Recommendations for Standardizing Glucose Reporting and Analysis to Optimize Clinical Decision Making in Diabetes: The Ambulatory Glucose Profile (AGP)
Abstract
Abstract Underutilization of glucose data and lack of easy and standardized glucose data collection, analysis, visualization, and guided clinical decision making are key contributors to poor glycemic control among individuals with type 1 diabetes. An expert panel of diabetes specialists, facilitated by the International Diabetes Center and sponsored by the Helmsley Charitable Trust, met in 2012 to discuss recommendations for standardization of analysis and presentation of glucose monitoring data, with the initial focus on data derived from CGM systems. The panel members were introduced to a universal software report, the Ambulatory Glucose Profile (AGP), and asked to provide feedback on its content and functionality, both as a research tool and in clinical settings. This paper provides a summary of the topics and issues discussed during the meeting and presents recommendations from the expert panel regarding the need to standardize glucose profile summary metrics and the value of a uniform glucose report to aid clinicians, researchers, and patients.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Originally created by Mazze et. al. (68), the standardized AGP report was further developed by the International Diabetes Center and now incorporates all the core CGM metrics and targets along with a 14-day composite glucose profile as an integral component of clinical decision making (24).”
From Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range (2019) · cites this paper as [24]
In the citing paper
“The key benefits of CGM include real-time data monitoring for people with T1D and access for them and their HCPs to complete glucose datasets17; both of which help people to achieve their target time spent within the acceptable range (time in range [TIR]: 3.9–10.0 mmol/L [70–180 mg/dL]) and avoid hyper- and hypoglycemia.18”
From Increased Time in Range and Fewer Missed Bolus Injections After Introduction of a Smart Connected Insulin Pen (2020) · cites this paper as [18]
Citations
Cites
The Triple Aim: Care, Health, And Cost
Berwick, D. M., Nolan, T. W., & Whittington, J. (2008) · Health Affairs · doi:10.1377/hlthaff.27.3.759 · Cited by 5,136
Standards of medical care in diabetes—2013
American Diabetes Association (2013) · Diabetes Care · doi:10.2337/dc13-s011 · Cited by 4,369
Effect of a multifactorial intervention on mortality in type 2 diabetes
Gæde, P., Lund-Andersen, H., Parving, H.-H., & Pedersen, O. (2008) · New England Journal of Medicine · doi:10.1056/nejmoa0706245 · Cited by 3,465 · cited 1×
“37”
Monnier, L., Mas, E., Ginet, C., Michel, F., Villon, L., Cristol, J.-P., & Colette, C. (2006) · JAMA · doi:10.1001/jama.295.14.1681 · Cited by 2,444
Translating the A1C assay into estimated average glucose values
Nathan, D. M., Kuenen, J., Borg, R., Zheng, H., Schoenfeld, D., Heine, R. J., & for the A1c-Derived Average Glucose (ADAG) Study Group (2008) · Diabetes Care · doi:10.2337/dc08-0545 · Cited by 1,812
Severe hypoglycemia and risks of vascular events and death
Zoungas, S., Patel, A., Chalmers, J., de Galan, B. E., Li, Q., Billot, L., Woodward, M., Ninomiya, T., Neal, B., MacMahon, S., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1003795 · Cited by 1,471
Defining and Reporting Hypoglycemia in Diabetes
American Diabetes Association Workgroup on Hypoglycemia (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1245 · Cited by 1,146
Hypoglycaemia: the limiting factor in the glycaemic management of type I and type II diabetes.
Cryer, P. (2002) · Diabetologia · doi:10.1007/s00125-002-0822-9 · Cited by 794
Wild, D., von Maltzahn, R., Brohan, E., Christensen, T., Clauson, P., & Gonder-Frederick, L. (2007) · Patient Education and Counseling · doi:10.1016/j.pec.2007.05.003 · Cited by 545
Pickup, J. C., Freeman, S. C., & Sutton, A. J. (2011) · BMJ · doi:10.1136/bmj.d3805 · Cited by 541
Flagged first, then by citations
Research works printed but not linked: 4
- McGill J.B, , Chen V, , Beck RW. Diabetes management among adults under excellent control in the Type 1 Diabetes Exchange Clinic Registry: how do they do it? Oral presentation #186 at the European Association for the Study of Diabetes Meeting, Berlin, 2012.
- Self-monitoring of blood glucose among adults with diabetes—United States, 1997–2006. MMWR Morb Mortal Wkly Rep, 2007; 56:1133–1137.
