Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants
Open accessAbstract
For nearly 55 years, tungsten microwires have been widely used in neurophysiological experiments in animal models to chronically record neuronal activity. While tungsten microwires initially provide stable recordings, their inability to reliably record high-quality neural signals for tens of years has limited their efficacy for neuroprosthetic applications in humans. Comprehensive understanding of the mechanisms of electrode performance and failure is necessary for developing next generation neural interfaces for humans. In this study, we evaluated the abiotic (electrophysiology, impedance, electrode morphology) and biotic (microglial reactivity, blood-brain barrier disruption, biochemical markers of axonal injury) effects of 16-channel, 50 µm diameter, polyimide insulated tungsten microwires array for implant durations that ranged from acute to up to 9 months in 25 rats. Daily electrode impedance spectroscopy, electrophysiological recordings, blood and cerebrospinal fluid (CSF) withdrawals, and histopathological analysis were performed to study the time-varying effects of chronic electrode implantation. Structural changes at the electrode recording site were observed as early as within 2-3 h of electrode insertion. Abiotic analysis indicated the first 2-3 weeks following surgery was the most dynamic period in the chronic electrode lifetime as there were greater variations in the electrode impedance, functional electrode performance, and the structural changes occurring at the electrode recording tips. Electrode recording site deterioration continued for the long-term chronic animals as insulation damage occurred and recording surface became more recessed over time. In general, electrode impedance and functional performance had smaller daily variations combined with reduced electrode recording site changes during the chronic phase. Histopathological studies were focused largely on characterizing microglial cell responses to electrode implantation. We found that activated microglia were present near the electrode tracks in all non-acute animals studied, thus indicating presence of a neuroinflammatory response regardless of post-implantation survival times and electrode performance. Conversely, dystrophic microglia detectable as fragmented cells were found almost exclusively in acute animals surviving only few hours after implantation. While there was no consistent relationship between microglial cell responses and electrode performance, we noticed co-occurrence of high ferritin expression, intraparenchymal bleeding, and microglial degeneration suggesting presence of excessive oxidative stress via Fenton chemistry. Biochemical analysis indicated that these electrodes always caused a persistent release of axonal injury biomarkers even several months after implantation suggesting persistent tissue damage. Our study suggests that mechanisms of electrode failure are multi-factorial involving both abiotic and biotic parameters. Since these failure modes occur concurrently and cannot be isolated from one another, the lack of consistent relationship between electrode performance and microglial responses in our results suggest that one or more of the abiotic factors were equally responsible for degradation in electrode performance over long periods of time.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Neural injury and inflammation that accompany device implantation are key contributing factors to signal loss and instability.
“but the neural injury and inflammation that accompany device implantation are key contributing factors.”
From Regenerative Electrode Interfaces for Neural Prostheses. (2016) · cites this paper as [16]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
The rat brain in stereotaxic coordinates
Palkovits, M. (1983) · Neuropeptides · doi:10.1016/0143-4179(83)90049-5 · Cited by 28,959 · see citing works
The Rat Brain in Stereotaxic Coordinates
Paxinos, G., & Watson, C. (1983) · Elsevier eBooks · doi:10.1016/c2009-0-63235-9 · Cited by 28,585 · see citing works
Microglia: a sensor for pathological events in the CNS
Kreutzberg, G. W. (1996) · Trends in Neurosciences · doi:10.1016/0166-2236(96)10049-7 · Cited by 4,396 · see citing works
Neuronal ensemble control of prosthetic devices by a human with tetraplegia
Hochberg, L. R., Serruya, M. D., Friehs, G. M., Mukand, J. A., Saleh, M., Caplan, A. H., Branner, A., Chen, D., Penn, R. D., & Donoghue, J. P. (2006) · Nature · doi:10.1038/nature04970 · Cited by 3,295 · see citing works
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · see citing works
Learning to control a brain–machine interface for reaching and grasping by primates
Carmena, J. M., Lebedev, M. A., Crist, R. E., O'Doherty, J. E., Santucci, D. M., Dimitrov, D. F., Patil, P. G., Henriquez, C. S., & Nicolelis, M. A. L. (2003) · PLoS Biology · doi:10.1371/journal.pbio.0000042 · Cited by 1,761 · see citing works
Astrocyte activation and reactive gliosis
