Pathways for training and accreditation for transvenous lead extraction: a European Heart Rhythm Association position paper
Open accessThe DOI matches its Crossref record; 73 of the 73 references it cites were found in a registry or an index.
Abstract
The European Heart Rhythm Association charged the present writing committee with the task of producing a consensus document on training and accreditation for transvenous extraction of chronically implanted pacing and defibrillator leads.1,2 The core curriculum for the European Heart Rhythm Specialist includes implantation of heart rhythm devices but does not specifically mandate learning and accreditation for extraction techniques.3 The lack of such recommendations is related to the limited number of extractions available for training purposes and attaining competency. The present document focuses on the rising burden and the increasing complexity of techniques of lead extraction with an emphasis on the critical issues of training, accreditation, and documentation of results. There is also an educational component felt necessary to include by the Task Force in view of the specialized and emerging nature of this field. Where appropriate, a European perspective is presented and paediatric aspects are treated separately. The aim is to complement two recently published documents, one from the Heart Rhythm Society (HRS)4 on facilities, training, indications, and management of transvenous lead extraction and the other from the American Heart Association on device-related infections.5 The indications for lead extraction have not changed since these publications and are therefore not covered in this present document. Due to improving recognition of clinical need and wider indications, the implant rate of Cardiovascular Implantable Electronic Devices (CIED) continues to rise in most countries.6,7 The number of leads per patient is increasing with cardiac resynchronisation therapy–pacemaker/defibrillator, upgrades and a higher proportion of dual vs. single-chamber devices. As life expectancy has risen, so have the number of generator and lead changes despite advances in technology. Product advisories are inevitable despite overall improvements in reliability and have led to surges in extraction.8–10 Currently, infection accounts for approximately two-thirds of all extractions.2,11–15 Lead revisions and generator changes carry a greater risk of infection than new implants. Ageing patients are more likely to be immuno-compromised, to have septicaemia16 and to be on anti-thrombotic agents and anticoagulants. These potential risk factors for CIED complication are becoming more prevalent with time along with the risk of hospital acquired infections. The rates of CIED infection are growing out of proportion to the rise in new implants.17 Clinical manifestations of cardiac device infections may be local or systemic, and primary or secondary to septicaemia, particularly with staphylococcus.18–20 Although the majority present as a pocket infection or erosion in the absence of septicaemia,11,19 the intravascular portion of the lead is usually culture positive.5,21 Superficial skin infection overlying but not adherent to the pocket, that responds rapidly to antibiotics, may not require further action. This is in contrast to the treatment needed for pocket infections or erosions. Erosion may be caused by infection or mechanical factors or a combination of the two. Adherence of the generator or leads to the skin often precedes erosion. Most CIED national databases or registries do not detail the number of extractions, may underreport infections and do not identify or separate out erosions. Where data are available on complications and their treatment, the usual estimate is of prevalence using the number of new implants as a denominator. True incidence calculated from the number of complications expressed as a proportion of the time of exposure to the risk, in this case lead exposure, has rarely been calculated.22,23 Local extractions rates will vary on case mix and in-house expertise. Reports from specialized centres are often biased towards more complex systems and include external referrals. The majority of extraction procedures are transvenous and performed by cardiologists and/or cardiac surgeons.2,24 Taking all these considerations and limitations in mind, the Task Force has attempted to calculate the total transvenous extraction need based on 1.5 times the prevalence of reported infection in patients implanted with a CIED (Table 1). This broad approximation holds out the possibility of assessing trends and national differences in extraction requirements. The incidence of both infectious and non-infectious cases appears to be rising. It is assumed that the ratio is stable, but in future, the proportion of extractions that are for non-infectious causes may increase reflecting greater extractor experience and confidence. The prevalence of infection from registries and national databases varies widely but mostly within the range of 1–4%. Estimated need for transvenous lead extraction Estimated need for transvenous lead extraction Details of the criteria of clinical and technical success in lead extraction have recently been rev
