Potential errors in pulse oximetry: II. Effects of changes in saturation and signal quality*
Open accessThe DOI matches its Crossref record; 43 of the 43 references it cites were found in a registry or an index.
Abstract
Summary: The published studies of pulse oximeter performance under conditions of normal, high and low saturation, exercise, poor signal quality and cardiac arrhythmia are reviewed. Most pulse oximeters have an absolute mean error of less than 2% at normal saturation and perfusion; two‐thirds have a standard deviation (SD) of less than 2%, and the remainder an SD of less than 3%. Some pulse oximeters tend to read 100% with fractional saturations of 97–98%. Pulse oximeters may be suitable hyperoxic alarms for neonates if the alarm limit chosen is directly validated for each device. Pulse oximeters are poorly calibrated at low saturations and are generally less accurate and less precise than at normal saturations; nearly 30% of 244 values reviewed were in error by more than 5% at saturations of less than 80%. Ear, nose and forehead probes respond more rapidly to rapid desaturation than finger probes, but are generally less accurate and less precise. Ear oximetry may be inaccurate during exercise. Low signal quality can result in failure to present a saturation reading, but data given with low signal quality warning messages are generally no less accurate than those without. Cardiac arrhythmias do not decrease accuracy of pulse oximeters so long as saturation readings are steady.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
YELDERMAN, M., & NEW, W. (1983) · Anesthesiology · doi:10.1097/00000542-198310000-00015 · Cited by 486
Effects of Methemoglobinemia on Pulse Oximetry and Mixed Venous Oximetry
Barker, S. J., Tremper, K. K., & Hyatt, J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00021 · Cited by 382
Accuracy of Response of Six Pulse Oximeters to Profound Hypoxia
Severinghaus, J. W., & Naifeh, K. H. (1987) · Anesthesiology · doi:10.1097/00000542-198710000-00017 · Cited by 312
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring · doi:10.1007/bf01617328 · Cited by 306
Errors in 14 pulse oximeters during profound hypoxia
Severinghaus, J. W., Naifeh, K. H., & Koh, S. O. (1989) · Journal of Clinical Monitoring · doi:10.1007/bf01617877 · Cited by 217
Potential errors in pulse oximetry: I. Pulse oximeter evaluation*
RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09410.x · Cited by 166
Self-citationNoninvasive Detection of Profound Arterial Desaturations Using a Pulse Oximetry Device
Mihm, F. G., & Halperin, B. D. (1985) · Anesthesiology · doi:10.1097/00000542-198501000-00020 · Cited by 153
Neonatal Pulse Oximetry: Accuracy and Reliability
Hay, W. W., Brockway, J. M., & Eyzaguirre, M. (1989) · Pediatrics · doi:10.1542/peds.83.5.717 · Cited by 141
Blood Flow Limits and Pulse Oximeter Signal Detection
Lawson, D., Norley, I., Korbon, G., Loeb, R. G., & Ellis, J. B. (1987) · Anesthesiology · doi:10.1097/00000542-198710000-00032 · Cited by 137
Bias and Precision of Pulse Oximeters and Arterial Oximeters
Nickerson, B. G., Sarkisian, C. A., & Tremper, K. K. (1988) · CHEST Journal · doi:10.1378/chest.93.3.515 · Cited by 132
Pulse Oximetry: Uses and Limitations
DeMeulenaere, S. (2007) · The Journal for Nurse Practitioners · doi:10.1016/j.nurpra.2007.02.021 · Cited by 112
A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion
Clayton, D. G., Webb, R. K., Ralston, A., Duthie, D. J. R., & Runciman, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09303.x · Cited by 112
Self-citationPulse Oximetry: Technical Aspects of Machine Design
Pologe, J. A. (1987) · International Anesthesiology Clinics · doi:10.1097/00004311-198702530-00009 · Cited by 110
Evaluation of the Ohmeda 3700 Pulse Oximeter: Steady-state and Transient Response Characteristics
Kagle, D. M., Alexander, C. M., Berko, R. S., Gluffre, M., & Gross, J. B. (1987) · Anesthesiology · doi:10.1097/00000542-198703000-00018 · Cited by 109
Hemoglobin desaturation in highly trained athletes during heavy exercise
WILLIAMS, J. H., POWERS, S. K., & STUART, M. K. (1986) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-198604000-00004 · Cited by 107
Siggaard‐Andersen, O., Nørgaard‐Pedersen, B., & Rem, J. (1972) · Clinica Chimica Acta · doi:10.1016/0009-8981(72)90380-4 · Cited by 92
CLAYTON, D. G., WEBB, R. K., RALSTON, A. C., DUTHIE, D., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11492.x · Cited by 90
Self-citationHyperoxemia in Newborn Infants: Detection by Pulse Oximetry
Bucher, H.-U., Fanconi, S., Baeckert, P., & Duc, G. V. (1989) · PEDIATRICS · doi:10.1542/peds.84.2.226 · Cited by 84
The accuracy of pulse oximeters: A comparative clinical evaluation of five pulse oximeters
Taylor, M. B., & WHITWAM, J. G. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb05549.x · Cited by 82
Ear oximetry during combined hypoxia and exercise
Smyth, R. J., D'Urzo, A. D., Slutsky, A. S., Galko, B. M., & Rebuck, A. S. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.60.2.716 · Cited by 80
Validity of Ear Oximetry in Clinical Exercise Testing
Hansen, J. E., & Casaburi, R. H. (1987) · CHEST Journal · doi:10.1378/chest.91.3.333 · Cited by 75
Design and evaluation of a new reflectance pulse oximeter sensor.
