMethodAhuja, A. K., Dorn, J. D., Caspi, A., McMahon, M. J., Dagnelie, G., daCruz, L., Stanga, P., Humayun, M. S., Greenberg, R. J., & Argus II Study Group (2011) · British Journal of Ophthalmology · doi:10.1136/bjo.2010.179622 · Cited by 289 · cited 7×
“MEAs were fabricated by the authors at the W M Keck Photonics Laboratory at USC as described previously [18, 19].”
MethodInfluentialSelf-citationFactors affecting perceptual thresholds in epiretinal prostheses
de Balthasar, C., Patel, S., Roy, A., Freda, R., Greenwald, S., Horsager, A., Mahadevappa, M., Yanai, D., McMahon, M. J., Humayun, M. S., et al. (2008) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.07-0696 · Cited by 204
Self-citationPrado-Guitierrez, P., Fewster, L. M., Heasman, J. M., McKay, C. M., & Shepherd, R. K. (2006) · Hearing Research · doi:10.1016/j.heares.2006.03.006 · Cited by 173
Nanduri, D., Fine, I., Horsager, A., Boynton, G. M., Humayun, M. S., Greenberg, R. J., & Weiland, J. D. (2012) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.11-8401 · Cited by 163 · cited 1×
“Clinical testing has shown that phosphenes tend to get larger and brighter as stimulus amplitude increases [14, 15, 33, 34].”
Background
Cited by
Piech, D. K., Johnson, B. C., Shen, K., Ghanbari, M. M., Li, K. Y., Neely, R. M., Kay, J. E., Carmena, J. M., Maharbiz, M. M., & Muller, R. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-0518-9 · Cited by 442
Retinal stimulation strategies to restore vision: fundamentals and systems
Yue, L., Weiland, J. D., Roska, B., & Humayun, M. S. (2016) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2016.05.002 · Cited by 272
“Addition of an interphase gap between the leading cathodic phase and the following anodic phase has been demonstrated to be effective in reducing electrical stimulation thresholds for the epiretinal Argus II prosthesis (Weitz et al., 2014).”
Self-citationImproving the spatial resolution of epiretinal implants by increasing stimulus pulse duration
Weitz, A. C., Nanduri, D., Behrend, M. R., Gonzalez-Calle, A., Greenberg, R. J., Humayun, M. S., Chow, R. H., & Weiland, J. D. (2015) · Science Translational Medicine · doi:10.1126/scitranslmed.aac4877 · Cited by 154
Self-citationFactors Affecting Perceptual Thresholds in a Suprachoroidal Retinal Prosthesis
Shivdasani, M. N., Sinclair, N. C., Dimitrov, P. N., Varsamidis, M., Ayton, L. N., Luu, C. D., Perera, T., McDermott, H. J., & Blamey, P. J. (2014) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.14-14396 · Cited by 78
Stimulation strategies for improving the resolution of retinal prostheses
Tong, W., Meffin, H., Garrett, D. J., & Ibbotson, M. R. (2020) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.00262 · Cited by 66
Metadata
- Authors
- Weitz, A. C., Behrend, M. R., Ahuja, A. K., Christopher, P., Wei, J., Wuyyuru, V., Patel, U., Greenberg, R. J., Humayun, M. S., Chow, R. H., & Weiland, J. D.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2014-01-08
- Volume
- 11
- Issue
- 1
- Pages
- 016007
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W1994560947
- Cited by
- 50
- References
- 13
- FWCI
- 2.42
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Interphase gap as a means to reduce electrical stimulation thresholds for epiretinal prostheses. https://app.cochat.ai/papers/10.1088/1741-2560/11/1/016007
- The paper
- Weitz, A. C., Behrend, M. R., Ahuja, A. K., Christopher, P., Wei, J., Wuyyuru, V., Patel, U., Greenberg, R. J., Humayun, M. S., Chow, R. H., & Weiland, J. D. (2014). Interphase gap as a means to reduce electrical stimulation thresholds for epiretinal prostheses. Journal of Neural Engineering. https://doi.org/10.1088/1741-2560/11/1/016007
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.