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Baddour, L. M., Epstein, A. E., Erickson, C. C., Knight, B. P., Levison, M. E., Lockhart, P. B., Masoudi, F. A., Okum, E. J., Wilson, W. R., Beerman, L. B., et al. (2010) · Circulation · doi:10.1161/circulationaha.109.192665 · Cited by 1,030 · cited 8×
“The original indications for device implantation may no longer be present and re-implantation may not be required.(5,19,78,79) In patients with non-infectious indications for lead extraction, re-implantation can be performed either during the same session, or shortly afterwards, using the same or alternative venous ro…”
BackgroundInfluentialWilkoff, B. L., Love, C. J., Byrd, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., Epstein, L. M., Friedman, R. A., Kennergren, C. E. H., Mitkowski, P., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.05.020 · Cited by 1,024
Self-citationAn Entirely Subcutaneous Implantable Cardioverter–Defibrillator
Bardy, G. H., Smith, W. M., Hood, M. A., Crozier, I. G., Melton, I. C., Jordaens, L., Theuns, D., Park, R. E., Wright, D. J., Connelly, D. T., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa0909545 · Cited by 780 · cited 1×
“For ICD patients, an alternative re-implant strategy may become available in the absence of adequate venous access, that of an entirely subcutaneous ICD system.(80)”
BackgroundManagement and Outcome of Permanent Pacemaker and Implantable Cardioverter-Defibrillator Infections
Sohail, M. R., Uslan, D. Z., Khan, A. H., Friedman, P. A., Hayes, D. L., Wilson, W. R., Steckelberg, J. M., Stoner, S., & Baddour, L. M. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.01.072 · Cited by 751
Systemic Infection Related to Endocarditis on Pacemaker Leads: Clinical Presentation and Management
Klug, D., Lacroix, D., Savoye, C., Goullard, L., Grandmougin, D., Hennequin, J. L., Kacet, S., & Lekieffre, J. (1997) · Circulation · doi:10.1161/01.cir.95.8.2098 · Cited by 593
Lead Extraction in the Contemporary Setting: The LExICon Study
Wazni, O., Epstein, L. M., Carrillo, R. G., Love, C., Adler, S. W., Riggio, D. W., Karim, S. S., Bashir, J., Greenspon, A. J., DiMarco, J. P., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.08.070 · Cited by 584 · cited 2×
“but there does not appear to be any increased mortality associated with greater vegetation size.(34,66,67)”
BackgroundSelf-citationDiagnosis and Management of Infections Involving Implantable Electrophysiologic Cardiac Devices
Chua, J. D., Wilkoff, B. L., Lee, I., Juratli, N., Longworth, D. L., & Gordon, S. M. (2000) · Annals of Internal Medicine · doi:10.7326/0003-4819-133-8-200010170-00011 · Cited by 511 · cited 1×
“The original indications for device implantation may no longer be present and re-implantation may not be required.(5,19,78,79) In patients with non-infectious indications for lead extraction, re-implantation can be performed either during the same session, or shortly afterwards, using the same or alternative venous ro…”
BackgroundSelf-citationWilkoff, B. L., Byrd, C. L., Love, C. J., Hayes, D. L., Sellers, T. D., Schaerf, R., Parsonnet, V., Epstein, L. M., Sorrentino, R. A., & Reiser, C. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00074-1 · Cited by 438 · cited 2×
“Perceptions of lead extraction safety and effectiveness and the outcomes of patients undergoing transvenous lead extraction have been based on this and following reports.(32,42,87) This information prompted and guided the development of new extraction tools and techniques.”
BackgroundSelf-citationVOIGT, A., SHALABY, A., & SABA, S. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02569.x · Cited by 396 · cited 2×
“The rates of CIED infection are growing out of proportion to the rise in new implants.(17) Clinical manifestations of cardiac device infections may be local or systemic, and primary or secondary to septicaemia, particularly with staphylococcus.”
BackgroundPermanent Pacemaker and Implantable Cardioverter Defibrillator Infection: A Population-Based Study
Uslan, D. Z. (2007) · Archives of Internal Medicine · doi:10.1001/archinte.167.7.669 · Cited by 391 · cited 1×
“True incidence calculated from the number of complications expressed as a proportion of the time of exposure to the risk, in this case lead exposure, has rarely been calculated.(22,23) Local extractions rates will vary on case mix and in-house expertise.”