Mendelson, Y., Kent, J. C., Yocum, B. L., & Birle, M. J. (1988) · PubMed · Cited by 63
GABRIELCZYK, M. R., & BUIST, R. J. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb09025.x · Cited by 59
Morris, R., Nairn, M., & Torda, T. A. (1989) · Anaesthesia and Intensive Care · doi:10.1177/0310057x8901700113 · Cited by 57
COMPARISON OF PULSE OXIMETERS: EFFECTS OF VASOCONSTRICTION AND VENOUS ENGORGEMENT
Wilkins, C. J., MOORES, M., & Hanning, C. D. (1989) · British Journal of Anaesthesia · doi:10.1093/bja/62.4.439 · Cited by 56
West, P., George, C. F., & Kryger, M. H. (1987) · SLEEP · doi:10.1093/sleep/10.3.263 · Cited by 55
Accuracy of pulse oximetry to estimate HbO2 fraction of total Hb during exercise
Powers, S. K., Dodd, S., Freeman, J., Ayers, G. D., Samson, H., & McKnight, T. (1989) · Journal of Applied Physiology · doi:10.1152/jappl.1989.67.1.300 · Cited by 53
NONINVASIVE PULSE OXIMETRY IN CHILDREN WITH CYANOTIC CONGENITAL HEART DISEASE
Boxer, R. A., GOTTESFELD, I. B., Singh, S., LaCorte, M. A., Parnell, V. A., & WALKER, P. (1987) · Critical Care Medicine · doi:10.1097/00003246-198711000-00015 · Cited by 49
Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery
PALVE, H., & Vuori, A. (1989) · Critical Care Medicine · doi:10.1097/00003246-198901000-00014 · Cited by 49
ACCURACY OF A PULSE OXIMETER IN THE CRITICALLY ILL ADULT
Tremper, K. K., Hufstedler, S. M., Barker, S. J., Adams, A. L., Wong, D. H. W., Zaccari, J., Benik, K., & Lemons, V. (1985) · Anesthesiology · doi:10.1097/00000542-198509001-00175 · Cited by 37
Relationship of pulse oximetry to arterial oxygen tension in infants
Walsh, M. C., NOBLE, L. M., Carlo, W. A., & Martin, R. J. (1987) · Critical Care Medicine · doi:10.1097/00003246-198712000-00005 · Cited by 35
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
An assessment of blood vessel vasoactivity using photoplethysmography.
Evans, M., & Geddes, L. A. (1988) · PubMed · Cited by 28
Comparison of recorded values from six pulse oximeters
Choe, H., Tashiro, C., Fukumitsu, K., YAGI, M., & Yoshiya, I. (1989) · Critical Care Medicine · doi:10.1097/00003246-198907000-00016 · Cited by 28
Ridley, S. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb05484.x · Cited by 26
Definition and notation of hemoglobin oxygen saturation.
Zijlstra, W. G., & Oeseburg, B. (1989) · PubMed · Cited by 22
Reusing the Nellcor Pulse Oximeter Probe: Is it Worth the Savings?
Racys, V. A., & Nahrwold, M. L. (1987) · Anesthesiology · doi:10.1097/00000542-198705000-00039 · Cited by 16
VENOUS CONGESTION AFFECTS ARTERIAL HEMOGLOBIN SATURATION MEASURED BY THE PULSE OXIMETER
Kim, J.-M., & Mathewson, H. S. (1985) · Anesthesiology · doi:10.1097/00000542-198509001-00174 · Cited by 15
ACCURACY OF PULSE OIIMETRY IN CYANOTIC CONGENITAL HEART DISEASE
Lazzell, V. A., & Jopling, M. W. (1987) · Anesthesiology · doi:10.1097/00000542-198709001-00169 · Cited by 14
Cool Fingers and Pulse Oximetry
Paulus, D. A., & Monroe, M. (1989) · Anesthesiology · doi:10.1097/00000542-198907000-00034 · Cited by 11
A TEST OBJECT FOR ASSESSING PULSE OXIMETERS
Munley, A. J., Sik, M. J., & Shaw, A. G. (1989) · The Lancet · doi:10.1016/s0140-6736(89)92447-1 · Cited by 8
Reusing the Nellcor Pulse Probe
KAO, Y. J., & Badgwell, J. M. (1987) · Anesthesiology · doi:10.1097/00000542-198711000-00059 · Cited by 6
Cited by
Rice, T. W., Wheeler, A. P., Bernard, G. R., Hayden, D. L., Schoenfeld, D. A., & Ware, L. B. (2007) · Chest · doi:10.1378/chest.07-0617 · Cited by 820
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 491
Jubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 401
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243
RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11501.x · Cited by 192
Self-citation
Metadata
- Authors
- WEBB, R. K., RALSTON, A. C., & RUNCIMAN, W. B.
- Venue
- Anaesthesia
- Publisher
- Wiley
- Published
- 1991-03
- Volume
- 46
- Issue
- 3
- Pages
- 207-212
- ISSN
- 0003-2409 (ISSN-L)
- 1365-2044
- OpenAlex
- W1567738293
- Cited by
- 213
- References
- 43
- FWCI
- 9.78
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.