Pekny, M., & Nilsson, M. (2005) · Glia · doi:10.1002/glia.20207 · Cited by 1,611 · see citing works
Dijkstra, C. D., Döpp, E. A., Joling, P., & Kraal, G. (1985) · Springer eBooks · doi:10.1007/978-1-4613-2463-8_50 · Cited by 1,598 · see citing works
Cortical control of a prosthetic arm for self-feeding
Velliste, M., Perel, S., Spalding, M. C., Whitford, A. S., & Schwartz, A. B. (2008) · Nature · doi:10.1038/nature06996 · Cited by 1,591 · see citing works
Instant neural control of a movement signal
Serruya, M. D., Hatsopoulos, N. G., Paninski, L., Fellows, M. R., & Donoghue, J. P. (2002) · Nature · doi:10.1038/416141a · Cited by 1,415 · see citing works
A review of methods for spike sorting: the detection and classification of neural action potentials
Lewicki, M. S. (1998) · Network Computation in Neural Systems · doi:10.1088/0954-898x_9_4_001 · Cited by 1,372 · see citing works
Biran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893 · see citing works
Tungsten Microelectrode for Recording from Single Units
Hubel, D. H. (1957) · Science · doi:10.1126/science.125.3247.549 · Cited by 852 · see citing works
Raivich, G., Bohatschek, M., Kloss, C. U. A., Werner, A., Jones, L. L., & Kreutzberg, G. W. (1999) · Brain Research Reviews · doi:10.1016/s0165-0173(99)00007-7 · Cited by 843 · see citing works
Brain responses to micro-machined silicon devices
Szarowski, D. H., Andersen, M. D., Retterer, S. T., Spence, A. J., Isaacson, M. S., Craighead, H. G., Turner, J. N., & Shain, W. G. (2003) · Brain Research · doi:10.1016/s0006-8993(03)03023-3 · Cited by 798 · see citing works
Brain-Controlled Interfaces: Movement Restoration with Neural Prosthetics
Schwartz, A. B., Cui, X. T., Weber, D. J., & Moran, D. W. (2006) · Neuron · doi:10.1016/j.neuron.2006.09.019 · Cited by 792 · see citing works
Chronic, multisite, multielectrode recordings in macaque monkeys
Nicolelis, M. A. L., Dimitrov, D., Carmena, J. M., Crist, R., Lehew, G., Kralik, J. D., & Wise, S. P. (2003) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1934665100 · Cited by 740 · see citing works
Microglia: biology and pathology
Graeber, M., & Streit, W. (2009) · Acta Neuropathologica · doi:10.1007/s00401-009-0622-0 · Cited by 729 · see citing works
Self-citationStreit, W. J., Braak, H., Xue, Q.-S., & Bechmann, I. (2009) · Acta Neuropathologica · doi:10.1007/s00401-009-0556-6 · Cited by 727 · see citing works
Self-citationDystrophic microglia in the aging human brain
Streit, W. J., Sammons, N. W., Kuhns, A. J., & Sparks, D. L. (2004) · Glia · doi:10.1002/glia.10319 · Cited by 705 · see citing works
Self-citationNicolelis, M. A. L. (2001) · Nature · doi:10.1038/35053191 · Cited by 696 · see citing works
Chronic recording capability of the Utah Intracortical Electrode Array in cat sensory cortex
Rousche, P. J., & Normann, R. A. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(98)00031-4 · Cited by 646 · see citing works
Ludwig, K. A., Uram, J. D., Yang, J., Martin, D. C., & Kipke, D. R. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/1/007 · Cited by 638 · see citing works
Cerebral Astrocyte Response to Micromachined Silicon Implants
Turner, J. N., Shain, W. G., Szarowski, D. H., Andersen, M., Martins, S., Isaacson, M. S., & Craighead, H. G. (1999) · Experimental Neurology · doi:10.1006/exnr.1998.6983 · Cited by 593 · see citing works
Wadhwa, R., Lagenaur, C. F., & Cui, X. T. (2005) · Journal of Controlled Release · doi:10.1016/j.jconrel.2005.10.027 · Cited by 498 · see citing works
Simeral, J. D., Kim, S.-P., Black, M. J., Donoghue, J. P., & Hochberg, L. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/2/025027 · Cited by 496 · see citing works
McConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480 · see citing works
Factors influencing the biocompatibility of insertable silicon microshafts in cerebral cortex
Edell, D. J., Toi, V. V., McNeil, V. M., & Clark, L. D. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.141202 · Cited by 479 · see citing works
Principles of neural ensemble physiology underlying the operation of brain–machine interfaces
Nicolelis, M. A. L., & Lebedev, M. (2009) · Nature reviews. Neuroscience · doi:10.1038/nrn2653 · Cited by 477 · see citing works
Chronic Neural Recording Using Silicon-Substrate Microelectrode Arrays Implanted in Cerebral Cortex
Vetter, R. J., Williams, J. C., Hetke, J. F., Nunamaker, E. A., & Kipke, D. R. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.826680 · Cited by 477 · see citing works
The electrical properties of metal microelectrodes
Robinson, D. A. (1968) · Proceedings of the IEEE · doi:10.1109/proc.1968.6458 · Cited by 459 · see citing works
Suner, S., Fellows, M. R., Vargas-Irwin, C., Nakata, G. K., & Donoghue, J. P. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.857687 · Cited by 415 · cited 1× · see citing works
“Repeatability of such spikes and their peak-to-peak amplitudes relative to the background noise was used to classify them as coming from a single-unit [74, 75].”
