Positron Emission Tomography Partial Volume Correction: Estimation and Algorithms
Open accessThe DOI matches its Crossref record; 37 of the 37 references it cites were found in a registry or an index.
Abstract
Partial volume effects in positron emission tomography (PET) lead to quantitative under- and over-estimations of the regional concentrations of radioactivity in reconstructed images and corresponding errors in derived functional or parametric images. The limited resolution of PET leads to “tissue-fraction” effects, reflecting underlying tissue heterogeneity, and “spillover” effects between regions. Addressing the former problem in general requires supplementary data, for example, coregistered high-resolution magnetic resonance images, whereas the latter effect can be corrected for with PET data alone if the point-spread function of the tomograph has been characterized. Analysis of otherwise homogeneous region-of-interest data ideally requires a combination of tissue classification and correction for the point-spread function. The formulation of appropriate algorithms for partial volume correction (PVC) is dependent on both the distribution of the signal and the distribution of the underlying noise. A mathematical framework has therefore been developed to accommodate both of these factors and to facilitate the development of new PVC algorithms based on the description of the problem. Several methodologies and algorithms have been proposed and implemented in the literature in order to address these problems. These methods do not, however, explicitly consider the noise model while differing in their underlying assumptions. The general theory for estimation of regional concentrations, associated error estimation, and inhomogeneity tests are presented in a weighted least squares framework. The analysis has been validated using both simulated and real PET data sets. The relations between the current algorithms and those published previously are formulated and compared. The incorporation of tensors into the formulation of the problem has led to the construction of computationally rapid algorithms taking into account both tissue-fraction and spillover effects. The suitability of their application to dynamic and static images is discussed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Golub, G. (2012) · doi:10.56021/9781421407944 · Cited by 34,505
(2005) · Statistics for Biology and Health · doi:10.1007/0-387-27255-0_9 · Cited by 20,133
McCullagh, P., & Nelder, J. A. (1989) · doi:10.1007/978-1-4899-3242-6 · Cited by 16,836
Lohrli, A. (1985) · Notes and Queries · doi:10.1093/notesj/32.3.377-a · Cited by 4,352
Generalized Inverse of Matrices and Its Applications
Banerjee, K. S. (1973) · Technometrics · doi:10.1080/00401706.1973.10489026 · Cited by 2,310
Axelsson, O. (1994) · Cambridge University Press eBooks · doi:10.1017/cbo9780511624100 · Cited by 2,085
Nonlinear spatial normalization using basis functions
Ashburner, J., & Friston, K. J. (1999) · Human Brain Mapping · doi:10.1002/(sici)1097-0193(1999)7:4<254::aid-hbm4>3.0.co;2-g · Cited by 1,884 · cited 2×
“This was performed using linear and nonlinear transformations (Ashburner and Friston, 1999) with parameters derived from the MRI data set associated with the template (Hammers et al.”
MethodEnhancement of MR Images Using Registration for Signal Averaging
Holmes, C. J., Hoge, R., Collins, L., Woods, R., Toga, A. W., & Evans, A. C. (1998) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-199803000-00032 · Cited by 1,417
Self-citationQuantitation in Positron Emission Computed Tomography: 1. Effect of Object Size
Hoffman, E. J., Huang, S.-C., & Phelps, M. E. (1979) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-197906000-00001 · Cited by 1,143 · cited 2×
“The second is the “point-spread effect,” or spillover between regions, again reflecting the limited resolution of the tomograph (Hoffman et al., 1979; Mazziotta et al., 1981; Leahy and Qi, 2000).”
BackgroundCorrection for partial volume effects in PET: principle and validation.
ROUSSET, O. G., Ma, Y., & Evans, A. C. (1998) · PubMed · Cited by 1,062
Self-citationMultimodal Image Coregistration and Partitioning—A Unified Framework
Ashburner, J., & Friston, K. (1997) · NeuroImage · doi:10.1006/nimg.1997.0290 · Cited by 927 · cited 1×
“The transformed region template and the MRI data sets were then coregistered with the PET data set (Ashburner and Friston, 1997).”