BackgroundJohansen, J. B., Jorgensen, O. D., Moller, M., Arnsbo, P., Mortensen, P. T., & Nielsen, J. C. (2011) · European Heart Journal · doi:10.1093/eurheartj/ehq497 · Cited by 376 · cited 1×
“True incidence calculated from the number of complications expressed as a proportion of the time of exposure to the risk, in this case lead exposure, has rarely been calculated.(22,23) Local extractions rates will vary on case mix and in-house expertise.”
BackgroundPacemaker and ICD Generator Malfunctions: Analysis of Food and Drug Administration Annual Reports
Maisel, W. H., Moynahan, M., Zuckerman, B. D., Gross, T. P., Tovar, O. H., Tillman, D.-B., & Schultz, D. B. (2006) · JAMA · doi:10.1001/jama.295.16.1901 · Cited by 353
NORTH AMERICAN SOCIETY OF PACING AND ELECTROPHYSIOLOGY LEAD EXTRACTION CONFERENCE FACULTY, LOVE, C. J., WILKOFF, B. L., BYRD, C. L., BELOTT, P. H., BRINKER, J. A., FEARNOT, N. E., FRIEDMAN, R. A., FURMAN, S., GOODE, L. B., et al. (2000) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2000.tb00845.x · Cited by 353
Self-citationClinical Study of the Laser Sheath for Lead Extraction: The Total Experience in the United States
BYRD, C. L., WILKOFF, B. L., LOVE, C. J., SELLERS, T. D., & REISER, C. (2002) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2002.t01-1-00804.x · Cited by 308 · cited 1×
“In an early study involving mainly laser extraction, the complication rate declined substantially during the first 30 clinical cases, with further improvements till 400 cases.(33) Complete procedural success was reduced when 60 or fewer laser assisted lead extraction procedures were accomplished over the prior 4 years…”
BackgroundSelf-citationChamis, A. L., Peterson, G. E., Cabell, C. H., Corey, G. R., Sorrentino, R. A., Greenfield, R. A., Ryan, T., Reller, L. B., & Fowler, V. G. (2001) · Circulation · doi:10.1161/hc3401.095097 · Cited by 292
Gould, P. A. (2006) · JAMA · doi:10.1001/jama.295.16.1907 · Cited by 288
Intravascular Extraction of Problematic or Infected Permanent Pacemaker Leads: 1994–1996
BYRD, C. L., WILKOFF, B. L., LOVE, C. J., SELLERS, T. D., TURK, K. T., REEVES, R., YOUNG, R., CREVEY, B., KUTALEK, S. P., FREEDMAN, R., et al. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00628.x · Cited by 279
Self-citationFive‐Years Experience with Intravascular Lead Extraction
SMITH, H. J., FEARNOT, N. E., BYRD, C. L., WILKOFF, B. L., LOVE, C. J., SELLERS, T. D., & DATABASE, F. T. U. S. L. E. (1994) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1994.tb03792.x · Cited by 278
Self-citationUslan, D. Z. (2007) · Archives of Internal Medicine · doi:10.1001/archinte.167.8.834 · Cited by 274
Hauser, R. G., Katsiyiannis, W. T., Gornick, C. C., Almquist, A. K., & Kallinen, L. M. (2010) · Europace · doi:10.1093/europace/eup375 · Cited by 264
Pacemaker lead infection: echocardiographic features, management, and outcome
Victor, F., De Place, C., Camus, C., Le Breton, H., Leclercq, C., Pavin, D., Mabo, P., & Daubert, C. (1999) · Heart · doi:10.1136/hrt.81.1.82 · Cited by 221
KURTZ, S. M., OCHOA, J. A., LAU, E., SHKOLNIKOV, Y., PAVRI, B. B., FRISCH, D., & GREENSPON, A. J. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02670.x · Cited by 217 · cited 1×
“The population requiring extraction is getting older with more complex device systems and prior procedures and more co-morbidities.(45) For the largest published single or multicentre series, each reporting on 975 or more leads, major life threatening complications occurred in 0.”