MethodAbidian, M. R., Corey, J. M., Kipke, D. R., & Martin, D. C. (2010) · Small · doi:10.1002/smll.200901868 · Cited by 405 · see citing works
Massively Parallel Recording of Unit and Local Field Potentials With Silicon-Based Electrodes
Csicsvari, J., Henze, D. A., Jamieson, B., Harris, K. D., Sirota, A., Barthó, P., Wise, K. D., & Buzsáki, G. (2003) · Journal of Neurophysiology · doi:10.1152/jn.00116.2003 · Cited by 402 · see citing works
Complex impedance spectroscopy for monitoring tissue responses to inserted neural implants
Williams, J. C., Hippensteel, J. A., Dilgen, J., Shain, W. G., & Kipke, D. R. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/4/007 · Cited by 399 · see citing works
Williams, J. C., Rennaker, R. L., & Kipke, D. R. (1999) · Brain Research Protocols · doi:10.1016/s1385-299x(99)00034-3 · Cited by 396 · see citing works
Biological Performance Of Materials - Fundamentals Of Biocompatibility
Black, J. T. (1992) · Cited by 389 · see citing works
Drake, K. L., Wise, K. D., Farraye, J., Anderson, D. J., & BeMent, S. L. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7273 · Cited by 364 · see citing works
Kipke, D. R., Vetter, R. J., Williams, J. C., & Hetke, J. F. (2003) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2003.814443 · Cited by 364 · see citing works
Selective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 358 · cited 1× · see citing works
“While innovative developments of high-density microelectrode probes [15–24] have enabled the recording of signals from large neuronal ensembles, the current systems have been unable to access such signals for more than several years [16].”
BackgroundAbidian, M. R., & Martin, D. C. (2008) · Biomaterials · doi:10.1016/j.biomaterials.2007.11.022 · Cited by 339 · see citing works
Biran, R., Martin, D. C., & Tresco, P. A. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31138 · Cited by 336 · see citing works
Neuroglia in neurodegeneration
Heneka, M. T., Rodriguez, J. J., & Verkhratsky, A. (2009) · Brain Research Reviews · doi:10.1016/j.brainresrev.2009.11.004 · Cited by 303 · see citing works
Toward a comparison of microelectrodes for acute and chronic recordings
Ward, M. P., Rajdev, P., Ellison, C., & Irazoqui, P. P. (2009) · Brain Research · doi:10.1016/j.brainres.2009.05.052 · Cited by 296 · see citing works
Ludwig, K. A., Langhals, N. B., Joseph, M. D., Richardson-Burns, S. M., Hendricks, J. L., & Kipke, D. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/1/014001 · Cited by 283 · see citing works
Biomechanical analysis of silicon microelectrode-induced strain in the brain
Lee, H., Bellamkonda, R. V., Sun, W., & Levenston, M. E. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/4/003 · Cited by 281 · see citing works
Neuronal Interactions Improve Cortical Population Coding of Movement Direction
Maynard, E. M., Hatsopoulos, N. G., Ojakangas, C. L., Acuna, B. D., Sanes, J. N., Normann, R. A., & Donoghue, J. P. (1999) · Journal of Neuroscience · doi:10.1523/jneurosci.19-18-08083.1999 · Cited by 281 · see citing works
Primary Motor Cortex Tuning to Intended Movement Kinematics in Humans with Tetraplegia
Truccolo, W., Friehs, G. M., Donoghue, J. P., & Hochberg, L. R. (2008) · Journal of Neuroscience · doi:10.1523/jneurosci.4415-07.2008 · Cited by 273 · see citing works
Liu, X., McCreery, D. B., Carter, R. R., Bullara, L. A., Yuen, T. G. H., & Agnew, W. F. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.788468 · Cited by 272 · see citing works
Raivich, G., & Banati, R. (2004) · Brain research. Brain research reviews · doi:10.1016/j.brainresrev.2004.06.006 · Cited by 263 · see citing works
