Active tremor compensation in microsurgery
The DOI matches its Crossref record; 17 of the 22 references it cites were found in a registry or an index; 1 of the 15 sources its claims cite were checked for support.
Abstract
This work presents the development of an intelligent microsurgical instrument to perform real-time tremor compensation within a handheld tool. The intelligent instrument senses its own motion, distinguishes between voluntary and erroneous motion, and manipulates its tip to cancel the undesired component in real-time. The on-board sensing unit is made up of a magnetometer-aided all-accelerometer inertial measurement unit and sensor fusion is performed via a quaternion-based Kalman filtering. Tremor is modeled and filtered by an adaptive zero-phase notch filter. The intraocular shaft manipulator is a three DOF piezoelectric actuated mechanism driven by a feedforward controller with inverse rate-dependent hysteresis model. Laboratory experimental results of the system are presented.
Overview
CoChat Summary
## TL;DR
This paper presents “Micron,” a handheld microsurgical instrument that senses its own motion and actively cancels tremor at the tool tip in real time. In laboratory testing, the system reduced simulated tremulous motion by 51.9%.
## Key findings
- The instrument uses a magnetometer-aided all-accelerometer IMU with quaternion-based Kalman filtering for sensing, and an adaptive zero-phase notch/WFLC-based approach for tremor estimation and filtering.
- The intraocular shaft manipulator is a 3-DOF piezoelectric mechanism with mechanical stroke amplification of 8×, yielding a total stroke of >100 µm per axis.
- In a motion-canceling experiment with a 9 Hz testbed oscillator, the tip’s RMS motion was 24.7 µm and the compensated RMS motion was 11.9 µm, a 51.9% reduction over 10 runs.
- The paper also reports that the inverse feedforward controller for piezoelectric actuation tracked a negated tremor profile with RMS tracking error of 0.6%.
## Methods
- Design paper describing an active handheld microsurgical instrument (“Micron”) composed of three subsystems: motion sensing, tremor estimation/filtering, and a 3-DOF intraocular shaft manipulator.
- Sensing used three dual-axis accelerometers and a three-axis magnetometer; orientation and motion estimates were fused with an augmented-state quaternion-based Kalman filter.
- Tremor estimation used the weighted-frequency Fourier linear combiner (WFLC), and the piezoelectric actuator control used an inverse rate-dependent hysteresis model in an open-loop feedforward controller.
- Experimental evaluation used a testbed oscillator simulating hand tremor, with optical motion tracking of the tool tip; tests were run in one, two, or all three Cartesian coordinates.
## Limitations
- The paper states that “more engineering fine tuning” is needed to improve system performance; this is a stated limitation.
- The instrument’s active compensation depends on accurate estimation of undesired motion rather than merely filtering it out, which the paper describes as a technical challenge; this is evident from the text.
- The reported experiments are laboratory tests on simulated tremor, with future work proposed for mocked surgery and animal trials; this indicates the paper does not yet provide clinical validation.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
1 Valid · 1 Not valid · 13 Not checked · 19 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Even for microsurgeons in their prime years, fatigue, alcohol and caffeine consumption, and other factors affect the manual stability [9], [10].”
What the sources say
- Not checked — no full text of the sourcecites [9] [9] M. U. Humayun, R. S. Rader, D. J. Pieramici, C. C. Awh, E. de Juan, Jr., 'Quantitative measurement of the effects of caffeine and propranolol on surgeon hand tremor,' Arch. Ophthalmol ., vol. 115, pp. 371-374, Mar. 1997. - source
- Partly supportedcites [10] [10] Lakie, M., Frymann, K., Villagra, F., Jakeman, P., 'The effect of alcohol on physiological tremor,' Experimental Physiology 79 (1994) 273-276 - sourceExperimental — accuracy not yet measured
Our copy supports only the alcohol component: the Abstract states ethanol 'causes a large decrease in the size of physiological tremor', and the Results (p. ~2, Fig. 1 discussion) report a highly significant reduction in wrist tremor acceleration/power at both doses in ten healthy social drinkers, with the Introduction linking tremor reduction to 'activities where tremor would be a handicap'. Nothing in the paper addresses caffeine or fatigue, and no passage concerns surgeons or microsurgeons — those elements of the citing sentence are not supported by this source. Note also that the paper's finding is that alcohol improves (reduces) physiological tremor, so it evidences an effect on manual stability rather than impairment. (LLM, conf=0.70)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“and the rms amplitude of an error compensated motion is 11.9 µ m”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Experimental result has shown that the controller is capable of tracking a negated tremor profiles with rms tracking error of 0.6%.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“or a reduction of 51.9% over a set of 10 runs.”