- Adi S, , Beck RW, , DuBose SN, , Bode BW, , Peters AL, , Hirsch IB, , Miller KM. T1D Exchange Clinic Network: A comparison of users, nonusers of real-time continuous glucose monitoring (CGM) with type 1 diabetes (T1D) in the T1D Exchange. Poster presentation at the Diabetes Technology Society Meeting, October 27–292011.
- Dunn TC, , Crouther N. Assessment of the variance of the Ambulatory Glucose Profile over 3–20 days of continuous glucose monitoring. Abstract poster presented at the European Association for the Study of Diabetes Meeting, Stockholm, September 2010.
Cited by
Sorted by citations
Items 101–200 of 306
Time Savings Using a Standardized Glucose Reporting System and Ambulatory Glucose Profile
Mullen, D. M., Bergenstal, R., Criego, A., Arnold, K. C., Goland, R., & Richter, S. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817740592 · Cited by 22
Self-citationCrossen, S., Romero, C., Reggiardo, A., Michel, J., & Glaser, N. (2022) · JMIR Diabetes · doi:10.2196/33639 · Cited by 21
“The advent of continuous glucose monitoring (CGM) devices and insulin pumps has dramatically increased the volume of patient-generated health data (PGHD) available for T1D management [1,2], and mobile health apps have begun to make these data accessible remotely.”
BackgroundCurrent landscape and comprehensive management of glycemic variability in diabetic retinopathy
Chen, B., Shen, C., & Sun, B. (2024) · Journal of Translational Medicine · doi:10.1186/s12967-024-05516-w · Cited by 21
“In recent years, interquartile range (IQR), an important indicator in the ambulatory glucose profile, was proposed as a powerful tool for assessing GV, primarily reflecting interday variability [27, 36–38].”
MethodBeck, R. W., & Bergenstal, R. M. (2021) · Diabetes Spectrum · doi:10.2337/ds20-0090 · Cited by 21
Time above range for predicting the development of type 2 diabetes
Marco, A., Pazos-Couselo, M., Moreno-Fernández, J., Díez‐Fernández, A., Alonso‐Sampedro, M., Fernández‐Merino, C., González‐Quintela, A., & Gudé, F. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.1005513 · Cited by 21
Rai, S., Hulse, A., & Kumar, P. K. M. (2016) · Indian Journal of Endocrinology and Metabolism · doi:10.4103/2230-8210.192894 · Cited by 21
“The usefulness of AGP in adults has been studied and guidelines have been recommended by experts in the field.[10,11] To date, there are no studies looking into the feasibility of this tool in children.”
BackgroundProfessional flash continuous glucose monitoring as a supplement to A1C in primary care
Hirsch, I. B. (2017) · Postgraduate Medicine · doi:10.1080/00325481.2017.1383137 · Cited by 24
“The classic AGP ‘dashboard,’ designed as one-page report, consists of three parts [15]:”
BackgroundMethodInfluentialIt's Time to Move from the A1c to Better Metrics for Diabetes Control
Kowalski, A. J., & Dutta, S. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0060 · Cited by 21
Is It Time to Replace Time-in-Range with Time-in-Tight-Range? Maybe Not
Beck, R. W. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0602 · Cited by 20
Forlenza, G. P., Pyle, L. L., Maahs, D. M., & Dunn, T. C. (2016) · Pediatric Diabetes · doi:10.1111/pedi.12474 · Cited by 20
“Numerically the AGP also presents average glucose, estimated HbA1c, glucose standard deviation and IQR, as well as percent time in various ranges and the option for “Close-Up” statistics on area under the curve (AUC) analysis, variability, and hyperglycemia/hypoglycemia episodes.(15) To evaluate the longitudinal benef…”
BackgroundMethodGómez, A. M., Carrillo, L. F. M., Correa, C. M. A., Múñoz, Ó., Rondón, M., Silva-Herrera, J. L., & Henao, D. C. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0030 · Cited by 20
The Ambulatory Glucose Profile: Opportunities for Enhancement
Rodbard, D. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0524 · Cited by 20
“8–25 This report summarizes a number of proposals for improvement of the effectiveness and informativeness of the AGP, some that had been presented years ago (e.g., log scale, 38 display of data points, 5,7 risks or rates of hypoglycemia and hyperglycemia, 9,17 and analysis by day of the week 7,16…”
