A Suprachoroidal Electrical Retinal Stimulator Design for Long‐Term Animal Experiments and In Vivo Assessment of Its Feasibility and Biocompatibility in Rabbits
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
This article reports on a retinal stimulation system for long‐term use in animal electrical stimulation experiments. The presented system consisted of an implantable stimulator which provided continuous electrical stimulation, and an external component which provided preset stimulation patterns and power to the implanted stimulator via a paired radio frequency (RF) coil. A rechargeable internal battery and a parameter memory component were introduced to the implanted retinal stimulator. As a result, the external component was not necessary during the stimulation mode. The inductive coil pair was used to pass the parameter data and to recharge the battery. A switch circuit was used to separate the stimulation mode from the battery recharging mode. The implantable stimulator was implemented with IC chips and the electronics, except for the stimulation electrodes, were hermetically packaged in a biocompatible metal case. A polyimide‐based gold electrode array was used. Surgical implantation into rabbits was performed to verify the functionality and safety of this newly designed system. The electrodes were implanted in the suprachoroidal space. Evoked cortical potentials were recorded during electrical stimulation of the retina. Long‐term follow‐up using OCT showed no chorioretinal abnormality after implantation of the electrodes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Visual perception in a blind subject with a chronic microelectronic retinal prosthesis
Humayun, M. S., Weiland, J. D., Fujii, G. Y., Greenberg, R., Williamson, R., Little, J., Mech, B., Cimmarusti, V., Van Boemel, G., Dagnelie, G., et al. (2003) · Vision Research · doi:10.1016/s0042-6989(03)00457-7 · Cited by 856 · cited 3×
“Although the electrical stimulation of the retina in preliminary clinical studies showed encouraging results such as the patient’s perception of a small spot of light or basic shapes according to the stimulation pattern [4], there is much to be investigated and revised regarding retinal prostheses.”
BackgroundMethodInfluentialRizzo, J. F., Wyatt, J., Loewenstein, J., Kelly, S., & Shire, D. (2003) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.02-0817 · Cited by 478 · cited 1×
“Hindawi Publishing Corporation Journal of Biomedicine and Biotechnology Volume 2008, Article ID 547428, 10 pages doi:10.1155/2008/547428 Methodology Report”
BackgroundInfluential
Cited by
Subretinal electronic chips allow blind patients to read letters and combine them to words
Zrenner, E., Bartz-Schmidt, K. U., Benav, H., Besch, D., Bruckmann, A., Gabel, V.-P., Gekeler, F., Greppmaier, U., Harscher, A., Kibbel, S., et al. (2011) · Proceedings of the Royal Society B: Biological Sciences · doi:10.1098/rspb.2010.1747 · Cited by 819
Bhowmick, A., & Hazarika, S. M. (2017) · Journal on Multimodal User Interfaces · doi:10.1007/s12193-016-0235-6 · Cited by 271
Fujikado, T., Kamei, M., Sakaguchi, H., Kanda, H., Morimoto, T., Ikuno, Y., Nishida, K., Kishima, H., Maruo, T., Konoma, K., et al. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-6836 · Cited by 216
“Based on the safety of this approach, the STS system has been adopted by several other groups.(22,23) In an acute clinical trial with STS prosthesis, localized phosphenes were perceived with safety currents ( 1 mA) in two patients with advanced RP.”
MethodVisual prostheses for the blind
Shepherd, R. K., Shivdasani, M. N., Nayagam, D. A. X., Williams, C. E., & Blamey, P. J. (2013) · Trends in Biotechnology · doi:10.1016/j.tibtech.2013.07.001 · Cited by 178
Wilke, R., Gabel, V.-P., Sachs, H., Bartz Schmidt, K.-U., Gekeler, F., Besch, D., Szurman, P., Stett, A., Wilhelm, B., Peters, T., et al. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-6946 · Cited by 155
Focal activation of the feline retina via a suprachoroidal electrode array
Wong, Y. T., Chen, S. C., Seo, J. M., Morley, J. W., Lovell, N. H., & Suaning, G. J. (2009) · Vision Research · doi:10.1016/j.visres.2009.02.018 · Cited by 109
“To-date, studies involving suprachoroidal electrode placement have all utilized monopolar stimulation in rabbits and rats with return electrodes placed in the vitreous, or on the sclera (Kanda et al., 2004; Yamauchi et al., 2005; Zhou et al., 2008).”
Self-citationCurrent steering in retinal stimulation via a quasimonopolar stimulation paradigm
Matteucci, P. B., Chen, S. C., Tsai, D., Dodds, C. W. D., Dokos, S., Morley, J. W., Lovell, N. H., & Suaning, G. J. (2013) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.13-11653 · Cited by 79
“64 lC cm (2)) using a 1 ms phase time and a 300 lm electrode.(23) Monopolar and hexapolar (128.”
BackgroundMethodVillalobos, J., Nayagam, D. A. X., Allen, P. J., McKelvie, P., Luu, C. D., Ayton, L. N., Freemantle, A. L., McPhedran, M., Basa, M., McGowan, C. C., et al. (2013) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.12-10843 · Cited by 63
Yan, Y., Lu, Y., Li, M., Ma, Z., Cao, P., Chen, Y., Sun, X., Chai, X., Ren, Q., & Li, L. (2016) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.15-17543 · Cited by 20
Embracing the irregular: A patient-specific image processing strategy for visual prostheses
Kiral-Kornek, F. I., Savage, C. O., O'Sullivan-Greene, E., Burkitt, A. N., & Grayden, D. B. (2013) · 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610312 · Cited by 9
Metadata
- Authors
- Zhou, J. A., Woo, S. J., Park, S. I., Kim, E. T., Seo, J. M., Chung, H., & Kim, S. J.
- Venue
- BioMed Research International
- Publisher
- Wiley
- Published
- 2008-01
- Volume
- 2008
- Issue
- 1
- Pages
- 547428-547428
- ISSN
- 1110-7243 (ISSN-L)
- 2314-6133
- 2314-6141
- OpenAlex
- W1982902008
- Cited by
- 73
- References
- 2
- FWCI
- 3.88
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.