PPG2ABP: Translating photoplethysmogram (PPG) signals to arterial blood pressure (ABP) waveforms
Open accessAbstract
Cardiovascular diseases are one of the most severe causes of mortality, annually taking a heavy toll on lives worldwide. Continuous monitoring of blood pressure seems to be the most viable option, but this demands an invasive process, introducing several layers of complexities and reliability concerns due to non-invasive techniques not being accurate. This motivates us to develop a method to estimate the continuous arterial blood pressure (ABP) waveform through a non-invasive approach using Photoplethysmogram (PPG) signals. We explore the advantage of deep learning, as it would free us from sticking to ideally shaped PPG signals only by making handcrafted feature computation irrelevant, which is a shortcoming of the existing approaches. Thus, we present PPG2ABP, a two-stage cascaded deep learning-based method that manages to estimate the continuous ABP waveform from the input PPG signal with a mean absolute error of 4.604 mmHg, preserving the shape, magnitude, and phase in unison. However, the more astounding success of PPG2ABP turns out to be that the computed values of Diastolic Blood Pressure (DBP), Mean Arterial Pressure (MAP), and Systolic Blood Pressure (SBP) from the estimated ABP waveform outperform the existing works under several metrics (mean absolute error of 3.449 ± 6.147 mmHg, 2.310 ± 4.437 mmHg, and 5.727 ± 9.162 mmHg, respectively), despite that PPG2ABP is not explicitly trained to do so. Notably, both for DBP and MAP, we achieve Grade A in the BHS (British Hypertension Society) Standard and satisfy the AAMI (Association for the Advancement of Medical Instrumentation) standard.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Algorithm developmental biases stem from selecting features, adjusting training data composition, and model calibrations.
“The accuracy of heart rate measurements affects people with darker skin tones because of algorithm developmental biases that stem from selecting features and adjusting training data composition and model calibrations [23-26].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [24]
The Experiments sections (3.2–3.5) describe choices of loss functions, filters, and training parameters for estimating arterial blood pressure from PPG signals. A search for skin tone, heart-rate accuracy, and demographic algorithmic bias found no passage linking feature selection, training-data composition, or calibration to less accurate heart-rate measurements in people with darker skin. (LLM, conf=0.77)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
U-Net: Convolutional Networks for Biomedical Image Segmentation
Ronneberger, O., Fischer, P., & Brox, T. (2015) · Lecture notes in computer science · doi:10.1007/978-3-319-24574-4_28 · Cited by 99,404 · see citing works
Goldberger, A. L., Amaral, L. A. N., Glass, L., Hausdorff, J. M., Ivanov, P. C., Mark, R. G., Mietus, J. E., Moody, G. B., Peng, C.-K., & Stanley, H. E. (2000) · Circulation · doi:10.1161/01.cir.101.23.e215 · Cited by 14,809 · see citing works
De-noising by soft-thresholding
Donoho, D. L. (1995) · IEEE Transactions on Information Theory · doi:10.1109/18.382009 · Cited by 9,655 · see citing works
Ideal Spatial Adaptation by Wavelet Shrinkage
Donoho, D. L., & Johnstone, I. M. (1994) · Biometrika · doi:10.1093/biomet/81.3.425 · Cited by 8,730 · see citing works
Understanding bland altman analysis
Giavarina, D. (2015) · Biochemia Medica · doi:10.11613/bm.2015.015 · Cited by 4,022 · see citing works
MultiResUNet : Rethinking the U-Net architecture for multimodal biomedical image segmentation
Ibtehaz, N., & Rahman, M. S. (2019) · Neural Networks · doi:10.1016/j.neunet.2019.08.025 · Cited by 2,334 · see citing works
Self-citationCardiovascular disease in Europe: epidemiological update 2016
Townsend, N. P., Wilson, L., Bhatnagar, P., Wickramasinghe, K. K., Rayner, M., & Nichols, M. S. (2016) · European Heart Journal · doi:10.1093/eurheartj/ehw334 · Cited by 1,579 · see citing works
Laflamme, M. A., & Murry, C. E. (2011) · Nature · doi:10.1038/nature10147 · Cited by 1,231 · see citing works
Saeed, M., Villarroel, M., Reisner, A. T., Clifford, G., Lehman, L.-W., Moody, G., Heldt, T., Kyaw, T. H., Moody, B., & Mark, R. G. (2011) · Critical Care Medicine · doi:10.1097/ccm.0b013e31820a92c6 · Cited by 1,162 · see citing works
