Analysis of Time Lags and Other Sources of Error of the DexCom SEVEN Continuous Glucose Monitor
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Background: It has been assumed that continuous glucose sensors show substantial time lags versus blood glucose. This assumption has led to suggestions that sensors are less accurate during rapidly changing glucose levels and that sensors should only be calibrated when glucose levels are stable. The analysis presented here tests the assumption of substantial sensor time lag and its suggested effects using clinical data from the DexCom™ (San Diego, CA) SEVEN ® . Methods: Sensor and blood glucose data were collected from 117 adult subjects with insulin-dependent diabetes. Each subject wore the sensor for 7 days and underwent an 8–10-h in-clinic tracking study during which blood glucose was measured every 15–20 min. Accuracy (absolute relative difference [ARD]) versus blood glucose rate of change was evaluated on the in-clinic data set. The effect on accuracy of calibration during rapid rates of change was evaluated on the combined home-use and in-clinic data set. Results: Average sensor time lag versus blood glucose was 5.7 min. Mean ARD versus rate of change (less than −2 to >2 mg/dL/minute) ranged between 15.0% to 16.3%. Across rates of change during calibration, mean ARD after calibration ranged between 13.2% and 16.0%. Calibration with reference measurements instead of patient measurements improved overall mean ARD from 16.0% to 8.5%. Conclusions: For this sensor, the assumption of substantial time lag and its suggested effects may be incorrect. The main source of error is the calibration process.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Inflammatory Response to Implants
Anderson, J. M. (1988) · ASAIO Journal · doi:10.1097/00002480-198804000-00005 · Cited by 591
A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.
Updike, S. J., Shults, M. C., Gilligan, B. J., & Rhodes, R. K. (2000) · Diabetes Care · doi:10.2337/diacare.23.2.208 · Cited by 135
Feasibility of Continuous Long-Term Glucose Monitoring from a Subcutaneous Glucose Sensor in Humans
Gilligan, B. C., Shults, M., Rhodes, R. K., Jacobs, P. G., Brauker, J. H., Pintar, T. J., & Updike, S. J. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504774198089 · Cited by 71 · cited 1×
“long-term implantation of sensors have been achieved in humans, 18,19 but these techniques are not used in the trans-cutaneous glucose sensor evaluated here.”
Self-citation
Cited by
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202
Modeling the glucose sensor error
Facchinetti, A., Del Favero, S., Sparacino, G., Castle, J. R., Ward, W. K., & Cobelli, C. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2284023 · Cited by 130
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Facchinetti, A., Del Favero, S., Sparacino, G., & Cobelli, C. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1226-y · Cited by 80
Clinical Implications of Accuracy Measurements of Continuous Glucose Sensors
Bailey, T. S. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0050 · Cited by 60
“24 However, improvements in software alone have also demonstrated significant improvement in MARD.”
Influential
Metadata
- Authors
- Kamath, A., Mahalingam, A., & Brauker, J.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2009-11
- Volume
- 11
- Issue
- 11
- Pages
- 689-695
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W1975330563
- Cited by
- 69
- References
- 4
- FWCI
- 3.91
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.