Hirsch, I. B., & Verderese, C. A. (2017) · Endocrine Practice · doi:10.4158/ep171962.ra · Cited by 22
Reutrakul, S., Irsheed, G. A., Park, M., Steffen, A. D., Burke, L., Pratuangtham, S., Baron, K. G., Duffecy, J. L., Perez, R. I., Quinn, L., et al. (2023) · Sleep Health · doi:10.1016/j.sleh.2023.07.007 · Cited by 19
Kohnert, К.-D., Heinke, P., Vogt, L., Augstein, P., Thomas, A., & Salzsieder, E. (2017) · Journal of Endocrinological Investigation · doi:10.1007/s40618-017-0682-2 · Cited by 19
Thyde, D. N., Mohebbi, A., Bengtsson, H. L., Jensen, M. L., & Mørup, M. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820912411 · Cited by 24
Aleppo, G., Chmiel, R., Zurn, A., Bandoske, R., Creamer, P., Neubauer, N., Wong, J., Andrade, S. B., & Hauptman, A. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231196168 · Cited by 19
Ambulatory glucose profile (AGP): utility in UK clinical practice
Evans, M. L., CRANSTON, I. C., & Bailey, C. J. (2017) · British Journal of Diabetes · doi:10.15277/bjd.2017.121 · Cited by 18
Kesavadev, J., Ramachandran, L., & Krishnan, G. (2017) · Journal of Diabetology · doi:10.4103/jod.jod_14_17 · Cited by 20
Continuous glucose monitoring: a systematic review of sensor systems and prospects
Ray, P. P. (2018) · Sensor Review · doi:10.1108/sr-12-2017-0268 · Cited by 18
Gómez, A. M., Henao, D. C., Madero, A. I., Múñoz, Ó., Rondón, M., Kattah, L., Guerrero, J. S., Morros-González, E., Llano, J. P., Maira, G., et al. (2018) · Endocrinología Diabetes y Nutrición · doi:10.1016/j.endinu.2018.03.009 · Cited by 18
Nimri, R., Yakoob, H., Schoenberg, B. E., & Dassau, E. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.2504 · Cited by 20
Mader, J. K., Lilly, L. C., Aberer, F., Poettler, T., Johns, D., TRAUTMANN, M. E., WARNER, J. L., & Pieber, T. R. (2018) · Diabetic Medicine · doi:10.1111/dme.13708 · Cited by 18
Ekhlaspour, L., Tabatabai, I., & Buckingham, B. A. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819851790 · Cited by 16
Glycemic Status Assessment by the Latest Glucose Monitoring Technologies
Malandrucco, I., Russo, B., Picconi, F., Menduni, M., & Frontoni, S. (2020) · International Journal of Molecular Sciences · doi:10.3390/ijms21218243 · Cited by 17
“The AGP report incorporates all the core CGM metrics and targets along with a 14-day composite glucose profile as an integral component of clinical decision making [19].”
BackgroundHealth-System-Based Interventions to Improve Care in Pediatric and Adolescent Type 1 Diabetes
Corathers, S. D., Schoettker, P. J., Clements, M. A., List, B. A., Mullen, D. M., OHMER, A. L., Shah, A. C., & Lee, J. M. (2015) · Current Diabetes Reports · doi:10.1007/s11892-015-0664-8 · Cited by 17
Vianna, A. G. D., de Lacerda, C. S., Pechmann, L. M., Polesel, M. G., Marino, E. C., & Faria‐Neto, J. R. (2018) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2018.03.035 · Cited by 17
Association Between Glycated Hemoglobin and Health Utility for Type 1 Diabetes
McQueen, R. B., Ellis, S. L., Maahs, D. M., Anderson, H. D., Nair, K. V., Libby, A. M., & Campbell, J. D. (2014) · Patient · doi:10.1007/s40271-014-0045-4 · Cited by 17
“Leaders in the type 1 diabetes research field have suggested using alternative metrics to describe glycemic control such as glycemic variability (GV) [32].”
BackgroundProfiling intra-patient type I diabetes behaviors
Contreras, I., Quirós, C., Giménez, M., Conget, I., & Vehı́, J. (2016) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2016.08.022 · Cited by 17
The ambulatory glucose profile and its interpretation
Lin, R., Brown, F., & Ekinci, E. I. (2022) · The Medical Journal of Australia · doi:10.5694/mja2.51666 · Cited by 15
Using Flash Continuous Glucose Monitoring in Primary Practice
Hirsch, I. B., & Wright, E. E. (2018) · Clinical Diabetes · doi:10.2337/cd18-0054 · Cited by 15
“FCGM personal and professional devices were among the first to feature the AGP, which is gradually being incorporated into other CGM systems (84,85).”