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945 · see citing works
Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0287-x · Cited by 895 · see citing works
The velocity of pulse wave in man
Bramwell, J. C., & Hill, A. V. (1922) · Proceedings of the Royal Society of London Series B Containing Papers of a Biological Character · doi:10.1098/rspb.1922.0022 · Cited by 810 · see citing works
Catheter-Based Renal Sympathetic Denervation for Resistant Hypertension
Symplicity HTN-1 Investigators (2011) · Hypertension · doi:10.1161/hypertensionaha.110.163014 · Cited by 743 · see citing works
Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621 · see citing works
Brien, E. O., Petrie, J. C., Little, W. C., de Swiet, M., Padfield, P. L., Altma, D. G., Bland, M., Coats, A. J. S., & Atkins, N. (1993) · Journal of Hypertension · doi:10.1097/00004872-199306000-00013 · Cited by 518 · see citing works
Optimal selection of wavelet basis function applied to ECG signal denoising
Singh, B. N., & Tiwari, A. K. (2006) · Digital Signal Processing · doi:10.1016/j.dsp.2005.12.003 · Cited by 511 · see citing works
Gesche, H., Grosskurth, D., Küchler, G., & Patzak, A. R. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1983-3 · Cited by 472 · see citing works
Flexible Piezoresistive Sensor Patch Enabling Ultralow Power Cuffless Blood Pressure Measurement
Luo, N., Dai, W., Li, C., Zhou, Z., Lu, L., Poon, C. C. Y., Chen, S., Zhang, Y., & Zhao, N. (2015) · Advanced Functional Materials · doi:10.1002/adfm.201504560 · Cited by 466 · see citing works
Blood Pressure Estimation from Photoplethysmogram Using a Spectro-Temporal Deep Neural Network
Slapničar, G., Mlakar, N., & Luštrek, M. (2019) · Sensors · doi:10.3390/s19153420 · Cited by 387 · see citing works
Cuff-less high-accuracy calibration-free blood pressure estimation using pulse transit time
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2015) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2015.7168806 · Cited by 351 · see citing works
Optimal Signal Quality Index for Photoplethysmogram Signals
Elgendi, M. (2016) · Bioengineering · doi:10.3390/bioengineering3040021 · Cited by 338 · see citing works
Monte-Moreno, E. (2011) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2011.05.001 · Cited by 311 · see citing works
Pulse Transit Time as an Indicator of Arterial Blood Pressure
Geddes, L. A., Voelz, M. H., Babbs, C. F., Bourland, J. D., & Tacker, W. A. (1981) · Psychophysiology · doi:10.1111/j.1469-8986.1981.tb01545.x · Cited by 293 · see citing works
(2012) · PubMed · Cited by 264 · see citing works
Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data
Krishnan, R., Natarajan, B., & Warren, S. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2039568 · Cited by 263 · see citing works
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263 · see citing works
Peter, L., Noury, N., & Cerny, M. (2014) · IRBM · doi:10.1016/j.irbm.2014.07.002 · Cited by 261 · see citing works
An optimal filter for short photoplethysmogram signals
Liang, Y., Elgendi, M., Chen, Z., & Ward, R. (2018) · Scientific Data · doi:10.1038/sdata.2018.76 · Cited by 242 · see citing works
Blood pressure estimation from appropriate and inappropriate PPG signals using A whole-based method
Mousavi, S. S., Firouzmand, M., Charmi, M., Hemmati, M. R., Moghadam, M., & Ghorbani, Y. (2018) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2018.08.022 · Cited by 234 · see citing works
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226 · see citing works
Kim, S.-H., Lilot, M., Sidhu, K. S., Rinehart, J. B., Yu, Z., Canales, C., & Cannesson, M. P. (2014) · Anesthesiology · doi:10.1097/aln.0000000000000226 · Cited by 206 · see citing works
Wong, M. Y. M., Poon, C. C. Y., & Zhang, Y. (2009) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-009-9070-7 · Cited by 191 · see citing works
Esmaili, A., Kachuee, M., & Shabany, M. (2017) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2017.2745081 · Cited by 155 · see citing works
Real-Time Cuffless Continuous Blood Pressure Estimation Using Deep Learning Model
Li, Y., Harfiya, L. N., Purwandari, K., & Lin, Y. (2020) · Sensors · doi:10.3390/s20195606 · Cited by 155 · see citing works
Miao, F., Wen, B., Hu, Z., Fortino, G., Wang, X., Liu, Z.-D., Tang, M., & Li, Y. (2020) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2020.101919 · Cited by 145 · see citing works