- Also stated in: § IX. CONCLUSION · boosted
“Laboratory testing has shown that the instrument successfully reduces simulated tremulous motion by 51.9%.”
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § IX. CONCLUSION · boosted
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The rms amplitude of the tip's motion generated by the testbed oscillator is 24.7 µ m at 9 Hz”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
15 more claims
Create a free account to see all 34 claims and every source verdict.
- It may state two separate claims as one. 8
- Probably not a substantial claim. 8
Citations
Cites
Prevalence and Associations of Retinal Vein Occlusion in Australia: The Blue Mountains Eye Study
Mitchell, P. (1996) · Archives of Ophthalmology · doi:10.1001/archopht.1996.01100140443012 · Cited by 545 · cited 1×
“6% of persons aged 49 years or older [8].”
BackgroundA Steady-Hand Robotic System for Microsurgical Augmentation
Taylor, R., Jensen, P., Whitcomb, L., Barnes, A., Kumar, R., Stoianovici, D., Gupta, P., Wang, Z., Dejuan, E., & Kavoussi, L. (1999) · The International Journal of Robotics Research · doi:10.1177/02783649922067807 · Cited by 528
Adaptive cancelling of physiological tremor for improved precision in microsurgery
Riviere, C. N., Rader, R. S., & Thakor, N. V. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.686791 · Cited by 260 · cited 1×
“Estimation of tremor will be performed by a system based on the weighted-frequency Fourier linear combiner (WFLC) algorithm [17].”
MethodSelf-citationA teleoperated microsurgical robot and associated virtual environment for eye surgery
Hunter, I. W., Doukoglou, T. D., Lafontaine, S. R., Charette, P. G., Jones, L. A., Sagar, M. A., Mallinson, G. D., & Hunter, P. J. (1993) · Presence: Teleoperators and Virtual Environments · doi:10.1162/pres.1993.2.4.265 · Cited by 164
A Steady-Hand Robotic System for Microsurgical Augmentation
Taylor, R. H., Jensen, P. S., Whitcomb, L. L., Barnes, A. C., Kumar, R., Stoianovici, D. S., Gupta, P., Wang, Z., deJuan, E., & Kavoussi, L. R. (1999) · Lecture notes in computer science · doi:10.1007/10704282_112 · Cited by 154
Physiologic tremor and microsurgery
Harwell, R. C., & Ferguson, R. L. (1983) · Microsurgery · doi:10.1002/micr.1920040310 · Cited by 131
Physiological tremor amplitude during retinal microsurgery
Singh, S. P. N., & Riviere, C. N. (2002) · Proceedings of the IEEE 28th Annual Northeast Bioengineering Conference (IEEE Cat. No.02CH37342) · doi:10.1109/nebc.2002.999520 · Cited by 131
Self-citationEvaluation of a Telerobotic System to Assist Surgeons in Microsurgery
Das, H., Zak, H., Johnson, J., Crouch, J., & Frambach, D. (1999) · Computer Aided Surgery · doi:10.3109/10929089909148155 · Cited by 116
Weiss, J. N. (1998) · American Journal of Ophthalmology · doi:10.1016/s0002-9394(98)00086-5 · Cited by 100 · cited 1×
“Another example is the treatment of retinal vein occlusions by injection of anticoagulants using intraocular cannulation [6]-[7].”