BackgroundHow use of continuous glucose monitoring can address therapeutic inertia in primary care
Martens, T. W., & Parkin, C. G. (2022) · Postgraduate Medicine · doi:10.1080/00325481.2022.2080419 · Cited by 16
Individual categorisation of glucose profiles using compositional data analysis
Biagi, L., Bertachi, A. H., Giménez, M., Conget, I., Bondía, J., Fernández, J. A. M., & Vehı́, J. (2018) · Statistical Methods in Medical Research · doi:10.1177/0962280218808819 · Cited by 15
Standardization Versus Customization of Glucose Reporting
Rodbard, D. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0116 · Cited by 15
Glucodensity functional profiles outperform traditional continuous glucose monitoring metrics
Matabuena, M., Ghosal, R., Aguilar, J. E., Keshet, A., Wagner, R. J., Fernández‐Merino, C., Castro, J. S., Zipunnikov, V., Onnela, J.-P. “., & Gudé, F. (2025) · Scientific Reports · doi:10.1038/s41598-025-18119-2 · Cited by 15
“The traditional CGM and non-CGM models include: (i) baseline FPG and HbA1c; and (ii) baseline FPG and HbA1c, along with CGM metrics like the Area Under the Curve (AUC), Mean Amplitude of Glycemic Excursions (MAGE), Continuous Overall Net Glycemic Action (CONGA), and hyperglycemia time-in-range.”
McQueen, R. B., Ellis, S. L., Maahs, D. M., Anderson, H. D., Nair, K. V., & Campbell, J. D. (2014) · Endocrine Practice · doi:10.4158/ep14027.or · Cited by 14
Use of continuous glucose monitoring trend arrows in the younger population with type 1 diabetes
Elbarbary, N., Moser, O., Yaarubi, S. A., Alsaffar, H., Shaikh, A. A., Ajjan, R., & Deeb, A. (2021) · Diabetes & vascular disease research · doi:10.1177/14791641211062155 · Cited by 14
Olsen, M. T., Klarskov, C. K., Dungu, A. M., Hansen, K. B., Pedersen‐Bjergaard, U., & Kristensen, P. L. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231221803 · Cited by 14
“38 The AGP report has three major components: (1) glucose statistics and time in different glucose ranges, (2) a single 24-hour “modal” day that combines all daily CGM glucose data, referred to as AGP, and (3) daily glucose profiles.”
Toschi, E., Fisher, L., Wolpert, H. A., Love, M., Dunn, T. C., & Hayter, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818791534 · Cited by 15
Martyn-Nemeth, P. A., Duffecy, J. L., Quinn, L., Park, C., Mihailescu, D. V., & Penckofer, S. M. (2019) · Trials · doi:10.1186/s13063-019-3876-4 · Cited by 13
Standardized Hybrid Closed-Loop System Reporting
Shah, V. N., & Garg, S. K. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0622 · Cited by 14
“26 In addition, a recent study showed that mean glucose has the highest discrimination power, suggesting that mean glucose is the easiest and best indicator to discriminate between individuals.”
InfluentialRodbard, D., & Garg, S. K. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0030 · Cited by 13
Electrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 13
“The development of standardized, automated data analysis and display tools is enabling HCPs to evaluate the status of a patient in a convenient way [272].”
BackgroundMethodHulse, A., Rai, S., & KUMAR, K. M. P. (2016) · Indian Journal of Endocrinology and Metabolism · doi:10.4103/2230-8210.190546 · Cited by 12
Relationships Between Glycemic Control and Cardiovascular Fitness
Moxley, E. W., SMITH, D. P., Quinn, L. T., & Park, C. (2018) · Biological Research For Nursing · doi:10.1177/1099800418767572 · Cited by 13
Hirsch, I. B., Nardacci, E., & Verderese, C. A. (2019) · Diabetes Spectrum · doi:10.2337/ds18-0059 · Cited by 12
“FCGM personal and professional devices were among the first to feature the AGP, which is now available with other CGM systems as well (88,89).”