Esmaelpoor, J., Moradi, M. H., & Kadkhodamohammadi, A. (2020) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2020.103719 · Cited by 144 · see citing works
Multi-Sensor Fusion Approach for Cuff-Less Blood Pressure Measurement
Miao, F., Liu, Z.-D., Liu, J., Wen, B., He, Q., & Li, Y. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2901724 · Cited by 136 · see citing works
Athaya, T., & Choi, S. (2021) · Sensors · doi:10.3390/s21051867 · Cited by 136 · see citing works
Ilies, C., Bauer, M., van den Berg, P., Rosenberg, J., Hedderich, J., Bein, B., Hinz, J., & Hanß, R. (2011) · British Journal of Anaesthesia · doi:10.1093/bja/aer394 · Cited by 128 · see citing works
Kavsaoğlu, A. R., Polat, K., & Hariharan, M. (2015) · Applied Soft Computing · doi:10.1016/j.asoc.2015.04.008 · Cited by 114 · see citing works
Harfiya, L. N., Chang, C.-C., & Li, Y. (2021) · Sensors · doi:10.3390/s21092952 · Cited by 106 · see citing works
Thambiraj, G., Gandhi, U., Umapathy, M., Jose, V. J. M., & Anand, M. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101942 · Cited by 99 · see citing works
Zhang, B., Ren, J., Cheng, Y., Wang, B., & Wei, Z. (2019) · IEEE Access · doi:10.1109/access.2019.2902217 · Cited by 93 · see citing works
Proença, J., Muehlsteff, J., Aubert, X. L., & Carvalho, P. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5626627 · Cited by 92 · see citing works
Hsu, Y.-C., Li, Y., Chang, C.-C., & Harfiya, L. N. (2020) · Sensors · doi:10.3390/s20195668 · Cited by 92 · see citing works
Baker, S., Xiang, W., & Atkinson, I. (2021) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2021.106191 · Cited by 90 · see citing works
Mahmud, S., Ibtehaz, N., Khandakar, A. A., Tahir, A., Rahman, T., Islam, K. R., Hossain, M. S., Rahman, M. S., Musharavati, F., Ayari, M. A., et al. (2022) · Sensors · doi:10.3390/s22030919 · Cited by 86 · see citing works
Self-citationRong, M., & Li, K. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102772 · Cited by 85 · see citing works
Feature-Based Neural Network Approach for Oscillometric Blood Pressure Estimation
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolić, M., & Rajan, S. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2123210 · Cited by 80 · see citing works
Cuffless Blood Pressure Monitoring Using Hydrostatic Pressure Changes
Shaltis, P., Reisner, A. T., & Asada, H. H. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.919142 · Cited by 71 · see citing works
Hill, B. L., Rakocz, N., Rudas, Á., Chiang, J. N., Wang, S., Hofer, I. S., Cannesson, M. P., & Halperin, E. (2021) · Scientific Reports · doi:10.1038/s41598-021-94913-y · Cited by 71 · see citing works
Hahn, R., Rinösl, H., Neuner, M. P., et al. (2012) · British Journal of Anaesthesia · doi:10.1093/bja/aer499 · Cited by 70 · see citing works
Continuous cuffless blood pressure monitoring based on PTT
Shriram, R., Wakankar, A., Daimiwal, N., & Ramdasi, D. S. (2010) · International Conference on Bioinformatics and Biomedical Technology (ICBBT) · doi:10.1109/icbbt.2010.5479013 · Cited by 62 · see citing works
Qin, K., Huang, W. X., & Zhang, T. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102972 · Cited by 60 · see citing works
Mahmud, S., Ibtehaz, N., Khandakar, A. A., Rahman, M. S., Gonzales, A. J., Rahman, T., Hossain, M. S., Hossain, M. S. A., Faisal, M. A. A., Abir, F. F., et al. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.104247 · Cited by 60 · see citing works
Self-citationEnhancing the estimation of blood pressure using pulse arrival time and two confounding factors
Baek, H. J., Kim, K. K., Kim, J. S., Lee, B., & Park, K. S. (2009) · Physiological Measurement · doi:10.1088/0967-3334/31/2/002 · Cited by 59 · see citing works
A Deep Learning Approach to Predict Blood Pressure from PPG Signals
Tazarv, A., & Levorato, M. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9629687 · Cited by 58 · see citing works
Li, P., & Laleg‐Kirati, T. (2021) · IEEE Access · doi:10.1109/access.2021.3065576 · Cited by 56 · see citing works
Mehrabadi, M. A., Aqajari, S. A. H., Zargari, A. H. A., Dutt, N. D., & Rahmani, A. M. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871962 · Cited by 48 · see citing works
PPG-Based Systolic Blood Pressure Estimation Method Using PLS and Level-Crossing Feature
Fujita, D., Suzuki, A., & Ryu, K. (2019) · Applied Sciences · doi:10.3390/app9020304 · Cited by 46 · see citing works
Monitoring heart and respiratory rates at radial artery based on PPG
