A Model for Cochlear Implant Electrode Insertion and Force Evaluation: Results with a New Electrode Design and Insertion Technique
The DOI matches its Crossref record; 31 of the 31 references it cites were found in a registry or an index.
Abstract
Objectives and Hypothesis: This study has the specific aim of evaluating the insertion characteristics of a new cochlear implant electrode. Techniques for evaluation of fluoroscopic real time mechanical insertion dynamics, histologic electrode position and trauma results, hydraulic force, and mechanical insertion forces are presented. In addition, this study should serve to present a novel model for cochlear implant electrode insertion evaluations. Study Design: Prospective analysis using a series of analytical techniques. Methods: All studies are conducted in fixed cadaveric temporal bones. Real time fluoroscopic insertion evaluations, histologic evaluations for trauma and electrode position in embedded bones, hydraulic measures, and mechanical intracochlear force measurements are conducted with a current and new electrode. Results: The Contour Advance electrode provides a more reliable and less traumatic insertion when deployed with the Advance Off Stylet technique. This is largely because of a reduction in intracochlear outer wall force generation. Fluoroscopic and histologic analysis reveal a smooth insertion without reliance on cochlear outer wall contact. No hydraulic forces were detected when measured from the superior semicircular canal ampulla. Conclusion: The model used for this study provides valuable information to cochlear implant surgeons and design engineers. The Contour Advance electrode, inserted with the Advance Off Stylet technique, represents an improvement over the Contour electrode inserted with the standard insertion technique.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Charge density and charge per phase as cofactors in neural injury induced by electrical stimulation
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.102812 · Cited by 708
A model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526
Electrical stimulation of the auditory nerve: The effect of electrode position on neural excitation
Shepherd, R. K., Hatsushika, S., & Clark, G. M. (1993) · Hearing Research · doi:10.1016/0378-5955(93)90265-3 · Cited by 360
Histopathology of Cochlear Implants in Humans
Nadol, J. B., Burgess, B. J., Gantz, B. J., Coker, N. J., Ketten, D. R., Kos, I., Roland, J. T., Shiao, J. Y., Eddington, D. K., Montandon, P., et al. (2001) · Annals of Otology, Rhinology & Laryngology · doi:10.1177/000348940111000914 · Cited by 292
Self-citationLeake, P. A., Hradek, G. T., & Snyder, R. L. (1999) · The Journal of Comparative Neurology · doi:10.1002/(sici)1096-9861(19991004)412:4<543::aid-cne1>3.0.co;2-3 · Cited by 276
Comparative Study of Cochlear Damage With Three Perimodiolar Electrode Designs
Eshraghi, A. A., Yang, N. W., & Balkany, T. J. (2003) · The Laryngoscope · doi:10.1097/00005537-200303000-00005 · Cited by 257
Thorne, M., Salt, A. N., DeMott, J. E., Henson, M. M., Henson, O. W., & Gewalt, S. L. (1999) · The Laryngoscope · doi:10.1097/00005537-199910000-00021 · Cited by 254
Leake, P. A., Hradek, G. T., Rebscher, S. J., & Snyder, R. L. (1991) · Hearing Research · doi:10.1016/0378-5955(91)90120-x · Cited by 206
Threshold, Comfortable Level and Impedance Changes as a Function of Electrode-Modiolar Distance
Saunders, E., Cohen, L., Aschendorff, A., Shapiro, W., Knight, M., Stecker, M., Richter, B., Waltzman, S., Tykocinski, M., Roland, T., et al. (2002) · Ear and Hearing · doi:10.1097/00003446-200202001-00004 · Cited by 197
Frijns, J. H. M., Briaire, J. J., & Grote, J. J. (2001) · Otology & Neurotology · doi:10.1097/00129492-200105000-00012 · Cited by 155
A multiple electrode cochlear implant
Clark, G. M., Tong, Y. C., Black, R., Forster, I. C., Patrick, J. F., & Dewhurst, D. J. (1977) · The Journal of Laryngology & Otology · doi:10.1017/s0022215100084607 · Cited by 147
Cochlear implant histopathology.