BackgroundA Study of Surgical Approaches to Retinal Vascular Occlusions
Tang, W. M. (2000) · Archives of Ophthalmology · doi:10.1001/archopht.118.1.138 · Cited by 84
Research works printed but not linked: 3
- [12] S. Charles, Dexterity enhancement for surgery. In: R.H. Taylor et al. (eds.): Computer Integrated Surgery: Technology and Clinical Applications. MIT Press, Cambridge (1996) 467-471. -
- [18] W. T. Ang, P. K. Khosla, and C. N. Riviere, 'Inverse Feedforward Controller with Rate-dependent Hysteresis Model for Piezoelectric Actuators', Submitted to IEEE Trans Control Systems & Tech . -
- [3] R. J. Elble and W. C. Koller, Tremor . Baltimore: Johns Hopkins, 1990. -
Cited by
Comparative study on classifying human activities with miniature inertial and magnetic sensors
Altun, K., Barshan, B., & Tunçel, O. (2010) · Pattern Recognition · doi:10.1016/j.patcog.2010.04.019 · Cited by 564
“For several decades, inertial sensors have been used for navigation of aircraft [1,2], ships, land vehicles, and robots [3–5], for state estimation and dynamic modeling of legged robots [6,7], for shock and vibration analysis in the automotive industry, and in telesurgery [8,9].”
BackgroundA review of medical robotics for minimally invasive soft tissue surgery
Dogangil, G., Davies, B. L., & Rodriguez y Baena, F. (2010) · Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · doi:10.1243/09544119jeim591 · Cited by 225
Adaptive least squares support vector machines filter for hand tremor canceling in microsurgery
Liu, Z., Wu, Q., Zhang, Y., & Philip Chen, C. L. (2011) · International Journal of Machine Learning and Cybernetics · doi:10.1007/s13042-011-0012-5 · Cited by 214
“The LS-SVMAF is composed of inertial measurement unit (IMU), tremor filtering unit, calculating and control unit, and driver unit [ 8 ].”
MethodThe milliDelta: A high-bandwidth, high-precision, millimeter-scale Delta robot
McClintock, H., Temel, F. Z., Doshi, N., Koh, J.-S., & Wood, R. J. (2018) · Science Robotics · doi:10.1126/scirobotics.aar3018 · Cited by 190
Enhanced sliding mode motion tracking control of piezoelectric actuators
Liaw, H. C., Shirinzadeh, B., & Smith, J. A. (2007) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2007.04.062 · Cited by 140
Liaw, H. C., Shirinzadeh, B., & Smith, J. A. (2007) · Mechatronics · doi:10.1016/j.mechatronics.2007.09.002 · Cited by 125
Liaw, H. C., Shirinzadeh, B., & Smith, J. A. (2008) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2007.916301 · Cited by 122
“Increasingly, piezoelectric actuator-driven manipulation systems are employed in many advanced areas, particularly in the biomedical engineering and microsurgery applications [8]–[11].”
BackgroundThe introduction of a new robot for assistance in ophthalmic surgery
Nasseri, M. A., Eder, M., Nair, S., Dean‐Leon, E., Maier, M. M., Zapp, D., Lohmann, C. P., & Knoll, A. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610840 · Cited by 131
Payne, C. J., & Yang, G. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1042-4 · Cited by 119
Bilateral Teleoperation Over Networks Based on Stochastic Switching Approach
Walker, K. C., Pan, Y., & Gu, J. J. (2009) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2009.2007126 · Cited by 72
“The input force is considered band-limited based on [16] where it was shown that voluntary human motion is limited to 6 Hz.”