BackgroundJones, S. O., Eckart, R. E., Albert, C. M., & Epstein, L. M. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.01.009 · Cited by 216
A single-centre experience of over one thousand lead extractions
Kennergren, C., Bjurman, C., Wiklund, R., & Gabel, J. (2009) · Europace · doi:10.1093/europace/eup054 · Cited by 205
Self-citationLocal symptoms at the site of pacemaker implantation indicate latent systemic infection
Klug, D., Wallet, F., Lacroix, D., Marquié, C., Kouakam, C., Kacet, S., & Courcol, R. (2004) · Heart · doi:10.1136/hrt.2003.010595 · Cited by 195 · cited 2×
“20 Although the majority present as a pocket infection or erosion in the absence of septicaemia,(11,19) the intravascular portion of the lead is usually culture positive.(5,21) Superficial skin infection overlying but not”
BackgroundPacemaker and ICD Generator Reliability: Meta-analysis of Device Registries
Maisel, W. H. (2006) · JAMA · doi:10.1001/jama.295.16.1929 · Cited by 170
Extraction of Transvenous Pacing and ICD Leads
SMITH, M. C., & LOVE, C. J. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01079.x · Cited by 167 · cited 1×
“The usual approach is a long sheath introduced through the femoral vein and used in conjunction with any number of positioning and grasping tools.(38)”
MethodSelf-citationMOND, H. G., IRWIN, M., ECTOR, H., & PROCLEMER, A. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01164.x · Cited by 165 · cited 1×
“Due to improving recognition of clinical need and wider indications, the implant rate of Cardiovascular Implantable Electronic Devices (CIED) continues to rise in most countries.(6,7) The number of leads per patient is increasing with cardiac resynchronisation therapy–pacemaker/defibrillator, upgrades and a higher pro…”
BackgroundHauser, R. G., Hayes, D. L., Kallinen, L. M., Cannom, D. S., Epstein, A. E., Almquist, A. K., Song, S. L., Tyers, G. F. O., Vlay, S. C., & Irwin, M. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2006.10.009 · Cited by 165 · cited 1×
“Such malfunction may exist at implant or develop later.(25,26) Despite meticulous preparation of the skin, bacterial contamination of the generator occurs in up to one-half of all implants.”
BackgroundGrammes, J. A., Schulze, C., Al-Bataineh, M., Yesenosky, G. A., Saari, C. S., Vrabel, M. J., Horrow, J. C., Chowdhury, M., Fontaine, J. M., & Kutalek, S. P. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.11.034 · Cited by 156 · cited 2×
“but there does not appear to be any increased mortality associated with greater vegetation size.(34,66,67)”
BackgroundMethodRisk Factors for Mortality in Patients With Cardiac Device-Related Infection
Baman, T. S., Gupta, S. K., Valle, J. A., & Yamada, E. (2009) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.108.816868 · Cited by 147
Laser-assisted lead extraction: the European experience
Kennergren, C. E. H., Bucknall, C. A., Butter, C., CHARLES, R. G., Fuhrer, J., Grosfeld, M. D., Tavernier, R., Morgado, T. B., Mortensen, P. T., Paul, V. E., et al. (2007) · EP Europace · doi:10.1093/europace/eum098 · Cited by 147 · cited 1×
“Perceptions of lead extraction safety and effectiveness and the outcomes of patients undergoing transvenous lead extraction have been based on this and following reports.(32,42,87) This information prompted and guided the development of new extraction tools and techniques.”
BackgroundSelf-citationRisk Factors for Venous Obstruction in Children with Transvenous Pacing Leads
Figa, F. H., McCrindle, B. W., Bigras, J., Hamilton, R. M., & Gow, R. M. (1997) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1997.tb03594.x · Cited by 143 · cited 1×
“Venous status Occluded or severely stenosed Higher risk with greater lead cross-sectional area in the young.(48) Limited access for additional procedures”
BackgroundThe Role of Swab and Tissue Culture in the Diagnosis of Implantable Cardiac Device Infection
Chua, J. D., Abdul‐Karim, A., Mawhorter, S. D., Procop, G. W., TCHOU, P. J., Niebauer, M. J., Saliba, W. I., Schweikert, R. A., & Wilkoff, B. L. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.00268.x · Cited by 137 · cited 2×
“Pocket swabs are less likely to be positive than culturing of pocket tissue.(81) Extravascular and intravascular lead segments (the pin and lead tip) are more likely to be positive than pocket tissue(82,83) and more so than blood cultures.”
