Nonlinear Cuffless Blood Pressure Estimation of Healthy Subjects Using Pulse Transit Time and Arrival Time
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
This paper presents a novel blood pressure (BP) estimation method based on pulse transit time (PTT) and pulse arrival time (PAT) to estimate the systolic BP (SBP) and the diastolic BP (DBP). A data acquisition hardware is designed for high-resolution sampling of phonocardiogram (PCG), photoplethysmogram, and electrocardiogram (ECG). PCG and ECG perform as the proximal timing reference to obtain PTT and PAT indices, respectively. In order to derive a BP estimator model, a calibration procedure, including a supervised physical exercise, is conducted for each individual, which causes changes in their BP, and then, a number of reference BPs are measured alongside the acquisition of the signals per subject. It is suggested to use a force-sensing resistor that is placed under the cuff of the BP reference device to mark the exact moments of reference BP measurements, which are corresponding to the inflation of the cuff. Additionally, a novel BP estimator nonlinear model, based on the theory of elastic tubes, is introduced to estimate the BP using PTT/PAT values precisely. The proposed method is evaluated on 32 subjects. Using the PTT index, the correlation coefficients for SBP and DBP estimation are 0.89 and 0.84, respectively. Using the PAT index, the correlation coefficients for SBP and DBP estimation are 0.95 and 0.84, respectively. The results show that the proposed method, exploiting the introduced nonlinear model with the use of PAT index or PTT index, provides a reliable estimation of SBP and DBP.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · cited 1×
“Among common interventions proposed in the literature for perturbing BP [12], [22]–[26], here, physical exercise is employed, since it has been shown to cause a sensible increase in both SBP and DBP up to 40 mmHg [27].”
BackgroundParati, G., Casadei, R., Groppelli, A., Di Rienzo, M., & Mancia, G. (1989) · Hypertension · doi:10.1161/01.hyp.13.6.647 · Cited by 899 · cited 1×
“Among common interventions proposed in the literature for perturbing BP [22], [23], [24], [25], [26], [12], here, physical exercise is employed since it has been shown to cause a sensible increase in both SBP and DBP up to 40 mmHg [27].”
BackgroundCuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621
Self-citationChen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Evaluation of blood pressure changes using vascular transit time
Foo, J. Y. A., Lim, C. S., & Wang, P. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/8/003 · Cited by 62
Cited by
ECG Heartbeat Classification: A Deep Transferable Representation
Kachuee, M., Fazeli, S., & Sarrafzadeh, M. (2018) · 2018 IEEE International Conference on Healthcare Informatics (ICHI) · doi:10.1109/ichi.2018.00092 · Cited by 447
Self-citationWearable Sensing and Telehealth Technology with Potential Applications in the Coronavirus Pandemic
Ding, X., Clifton, D., Ji, N., Lovell, N. H., Bonato, P., Chen, W., Yu, X., Xue, Z., Xiang, T., Long, X., et al. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2020.2992838 · Cited by 281
“Several studies use alternative cardiac signals for ECG, such as impedance plethysmography (IPG) [99], [100], phonocardiogram (PCG) [101], [102], or ballistocardiogram”
BackgroundWearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250
PPG2ABP: Translating photoplethysmogram (PPG) signals to arterial blood pressure (ABP) waveforms
Ibtehaz, N., Mahmud, S., Chowdhury, M. E. H., Khandakar, A., Salman Khan, M., Ayari, M. A., Tahir, A. M., & Rahman, M. S. (2022) · Bioengineering · doi:10.3390/bioengineering9110692 · Cited by 133
Cuffless Blood Pressure Monitoring: Promises and Challenges
Pandit, J. A., Lores, E., & Batlle, D. (2020) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.03680320 · Cited by 101
Metadata
- Authors
- Esmaili, A., Kachuee, M., & Shabany, M.
- Venue
- IEEE Transactions on Instrumentation and Measurement
- Published
- 2017-12
- Volume
- 66
- Issue
- 12
- Pages
- 3299-3308
- ISSN
- 0018-9456 (ISSN-L)
- 1557-9662
- OpenAlex
- W2755048715
- Cited by
- 140
- References
- 5
- FWCI
- 4.07
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Nonlinear Cuffless Blood Pressure Estimation of Healthy Subjects Using Pulse Transit Time and Arrival Time. https://app.cochat.ai/papers/10.1109/tim.2017.2745081
- The paper
- Esmaili, A., Kachuee, M., & Shabany, M. (2017). Nonlinear Cuffless Blood Pressure Estimation of Healthy Subjects Using Pulse Transit Time and Arrival Time. IEEE Transactions on Instrumentation and Measurement. https://doi.org/10.1109/tim.2017.2745081
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.