Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin
The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
We determined the millimolar absorptivities of the four clinically relevant derivatives of fetal and adult human hemoglobin in the visible and near-infrared spectral range (450-1000 nm). As expected, spectral absorption curves of similar shape were found, but the small differences between fetal and adult hemoglobin absorptivity were important enough that they should be taken into account in multicomponent analysis of hemoglobin derivatives. Common pulse oximeters, however, involving light of 660 and 940 nm, are so insensitive to the presence of fetal hemoglobin that they can be used safely in neonates. The error in pulse oximetry caused by the presence of carboxyhemoglobin is insubstantial, but methemoglobin gives either an understimation or an overestimation at high or low oxygen saturation, respectively, the turning point being near 70% saturation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Neonatal Pulse Oximetry: Accuracy and Reliability
Hay, W. W., Brockway, J. M., & Eyzaguirre, M. (1989) · Pediatrics · doi:10.1542/peds.83.5.717 · Cited by 141
Pulse Oximeter Desaturation Due to Methemoglobinemia
Eisenkraft, J. B. (1988) · Anesthesiology · doi:10.1097/00000542-198802000-00018 · Cited by 116
Pulse Oximetry: An Alternative Method for the Assessment of Oxygenation in Newborn Infants
Jennis, M. S., & Peabody, J. L. (1987) · PEDIATRICS · doi:10.1542/peds.79.4.524 · Cited by 111
de Zwart, A. E. S., Buursma, A., Oeseburg, B., & Zijlstra, W. G. (1981) · Clinical Chemistry · doi:10.1093/clinchem/27.11.1903 · Cited by 77
Self-citationPulse Oximetry in Methemoglobinemia
Watcha, M. F., Connor, M. T., & Hing, A. V. (1989) · Archives of Pediatrics and Adolescent Medicine · doi:10.1001/archpedi.1989.02150190095030 · Cited by 69
Mendelson, Y., & Kent, J. C. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.30810 · Cited by 60
Absorption characteristics of human fetal hemoglobin at wavelengths used in pulse oximetry
Harris, A. P., SENDAK, M. J., Donham, R. T., Thomas, M., & Duncan, D. D. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01621812 · Cited by 56
Definition and notation of hemoglobin oxygen saturation.
Zijlstra, W. G., & Oeseburg, B. (1989) · PubMed · Cited by 22
Self-citation
Cited by
Non-contact, automated cardiac pulse measurements using video imaging and blind source separation
Poh, M.-Z., McDuff, D. J., & Picard, R. W. (2010) · Optics Express · doi:10.1364/oe.18.010762 · Cited by 1,643
Advancements in Noncontact, Multiparameter Physiological Measurements Using a Webcam
Poh, M.-Z., McDuff, D. J., & Picard, R. W. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2086456 · Cited by 1,486
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
“However, green-wavelength PPG devices are becoming increasingly popular due to the large intensity variations in modulation observed during the cardiac cycle for these wavelengths [7–9].”
BackgroundAll-organic optoelectronic sensor for pulse oximetry
Lochner, C. M., Khan, Y., Pierre, A., & Arias, A. C. (2014) · Nature Communications · doi:10.1038/ncomms6745 · Cited by 726
Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
Chan, E. D., Chan, M. M., & Chan, M. M. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.02.004 · Cited by 710
Optical properties of human skin
Lister, T., Wright, P. A., & Chappell, P. H. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.9.090901 · Cited by 467
Fiber optic probes for biomedical optical spectroscopy
Utzinger, U., & Richards-Kortum, R. R. (2003) · Journal of Biomedical Optics · doi:10.1117/1.1528207 · Cited by 443
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438
“The extinction coefficients values for HbO 2 and Hb can be found in the literature 61–64 for the wavelengths in the visible and infrared regions, but the discrepancy between the different sources is significant when aiming for accuracy of about 1%.”
Photoplethysmogram Analysis and Applications: An Integrative Review
Park, J., Seok, H. S., Kim, S.-S., & Shin, H. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2021.808451 · Cited by 422
Battery-free, stretchable optoelectronic systems for wireless optical characterization of the skin
Kim, J., Salvatore, G. A., Araki, H., Chiarelli, A. M., Xie, Z., Banks, A., Sheng, X., Liu, Y., Lee, J. W., Jang, K.-I., et al. (2016) · Science Advances · doi:10.1126/sciadv.1600418 · Cited by 403
A review of in‐vivo optical properties of human tissues and its impact on PDT
Sandell, J. L., & Zhu, T. C. (2011) · Journal of Biophotonics · doi:10.1002/jbio.201100062 · Cited by 384
Non‐invasive Assessment of Skin Microvascular Function in Humans: An Insight Into Methods
ROUSTIT, M., & CRACOWSKI, J. (2012) · Microcirculation · doi:10.1111/j.1549-8719.2011.00129.x · Cited by 373
Recent developments in optical coherence tomography for imaging the retina
van Velthoven, M. E. J., Faber, D. J., Verbraak, F. D., van Leeuwen, T. G., & de Smet, M. D. (2007) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2006.10.002 · Cited by 343
Miniaturized Battery‐Free Wireless Systems for Wearable Pulse Oximetry