BackgroundSelf-citationRohacek, M., Weisser, M., Kobza, R., Schoenenberger, A. W., Pfyffer, G. E., Frei, R., Erné, P., & Trampuž, A. (2010) · Circulation · doi:10.1161/circulationaha.109.906461 · Cited by 134
Lead Extraction Is Preferred for Lead Revisions and System Upgrades: When Less Is More
Maytin, M., Epstein, L. M., & Henrikson, C. A. (2010) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.110.954107 · Cited by 131
Venous Obstruction After Pacemaker Implantation
Korkeila, P., Nyman, K., Ylitalo, A., Koistinen, J. M., Karjalainen, P. P., Lund, J., & Airaksinen, J. K. E. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00650.x · Cited by 125 · cited 1×
“During the chronic phase of lead implant, thrombus may organize, leading to subtotal or total occlusion along the venous route, and formation of fibrous bridges or tunnels containing the pacing lead at points of adherence.(30,31) The commonest binding sites are located at the venous entry, especially under the clavicl…”
BackgroundEpstein, L. M., BYRD, C. L., Wilkoff, B. L., Love, C. J., Sellers, T. D., Hayes, D. L., & Reiser, C. (1999) · Circulation · doi:10.1161/01.cir.100.5.516 · Cited by 125
Self-citationKay, G. N., Brinker, J. A., Kawanishi, D. T., Love, C. J., Lloyd, M. A., Reeves, R. C., Pioger, G., Fee, J., Overland, M. K., Ensign, L. G., et al. (1999) · Circulation · doi:10.1161/01.cir.100.23.2344 · Cited by 118
Self-citationMeier-Ewert, H. K., Gray, M. E., & John, R. M. (2003) · American Heart Journal · doi:10.1016/s0002-8703(03)00188-1 · Cited by 118 · cited 1×
“but there does not appear to be any increased mortality associated with greater vegetation size.(34,66,67)”
BackgroundLead Extraction in Pediatric and Congenital Heart Disease Patients
Cecchin, F., Atallah, J., Walsh, E. P., Triedman, J. K., Alexander, M. E., & Berul, C. I. (2010) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.110.957324 · Cited by 112
Agarwal, S. K., Kamireddy, S., Němec, J., Voigt, A., & Saba, S. (2008) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2008.01283.x · Cited by 109
Risk of Failure of Transvenous Implantable Cardioverter-Defibrillator Leads
Borleffs, C. J. W., van Erven, L., van Bommel, R. J., van der Velde, E. T., van der Wall, E. E., Bax, J. J., Rosendaal, F. R., & Schalij, M. J. (2009) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.108.834093 · Cited by 108
Uslan, D. Z., Dowsley, T. F., Sohail, M. R., Hayes, D. L., Friedman, P. A., Wilson, W. R., Steckelberg, J. M., & Baddour, L. M. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02565.x · Cited by 105
Laser Lead Extraction: Predictors of Success and Complications
Roux, J.-F., Pagé, P., Dubuc, M., Thibault, B., Guerra, P. G., Macle, L., Roy, D., Talajic, M. R., & Khairy, P. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00652.x · Cited by 105
Cooper, J. M., Stephenson, E. A., Berul, C. I., Walsh, E. P., & Epstein, L. M. (2003) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2003.02500.x · Cited by 102
Novák, M., Dvorak, P., Kamarýt, P., Slana, B., & Lipoldová, J. (2009) · EP Europace · doi:10.1093/europace/eup216 · Cited by 99 · cited 1×
“During the chronic phase of lead implant, thrombus may organize, leading to subtotal or total occlusion along the venous route, and formation of fibrous bridges or tunnels containing the pacing lead at points of adherence.(30,31) The commonest binding sites are located at the venous entry, especially under the clavicl…”
BackgroundNeužil, P., Táborský, M., Rezek, Z., Vopálka, R., Šedivá, L., Niederle, P., & Reddy, V. Y. (2007) · EP Europace · doi:10.1093/europace/eul171 · Cited by 96 · cited 1×
“–34 The electrosurgical dissection sheath utilizes radiofrequency energy produced by a standard electrosurgical unit to cut through fibrous tissues.(35)”
MethodBracke, F. A. L. E., Meijer, A., & van Gelder, B. (2002) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2002.01037.x · Cited by 92