BackgroundLiaw, H. C., Shirinzadeh, B., & Smith, J. A. (2009) · IEEE Transactions on Neural Networks · doi:10.1109/tnn.2008.2004406 · Cited by 69
Classifying Human Leg Motions with Uniaxial Piezoelectric Gyroscopes
Tunçel, O., Altun, K., & Barshan, B. (2009) · Sensors · doi:10.3390/s91108508 · Cited by 72
“…inertial sensors have been used in various applications such as for navigation of aircraft [4–6], ships, land vehicles and robots [7–9], for state estimation and dynamic modeling of legged robots [10, 11], in the automotive industry, for shock and vibration analysis, and in telesurgery [12, 13].”
MethodLiaw, H. C., & Shirinzadeh, B. (2008) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2008.03.020 · Cited by 63
Nasseri, M. A., Eder, M., Eberts, D., Nair, S., Maier, M. M., Zapp, D., Lohmann, C. P., & Knoll, A. (2013) · IEEE/ASME International Conference on Advanced Intelligent Mechatronics · doi:10.1109/aim.2013.6584107 · Cited by 54
“…q̇1 q̇2 ] = [ 1 2 1 2 −d d2+(L2−L1)2 d d2+(L2−L1)2 ][ L̇1 L̇2 ] It can be observed that the determinant of the Jacobian matrix (JL) is 0, only if d = 0 which implies that for the proposed configuration, where the distance of two prismatic parallel actuators is more than 0, JL matrix is full…”
BackgroundFlexure-based Manipulator for Active Handheld Microsurgical Instrument
Choi, D. Y., & Riviere, C. N. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1616932 · Cited by 49
Self-citationAccurate real-time depth control for CP-SSOCT distal sensor based handheld microsurgery tools
Cheon, G. W., Huang, Y., Cha, J., Gehlbach, P. L., & Kang, J. U. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.001942 · Cited by 45
Development of a medical robot system for minimally invasive surgery
Feng, M., Fu, Y., Pan, B., & Liu, C. (2011) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.440 · Cited by 38
“(37) proposed an intelligent instrument to perform real-time tremor compensation within a handheld tool, which successfully reduced the simulated tremulous motion by 51.”
BackgroundMethodFrom Teleoperation to Autonomous Robot-assisted Microsurgery: A Survey
Zhang, D., Si, W., Fan, W., Guan, Y. H., & Yang, C. (2022) · Machine Intelligence Research · doi:10.1007/s11633-022-1332-5 · Cited by 50
“system “A 73” was developed for paranasal sinus surgery[22], which was tested on macerated cadaveric”
BackgroundThe Development of Robotics for Interventional MRI
Lwu, S., & Sutherland, G. R. (2009) · Neurosurgery Clinics of North America · doi:10.1016/j.nec.2009.04.011 · Cited by 36
Plastic and reconstructive robotic microsurgery – a review of current practices
Saleh, D. B., Syed, M., Kulendren, D., et al. (2015) · Annales de Chirurgie Plastique Esthétique · doi:10.1016/j.anplas.2015.03.005 · Cited by 30
Saxena, A., & Patel, R. V. (2013) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2013.6696969 · Cited by 31
“One of the most significant contributions towards an active handheld microsurgical device is the Micron [7][8].”
BackgroundDesign and control of a parallel linkage wrist for robotic microsurgery
Değirmenci, A., Hammond, F. L., Gafford, J. B., Walsh, C. J., Wood, R. J., & Howe, R. D. (2015) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2015.7353378 · Cited by 29
“For instance, in vitroretinal surgery, a 10 μm accuracy is desired [1]; however even trained surgeons have a root mean square (RMS) tremor between 49 - 133 μm at the tip of their surgical tool [2].”
BackgroundAdaptive fuzzy wavelet neural network filter for hand tremor canceling in microsurgery
Liu, Z., Wu, Q., Zhang, Y., Wang, Y., & Chen, C. L. P. (2011) · Applied Soft Computing · doi:10.1016/j.asoc.2011.05.027 · Cited by 27
“imately rhythmic, and roughly sinusoidal movement [8,9].”