BackgroundAdvances in glucose monitoring: Improving diabetes management through evidence-based medicine
Mazze, R. S. (2020) · Primary care diabetes · doi:10.1016/j.pcd.2020.03.001 · Cited by 12
Fonseca, V. A., & Grunberger, G. (2017) · Endocrine Practice · doi:10.4158/1934-2403-23.5.629 · Cited by 12
DeVries, J. H., Meneghini, L., Barnett, A. H., Reid, T., Dain, M., Sanofi, P., Landgraf, W., Vlajnic, A., Traylor, L., & Bergenstal, R. M. (2014) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2014.10.01.23 · Cited by 11
Monnier, L., & Colette, C. (2015) · Diabetes & Metabolism · doi:10.1016/j.diabet.2015.03.004 · Cited by 11
Simonson, G. D., Criego, A. B., Battelino, T., Carlson, A. L., Choudhary, P., Franc, S., Gershenoff, D. C., Grunberger, G., Hirsch, I. B., Isaacs, D. M., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0107 · Cited by 11
Kingsnorth, A. P., Whelan, M., Sanders, J. P., Sherar, L. B., & Esliger, D. (2018) · JMIR mhealth and uhealth · doi:10.2196/mhealth.9471 · Cited by 11
“Mean glucose and SD of glucose were indicated as the most common and easily interpreted metrics [24], and MAGE is considered the gold standard for glucose variability measurement [25].”
Investigating Temporal Patterns of Glycemic Control around Holidays
Belsare, P. V., Lu, B., Bartolome, A., & Prioleau, T. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871646 · Cited by 10
Kovatchev, B. P., & Lobo, B. J. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0117 · Cited by 10
“19 This format was endorsed by the international consensus on TIR 17 and was also referenced by the American Diabetes Association 2023 Standards of Care 20 and the American Association of Clinical Endocrinology (AACE) consensus on the use of CGM. 21 The AGP report is now adopted by many CGM device…”
Zhang, Y., Bartolomeo, S. D., Sheng, F., Jimison, H. B., & Dunne, C. (2019) · Visualization · doi:10.1109/visual.2019.8933584 · Cited by 10
“2, and taskonomies for Type 1 diabetes management [1, 15].”
BackgroundUnderstanding Reflection Needs for Personal Health Data in Diabetes
Prioleau, T., Sabharwal, A., & Vasudevan, M. M. (2020) · International Conference on Pervasive Computing Technologies for Healthcare · doi:10.1145/3421937.3421972 · Cited by 10
“hen there is no supporting data of a patient’s blood glucose trends. (2) Low Blood Glucose is directly associated with acute complications such as loss of consciousness, seizures, and even death [4], [7], [23]. Following analysis for data sufficiency, identifying hypoglycemic (i.e. low blood glucose) events is a prior…”
BackgroundMethodInfluentialIrace, C., Cutruzzolà, A., Tweden, K. S., & KAUFMAN, F. R. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.1982380 · Cited by 12
Camerlingo, N., Vettoretti, M., Sparacino, G., Facchinetti, A., Mader, J. K., Choudhary, P., & Favero, S. D. (2021) · Diabetic Medicine · doi:10.1111/dme.14758 · Cited by 10
Wu, W., Chai, Q., & Zhang, Z. (2021) · Molecular and Cellular Biology · doi:10.1128/mcb.00282-21 · Cited by 10
“ability is closer to the actual changes in blood sugar level in diabetic patients (5).”
BackgroundIs Continuous Glucose Monitoring a Tool, an Intervention, or Both?
Wright, E. E., & Subramanian, S. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0025 · Cited by 10
Glucose concentrations in blood and tissue - a pilot study on variable time lag
Chlup, R., Krejčí, J., O’Connell, M. T., Sebestova, B., Plička, R., Ježová, L., Brozova, T., Doubravová, B., Zálešáková, H., Ďurajková, E., et al. (2015) · Biomedical Papers · doi:10.5507/bp.2015.007 · Cited by 10
Development of the Likelihood of Low Glucose (LLG) Algorithm for Evaluating Risk of Hypoglycemia
Dunn, T. C., Hayter, G., Doniger, K., & Wolpert, H. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532200 · Cited by 11
“There is a need for glucose reports that can provide standardized, efficient output to effectively guide therapeutic decision making.(12-14) Key benefits of glucose reports include consistent view of glucose trends and patterns over the day, and showing the detail that A1C cannot.”
BackgroundMolfetta, S. D., Rossi, A., Assaloni, R., Cherubini, V., Consoli, A., Bartolo, P. D., Guardasole, V., Laurenzi, A., Lombardo, F., Maffeis, C., et al. (2022) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2022.109867 · Cited by 13
Chlup, R., Doubravová, B., Bartek, J., Zapletalová, J., Krystyník, O., & Procházka, V. (2013) · Disease Markers · doi:10.1155/2013/702427 · Cited by 10
Kohnert, K., Heinke, P., Vogt, L., Augstein, P., & Salzsieder, E. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01257 · Cited by 10
“We used TIR as an established and clinically useful indicator of the quality of glycemic control, reflecting the time in predefined target ranges (Bergenstal et al., 2013 and Rodbard, 2018).”
