Pathways for training and accreditation for transvenous lead extraction: a European Heart Rhythm Association position paper
Open accessThe DOI matches its Crossref record; 73 of the 73 references it cites were found in a registry or an index.
Abstract
The European Heart Rhythm Association charged the present writing committee with the task of producing a consensus document on training and accreditation for transvenous extraction of chronically implanted pacing and defibrillator leads.1,2 The core curriculum for the European Heart Rhythm Specialist includes implantation of heart rhythm devices but does not specifically mandate learning and accreditation for extraction techniques.3 The lack of such recommendations is related to the limited number of extractions available for training purposes and attaining competency. The present document focuses on the rising burden and the increasing complexity of techniques of lead extraction with an emphasis on the critical issues of training, accreditation, and documentation of results. There is also an educational component felt necessary to include by the Task Force in view of the specialized and emerging nature of this field. Where appropriate, a European perspective is presented and paediatric aspects are treated separately. The aim is to complement two recently published documents, one from the Heart Rhythm Society (HRS)4 on facilities, training, indications, and management of transvenous lead extraction and the other from the American Heart Association on device-related infections.5 The indications for lead extraction have not changed since these publications and are therefore not covered in this present document. Due to improving recognition of clinical need and wider indications, the implant rate of Cardiovascular Implantable Electronic Devices (CIED) continues to rise in most countries.6,7 The number of leads per patient is increasing with cardiac resynchronisation therapy–pacemaker/defibrillator, upgrades and a higher proportion of dual vs. single-chamber devices. As life expectancy has risen, so have the number of generator and lead changes despite advances in technology. Product advisories are inevitable despite overall improvements in reliability and have led to surges in extraction.8–10 Currently, infection accounts for approximately two-thirds of all extractions.2,11–15 Lead revisions and generator changes carry a greater risk of infection than new implants. Ageing patients are more likely to be immuno-compromised, to have septicaemia16 and to be on anti-thrombotic agents and anticoagulants. These potential risk factors for CIED complication are becoming more prevalent with time along with the risk of hospital acquired infections. The rates of CIED infection are growing out of proportion to the rise in new implants.17 Clinical manifestations of cardiac device infections may be local or systemic, and primary or secondary to septicaemia, particularly with staphylococcus.18–20 Although the majority present as a pocket infection or erosion in the absence of septicaemia,11,19 the intravascular portion of the lead is usually culture positive.5,21 Superficial skin infection overlying but not adherent to the pocket, that responds rapidly to antibiotics, may not require further action. This is in contrast to the treatment needed for pocket infections or erosions. Erosion may be caused by infection or mechanical factors or a combination of the two. Adherence of the generator or leads to the skin often precedes erosion. Most CIED national databases or registries do not detail the number of extractions, may underreport infections and do not identify or separate out erosions. Where data are available on complications and their treatment, the usual estimate is of prevalence using the number of new implants as a denominator. True incidence calculated from the number of complications expressed as a proportion of the time of exposure to the risk, in this case lead exposure, has rarely been calculated.22,23 Local extractions rates will vary on case mix and in-house expertise. Reports from specialized centres are often biased towards more complex systems and include external referrals. The majority of extraction procedures are transvenous and performed by cardiologists and/or cardiac surgeons.2,24 Taking all these considerations and limitations in mind, the Task Force has attempted to calculate the total transvenous extraction need based on 1.5 times the prevalence of reported infection in patients implanted with a CIED (Table 1). This broad approximation holds out the possibility of assessing trends and national differences in extraction requirements. The incidence of both infectious and non-infectious cases appears to be rising. It is assumed that the ratio is stable, but in future, the proportion of extractions that are for non-infectious causes may increase reflecting greater extractor experience and confidence. The prevalence of infection from registries and national databases varies widely but mostly within the range of 1–4%. Estimated need for transvenous lead extraction Estimated need for transvenous lead extraction Details of the criteria of clinical and technical success in lead extraction have recently been rev