Linthicum, F. H., Fayad, J. N., Otto, S. R., Galey, F. R., & House, W. F. (1991) · PubMed · Cited by 143
Increased spiral ganglion cell survival in electrically stimulated, deafened guinea pig cochleae
Lousteau, R. J. (1987) · The Laryngoscope · doi:10.1288/00005537-198707000-00012 · Cited by 141
Banded Intracochlear Electrode Array: Evaluation of Insertion Trauma in Human Temporal Bones
Shepherd, R. K., Pyman, B. C., Clark, G. M., & Webb, R. L. (1985) · Annals of Otology Rhinology & Laryngology · doi:10.1177/000348948509400112 · Cited by 128
The Contour Electrode Array: Safety Study and Initial Patient Trials of a New Perimodiolar Design
Tykocinski, M., Saunders, E., Cohen, L. T., Treaba, C., Briggs, R. J. S., Gibson, P., Clark, G. M., & Cowan, R. S. C. (2001) · Otology & Neurotology · doi:10.1097/00129492-200101000-00007 · Cited by 114
Comparison of electrode position in the human cochlea using various perimodiolar electrode arrays
Tykocinski, M., Cohen, L. T., Pyman, B. C., ROLANDJR, T., Treaba, C. G., Palamara, J. E. A., DAHM, M., Shepherd, R. K., Xu, J., & Cowan, R. S. (2000) · American Journal of Otolaryngology · doi:10.1016/s0196-0709(00)80010-1 · Cited by 112
Cochlear Implant Deep Electrode Insertion: Extent of Insertional Trauma
Gstoettner, W. K., Plenk, H., Jr., Franz, P., Hamzavi, J. S., Baumgartner, W., Czerny, C., & Ehrenberger, K. (1997) · Acta Oto-Laryngologica · doi:10.3109/00016489709117786 · Cited by 105
Design and fabrication of an experimental cochlear prosthesis
Loeb, G. E., Byers, C. L., Rebscher, S. J., Casey, D., Fong, M., Schindler, R. A., Gray, R. F., & Merzenich, M. M. (1983) · Medical & Biological Engineering & Computing · doi:10.1007/bf02478489 · Cited by 86
Briggs, R., Tykocinski, M., Saunders, E., Hellier, W. P. L., Dahm, M. C., Pyman, B. C., & Clark, G. M. (2001) · Cochlear Implants International · doi:10.1179/cim.2001.2.2.135 · Cited by 83
An advanced multiple channel cochlear implant
McDermott, H. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32112 · Cited by 81
The Nucleus Contour Electrode Array: A Radiological and Histological Study
Richter, B., Aschendorff, A., Lohnstein, P., Husstedt, H. W., Nagursky, H., & Laszig, R. (2001) · The Laryngoscope · doi:10.1097/00005537-200103000-00023 · Cited by 75
Insertional trauma of multichannel cochlear implants
Welling, D. B., Hinojosa, R., Gantz, B. J., & Lee, J. (1993) · The Laryngoscope · doi:10.1288/00005537-199309000-00010 · Cited by 75
Electrode to modiolus proximity: A fluoroscopic and histologic analysis
Roland, J. T., Fishman, A. J., Alexiades, G., & Cohen, N. L. (2000) · American Journal of Otolaryngology · doi:10.1016/s0196-0709(00)80012-5 · Cited by 73
Self-citationStrategies to improve electrode positioning and safety in cochlear implants
Rebscher, S. J., Heilmann, M. R., Bruszewski, W., Talbot, N. H., Snyder, R. L., & Merzenich, M. M. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.748987 · Cited by 62
DESIGN AND FABRICATION OF THE BANDED ELECTRODE ARRAYa
Clark, G. M., Shepherd, R. K., Patrick, J. F., Black, R. C., & Tong, Y. C. (1983) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1983.tb31632.x · Cited by 54
The Clarion electrode positioner: Temporal bone studies
Fayad, J. N., Luxford, W. M., & Linthicum, F. H. (2000) · American Journal of Otolaryngology · doi:10.1016/s0196-0709(00)80013-7 · Cited by 47
Richter, B., Aschendorff, A., Lohnstein, P., Husstedt, H. W., Nagursky, H., & Laszig, R. (2002) · The Journal of Laryngology & Otology · doi:10.1258/002221502760132584 · Cited by 32
DESIGN AND FABRICATION OF MULTIWIRE SCALA TYMPANI ELECTRODESa
Hochmair-Desoyer, I. J., Hochmair, E. S., & Burian, K. (1983) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1983.tb31630.x · Cited by 27
Cylindrical Cochlear Electrode Array for Use in Humans
van der Puije, P. D., Pon, C. R., & Robillard, H. (1989) · Annals of Otology, Rhinology & Laryngology · doi:10.1177/000348948909800614 · Cited by 14
Influence of materials and geometry on fields produced by cochlear electrode arrays
Ruddy, H. A., & Loeb, G. E. (1995) · Medical & Biological Engineering & Computing · doi:10.1007/bf02523011 · Cited by 11
Experiments with A Thinfilm Multichannel Electrode for Cochlear Implantation
Lauridsen, O., Günthersen, C., Bonding, P., & Tos, M. (1983) · Acta Oto-Laryngologica · doi:10.3109/00016488309130938 · Cited by 5
Cited by
Cochlear Implants: System Design, Integration, and Evaluation
Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., & Feng, H. (2008) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2008.2008250 · Cited by 878
Role of Electrode Placement as a Contributor to Variability in Cochlear Implant Outcomes
Finley, C. C., Holden, T. A., Holden, L. K., Whiting, B. R., Chole, R. A., Neely, G. J., Hullar, T. E., & Skinner, M. W. (2008) · Otology & Neurotology · doi:10.1097/mao.0b013e318184f492 · Cited by 549
Impact of electrode design and surgical approach on scalar location and cochlear implant outcomes
Wanna, G. B., Noble, J. H., Carlson, M. L., Gifford, R. H., Dietrich, M. S., Haynes, D. S., Dawant, B. M., & Labadie, R. F. (2014) · The Laryngoscope · doi:10.1002/lary.24728 · Cited by 321
The cochlear implant: historical aspects and future prospects
Eshraghi, A. A., Nazarian, R., Telischi, F. F., Rajguru, S. M., Truy, E., & Gupta, C. (2012) · The Anatomical Record · doi:10.1002/ar.22580 · Cited by 285
“…trauma caused to the cochlea from different multichannel cochlear implant electrodes (O’Leary et al., 1991; Welling et al., 1993; Gstoettner, 1997; Nadol et al., 2001; Adunka et al., 2004; Gstoettner et al., 2004; Roland, 2005; Fraysse et al., 2006; Rebscher et al., 2008; Helbig et al., 2011).”
BackgroundThe Development of the Nucleus® Freedom™ Cochlear Implant System
Patrick, J. F., Busby, P. A., & Gibson, P. J. (2006) · Trends in Amplification · doi:10.1177/1084713806296386 · Cited by 196
Metadata
- Authors
- Roland, J. T.
- Venue
- The Laryngoscope
- Publisher
- Wiley
- Published
- 2005-08
- Volume
- 115
- Issue
- 8
- Pages
- 1325-1339
- ISSN
- 0023-852X (ISSN-L)
- 1531-4995
- OpenAlex
- W2143200041
- Cited by
- 192
- References
- 31
- FWCI
- 2.96
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.