MethodInfluentialA Markov model for inferring event types on diabetes patients data
de Carvalho, D. F., Kaymak, U., Gorp, P. M. E. V., & van Riel, N. A. W. (2022) · Healthcare Analytics · doi:10.1016/j.health.2022.100024 · Cited by 9
“To enable a less wide but still representative observation of glucose levels, the BG signal was discretized based on six recommended ranges according to clinical diagnosis and safety [34–36].”
MethodKong, L., Deng, B., Guo, M., Chen, M., Wang, X., Zhang, M., Tang, H., Wang, Q., Yang, L., & Xiong, Z. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1229494 · Cited by 10
Chen, M.-Y., Li, H., Shen, Y., Liu, B., Yan, R., Sun, X., Ye, L., Lee, K., Ma, J., & Su, X. (2020) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/a-0994-9850 · Cited by 9
O’Donnell, H. K., Vigers, T., Johnson, S. B., Pyle, L. L., Wright, N. A., Deeb, L. C., & Driscoll, K. A. (2021) · Contemporary Clinical Trials · doi:10.1016/j.cct.2021.106279 · Cited by 10
“In fact, some T1D providers are uncertain about how to interpret and use the reports in clinical care and may not be able to invest the time necessary to become proficient in interpreting the data [14,15].”
BackgroundEstock, J. L., Codario, R. A., Keddem, S., Zupa, M. F., Rodriguez, K. L., & DiNardo, M. M. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0498 · Cited by 9
Múñoz, Ó., Gómez, A. M., Maira, G., Vargas, F. M. L., & Ruiz, Á. J. (2018) · Diabetes & Metabolic Syndrome Clinical Research & Reviews · doi:10.1016/j.dsx.2018.03.028 · Cited by 9
“Thus, IQR is a component of the international proposal for the standardized reporting on the ambulatory glucose profile [17].”
BackgroundA Primary Care Perspective on the Use of Continuous Glucose Monitoring in Clinical Practice
Bhide, M., Grey, J. M. S., Moser, E. G., & Garg, S. K. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0169 · Cited by 9
Unleashing the Potential of Blood Glucose Monitoring Data With the Ambulatory Glucose Profile Report
Simonson, G. D., Holt, E. H., Grady, M., Hurrell, G., Gaudiani, L., & Bergenstal, R. M. (2024) · Clinical Diabetes · doi:10.2337/cd23-0092 · Cited by 10
Spartano, N. L., Prescott, B. R., Walker, M. E., Shi, E., Venkatesan, G., Fei, D., Lin, H., Murabito, J. M., Ahn, D. T., Battelino, T., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251315171 · Cited by 8
“15 We observed remarkable differences of opinion among clinicians in terms of how to interpret elevated post-prandial glucose levels (both brief and prolonged) or substantial time spent below range ( 70 mg/dL), including what glucose levels may be considered “elevated” or “low.””
Shah, V. N., Aktürk, H. K., Vigers, T., Pyle, L. L., Oliver, N. S., & Klonoff, D. C. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0365 · Cited by 8
Possible Impacts of New Accuracy Standards of Self-Monitoring of Blood Glucose
Garg, S. K., Reed, K., Grey, J. M. S., Westerman, A., & Pediatrics, P. O. M. A. (2013) · US Endocrinology · doi:10.17925/use.2013.09.01.16 · Cited by 8
Kingsnorth, A. P., Whelan, M., Orme, M. W., Routen, A. C., Sherar, L. B., & Esliger, D. (2020) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2020-0386 · Cited by 8
“Mean daily glucose as a measure of glucose exposure was calculated as the average of all datapoints for a given data file, which has been reported as a metric which both patients and clinicians can understand (Bergenstal et al. 2013), and characterises daily variations in glucose concentrations.”
MethodThe Future of Glucose Monitoring
Garg, S. K. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0421 · Cited by 8
Practical Approaches to Continuous Glucose Monitoring in Primary Care: A UK-Based Consensus Opinion
Fernando, K., Alabraba, V., Welsh, J. B., Seidu, S., Strain, W. D., Bell, H. E., Davies, S., & Evans, M. (2025) · Diabetes Therapy · doi:10.1007/s13300-025-01706-x · Cited by 7
“The ambulatory glucose profile (AGP) was introduced in 2013 as a standardised way to present CGM data [20, 21], and most CGM manufacturers license it for use with their systems’ data.”