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Baddour, L. M., Epstein, A. E., Erickson, C. C., Knight, B. P., Levison, M. E., Lockhart, P. B., Masoudi, F. A., Okum, E. J., Wilson, W. R., Beerman, L. B., et al. (2010) · Circulation · doi:10.1161/circulationaha.109.192665 · Cited by 1,030 · cited 8×
“The original indications for device implantation may no longer be present and re-implantation may not be required.(5,19,78,79) In patients with non-infectious indications for lead extraction, re-implantation can be performed either during the same session, or shortly afterwards, using the same or alternative venous ro…”
BackgroundInfluentialWilkoff, B. L., Love, C. J., Byrd, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., Epstein, L. M., Friedman, R. A., Kennergren, C. E. H., Mitkowski, P., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.05.020 · Cited by 1,024
Self-citationAn Entirely Subcutaneous Implantable Cardioverter–Defibrillator
Bardy, G. H., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R. E., Wright, D. J., Connelly, D. T., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa0909545 · Cited by 780 · cited 1×
“For ICD patients, an alternative re-implant strategy may become available in the absence of adequate venous access, that of an entirely subcutaneous ICD system.(80)”
BackgroundManagement and Outcome of Permanent Pacemaker and Implantable Cardioverter-Defibrillator Infections
Sohail, M. R., Uslan, D. Z., Khan, A. H., Friedman, P. A., Hayes, D. L., Wilson, W. R., Steckelberg, J. M., Stoner, S., & Baddour, L. M. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.01.072 · Cited by 751
Systemic Infection Related to Endocarditis on Pacemaker Leads: Clinical Presentation and Management
Klug, D., Lacroix, D., Savoye, C., Goullard, L., Grandmougin, D., Hennequin, J. L., Kacet, S., & Lekieffre, J. (1997) · Circulation · doi:10.1161/01.cir.95.8.2098 · Cited by 593
Lead Extraction in the Contemporary Setting: The LExICon Study
Wazni, O., Epstein, L. M., Carrillo, R. G., Love, C., Adler, S. W., Riggio, D. W., Karim, S. S., Bashir, J., Greenspon, A. J., DiMarco, J. P., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.08.070 · Cited by 584 · cited 2×
“but there does not appear to be any increased mortality associated with greater vegetation size.(34,66,67)”
BackgroundSelf-citationDiagnosis and Management of Infections Involving Implantable Electrophysiologic Cardiac Devices
Chua, J. D., Wilkoff, B. L., Lee, I., Juratli, N., Longworth, D. L., & Gordon, S. M. (2000) · Annals of Internal Medicine · doi:10.7326/0003-4819-133-8-200010170-00011 · Cited by 511 · cited 1×
“The original indications for device implantation may no longer be present and re-implantation may not be required.(5,19,78,79) In patients with non-infectious indications for lead extraction, re-implantation can be performed either during the same session, or shortly afterwards, using the same or alternative venous ro…”
BackgroundSelf-citationWilkoff, B. L., Byrd, C. L., Love, C. J., Hayes, D. L., Sellers, T. D., Schaerf, R., Parsonnet, V., Epstein, L. M., Sorrentino, R. A., & Reiser, C. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00074-1 · Cited by 438 · cited 2×
“Perceptions of lead extraction safety and effectiveness and the outcomes of patients undergoing transvenous lead extraction have been based on this and following reports.(32,42,87) This information prompted and guided the development of new extraction tools and techniques.”
BackgroundSelf-citationVOIGT, A., SHALABY, A., & SABA, S. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02569.x · Cited by 396 · cited 2×
“The rates of CIED infection are growing out of proportion to the rise in new implants.(17) Clinical manifestations of cardiac device infections may be local or systemic, and primary or secondary to septicaemia, particularly with staphylococcus.”
BackgroundPermanent Pacemaker and Implantable Cardioverter Defibrillator Infection: A Population-Based Study
Uslan, D. Z. (2007) · Archives of Internal Medicine · doi:10.1001/archinte.167.7.669 · Cited by 391 · cited 1×
“True incidence calculated from the number of complications expressed as a proportion of the time of exposure to the risk, in this case lead exposure, has rarely been calculated.(22,23) Local extractions rates will vary on case mix and in-house expertise.”