Lopes, K. O., Sparks, D. L., & Streit, W. J. (2008) · Glia · doi:10.1002/glia.20678 · Cited by 262 · see citing works
Self-citationRole of microglia in the central nervous system's immune response
Streit, W. J., Conde, J. R., Fendrick, S. E., Flanary, B. E., & Mariani, C. L. (2005) · Neurological Research · doi:10.1179/016164105x49463a · Cited by 258 · see citing works
Self-citationAbidian, M. R., Ludwig, K. A., Marzullo, T., Martin, D. C., & Kipke, D. R. (2009) · Advanced Materials · doi:10.1002/adma.200900887 · Cited by 256 · see citing works
Biological performance of materials — fundamentals of biocompatibility
Bruck, S. D. (1982) · Biomaterials · doi:10.1016/0142-9612(82)90013-8 · Cited by 246 · see citing works
Criteria for the Selection of Materials for Implanted Electrodes
Geddes, L. A., & Roeder, R. (2003) · Annals of Biomedical Engineering · doi:10.1114/1.1581292 · Cited by 240 · see citing works
A multielectrode array for intrafascicular recording and stimulation in sciatic nerve of cats
Branner, A., & Normann, R. A. (2000) · Brain Research Bulletin · doi:10.1016/s0361-9230(99)00231-2 · Cited by 234 · see citing works
Microglial Response to Brain Injury: A Brief Synopsis
Streit, W. J. (2000) · Toxicologic Pathology · doi:10.1177/019262330002800104 · Cited by 230 · see citing works
Self-citationHyperphosphorylated neurofilament NF-H is a serum biomarker of axonal injury
Shaw, G., Yang, C., Ellis, R. C., et al. (2005) · Biochemical and Biophysical Research Communications · doi:10.1016/j.bbrc.2005.08.252 · Cited by 229 · see citing works
Self-citationSpataro, L. E., Dilgen, J., Retterer, S. T., Spence, A. J., Isaacson, M. S., Turner, J. N., & Shain, W. G. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2004.08.037 · Cited by 215 · see citing works
Diffusion barriers evoked in the rat cortex by reactive astrogliosis
Roitbak, T., & Sykov�, E. (1999) · Glia · doi:10.1002/(sici)1098-1136(199910)28:1 · Cited by 208 · see citing works
Implant Size and Fixation Mode Strongly Influence Tissue Reactions in the CNS
Thelin, J., Jörntell, H., Psouni, E., Garwicz, M., Schouenborg, J., Danielsen, N., & Linsmeier, C. E. (2011) · PLoS ONE · doi:10.1371/journal.pone.0016267 · Cited by 188 · see citing works
Biocompatibility of silicon‐based electrode arrays implanted in feline cortical tissue
Schmidt, S., Horch, K., & Normann, R. A. (1993) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820271106 · Cited by 182 · see citing works
Anderson, K. J., Scheff, S. W., Miller, K. M., Roberts, K. N., Gilmer, L. K., Yang, C., & Shaw, G. (2008) · Journal of Neurotrauma · doi:10.1089/neu.2007.0488 · Cited by 180 · see citing works
Self-citationWinslow, B. D., & Tresco, P. A. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2009.11.049 · Cited by 178 · see citing works
Histopathological evaluation of materials implanted in the cerebral cortex
Stensaas, S. S., & Stensaas, L. J. (1978) · Acta Neuropathologica · doi:10.1007/bf00689766 · Cited by 177 · see citing works
Histopathologic evaluation of prolonged intracortical electrical stimulation
Agnew, W. F., Yuen, T. G. H., McCreery, D. B., & Bullara, L. A. (1986) · Experimental Neurology · doi:10.1016/0014-4886(86)90132-9 · Cited by 173 · see citing works
Corrosion of tungsten microelectrodes used in neural recording applications
Patrick, E. E., Orazem, M. E., Sanchez, J. C., & Nishida, T. (2011) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2011.03.012 · Cited by 168 · see citing works