Wang, C., Li, Z., & Wei, X. (2013) · Optik · doi:10.1016/j.ijleo.2012.11.044 · Cited by 44 · see citing works
Monitoring of reactive hyperemia using photoplethysmographic pulse amplitude and transit time
Selvaraj, N., Jaryal, A. K., Santhosh, J., Anand, S., & Deepak, K. K. (2009) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-009-9199-3 · Cited by 38 · see citing works
Beat-Based PPG-ABP Cleaning Technique for Blood Pressure Estimation
Salah, M., Omer, O. A., Hassan, L., Ragab, M., Hassan, A. M., & Abdelreheem, A. A. (2022) · IEEE Access · doi:10.1109/access.2022.3175436 · Cited by 29 · see citing works
Hossain, M. S., Chowdhury, M. E. H., Reaz, M. B. I., Ali, S. H. M., Bakar, A. A. A., Serkan Kiranyaz, B., Khandakar, A. A., Alhatou, M. I., Habib, R., & Hossain, M. (2022) · Sensors · doi:10.3390/s22093169 · Cited by 27 · see citing works
Self-citationMarcinkevičs, Z., Grēve, M., Aivars, J. I., Erts, R., & Zehtabi, A. H. (2009) · Cited by 24 · see citing works
Thambiraj, G., Gandhi, U., Devanand, V., & Umapathy, M. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab1f17 · Cited by 20 · see citing works
Kamboj, N., Chang, K., Metcalfe, K. A., Chu, C. H., & Conway, A. (2021) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2021.103091 · Cited by 17 · see citing works
A fusion approach to improve accuracy and estimate uncertainty in cuffless blood pressure monitoring
Landry, C., Peterson, S. D., & Arami, A. (2022) · Scientific Reports · doi:10.1038/s41598-022-12087-7 · Cited by 16 · see citing works
Bose, S. S. N., & Kandaswamy, A. (2017) · International Conference on Advanced Computing and Communication Systems (ICACCS) · doi:10.1109/icaccs.2017.8014669 · Cited by 12 · see citing works
HYPE: Predicting Blood Pressure from Photoplethysmograms in a Hypertensive PopulationPreprint
Sasso, A. M., Datta, S., Jeitler, M., et al. (2020) · medRxiv · doi:10.1101/2020.05.27.20107243 · Cited by 4 · see citing works
Pradenas, L. (2020) · Biomedical Journal of Scientific & Technical Research · doi:10.26717/bjstr.2020.30.004883 · Cited by 3 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- [68]Lee C.-Y. Xie S. Gallagher P. Zhang Z. Tu Z. Deeply-supervised nets Proceedings of the Artificial Intelligence and Statistics Valencia, Spain 28–30 March 2015 562 570
- [71]Kingma D.P. Ba J. Adam: A method for stochastic optimization arXiv 2014 1412.6980
- [76]Ibtehaz N. Mahmud S. Chowdhury M.E.H. Khandakar A. Ayari M.A. Tahir A. Rahman M.S. Ppg2abp: Translating photoplethysmogram (ppg) signals to arterial blood pressure (abp) waveforms using fully convolutional neural networks arXiv 2020 2005.01669 PMC9687508 36421093
Cited by
Sorted by citations
Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring
Asif, S., AlSaafeen, A., Nadar, S., Nambiar, S., Dannawi, J., Korrapati, H., & Wilkhoo, H. S. (2025) · Cureus · doi:10.7759/cureus.94074 · Cited by 9 · see citing works
- Authors
- Ibtehaz, N., Mahmud, S., Chowdhury, M. E. H., Khandakar, A., Salman Khan, M., Ayari, M. A., Tahir, A. M., & Rahman, M. S.
- Affiliations
- Purdue University West Lafayette (US)
- Qatar University (QA)
- Department of Electrical Engineering, Qatar University, Doha 2713, Qatar
- University of British Columbia (CA)
- Bangladesh University of Engineering and Technology (BD)
- Corresponding author
- Chowdhury, Muhammad E. H.
- Rahman, M. Sohel
- ORCID
- Venue
- Bioengineering
- Publisher
- MDPI AG
- Published
- 2022-11-15
- Volume
- 9
- Issue
- 11
- Pages
- 692
- ISSN
- 2306-5354 (ISSN-L)
- PMC
- PMC9687508
- OpenAlex
- W4309119158
- Cited by
- 133
- References
- 71
- Citations by year
- 2026: 27 · 2025: 36 · 2024: 37 · 2023: 21 · 2022: 10 · 2021: 1 · 2020: 1
- FWCI
- 7.56
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Bioengineering (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- cIRcle (University of British Columbia) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Qatar University QSpace (Qatar University) (submittedVersion)
- Open Collections
- Funders
- Qatar National Research Fund (NPRP12S-0227-190164, QUST-2-CENG-2022-711)
- Qatar University (NPRP12S-0227-190164, QUST-2-CENG-2022-711)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- photoplethysmogram, arterial blood pressure estimation, non-invasive blood pressure monitoring, deep learning, systolic blood pressure, diastolic blood pressure, mean arterial pressure, AAMI standards
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review