MethodGeneralized Inverse of Matrices and Its Applications
Mayne, A. J., Rao, C. R., & Mitra, S. K. (1972) · Operational Research Quarterly (1970-1977) · doi:10.2307/3007981 · Cited by 817
Analytic 3D image reconstruction using all detected events
Kinahan, P. E., & Rogers, J. G. (1989) · IEEE Transactions on Nuclear Science · doi:10.1109/23.34585 · Cited by 767 · cited 1×
“The images were reconstructed using the reprojection algorithm (Kinahan and Rogers, 1989) with the ramp and Colscher filters set at Nyquist cut-off frequency.”
MethodMüller-Gärtner, H. W., Links, J. M., Prince, J. L., Bryan, R. N., McVeigh, E., Leal, J. P., Davatzikos, C., & Frost, J. J. (1992) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1038/jcbfm.1992.81 · Cited by 756
Huang, S.-C., Hoffman, E. J., Phelps, M. E., & Kuhl, D. E. (1979) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-197912000-00018 · Cited by 443
Quantitation in Positron Emission Computed Tomography: 5. Physical–Anatomical Effects
Mazziotta, J. C., Phelps, M. E., Plummer, D., & Kuhl, D. E. (1981) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198110000-00029 · Cited by 337 · cited 2×
“The second is the “point-spread effect,” or spillover between regions, again reflecting the limited resolution of the tomograph (Hoffman et al., 1979; Mazziotta et al., 1981; Leahy and Qi, 2000).”
BackgroundA Single Scatter Simulation Technique for Scatter Correction in 3D PET
Watson, C. C., Newport, D., & Casey, M. E. (1996) · Computational Imaging and Vision · doi:10.1007/978-94-015-8749-5_18 · Cited by 333 · cited 2×
“A calculated attenuation correction and a modelbased scatter correction (Watson et al., 1996) were applied to the data during reconstruction.”
MethodCorrection of PET Data for Partial Volume Effects in Human Cerebral Cortex by MR Imaging
Meltzer, C. C., Leal, J. P., Mayberg, H. S., Wagner, H. N., & Frost, J. J. (1990) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-199007000-00011 · Cited by 329
Comparative evaluation of MR-based partial-volume correction schemes for PET.
MELTZER, C. C., Kinahan, P. E., Greer, P. J., Nichols, T. E., Comtat, C., Cantwell, M. N., Lin, M., & Price, J. C. (1999) · PubMed · Cited by 254
A convolution-subtraction scatter correction method for 3D PET
Bailey, D. L., & Meikle, S. R. (1994) · Physics in Medicine and Biology · doi:10.1088/0031-9155/39/3/009 · Cited by 228
Statistical approaches in quantitative positron emission tomography
Leahy, R. M., & Qi, J. (2000) · Statistics and Computing · doi:10.1023/a:1008946426658 · Cited by 215 · cited 4×
“The second is the “point-spread effect,” or spillover between regions, again reflecting the limited resolution of the tomograph (Hoffman et al., 1979; Mazziotta et al., 1981; Leahy and Qi, 2000).”
BackgroundMethodInfluentialDoes cerebral blood flow decline in healthy aging? A PET study with partial-volume correction.
MELTZER, C. C., Cantwell, M. N., Greer, P. J., Ben-Eliezer, D., Smith, G. S., Frank, G., Kaye, W. H., Houck, P. R., & Price, J. C. (2000) · PubMed · Cited by 208
Fast, accurate, and reproducible automatic segmentation of the brain in T1-weighted volume MRI data
Lemieux, L., Hagemann, G., Krakow, K., & Woermann, F. G. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199907)42:1<127::aid-mrm17>3.0.co;2-o · Cited by 186 · cited 2×
“, in press) The subject’s own MRI was also segmented (Lemieux et al., 1999) into grey matter, white matter, and CSF.”