Kim, J., Gutruf, P., Chiarelli, A. M., Heo, S. Y., Cho, K., Xie, Z., Banks, A., Han, S., Jang, K., Lee, J. W., et al. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201604373 · Cited by 334
Illuminating the BOLD signal: combined fMRI–fNIRS studies
Steinbrink, J., Villringer, A., Kempf, F., Haux, D., Boden, S., & Obrig, H. (2006) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2005.12.034 · Cited by 334
Cheung, C., Culver, J. P., Takahashi, K., Greenberg, J. H., & Yodh, A. G. (2001) · Physics in Medicine and Biology · doi:10.1088/0031-9155/46/8/302 · Cited by 308
Low, F. M., Hampton, M. B., Peskin, A. V., & Winterbourn, C. C. (2007) · Blood · doi:10.1182/blood-2006-09-048728 · Cited by 282
Oxygen Saturation-Dependent Absorption and Scattering of Blood
Faber, D. J., Aalders, M. C. G., Mik, E. G., Hooper, B. A., van Gemert, M. J. C., & van Leeuwen, T. G. (2004) · Physical Review Letters · doi:10.1103/physrevlett.93.028102 · Cited by 264
Zijlstra, W. G., & Buursma, A. (1997) · Comparative Biochemistry and Physiology Part B: Biochemistry and Molecular Biology · doi:10.1016/s0305-0491(97)00230-7 · Cited by 232
Self-citationNon‐Invasive Measurements of Skin Pigmentation In Situ
Stamatas, G. N., Zmudzka, B. Z., Kollias, N., & Beer, J. Z. (2004) · Pigment Cell Research · doi:10.1111/j.1600-0749.2004.00204.x · Cited by 222
Myers, D. E., Anderson, L. D., Seifert, R. P., Ortner, J. P., Cooper, C. E., Beilman, G. J., & Mowlem, J. D. (2005) · Journal of Biomedical Optics · doi:10.1117/1.1925250 · Cited by 187
Iftikar, F. I., & Hickey, A. J. R. (2013) · PLoS ONE · doi:10.1371/journal.pone.0064120 · Cited by 180
“As oxygenated haemoglobin absorbs at near infrared wavelengths around 940 nm and deoxygenated haemoglobin absorbs at 600 nm [55], a 3 Watt infrared LED (peak wavelength 940 nm) and a 3 Watt red light-emitting LED (peak wavelength 600 nm) were placed on one side of the fish near the base of the pectoral fins.”
BackgroundComputer simulation of the skin reflectance spectra
Meglinski, I. V., & Matcher, S. J. (2003) · Computer Methods and Programs in Biomedicine · doi:10.1016/s0169-2607(02)00099-8 · Cited by 179
“The absorption factors for oxy-and deoxy-hemoglobin and water are re-calculated from their spectra of extinction coefficients [24,25] with respect to their relati v e concentrations.”
BackgroundAssessment of Infant Brain Development With Frequency-Domain Near-Infrared Spectroscopy
Franceschini, M. A., Thaker, S., Themelis, G., Krishnamoorthy, K. K., Bortfeld, H., Diamond, S. G., Boas, D. A., Arvin, K., & Grant, P. E. (2007) · Pediatric Research · doi:10.1203/pdr.0b013e318045be99 · Cited by 177
“We compared the published spectra of adult and fetal Hb (31) and saw no significant differences in the spectral range used.”
MethodWireless, battery-free optoelectronic systems as subdermal implants for local tissue oximetry
Zhang, H., Gutruf, P., Meacham, K., Montana, M. C., Zhao, X., Chiarelli, A. M., Vázquez-Guardado, A., Norris, A., Lu, L., Guo, Q., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw0873 · Cited by 173
“e(l) of HbO2 and Hb are available from previous reports (27, 55).”
MethodBale, G., Elwell, C. E., & Tachtsidis, I. (2016) · Journal of Biomedical Optics · doi:10.1117/1.jbo.21.9.091307 · Cited by 169
“The hemoglobin species extinction spectra are included for reference (450 to 740 nm from Zijlstra et al.(27) and 740 to 915 nm from Mark Cope,(28) causing the discontinuity at 740 nm).”
Forensic quest for age determination of bloodstains
Bremmer, R. H., de Bruin, K. G., van Gemert, M. J. C., van Leeuwen, T. G., & Aalders, M. C. G. (2012) · Forensic Science International · doi:10.1016/j.forsciint.2011.07.027 · Cited by 168
(2011) · PubMed · Cited by 142
Algorithms for Monitoring Heart Rate and Respiratory Rate From the Video of a User’s Face
Sanyal, S., & Nundy, K. K. (2018) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2018.2818687 · Cited by 125
“as a sensor and a LED as an illuminant with appropriate illumination frequency [15], both in the case of transmission or reflectance mode.”
BackgroundRacial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120
“In some investigations, the absorption spectra of fetal and adult hemoglobin were found to be identical [26, 27], however, there is some evidence that an admixture of HbF and HbA results in impaired pulse oximetry accuracy (underestimation) by 3–4% [28, 29], although this effect occurs primarily at lower oxygen satura…”
BackgroundQuantification of ischemic muscle deoxygenation by near infrared time-resolved spectroscopy
Hamaoka, T., Katsumura, T., Murase, N., Nishio, S., Osada, T., Sako, T., Higuchi, H., Kurosawa, Y., Shimomitsu, T., Miwa, M., et al. (2000) · Journal of Biomedical Optics · doi:10.1117/1.429975 · Cited by 97
Metadata
- Authors
- Zijlstra, W. G., Buursma, A., & Meeuwsen-van der Roest, W. P.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 1991-09-01
- Volume
- 37
- Issue
- 9
- Pages
- 1633-1638
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W1950905025
- Cited by
- 665
- References
- 8
- FWCI
- 2.89
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.