A CMOS Current Steering Neurostimulation Array With Integrated DAC Calibration and Charge Balancing
Open accessThe DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Abstract
An 8-channel current steerable, multi-phasic neural stimulator with on-chip current DAC calibration and residue nulling for precise charge balancing is presented. Each channel consists of two sub-binary radix DACs followed by wide-swing, high output impedance current buffers providing time-multiplexed source and sink outputs for anodic and cathodic stimulation. A single integrator is shared among channels and serves to calibrate DAC coefficients and to closely match the anodic and cathodic stimulation phases. Following calibration, the differential non-linearity is within ±0.3 LSB at 8-bit resolution, and the two stimulation phases are matched within 0.3%. Individual control in digital programming of stimulation coefficients across the array allows altering the spatial profile of current stimulation for selection of stimulation targets by current steering. Combined with the self-calibration and current matching functions, the current steering capabilities integrated on-chip support use in fully implanted neural interfaces with autonomous operation for and adaptive stimulation under variations in electrode and tissue conditions. As a proof-of-concept we applied current steering stimulation through a multi-channel cuff electrode on the sciatic nerve of a rat.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 1×
“Chronic electrical stimulation can be damaging to both the electrode and tissue either through excitotoxic or electrochemical means [11].”
BackgroundA jump-start for electroceuticals
Famm, K., Litt, B., Tracey, K. J., Boyden, E. S., & Slaoui, M. (2013) · Nature · doi:10.1038/496159a · Cited by 554 · cited 1×
“Emerging applications of PNS stimulation include treatment of hypertension and inflammatory disorders [2].”
BackgroundSelective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · cited 1×
“used cuff electrodes with contacts spanning the entire circumference of the nerve [5], [33], [34], whereas here the arrays used covered 75% of the nerve circumference, in the future, histo-”
MethodSelective activation of small motor axons by quasitrapezoidal current pulses
Fang, Z.-P., & Mortimer, J. T. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.76383 · Cited by 222 · cited 1×
“Modifying the stimulus pulse shape or duration is one method for achieving more selective stimulation [3], [4].”
MethodThe effect of stimulus pulse duration on selectivity of neural stimulation
Grill, W. M., & Mortimer, J. T. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.481985 · Cited by 215 · cited 1×
“Modifying the stimulus pulse shape or duration is one method for achieving more selective stimulation [3], [4].”
MethodNoorsal, E., Sooksood, K., Xu, H., Hornig, R., Becker, J., & Ortmanns, M. (2012) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2164667 · Cited by 212 · cited 2×
“Parameter This Work [37] [10] [18] [28] [38] [27] [14] [13]”
MethodA nerve cuff technique for selective excitation of peripheral nerve trunk regions
Sweeney, J. D., Ksienski, D. A., & Mortimer, J. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55681 · Cited by 196 · cited 2×
“used cuff electrodes with contacts spanning the entire circumference of the nerve [5], [33], [34], whereas here the arrays used covered 75% of the nerve circumference, in the future, histo-”
MethodSit, J.-J., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.911631 · Cited by 155 · cited 3×
“The use of an additional “shorting” switch in each channel allows for DC current levels to be brought well below safety-limits [13], [18], [28].”
BackgroundTarler, M. D., & Mortimer, J. T. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.828415 · Cited by 110
A Complete 256-Electrode Retinal Prosthesis Chip
Tran, N., Bai, S., Yang, J., Chun, H., Kavehei, O., Yang, Y., Muktamath, V., Ng, D., Meffin, H., Halpern, M., et al. (2014) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2298037 · Cited by 90 · cited 4×
“The use of an additional “shorting” switch in each channel allows for DC current levels to be brought well below safety-limits [13], [18], [28].”
BackgroundMethodInfluential
Cited by
Zhang, M., Tang, Z., Liu, X., & Van der Spiegel, J. (2020) · Nature Electronics · doi:10.1038/s41928-020-0390-3 · Cited by 231
A mm-Sized Wireless Implantable Device for Electrical Stimulation of Peripheral Nerves
Charthad, J., Chang, T. C., Liu, Z., Sawaby, A., Weber, M. J., Baker, S., Gore, F., Felt, S. A., & Arbabian, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2799623 · Cited by 230
“tor architectures, have been demonstrated by other researchers [60], [61], and can be considered for future implementations of this implant.”
BackgroundAn ASIC for Recording and Stimulation in Stacked Microchannel Neural Interfaces
Lancashire, H. T., Jiang, D., Demosthenous, A., & Donaldson, N. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2891284 · Cited by 15
Metadata
- Authors
- Greenwald, E., Maier, C., Wang, Q., Beaulieu, R., Etienne-Cummings, R., Cauwenberghs, G., & Thakor, N.
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2017-04
- Volume
- 11
- Issue
- 2
- Pages
- 324-335
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W2575060568
- Cited by
- 53
- References
- 11
- FWCI
- 2.10
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.