MethodActive vibration control in human forearm model using paired piezoelectric sensor and actuator
Hosseini, S. M., Kalhori, H., & Al-Jumaily, A. A. (2020) · Journal of Vibration and Control · doi:10.1177/1077546320957533 · Cited by 25
Biomimetic grasp planning for cortical control of a robotic hand
Ciocarlie, M., Clanton, S. T., Spalding, M. C., & Allen, P. K. (2008) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2008.4651179 · Cited by 22
“Finally, the idea of human-robotic shared control is a common theme in robotic-assisted surgery, as in work presented by Kragic et al.[19] and Ang et al.[20].”
MethodNasseri, M. A., Maier, M. M., & Lohmann, C. P. (2017) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2017.8037815 · Cited by 20
Finger tremor can be voluntarily reduced during a tracking task
Daneault, J., Carignan, B., & Duval, C. (2010) · Brain Research · doi:10.1016/j.brainres.2010.11.047 · Cited by 25
Ebihara, Y., Hirano, S., Takano, H., et al. (2023) · Journal of Robotic Surgery · doi:10.1007/s11701-023-01522-0 · Cited by 14
“Tremor compensation [22, 23], virtual fixtures [24, 25], and robotic arm motion scaling have been reported to be useful for improving the accuracy, safety, and control of telerobotic systems.”
BackgroundMode‐dependent switching control of bilateral teleoperation against random denial‐of‐service attacks
Hu, L., Wang, K., Hu, D.-X., et al. (2021) · IET Cyber-Physical Systems Theory & Applications · doi:10.1049/cps2.12015 · Cited by 11
“The input force is considered band ‐ limited [18], where it was shown that voluntary human motion is limited to 6 Hz.”
BackgroundDeep Learning Inspired Feature Engineering for Classifying Tremor Severity
Taee, A. A. A., Hosseini, S., Khushaba, R. N., Zia, T. A., Lin, C., & Al-Jumaily, A. A. (2022) · IEEE Access · doi:10.1109/access.2022.3210344 · Cited by 10
“INTRODUCTION Tremor is one of the most common neurological disorders that causes involuntary sinusoidal movements or shaking in one or more body parts, such as the head, hands, voice, trunk, or even eyes [1].”
BackgroundAdvances and Key Techniques of Ophthalmic Microsurgical Robots
Xiao, J. (2013) · Journal of Mechanical Engineering · doi:10.3901/jme.2013.01.015 · Cited by 9
Network Performance Measurement through Machine to Machine Communication in TeleRobotics System
Lokhande, M., & Patil, D. D. (2021) · Tehnički glasnik · doi:10.31803/tg-20210205092413 · Cited by 10
“In [31] developed a handheld robotic device called Micron for microsurgery.”
MethodNasseri, M. A., Dean‐Leon, E., Nair, S., Eder, M., Knoll, A., Maier, M. M., & Lohmann, C. P. (2012) · International Conference on Biomedical Engineering and Informatics (BMEI) · doi:10.1109/bmei.2012.6512987 · Cited by 8
“But the EMT method is not suitable to measure surgeon’s tremor precisely which according to the state of the art reported less than 200μm [1].”
BackgroundMethodTavallaeinejad, M., Eghtesad, M., Mahzoon, M., et al. (2014) · Journal of Vibration and Control · doi:10.1177/1077546314549587 · Cited by 7
“…particles separation, chromosome cutting or even genetic manipulation) in several works (Riviere et al., 1998; Tan and Ng, 2001; Imura et al., 2002; Ang et al., 2004) and also for a large number of micro-manufacturing applications (Cohn et al., 1998; Kim et al., 2005; Zhu and Tan, 2002).”