Background
Cited by
Habib, G., Lancellotti, P., Antunes, M. J., Bongiorni, M. G., Casalta, J.-P., Del Zotti, F., Dulgheru, R., El Khoury, G., Erba, P. A., Iung, B., et al. (2015) · European Heart Journal · doi:10.1093/eurheartj/ehv319 · Cited by 4,997
Self-citation2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
Self-citation2020 ESC guidelines on sports cardiology and exercise in patients with cardiovascular disease
Pelliccia, A., Sharma, S., Gati, S., Bäck, M., Börjesson, M., Caselli, S., Collet, J.-P., Corrado, D., Drezner, J. A., Halle, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehaa605 · Cited by 1,958
Kusumoto, F. M., Schoenfeld, M. H., Wilkoff, B. L., Berul, C. I., Birgersdotter-Green, U. M., Carrillo, R., Cha, Y.-M., Clancy, J., Deharo, J.-C., Ellenbogen, K. A., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.09.001 · Cited by 1,199
Self-citation2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
Self-citationBongiorni, M. G., Kennergren, C., Butter, C., Deharo, J. C., Kutarski, A., Rinaldi, C. A., Romano, S. L., Maggioni, A. P., Andarala, M., Auricchio, A., et al. (2017) · European Heart Journal · doi:10.1093/eurheartj/ehx080 · Cited by 496
Self-citationBlomström-Lundqvist, C., Traykov, V., Erba, P. A., Burri, H., Nielsen, J. C., Bongiorni, M. G., Poole, J., Boriani, G., Costa, R., Deharo, J.-C., et al. (2020) · EP Europace · doi:10.1093/europace/euz246 · Cited by 423
Self-citationBongiorni, M. G., Burri, H., Deharo, J. C., Starck, C., Kennergren, C., Saghy, L., Rao, A., Tascini, C., Lever, N., Kutarski, A., et al. (2018) · EP Europace · doi:10.1093/europace/euy050 · Cited by 401
Self-citationRaatikainen, M. J. P., Arnar, D. O., Merkely, B., Nielsen, J. C., Hindricks, G., Heidbuchel, H., & Camm, J. (2017) · EP Europace · doi:10.1093/europace/eux258 · Cited by 323
Blomström-Lundqvist, C., Traykov, V., Erba, P. A., Burri, H., Nielsen, J. C., Bongiorni, M. G., Poole, J., Boriani, G., Costa, R., Deharo, J.-C., et al. (2020) · European Journal of Cardio-Thoracic Surgery · doi:10.1093/ejcts/ezz296 · Cited by 318
Self-citationLinde, C., Bongiorni, M. G., Birgersdotter-Green, U., Curtis, A. B., Deisenhofer, I., Furokawa, T., Gillis, A. M., Haugaa, K. H., Lip, G. Y. H., Van Gelder, I., et al. (2018) · EP Europace · doi:10.1093/europace/euy067 · Cited by 296
Self-citationBlomström-Lundqvist, C., Traykov, V., Erba, P. A., Burri, H., Nielsen, J. C., Bongiorni, M. G., Poole, J., Boriani, G., Costa, R., Deharo, J.-C., et al. (2020) · European Heart Journal · doi:10.1093/eurheartj/ehaa010 · Cited by 222
Self-citationRaatikainen, M. J. P., Arnar, D. O., Zeppenfeld, K., Merino, J. L., Levya, F., Hindriks, G., & Kuck, K.-H. (2015) · Europace · doi:10.1093/europace/euu300 · Cited by 180
Metadata
- Authors
- Authors (EHRA Task Force Members), Deharo, J. C., Bongiorni, M. G., Rozkovec, A., Bracke, F., Defaye, P., Fernandez-Lozano, I., Golzio, P. G., Hansky, B., Kennergren, C., Manolis, A. S., Mitkowski, P., Platou, E. S., reviewers, E., Love, C., & Wilkoff, B.
- Venue
- Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2011-12-13
- Volume
- 14
- Issue
- 1
- Pages
- 124-134
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- OpenAlex
- W2011130343
- Cited by
- 210
- References
- 73
- FWCI
- 7.79
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.