Self-citationFranceschi, F., Thuny, F., Giorgi, R., Sanaa, I., Peyrouse, E., Assouan, X., Prévôt, S., Bastard, E., Habib, G., & Deharo, J. (2009) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.02.045 · Cited by 88
Self-citationKleemann, T., Becker, T., Strauss, M., Dyck, N., Weisse, U., Saggau, W., Burkhardt, U., & Seidl, K. (2009) · EP Europace · doi:10.1093/europace/eup334 · Cited by 87
Pacemaker Lead Infection in Young Patients
Klug, D., Vaksmann, G., Jarwe, M., WALLET, F., Francart, C., Kacet, S., & Rey, C. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.t01-1-00215.x · Cited by 78
Intravascular pacemaker and defibrillator lead extraction: A state-of-the-art review
Verma, A., & Wilkoff, B. L. (2004) · Heart Rhythm · doi:10.1016/j.hrthm.2004.09.020 · Cited by 74
Self-citationPacemaker and Defibrillator Lead Extraction: Predictors of Mortality during Follow-Up
Hamid, S., Arujuna, A., Ginks, M., McPhail, M. J. W., Patel, N., Bucknall, C. A., & Rinaldi, C. A. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02601.x · Cited by 69
High prevalence of asymptomatic bacterial colonization of rhythm management devices
Pichlmaier, M., Marwitz, V., Kühn, C., Niehaus, M., Klein, G. G., Bara, C., Haverich, A., & Abraham, W. (2008) · EP Europace · doi:10.1093/europace/eun191 · Cited by 63
Learning Curve Characteristics of Pacing Lead Extraction with a Laser Sheath
Bracke, F. A. L. E., Meijer, A., & van Gelder, B. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb01172.x · Cited by 59 · cited 1×
“As for all procedural techniques there is a learning curve for extraction.(68) To become an independent operator for transvenous extraction requires the ability to master some, if not all of the different extraction techniques in a sufficiently large number of patients of an appropriate case mix.”
MethodSelf-citationTask Force 6: Training in Specialized Electrophysiology, Cardiac Pacing, and Arrhythmia Management
Naccarelli, G. V., CONTI, J. B., Dimarco, J., & Tracy, C. M. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.11.014 · Cited by 58
Cori, A. D., Bongiorni, M. G., Zucchelli, G., Segreti, L., Viani, S. M., Paperini, L., & Soldati, E. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2010.02879.x · Cited by 57
Self-citationNonthoracotomy Implantable Defibrillator Lead Extraction:
Saad, E., Saliba, W. I., Schweikert, R. A., Al-Khadra, A. S., Abdul‐Karim, A., Niebauer, M. J., & Wilkoff, B. L. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.00300.x · Cited by 54
Self-citationLead‐Associated Endocarditis: The Important Role of Methicillin‐Resistant Staphylococcus aureus
Greenspon, A. J., Rhim, E. S., Mark, G. E., DESIMONE, J., & Ho, R. T. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01039.x · Cited by 54
Gula, L. J., Ames, A., WOODBURN, A., MAtkins, J., McCORMICK, M., BELL, J., SINK, D., McConville, J. W., & Epstein, L. M. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.00163.x · Cited by 53
How to select patients for lead extraction
Field, M. E., Jones, S. O., & Epstein, L. M. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2007.05.022 · Cited by 42
Franceschi, F., Dubuc, M., Deharo, J., Mancini, J., Pagé, P., Thibault, B., Koutbi, L., Prévôt, S., & Khairy, P. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.02.007 · Cited by 41
Self-citationBacteriology of infected extracted pacemaker and ICD leads
Anselmino, M., Vinci, M., Comoglio, C., Rinaldi, M., Bongiorni, M. G., Trevi, G. P., & Golzio, P. G. (2009) · Journal of Cardiovascular Medicine · doi:10.2459/jcm.0b013e32832b3585 · Cited by 40
Self-citationKlug, D., Wallet, F., Kacet, S., & Courcol, R. (2005) · American Heart Journal · doi:10.1016/j.ahj.2004.07.032 · Cited by 38 · cited 1×
“Extravascular and intravascular lead segments (the pin and lead tip) are more likely to be positive than pocket tissue(82,83) and more so than blood cultures.(82,83) Arrangements for appropriate storage and preservation of specimens until timely delivery to the microbiology lab should be established.”