BackgroundMethodBenzodiazepine receptors in focal epilepsy with cortical dysgenesis: An 11C‐flumazenil PET study
Richardson, M. P., Koepp, M. J., Duncan, J. S., Brooks, D. J., & Fish, D. R. (1996) · Annals of Neurology · doi:10.1002/ana.410400210 · Cited by 147 · cited 2×
“A [(11)C]flumazenil data set was obtained from an ongoing clinical study and acquired according to a protocol described previously (Koepp et al., 1996; Richardson et al., 1996).”
MethodCerebral benzodiazepine receptors in hippocampal sclerosis: An objective in vivo analysis
Koepp, M. J., Richardson, M. P., Brooks, D. J., Poline, J.-B., Paesschen, W. V., Friston, K. J., & Duncan, J. S. (1996) · Brain · doi:10.1093/brain/119.5.1677 · Cited by 115 · cited 2×
“A [(11)C]flumazenil data set was obtained from an ongoing clinical study and acquired according to a protocol described previously (Koepp et al., 1996; Richardson et al., 1996).”
MethodA Maximum Likelihood Method for Region-of-Interest Evaluation in Emission Tomography
Carson, R. E. (1986) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198607000-00021 · Cited by 113 · cited 1×
“Accurate characterization of the noise model in image space is difficult, but might be simplified by considering the problem in sinogram space (Carson, 1986).”
BackgroundImplementation and application of a brain template for multiple volumes of interest
Hammers, A., Koepp, M. J., Free, S. L., Brett, M., Richardson, M. P., Labbé, C., Cunningham, V. J., Brooks, D. J., & Duncan, J. (2002) · Human Brain Mapping · doi:10.1002/hbm.10016 · Cited by 100
Self-citationLabbé, C., Froment, J. C., Kennedy, A., Ashburner, J., & Cinotti, L. (1996) · Alzheimer Disease & Associated Disorders · doi:10.1097/00002093-199601030-00005 · Cited by 99 · cited 7×
“This assumption is J Cereb Blood Flow Metab, Vol. 22, No. 8, 2002 weaker than the assumptions that are made in some previous methods where stationarity was also assumed to facilitate computation (Labbé et al., 1996).”
BackgroundMethodResultInfluentialBailey, D. L., Young, H. E., Bloomfield, P. M., Meikle, S. R., Glass, D. M., Myers, M. J., Spinks, T. J., Watson, C. C., Luk, P., Peters, A. M., et al. (1997) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/bf01728302 · Cited by 96
A Statistical Method for the Analysis of Positron Emission Tomography Neuroreceptor Ligand Data
Aston, J. A. D., Gunn, R. N., Worsley, K. J., Ma, Y., Evans, A. C., & Dagher, A. (2000) · NeuroImage · doi:10.1006/nimg.2000.0620 · Cited by 87 · cited 2×
“Voxel-byvoxel parametric images of [(11)C]flumazenil volume of distribution (VD) were then produced from the brain uptake and arterial metabolite-corrected plasma input functions using a single compartmental tracer kinetic model and estimates of the variance of the parameters calculated (Aston et al., 2000) and is sho…”
MethodSelf-citationNoise characteristics of 3-D and 2-D PET images
Pajevic, S., Daube-Witherspoon, M. E., Bacharach, S. L., & Carson, R. E. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.668691 · Cited by 79 · cited 1×
“There is evidence that the variance is homogeneous across image (Pajevic et al., 1998).”
BackgroundIida, H., Law, I. G., Pakkenberg, B., Krarup‐Hansen, A., Eberl, S., Holm, S., Hansen, A. K., Gundersen, H. J. G., Thomsen, C. E., Svarer, C., et al. (2000) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1097/00004647-200008000-00009 · Cited by 73 · cited 6×
“The consequences of these effects on the signal have been investigated previously (Karp et al., 1991; Rousset et al., 1993; Iida et al., 2000), and a variety of correction methods, of differing complexities, have been proposed and applied to attempt accurate recovery of the signal in situations where regions are eithe…”
BackgroundMethodInfluentialRousset, O. G., Deep, P., Kuwabara, H., Evans, A. C., Gjedde, A. H., & Cumming, P. (2000) · Synapse · doi:10.1002/1098-2396(200008)37:2<81::aid-syn1>3.0.co;2-# · Cited by 71 · cited 2×
“Upwards of 300% increases in estimates of cerebral metabolism of [(18)F ]DOPA after PVC have been reported (Rousset et al., 2000).”
