In vitro measurement and characterization of current density profiles produced by nonrecessed, simple recessed, and radially varying recessed stimulating electrodes
Abstract
Potential fields induced by nonrecessed, simple recessed, and radially varying recessed electrode designs were measured in vitro. Comparison of experimental results with theoretical analyses substantiated the experimental measurement technique and emphasized the importance of considering both nonuniform charge injection and surface electrochemistry when designing implantable stimulating electrodes. Radially varying recesses produced uniform charge injection at the electrode surface and at the aperture-tissue interface. In general, the radially varying recessed electrodes provided a combination of uniform charge injection and flexibility in design and fabrication that warrants their incorporation into all appropriate planar stimulating electrode designs.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
For large electrodes used in some simulated studies, current intensity can be five-fold higher at the edge of the electrode than in the middle.
“However in practice current “pools” around the edge of electrodes87,88 – for the large electrodes used in some of the studies we simulated, current intensity can be five-fold higher at the edge of the electrode than in the middle.”
From A virtual patient simulation modeling the neural and perceptual effects of human visual cortical stimulation, from pulse trains to percepts (2024) · cites this paper as [88]
Cites
Electrodes and the measurement of bioelectric events
Shipton, H. W. (1973) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(73)90185-5 · Cited by 303 · see citing works
Chronic Electrical Stimulation of the Auditory Nerve in Cats
Shepherd, R. K., Clark, G. M., & Black, R. C. (1983) · Acta Oto-Laryngologica · doi:10.3109/00016488309105589 · Cited by 148 · see citing works
Brummer, S. B., & Turner, M. J. (1975) · Bioelectrochemistry and Bioenergetics · doi:10.1016/0302-4598(75)80002-x · Cited by 132 · see citing works
Brummer, S. B., McHardy, J., & Turner, M. J. (2008) · Brain Behavior and Evolution · doi:10.1159/000124611 · Cited by 106 · see citing works
Current Density Profiles of Surface Mounted and Recessed Electrodes for Neural Prostheses
Rubinstein, J. T., Spelman, F. A., Soma, M., & Suesserman, M. F. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.326007 · Cited by 89 · see citing works
Self-citationHambrecht, F. T. (1979) · Annual Review of Biophysics and Bioengineering · doi:10.1146/annurev.bb.08.060179.001323 · Cited by 75 · see citing works
Cochlear Pathology following Chronic Intracochlear Electrical Stimulation
Leake-Jones, P. A., Walsh, S. M., & Merzenich, M. M. (1981) · Annals of Otology Rhinology & Laryngology · doi:10.1177/00034894810902s202 · Cited by 25 · see citing works
A Simple Circuit to Reduce the Input Capacitance of Microelectrode Amplifiers
Scobey, R. P., Howard, D. L., & Gabor, A. J. (1981) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1981.324805 · Cited by 2 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · see citing works
A model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526 · see citing works
Tissue damage thresholds during therapeutic electrical stimulation
Cogan, S. F., Ludwig, K. A., Welle, C. G., & Takmakov, P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/021001 · Cited by 386 · see citing works
Retinal stimulation strategies to restore vision: fundamentals and systems
Yue, L., Weiland, J. D., Roska, B., & Humayun, M. S. (2016) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2016.05.002 · Cited by 272 · see citing works
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · see citing works
Factors affecting perceptual thresholds in epiretinal prostheses
de Balthasar, C., Patel, S., Roy, A., Freda, R., Greenwald, S., Horsager, A., Mahadevappa, M., Yanai, D., McMahon, M. J., Humayun, M. S., et al. (2008) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.07-0696 · Cited by 204 · see citing works
Metz, S., Bertsch, A., Bertrand, D., & Renaud, P. (2004) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2003.11.021 · Cited by 167 · see citing works
Resolution of the epiretinal prosthesis is not limited by electrode size
Behrend, M. R., Ahuja, A. K., Humayun, M. S., Chow, R. H., & Weiland, J. D. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2140132 · Cited by 102 · see citing works
Amorphous silicon carbide ultramicroelectrode arrays for neural stimulation and recording
Deku, F., Cohen, Y., Joshi-Imre, A., Kanneganti, A., Gardner, T. J., & Cogan, S. F. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa8f8b · Cited by 89 · see citing works
Fine, I., & Boynton, G. M. (2024) · Scientific Reports · doi:10.1038/s41598-024-65337-1 · Cited by 19 · see citing works
- Authors
- Suesserman, M. F., Spelman, F. A., & Rubinstein, J. T.
- Affiliations
- University of Washington (US)
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 1991-05
- Volume
- 38
- Issue
- 5
- Pages
- 401-408
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2059770295
- Cited by
- 83
- References
- 8
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 3 · 2023: 2 · 2021: 1 · 2020: 2 · 2019: 3 · 2018: 2 · 2017: 4 · 2016: 3 · 2015: 2 · 2014: 4 · 2013: 1 · 2012: 4
- FWCI
- 0.26
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institute of General Medical Sciences (GM07266, T32GM007266)
- National Institute of Neurological Disorders and Stroke (NS 13056)
- National Center for Research Resources (RR00166, P51RR000166)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Keywords
- stimulating electrodes, current density profile, charge injection, surface electrochemistry, implanted electrodes, electrode design, potential field, radial variation
- MeSH terms
- Electric Conductivity, Electric Stimulation, Electrodes, Implanted, Equipment Design, Surface Properties
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review