Self-citationAzemi, E., Lagenaur, C. F., & Cui, X. T. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2010.09.033 · Cited by 164 · see citing works
Techniques for Chronic, Multisite Neuronal Ensemble Recordings in Behaving Animals
Kralik, J. D., Dimitrov, D. F., Krupa, D. J., Katz, D. B., Cohen, D., & Nicolelis, M. A. L. (2001) · Methods · doi:10.1006/meth.2001.1231 · Cited by 162 · see citing works
The microglia/macrophage response in the neonatal rat facial nucleus following axotomy
Graeber, M. B., López, F. M., Ikoma, E., Ishikawa, M., Imai, Y., Nakajima, K., Kreutzberg, G. W., & Kohsaka, S. (1998) · Brain Research · doi:10.1016/s0006-8993(98)00859-2 · Cited by 159 · see citing works
Quantifying long-term microelectrode array functionality using chronic in vivo impedance testing
Prasad, A., & Sanchez, J. C. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/2/026028 · Cited by 157 · see citing works
Self-citationTansey, K. E., Mehta, S., Hill, D., Foley, N. C., Hsieh, J., Ethans, K. D., Potter, P. J., Baverstock, R., Teasell, R. W., Wolfe, D. L., et al. (2010) · Cited by 152 · see citing works
Brain tissue reaction to some chronically implanted metals
Dymond, A. M., Le, K., Jurist, J. M., & Crandall, P. H. (1970) · Journal of neurosurgery · doi:10.3171/jns.1970.33.5.0574 · Cited by 140 · see citing works
A floating metal microelectrode array for chronic implantation
Musallam, S., Bak, M. J., Troyk, P. R., & Andersen, R. A. (2006) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2006.09.005 · Cited by 135 · see citing works
Otto, K. J., Johnson, M. D., & Kipke, D. R. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.862530 · Cited by 132 · see citing works
Repeated voltage biasing improves unit recordings by reducing resistive tissue impedances
Johnson, M. D., Otto, K. J., & Kipke, D. R. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847373 · Cited by 131 · see citing works
McCreery, D. B., Yuen, T. G. H., Agnew, W. F., & Bullara, L. A. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.634645 · Cited by 130 · see citing works
Long-term chronic recording from cortical neurons
Schmidt, E. M., Bak, M. J., & McIntosh, J. S. (1976) · Experimental Neurology · doi:10.1016/0014-4886(76)90220-x · Cited by 129 · see citing works
Ceramic-Based Multisite Electrode Arrays for Chronic Single-Neuron Recording
Moxon, K. A., Leiser, S., Gerhardt, G. A., Barbee, K. A., & Chapin, J. K. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.821037 · Cited by 129 · see citing works
Neural Prostheses for Restoration of Sensory and Motor Function
Chapin, J. K., & Moxon, K. A. (2000) · CRC Press eBooks · doi:10.1201/9781420039054 · Cited by 122 · see citing works
Boylan, K., Yang, C., Crook, J. E., Overstreet, K., Heckman, M. G., Wang, Y., Borchelt, D., & Shaw, G. (2009) · Journal of Neurochemistry · doi:10.1111/j.1471-4159.2009.06386.x · Cited by 121 · see citing works
Self-citationMultielectrode recordings: the next steps
Nicolelis, M. A. L., & Ribeiro, S. T. G. (2002) · Current Opinion in Neurobiology · doi:10.1016/s0959-4388(02)00374-4 · Cited by 119 · see citing works
Skousen, J. L., Merriam, S. M. E., Srivannavit, O., Perlin, G., Wise, K. D., & Tresco, P. A. (2011) · Progress in Brain Research · doi:10.1016/b978-0-444-53815-4.00009-1 · Cited by 111 · see citing works
Minocycline increases quality and longevity of chronic neural recordings
Rennaker, R. L., Miller, J., Tang, H., & Wilson, D. A. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/2/l01 · Cited by 109 · see citing works
Consumer perspectives on mobility: implications for neuroprosthesis design.
