Zisser, Howard, Renard, Eric, Kovatchev, Boris, Cobelli, Claudio, Avogaro, Angelo, Nimri, Revital, Magni, Lalo, Buckingham, Bruce A., Chase, H. Peter, Doyle, Francis J., Lum, John, Calhoun, Peter, Kollman, Craig, Dassau, Eyal, Farret, Anne, Place, Jerome, Breton, Marc, Anderson, Stacey M., Dalla Man, Chiara, Del Favero, Simone, Bruttomesso, Daniela, Filippi, Alessio, Scotton, Rachele, Phillip, Moshe, Atlas, Eran, Muller, Ido, Miller, Shahar, Toffanin, Chiara, Raimondo, Davide Martino, De Nicolao, Giuseppe, Beck, Roy W., for the Control to Range Study Group · 2014 · Diabetes Technology & Therapeutics
Cited by 48

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

BACKGROUND: The Control to Range Study was a multinational artificial pancreas study designed to assess the time spent in the hypo- and hyperglycemic ranges in adults and adolescents with type 1 diabetes while under closed-loop control. The controller attempted to keep the glucose ranges between 70 and 180 mg/dL. A set of prespecified metrics was used to measure safety. RESEARCH DESIGN AND METHODS: We studied 53 individuals for approximately 22 h each during clinical research center admissions. Plasma glucose level was measured every 15-30 min (YSI clinical laboratory analyzer instrument [YSI, Inc., Yellow Springs, OH]). During the admission, subjects received three mixed meals (1 g of carbohydrate/kg of body weight; 100 g maximum) with meal announcement and automated insulin dosing by the controller. RESULTS: For adults, the mean of subjects' mean glucose levels was 159 mg/dL, and mean percentage of values 71-180 mg/dL was 66% overall (59% daytime and 82% overnight). For adolescents, the mean of subjects' mean glucose levels was 166 mg/dL, and mean percentage of values in range was 62% overall (53% daytime and 82% overnight). Whereas prespecified criteria for safety were satisfied by both groups, they were met at the individual level in adults only for combined daytime/nighttime and for isolated nighttime. Two adults and six adolescents failed to meet the daytime criterion, largely because of postmeal hyperglycemia, and another adolescent failed to meet the nighttime criterion. CONCLUSIONS: The control-to-range system performed as expected: faring better overnight than during the day and performing with variability between patients even after individualization based on patients' prior settings. The system had difficulty preventing postmeal excursions above target range.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2014.0066

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Cited by

Metadata

Authors
Zisser, H., Renard, E., Kovatchev, B., Cobelli, C., Avogaro, A., Nimri, R., Magni, L., Buckingham, B. A., Chase, H. P., Doyle, F. J., Lum, J., Calhoun, P., Kollman, C., Dassau, E., Farret, A., Place, J., Breton, M., Anderson, S. M., Dalla Man, C., Del Favero, S., Bruttomesso, D., Filippi, A., Scotton, R., Phillip, M., Atlas, E., Muller, I., Miller, S., Toffanin, C., Raimondo, D. M., De Nicolao, G., Beck, R. W., & for the Control to Range Study Group
Venue
Diabetes Technology & Therapeutics
Published
2014-10
Volume
16
Issue
10
Pages
613-622
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W1505393224
Cited by
48
References
10
FWCI
6.70
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.