BackgroundA European perspective on lead extraction: Part I
Kennergren, C. E. H. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2007.09.016 · Cited by 33 · cited 1×
“Chest computed tomography may be needed if extravascular coursing is suspected, but may not be diagnostic.(76,77) Echocardiography should be used to look for vegetations; their appearance and size will guide the decision to perform a transvenous or surgical approach.”
MethodSelf-citationHenrikson, C. A., & Brinker, J. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2007.10.018 · Cited by 32
Lead removal in young patients in view of lifelong pacing
Zartner, P. A., Wiebe, W., Toussaint-Goetz, N., & Schneider, M. (2010) · EP Europace · doi:10.1093/europace/euq059 · Cited by 31
Arrhythmia device lead extraction: Factors that necessitate laser assistance
Gula, L. J., Krahn, A. D., Yee, R. K., Skanes, A. C., Ghosh, N., & Klein, G. J. (2008) · Canadian Journal of Cardiology · doi:10.1016/s0828-282x(08)70681-0 · Cited by 30 · cited 1×
“For patients with infection, extraction has been reported to be easier.(43,55) There is limited experience and reporting of extraction of coronary sinus leads; the success rate for passive leads appears to be high but implant duration so far has been short.”
BackgroundPerformance and survival of transvenous defibrillation leads: need for a European data registry
Goette, A., Cantù, F., van Erven, L., Geelen, P., Halimi, F., Merino, J. L., & Morgan, J. M. (2008) · EP Europace · doi:10.1093/europace/eun301 · Cited by 28 · cited 1×
“The European Heart Rhythm Association charged the present writing committee with the task of producing a consensus document on training and accreditation for transvenous extraction of chronically implanted pacing and defibrillator leads.(1,2) The core curriculum for the European Heart Rhythm Specialist includes implan…”
MethodGolzio, P. G., Vinci, M., Anselmino, M., Comoglio, C., Rinaldi, M., Trevi, G. P., & Bongiorni, M. G. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.02257.x · Cited by 26
Self-citationCardiac Device Infection—or Not
Baddour, L. M. (2010) · Circulation · doi:10.1161/cir.0b013e3181de0334 · Cited by 20
Marijon, É., Guillebon, M. D., Bordachar, P., Jacob, S., Vahdat, O., Sidobre, L., Deplagne, A., Combes, N., Albenque, J., Clémenty, J., et al. (2009) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2009.01654.x · Cited by 10 · cited 1×
“The original indications for device implantation may no longer be present and re-implantation may not be required.(5,19,78,79) In patients with non-infectious indications for lead extraction, re-implantation can be performed either during the same session, or shortly afterwards, using the same or alternative venous ro…”
Background
Cited by
Habib, G., Lancellotti, P., Antunes, M. J., Bongiorni, M. G., Casalta, J.-P., Del Zotti, F., Dulgheru, R., El Khoury, G., Erba, P. A., Iung, B., et al. (2015) · European Heart Journal · doi:10.1093/eurheartj/ehv319 · Cited by 4,997
Self-citation2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
Self-citation2020 ESC guidelines on sports cardiology and exercise in patients with cardiovascular disease
Pelliccia, A., Sharma, S., Gati, S., Bäck, M., Börjesson, M., Caselli, S., Collet, J.-P., Corrado, D., Drezner, J. A., Halle, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehaa605 · Cited by 1,958
Kusumoto, F. M., Schoenfeld, M. H., Wilkoff, B. L., Berul, C. I., Birgersdotter-Green, U. M., Carrillo, R., Cha, Y.-M., Clancy, J., Deharo, J.-C., Ellenbogen, K. A., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.09.001 · Cited by 1,199
Self-citation2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
Self-citation
Metadata
- Authors
- Authors (EHRA Task Force Members), Deharo, J. C., Bongiorni, M. G., Rozkovec, A., Bracke, F., Defaye, P., Fernandez-Lozano, I., Golzio, P. G., Hansky, B., Kennergren, C., Manolis, A. S., Mitkowski, P., Platou, E. S., reviewers, E., Love, C., & Wilkoff, B.
- Venue
- Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2011-12-13
- Volume
- 14
- Issue
- 1
- Pages
- 124-134
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- OpenAlex
- W2011130343
- Cited by
- 210
- References
- 73
- FWCI
- 7.79
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.