Brown-Triolo, D. L., Roach, M. J., Nelson, K. M., & Triolo, R. J. (2002) · PubMed · Cited by 108 · see citing works
Technique for collection of cerebrospinal fluid from the cisterna magna in rat
Pegg, C. C., He, C., Stroink, A. R., et al. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.12.002 · Cited by 106 · see citing works
Freire, M. A. M., Morya, E., Faber, J., dos Santos, J. R., Guimarães, J. S., Lemos, N. A. M., Sameshima, K., Pereira, A., Ribeiro, S. T. G., & Nicolelis, M. A. L. (2011) · PLoS ONE · doi:10.1371/journal.pone.0027554 · Cited by 97 · see citing works
Electrochemical deposition of conducting polymer coatings on magnesium surfaces in ionic liquid
Luo, X., & Cui, X. T. (2010) · Acta Biomaterialia · doi:10.1016/j.actbio.2010.09.006 · Cited by 93 · see citing works
Gomes‐Leal, W., Corkill, D. J., Freire, M. A. M., Diniz, C. W. P., & Perry, V. H. (2004) · Experimental Neurology · doi:10.1016/j.expneurol.2004.06.028 · Cited by 93 · see citing works
Tissue response to potential neuroprosthetic materials implanted subdurally
Yuen, T. G. H., Agnew, W. F., & Bullara, L. A. (1987) · Biomaterials · doi:10.1016/0142-9612(87)90103-7 · Cited by 92 · see citing works
Fendrick, S. E., Xue, Q., & Streit, W. J. (2007) · Journal of Neuroinflammation · doi:10.1186/1742-2094-4-9 · Cited by 88 · see citing works
Self-citationThe reaction of the cerebral cortex to chronically implanted plastic needles.
Stensaas, S. S., & Stensaas, L. J. (1976) · PubMed · Cited by 88 · see citing works
Berger, T. W., Chapin, J. K., Gerhardt, G. A., McFarland, D. J., Príncipe, J. C., Soussou, W., Taylor, D. M., & Tresco, P. A. (2008) · Springer eBooks · doi:10.1007/978-1-4020-8705-9 · Cited by 82 · see citing works
Cox, J. D., Curry, M. S., Skirboll, S. K., et al. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00205-8 · Cited by 77 · see citing works
Streit (2010) · Frontiers in Aging Neuroscience · doi:10.3389/fnagi.2010.00022 · Cited by 75 · see citing works
Stability of chronic multichannel neural recordings: Implications for a long-term neural interface
Williams, J. C., Rennaker, R. L., & Kipke, D. R. (1999) · Neurocomputing · doi:10.1016/s0925-2312(99)00106-x · Cited by 70 · see citing works
Kam, L. C. (1999) · Biomaterials · doi:10.1016/s0142-9612(99)00163-5 · Cited by 69 · see citing works
Bioimpedance modeling to monitor astrocytic response to chronically implanted electrodes
McConnell, G. C., Butera, R. J., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/055005 · Cited by 67 · see citing works
The ultrastructure of the sheath around chronically implanted electrodes in brain
Schultz, R. L., & Willey, T. J. (1976) · Journal of Neurocytology · doi:10.1007/bf01181577 · Cited by 63 · see citing works
Seeding neural progenitor cells on silicon-based neural probes: Laboratory investigation
Azemi, E., Gobbel, G. T., & Cui, X. T. (2010) · Journal of Neurosurgery · doi:10.3171/2010.1.jns09313 · Cited by 47 · see citing works
Histological evaluation of polyesterimide-insulated gold wires in brain
Yuen, T. G. H., & Agnew, W. F. (1995) · Biomaterials · doi:10.1016/0142-9612(95)93121-s · Cited by 46 · see citing works
Structural modifications in chronic microwire electrodes for cortical neuroprosthetics: a case study
Sanchez, J. C., Alba, N. A., Nishida, T., Batich, C. D., & Carney, P. R. (2006) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2006.875581 · Cited by 45 · see citing works
Self-citationDecoding Movement Intent From Human Premotor Cortex Neurons for Neural Prosthetic Applications
Ojakangas, C. L., Shaikhouni, A., Friehs, G. M., Caplan, A. H., Serruya, M. D., Saleh, M., Morris, D. S., & Donoghue, J. P. (2006) · Journal of Clinical Neurophysiology · doi:10.1097/01.wnp.0000233323.87127.14 · Cited by 44 · see citing works
Sadowsky, C. L., Volshteyn, O., Schultz, L., et al. (2002) · Disability and Rehabilitation · doi:10.1080/09638280110110640 · Cited by 43 · see citing works
Effects of Glial Cells on Electrode Impedance Recorded from Neural Prosthetic Devices In Vitro
Frampton, J. P., Hynd, M. R., Shuler, M. L., & Shain, W. G. (2010) · Annals of Biomedical Engineering · doi:10.1007/s10439-010-9911-y · Cited by 37 · see citing works