BackgroundActive force control system for hand tremor suppression by different actuators
Hosseini, S., Al-Jumaily, A. A., & Abboud, S. A. (2016) · International Conference on Electronic Devices, Systems and Applications · doi:10.1109/icedsa.2016.7818487 · Cited by 8
Lin, C., Lin, C.-R., Yu, S.-K., et al. (2014) · Advances in Mechanical Engineering · doi:10.1155/2014/769192 · Cited by 6
Park, H.-S., Yeo, C. B. A., Gehlbach, P. L., & Song, C. O. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318247 · Cited by 7
“Various approaches to stabilize the tool tip of micro-surgical instruments have been proposed as follows: Micron [1-4], a steady handy robot [5-7], SMART [8-10], force sensing micro-forceps [11,12], iTrem [13], and each of these important approaches remains in continued development.”
MethodAnalytical solution for forced vibration of piezoelectrically actuated Timoshenko beam
Hosseini, S. M., & Al-Jumaily, A. A. (2019) · Journal of Intelligent Material Systems and Structures · doi:10.1177/1045389x19832118 · Cited by 6
Active force control system for tremor suppression in elbow joint
Hosseini, S., & Al-Jumaily, A. A. (2016) · IEEE International Symposium on Robotics and Intelligent Sensors (IRIS) · doi:10.1109/iris.2016.8066080 · Cited by 6
“The most famous characteristics for this disability are involuntary, roughly recurring, and approximately sinusoidal [1, 2].”
BackgroundControl gain design for bilateral teleoperation systems using linear matrix inequalities
Walker, K. C., Pan, Y., & Gu, J. J. (2007) · Conference proceedings/Conference proceedings - IEEE International Conference on Systems, Man, and Cybernetics · doi:10.1109/icsmc.2007.4413584 · Cited by 4
Bilateral Control and Stabilization of Asymmetric Teleoperators With Bounded Time-Varying Delays
Hilliard, T., & Pan, Y. (2014) · Journal of Dynamic Systems Measurement and Control · doi:10.1115/1.4026967 · Cited by 3
Master-slave control strategy for abdominal minimally invasive surgery robotic system
Fu, Y., Yu, Y., & Wang, S. (2011) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio.2011.6181329 · Cited by 3
Image-guided robotics for autonomous venipuncture
Chen, A. I. (2016) · doi:10.7282/t3gx4dt0 · Cited by 6
Intelligent sensing for robot mapping and simultaneous human localization and activity recognition
Altun, K. (2011) · Bilkent University Institutional Repository (Bilkent University) · Cited by 2
A Novel Stabilization Scheme for Asymmetric Bilateral Teleoperation Systems with Time Varying Delays
Hilliard, T., & Pan, Y. (2012) · Lecture notes in computer science · doi:10.1007/978-3-642-33509-9_63 · Cited by 1
“The first input considered is a modified step input, as a human operator cannot physically apply a true step input as human motion is frequency limited [6].”
BackgroundNiță, E. I., Comeaga, D. C., & Avram, M. (2024) · doi:10.1117/12.3021501 · Cited by 1
“The frequency of the tremor varies from person to person, depending on the biological motor system, being roughly located in a frequency band of 5-15Hz [7][8].”
BackgroundStochastic Switching Approach of Bilateral Teleoperation Systems: Part I - Theory
Walker, K. C., Pan, Y., & Gu, J. J. (2008) · IFAC Proceedings Volumes · doi:10.3182/20080706-5-kr-1001.00630 · Cited by 1
“Although no model is considered, the input force is considered band limited based on (Ang et al., 2004) where it was shown that voluntary motion is limited to 6Hz.”
BackgroundUngrounded haptic-feedback for hand-held surgical robots
Payne, C. J. (2015) · Spiral (Imperial College London) · doi:10.25560/26587 · Cited by 4
“The first implementation of a hand-held robotic tremor compensation was Micron [26][27], which was developed at Carnegie Mellon University by Riviere et al.”