BackgroundSelf-citationLaw, I. G., Iida, H., Holm, S., Nour, S., Rostrup, E., Svarer, C., & Paulson, O. B. (2000) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1097/00004647-200008000-00010 · Cited by 61 · cited 1×
“Such corrections can have significant impact on the quantitative analysis of PET studies (Labbé et al., 1996; Meltzer et al., 2000; Law et al., 2000), particularly when it is necessary to take into account the effects of pathological or structural changes.”
BackgroundLabbé, C., Koepp, M. J., Ashburner, J., Spinks, T. J., Richardson, M. P., Duncan, J. S., & Cunningham, V. J. (1998) · Elsevier eBooks · doi:10.1016/b978-012161340-2/50011-1 · Cited by 52 · cited 3×
“…methods, of differing complexities, have been proposed and applied to attempt accurate recovery of the signal in situations where regions are either defined by anatomy, by tissue type, or both (Meltzer et al., 1990; Muller-Gartner et al., 1992; Labbé et al., 1996, 1998; Rousset et al., 1998).”
MethodResultSelf-citationNumerical approximation of the product of the square root of a matrix with a vector
Allen, E. J., Baglama, J., & Boyd, S. (2000) · Linear Algebra and its Applications · doi:10.1016/s0024-3795(00)00068-9 · Cited by 41 · cited 1×
“In order to evaluate the square root matrix equation, we make use of the Lanczos algorithm (Golub and Van Loan, 1983) and its extensions to matrix square roots (Allen et al., 2000).”
MethodSchneider, M. K., & Willsky, A. S. (2001) · SIAM Journal on Scientific Computing · doi:10.1137/s1064827599357292 · Cited by 33 · cited 1×
“These methods have previously been rigorously analyzed for deconvolution (Schneider and Willsky, 2001), but not for inhomogeneity testing.”
Method
Cited by
Partial-Volume Effect in PET Tumor Imaging
Soret, M., Bacharach, S. L., & Buvat, I. (2007) · Journal of Nuclear Medicine · doi:10.2967/jnumed.106.035774 · Cited by 1,549
Erlandsson, K., Buvat, I., Pretorius, P. H., Thomas, B. A., & Hutton, B. F. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/21/r119 · Cited by 733
Latent feature representation with stacked auto-encoder for AD/MCI diagnosis
Initiative, T. A. D. N., Suk, H.-I., Lee, S.-W., & Shen, D. (2015) · Brain Structure and Function · doi:10.1007/s00429-013-0687-3 · Cited by 526
Advances in multimodal data fusion in neuroimaging: Overview, challenges, and novel orientation
Zhang, Y.-D., Dong, Z., Wang, S.-H., Yu, X., Yao, X., Zhou, Q., Hu, H., Li, M., Jiménez-Mesa, C., Ramirez, J., et al. (2020) · Information Fusion · doi:10.1016/j.inffus.2020.07.006 · Cited by 453
Imaging dopamine receptors in humans with [11C]-(+)-PHNO: Dissection of D3 signal and anatomy
Tziortzi, A. C., Searle, G. E., Tzimopoulou, S., Salinas, C., Beaver, J. D., Jenkinson, M., Laruelle, M., Rabiner, E. A., & Gunn, R. N. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2010.06.044 · Cited by 452
Self-citation
Metadata
- Authors
- Aston, J. A. D., Cunningham, V. J., Asselin, M.-C., Hammers, A., Evans, A. C., & Gunn, R. N.
- Venue
- Journal of Cerebral Blood Flow & Metabolism
- Publisher
- SAGE Publications
- Published
- 2002-08
- Volume
- 22
- Issue
- 8
- Pages
- 1019-1034
- ISSN
- 0271-678X (ISSN-L)
- 1559-7016
- OpenAlex
- W2120892458
- Cited by
- 180
- References
- 37
- FWCI
- 4.43
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.