Motor control of every day motor tasks: Guidance for neurological rehabilitation
Hirschfeld, H. (2007) · Physiology & Behavior · doi:10.1016/j.physbeh.2007.05.018 · Cited by 25 · see citing works
Ludwig, K. A., Miriani, R. M., Langhals, N. B., Marzullo, T., & Kipke, D. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/4/049801 · Cited by 19 · see citing works
Tresco, P. A., & Gerhardt, G. A. (2008) · Springer eBooks · doi:10.1007/978-1-4020-8705-9_3 · Cited by 8 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Canales, A., Jia, X., Froriep, U. P., Koppes, R. A., Tringides, C. M., Selvidge, J., Lu, C., Hou, C., Wei, L., Fink, Y., et al. (2015) · Nature Biotechnology · doi:10.1038/nbt.3093 · Cited by 668 · see citing works
Brain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies
Kozai, T. D. Y., Jaquins-Gerstl, A. S., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn500256e · Cited by 622 · see citing works
Glial responses to implanted electrodes in the brain
Salatino, J. W., Ludwig, K. A., Kozai, T. D. Y., & Purcell, E. K. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0154-1 · Cited by 642 · see citing works
Next-generation probes, particles, and proteins for neural interfacing
Rivnay, J., Wang, H., Fenno, L., Deisseroth, K., & Malliaras, G. G. (2017) · Science Advances · doi:10.1126/sciadv.1601649 · Cited by 514 · see citing works
A Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389 · see citing works
Restoring sensorimotor function through intracortical interfaces: progress and looming challenges
Bensmaia, S. J., & Miller, L. E. (2014) · Nature Reviews Neuroscience · doi:10.1038/nrn3724 · Cited by 368 · see citing works
Progress towards biocompatible intracortical microelectrodes for neural interfacing applications
Jorfi, M., Skousen, J. L., Weder, C., & Capadona, J. R. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/011001 · Cited by 384 · see citing works
Implantable devices: issues and challenges
Bazaka, K., & Jacob, M. (2012) · Electronics · doi:10.3390/electronics2010001 · Cited by 358 · see citing works
Streit, W. J., Xue, Q.-S., Tischer, J., & Bechmann, I. (2014) · Acta Neuropathologica Communications · doi:10.1186/s40478-014-0142-6 · Cited by 317 · see citing works
Self-citationMechanically-compliant intracortical implants reduce the neuroinflammatory response
Nguyen, J. K., Park, D. J., Skousen, J. L., Hess-Dunning, A. E., Tyler, D. J., Rowan, S. J., Weder, C., & Capadona, J. R. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/5/056014 · Cited by 274 · see citing works
A review on mechanical considerations for chronically-implanted neural probes
Lecomte, A., Descamps, E., & Bergaud, C. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa8b4f · Cited by 205 · see citing works
Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants
Prasad, A., Xue, Q.-S., Dieme, R., Sankar, V., Mayrand, R. C., Nishida, T., Streit, W. J., & Sanchez, J. C. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00002 · Cited by 178 · see citing works
Self-citationBorton, D., Micera, S., Millán, J. D. R., & Courtine, G. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3005968 · Cited by 171 · see citing works
Barrese, J. C., Aceros, J., & Donoghue, J. P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/026003 · Cited by 161 · see citing works
Mesh electronics: a new paradigm for tissue-like brain probes
Hong, G., Yang, X., Zhou, T., & Lieber, C. M. (2018) · Current Opinion in Neurobiology · doi:10.1016/j.conb.2017.11.007 · Cited by 159 · see citing works
Model validation of untethered, ultrasonic neural dust motes for cortical recording
Seo, D., Carmena, J. M., Rabaey, J. M., Maharbiz, M. M., & Alon, E. (2015) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2014.07.025 · Cited by 158 · see citing works
Luan, L., Robinson, J. T., Aazhang, B., Chi, T., Yang, K., Li, X., Rathore, H., Singer, A., Yellapantula, S., Fan, Y., et al. (2020) · Neuron · doi:10.1016/j.neuron.2020.10.011 · Cited by 150 · see citing works
Soft and MRI Compatible Neural Electrodes from Carbon Nanotube Fibers