Monzon, A. D., Majidi, S., Clements, M. A., & Patton, S. R. (2023) · International Journal of Behavioral Medicine · doi:10.1007/s12529-023-10159-0 · Cited by 7
“…child’s glucometer for the 2 weeks before their study visit, and researchers used the glucose value ranges recommended by Bergenstal and colleagues [35] to estimate the percentage of glucose levels that were below the target range (< 70 mg/dl), within the target range (70–180 mg/dl), and above the…”
Georga, E. I., Príncipe, J. C., Rizos, E. C., & Fotiadis, D. I. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8037430 · Cited by 7
“Ambulatory Glucose Profile (AGP) report [20] of a typical subject (Subject #1) is portrayed in Fig.”
BackgroundBlood Glucose Monitoring Before and After Type 1 Diabetes Clinic Visits
Driscoll, K. A., Johnson, S. B., Wang, Y., Wright, N. A., & Deeb, L. C. (2017) · Journal of Pediatric Psychology · doi:10.1093/jpepsy/jsx151 · Cited by 8
Sheng, T., Offringa, R., Kerr, D. E., Clements, M. A., Fischer, J., Parks, L. J., & Greenfield, M. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819873641 · Cited by 7
“While HbA1c is widely used as an indicator of recent (approximately three months) glycemic control, it is limited in its use to predict hypoglycemia risk(10) or as a rapid indicator following treatment and/or behavior change.(11) This is particularly important in the context of calculating safe and effective doses of…”
BackgroundTokutsu, A., Okada, Y., Torimoto, K., & Tanaka, Y. (2020) · Diabetology & Metabolic Syndrome · doi:10.1186/s13098-020-00577-5 · Cited by 5
“AGP is an excellent tool that clearly and visually presents the trends in interstitial glucose variability for the individual patient [9]; our study showed that AIQR is a valuable index for use in AGP.”
MethodDeeb, A., Muammar, T., Alsaffar, H., Sedaghat, S., Hassani, N. A., Odeh, R., Alkhayyat, H., Sinani, A. A., Attia, N. A., Adhami, S., et al. (2021) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2021.108671 · Cited by 7
Nimri, R., Murray, N., Ochs, A., Pinsker, J. E., & Dassau, E. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.2504 · Cited by 8
Practical Application of Continuous Glucose Monitoring in Clinical Practice: Case Studies
Unger, J., & Franco, D. R. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0080 · Cited by 7
“29 Originally created by Mazze et al., 30 the AGP was further refined by the International Diabetes Center (Min-neapolis, MN).”
Optimizing Diabetes Care With the Standardized Continuous Glucose Monitoring Report
Chun, J. H., & O’Neill, M. S. (2020) · Clinical Diabetes · doi:10.2337/cd19-0066 · Cited by 6
“A key contributor to clinicians’ reluctance to embrace CGM,particularly in the primary care setting, has been the large amount of data to interpret without a standardized report format to allow for efficient clinical application (7).”
BackgroundThe phenomenon of HbA1c stability and the risk of hypoglycemia in long-standing type 1 diabetes
Wolnik, B., Orłowska-Kunikowska, E., Błaszkowska, M., Graff, B., Wolf, J., Czupryniak, L., & Narkiewicz, K. (2019) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2019.05.010 · Cited by 7
Antsiferov, M. B., Galstyan, G. R., Zilov, A. V., Mayorov, A. Y., Markova, T. N., Demidov, N. A. D., Koteshkova, O. M. K., Лаптев, Д. Н., & Vitebskaya, A. V. (2019) · Diabetes Mellitus · doi:10.14341/dm10227 · Cited by 8
Lee, W., Kim, N., Jung, I., Park, S. Y., Yu, J. H., Seo, J. A., Kim, J., Kim, K. J., Kim, N. H., Yoo, H. J., et al. (2023) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2022.0273 · Cited by 6
“Currently, continuous glucose monitoring (CGM) has been recommended as an indicator of glycemic control and for decision making in treatment regimens [5] and as a tool to assess GV [6,7].”