BackgroundMethodInfluentialArmstrong, J. D. (2010) · UTA ResearchCommons (University of Texas Arlington)
Tan, U.-X. (2010) · DR-NTU (Nanyang Technological University) · doi:10.32657/10356/40882
A Virtual Technology Aiding for Hand-Impaired Persons
Wang, K., & Sun, Z. Y. (2018) · International Joint Conference on Information, Media and Engineering (IJCIME) · doi:10.1109/icime.2018.00057
“The common methods are based on the measurement of hand tremors, and (1) adding vibration oppose to tremor in hand interactive objects to offset hand tremors [14], (2) adjusting the conformation or interface of objects to balance hand tremors [15] [16] [17], and (3) filtering the effect of hand tremors on interactive…”
MethodAssistive control for non-contact machining of random shaped contours
Baran, A. E. (2014) · Sabanci University
Handle Navigation for a Smart Handheld Robot
Hung, S., & Yen, P. (2019) · EPiC series in health sciences · doi:10.29007/fhc6
“Among these are the bone mounted robot [2], or handheld robot [3], which have the superiority of compact installation space and compatible to daily surgical procedure.”
BackgroundLocalization Of Sensors In Presence Of Fading And Mobility
Omari, H. H. (2015)
Enhanced closed loop controller for electromagnetic actuators used in opto-mechatronic devices
Coandă, P., Niță, E.-I., & Comeaga, D. C. (2022) · Advances in 3OM: Opto-Mechatronics, Opto-Mechanics, and Optical Metrology (3OM) · doi:10.1117/12.2619497
Hosseini, S. M., Farashi, S., & Bashirian, S. (2023) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-023-01255-z · Cited by 1
“Inclusion and exclusion criteria The studies that satisfied the following inclusion criteria were included in this paper: [1] original research articles, clinical trials, randomized controlled trials, case studies, comparative studies, and pilot studies [2], studies involving PD patients affected by tremor [3], studie…”
BackgroundResultNiță, E. I., Comeaga, D. C., & Avram, M. (2024) · Advances in 3OM: Opto-Mechatronics, Opto-Mechanics, and Optical Metrology (3OM) · doi:10.1117/12.3021517
Lee, M. H., Im, J., & Song, C. O. (2024) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros58592.2024.10802576
“For example, an inertial measurement unit (IMU) [8] with an accelerometer, a gyroscope, and a geomagnetic sensor, a MICRON [9] that tracks LEDs attached to surgical instruments with a 3D camera, and an Electromagnetic (EM) [10] are used to compensate for hand tremors and the positions of the ends…”
Hysteresis modeling for smart materials and observer design for 2DOF robots
Carnevale, D. (2008) · Cineca Institutional Research Information System (Tor Vergata University) · doi:10.58015/carnevale-daniele_phd2008-09-02
Control of meso-robots for endoluminal surgery
Secades, L. A. S. (2013) · theses.fr (ABES) · doi:10.70675/21e02714z81d0z4bb8za425z0928a3c0ed09
Metadata
- Authors
- Ang, W. T., Pradeep, P. K., & Riviere, C. N.
- Affiliations
- Carnegie Mellon University (US)
- Venue
- The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
- Publisher
- IEEE
- Published
- 2005-03-21
- Volume
- 3
- Pages
- 2738-2741
- ISSN
- 2375-7477 (ISSN-L)
- OpenAlex
- W2548823874
- Cited by
- 61
- References
- 22
- Citations by year
- 2025: 1 · 2024: 3 · 2023: 2 · 2022: 3 · 2021: 2 · 2020: 1 · 2019: 2 · 2018: 2 · 2017: 1 · 2016: 3 · 2015: 6 · 2014: 5 · 2013: 5 · 2012: 2
- FWCI
- 6.16
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
- Physical Sciences › Engineering › Mechanical Engineering › Teleoperation and Haptic Systems
- Physical Sciences › Engineering › Control and Systems Engineering › Piezoelectric Actuators and Control
- Keywords
- tremor compensation, microsurgery, inertial measurement unit, magnetometer, sensor fusion, piezoelectric actuators, feedforward control, hysteresis compensation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.