Lu, L., Fu, X., Liew, Y., Zhang, Y., Zhao, S., Xu, Z., Zhao, J., Li, D., Li, Q., Stanley, G. B., et al. (2019) · Nano Letters · doi:10.1021/acs.nanolett.8b04456 · Cited by 146 · see citing works
Clinical applications of penetrating neural interfaces and Utah Electrode Array technologies
Normann, R. A., & Fernandez, E. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/6/061003 · Cited by 135 · see citing works
Kolarcik, C. L., Catt, K., Rost, E., Albrecht, I. N., Bourbeau, D., Du, Z., Kozai, T. D. Y., Luo, X., Weber, D. J., & Tracy Cui, X. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016008 · Cited by 126 · see citing works
Advances in Neuroprosthetic Learning and Control
Carmena, J. M. (2013) · PLoS Biology · doi:10.1371/journal.pbio.1001561 · Cited by 117 · see citing works
Recent advances in neural electrode–tissue interfaces
Woeppel, K., Yang, Q., & Cui, X. T. (2017) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2017.09.003 · Cited by 99 · see citing works
Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88 · see citing works
“The fundamental problem in cortical interfaces is longevity (Warren et al., 2016): The neurons (or neuronal tissues) at the implanted surface and the electrodes (implanted interface) degrade over time (McCreery et al., 2010; Prasad et al., 2012; Kane et al., 2013; Chen et al., 2014).”
Restoring tactile and proprioceptive sensation through a brain interface
Tabot, G. A., Kim, S. S., Winberry, J. E., & Bensmaia, S. J. (2015) · Neurobiology of Disease · doi:10.1016/j.nbd.2014.08.029 · Cited by 83 · see citing works
Regenerative Electrode Interfaces for Neural Prostheses.
Thompson, C. H., Zoratti, M. J., Langhals, N. B., & Purcell, E. K. (2016) · Tissue Engineering Part B: Reviews · doi:10.1089/ten.teb.2015.0279 · Cited by 67 · see citing works
Long-term recording reliability of liquid crystal polymer <i>µ</i> ECoG arrays
Woods, V., Trumpis, M., Bent, B., Palopoli-Trojani, K., Chiang, C.-H., Wang, C., Yu, C., Insanally, M. N., Froemke, R. C., & Viventi, J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aae39d · Cited by 63 · see citing works
A high-density carbon fiber neural recording array technology
Massey, T. L., Santacruz, S. R., Hou, J. F., Pister, K. S. J., Carmena, J. M., & Maharbiz, M. M. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/aae8d9 · Cited by 58 · see citing works
Fernández, E., Alfaro, A., & González-López, P. (2020) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.00681 · Cited by 35 · see citing works
Comparative analysis of energy transfer mechanisms for neural implants
Miziev, S., Pawlak, W. A., & Howard, N. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2023.1320441 · Cited by 21 · see citing works
Biotolerability of Intracortical Microelectrodes
Fernández, E., & Botella, P. (2018) · Advanced Biosystems · doi:10.1002/adbi.201700115 · Cited by 11 · see citing works
- Authors
- Prasad, A., Xue, Q.-S., Sankar, V., Nishida, T., Shaw, G., Streit, W. J., & Sanchez, J. C.
- Affiliations
- University of Miami (US)
- University of Florida (US)
- Corresponding author
- Prasad, Abhishek
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2012-09-25
- Volume
- 9
- Issue
- 5
- Pages
- 056015
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2155781262
- Cited by
- 298
- References
- 108
- Citations by year
- 2026: 9 · 2025: 15 · 2024: 18 · 2023: 17 · 2022: 15 · 2021: 25 · 2020: 30 · 2019: 21 · 2018: 27 · 2017: 28 · 2016: 21 · 2015: 20 · 2014: 36 · 2013: 14 · 2012: 1
- FWCI
- 9.36
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Defense Advanced Research Projects Agency (N66001)
- Advanced Research Projects Agency
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Cognitive Neuroscience › Undergraduate Neuroscience Education and Research
- Keywords
- electrode failure mechanisms, neural interfaces, microglial activation, neuroinflammation, blood-brain barrier disruption, ferritin, oxidative stress, Fenton reaction, electrode impedance, neuroprosthetics, rat model
- MeSH terms
- Animals, Cerebral Cortex, Electrodes, Implanted, Equipment Failure, Male, Microelectrodes, Tungsten, Electric Impedance, Rats, Sprague-Dawley, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review