BackgroundMethodTesting versus guessing blood glucose values: impact on self-care behaviors in type 2 diabetes
Pettus, J., Stenger, P., Schachner, H. C., Dunne, N., Parkes, J. L., Pardo, S. A., & Edelman, S. V. (2014) · Current Medical Research and Opinion · doi:10.1185/03007995.2014.929097 · Cited by 6
Faiola, A. J., Vatani, H., Greenhill, K., Bhuma, M., & Agarwal, M. (2018) · International Conference on Pervasive Computing Technologies for Healthcare · doi:10.1145/3240925.3240975 · Cited by 6
Gibb, F. W., Jennings, P. L., Leelarathna, L., & Wilmot, E. G. (2020) · British Journal of Diabetes · doi:10.15277/bjd.2020.240 · Cited by 6
FDA Supports Standardized Reporting and Analysis on CGM Devices
Serrano, K. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.8313 · Cited by 6
Chan, A. J., & Halperin, I. J. (2020) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2020.08.103 · Cited by 6
Martyn-Nemeth, P. A., Duffecy, J. L., Quinn, L., Reutrakul, S., Steffen, A. D., Burke, L., Withington, M. H. C., Irsheed, G. A., Perez, R., Park, M., et al. (2022) · Trials · doi:10.1186/s13063-022-06565-6 · Cited by 6
“Variables to be derived from the CGM are mean glucose level, standard deviation (SD), coefficient of variation (CV), percentage of time spent in range (70–180 mg/dL), percentage of time < 70 mg/dl, and percentage of time ≥ 180 mg/dl [65, 66].”
BackgroundCui, R., Nolan, C. J., Daskalaki, E., & Suominen, H. (2023) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc40787.2023.10340094 · Cited by 8
Thomas, A., Haak, T., Tombek, A., Kulzer, B., Ehrmann, D., Kordonouri, O., Kröger, J., Schubert-Olesen, O., Kolassa, R., Siegmund, T., et al. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241267768 · Cited by 5
“15,16 17-19 More recently, another parameter was proposed that overcomes some of the limitations of the other parameters, which eventually will become the preferred metric for outcomes trials using CGM, the glycemic risk index (GRI).”
Orszag, A., Falappa, C. M., Lovblom, L. E., Partridge, H., Menezes, H. T., Boulet, G., Picton, P. E., Cafazzo, J. A., & Perkins, B. A. (2015) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2015.01.005 · Cited by 5
Metadata
- Authors
- Bergenstal, R. M., Ahmann, A. J., Bailey, T., Beck, R. W., Bissen, J., Buckingham, B., Deeb, L., Dolin, R. H., Garg, S. K., Goland, R., Hirsch, I. B., Klonoff, D. C., Kruger, D. F., Matfin, G., Mazze, R. S., Olson, B. A., Parkin, C., Peters, A., Powers, M. A., Rodriguez, H., Southerland, P., Strock, E. S., Tamborlane, W., & Wesley, D. M.
- Affiliations
- Park Nicollet Clinic (US)
- Oregon Health & Science University (US)
- University of California San Diego (US)
- Jaeb Center for Health Research (US)
- Lucile Packard Children's Hospital (US)
- Florida State University (US)
- Lantana Consulting Group
- University of Colorado Denver (US)
- Columbia University Irving Medical Center (US)
- University of Washington (US)
- University of California, San Francisco (US)
- Henry Ford Health (US)
- CGParkin Communications, Inc.
- University of Southern California (US)
- University of South Florida (US)
- Team Type 1, Atlanta, Georiga.
- Yale University (US)
- Corresponding author
- Bergenstal, Richard M.
- ORCID
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2013-03
- Volume
- 15
- Issue
- 3
- Pages
- 198-211
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2388765709
- Cited by
- 307
- References
- 55
- Citations by year
- 2026: 17 · 2025: 13 · 2024: 14 · 2023: 21 · 2022: 24 · 2021: 25 · 2020: 26 · 2019: 31 · 2018: 31 · 2017: 33 · 2016: 22 · 2015: 20 · 2014: 16 · 2013: 14
- Type
- conference abstract
- Language
- en
- Open access
- closed
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Leona M. and Harry B. Helmsley Charitable Trust
- U.S. Food and Drug Administration
- Abbott Diabetes Care
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- ambulatory glucose profile, continuous glucose monitoring, type 1 diabetes, glycemic control, clinical decision-making, expert panel recommendations
- MeSH terms
- Blood Glucose, Decision Making, Diabetes Mellitus, Female, Humans, Hyperglycemia, Hypoglycemia, Male, Reference Standards, Software, United States, Blood Glucose Self-Monitoring, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review