A Two Centre Observational Study of Simultaneous Pulse Oximetry and Arterial Oxygen Saturation Recordings in Intensive Care Unit Patients
Open accessAbstract
The influence of variables that might affect the accuracy of pulse oximetry (SpO 2 ) recordings in critically ill patients is not well established. We sought to describe the relationship between paired SpO 2 /SaO 2 (oxygen saturation via arterial blood gas analysis) in adult intensive care unit (ICU) patients and to describe the diagnostic performance of SpO 2 in detecting low SaO 2 and PaO 2 . A paired SpO 2 /SaO 2 measurement was obtained from 404 adults in ICU. Measurements were used to calculate bias, precision, and limits of agreement. Associations between bias and variables including vasopressor and inotrope use, capillary refill time, hand temperature, pulse pressure, body temperature, oximeter model, and skin colour were estimated. There was no overall statistically significant bias in paired SpO 2 /SaO 2 measurements; observed limits of agreement were +/-4.4%. However, body temperature, oximeter model, and skin colour, were statistically significantly associated with the degree of bias. SpO 2 <89% had a sensitivity of 3/7 (42.9%; 95% confidence intervals, CI, 9.9% to 81.6%) and a specificity of 344/384 (89.6%; 95% CI 86.1% to 92.5%) for detecting SaO 2 <89%. The absence of statistically significant bias in paired SpO 2 /SaO 2 in adult ICU patients provides support for the use of pulse oximetry to titrate oxygen therapy. However, SpO 2 recordings alone should be used cautiously when SaO 2 recordings of 4.4% higher or lower than the observed SpO 2 would be of concern. A range of variables relevant to the critically ill had little or no effect on bias.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
SpO2 measurements provided by conventional pulse oximeters are susceptible to errors.
“SpO2 measurements provided by conventional pulse oximeters are susceptible to errors”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [143]
Objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [76]
A small but statistically significant difference between SaO2 and SpO2 in light and dark skin phototypes was found in 404 ICU patients classified by FSP.
“A small but statistically significant difference between SaO2 and SpO2 in light and dark skin phototypes [76] was found.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [76]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Knaus, W. A., Draper, E. A., Wagner, D. P., & Zimmerman, J. E. (1985) · Critical Care Medicine · doi:10.1097/00003246-198510000-00009 · Cited by 15,333 · see citing works
APACHE II-A Severity of Disease Classification System
Knaus, W. A., Draper, E. A., Wagner, D. P., & Zimmerman, J. E. (1986) · Critical Care Medicine · doi:10.1097/00003246-198608000-00028 · Cited by 13,379 · see citing works
Accuracy of pulse oximetry in the intensive care unit
de Louw, A. V., Cracco, C., Cerf, C., HARF, A., Duvaldestin, P., Lemaire, F., & Brochard, L. J. (2001) · Intensive Care Medicine · doi:10.1007/s001340101064 · Cited by 307 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Metrics of Arterial Hyperoxia and Associated Outcomes in Critical Care*
Helmerhorst, H. J. F., Arts, D. L., Schultz, M. J., van der Voort, P. H. J., Abu–Hanna, A., de Jonge, E., & van Westerloo, D. J. (2016) · Critical Care Medicine · doi:10.1097/ccm.0000000000002084 · Cited by 205 · see citing works
Neff, T. A. (1988) · CHEST Journal · doi:10.1378/chest.94.2.227a · Cited by 150 · see citing works
Wilson, B., Cowan, H. J., Lord, J. A., Zuege, D. J., & Zygun, D. A. (2010) · BMC Emergency Medicine · doi:10.1186/1471-227x-10-9 · Cited by 146 · see citing works
Perkins, G. D., McAuley, D. F., Giles, S., Routledge, H., & Gao, F. (2003) · Critical Care · doi:10.1186/cc2339 · Cited by 136 · see citing works
Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients
Hinkelbein, J., Genzwuerker, H. V., Sogl, R., & Fiedler, F. (2006) · Resuscitation · doi:10.1016/j.resuscitation.2006.06.024 · Cited by 100 · see citing works
Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit
Séguin, P., Rouzo, A. L., Tanguy, M., Guillou, Y. M., Feuillu, A., & Mallédant, Y. (2000) · Critical Care Medicine · doi:10.1097/00003246-200003000-00017 · Cited by 72 · see citing works
Artificial acrylic finger nails may alter pulse oximetry measurement
Hinkelbein, J., Koehler, H., Genzwuerker, H. V., & Fiedler, F. (2007) · Resuscitation · doi:10.1016/j.resuscitation.2006.11.018 · Cited by 69 · see citing works
Accuracy of pulse oximeters: the European multi-center trial.
Wouters, P. F., Gehring, H., Meyfroidt, G., et al. (2002) · PubMed · Cited by 47 · see citing works
Singh, A. (2017) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2017/24640.9961 · Cited by 45 · see citing works
Hyperoxia: At what level of SpO 2 is a patient safe? A study in mechanically ventilated ICU patients
Durlinger, E. M. J., de Man, A., Smit, B., de Grooth, H., Girbes, A. R. J., Straaten, H. M. O., & Smulders, Y. M. (2017) · Journal of Critical Care · doi:10.1016/j.jcrc.2017.02.031 · Cited by 44 · see citing works
Young, P. J., Beasley, R. W., Capellier, G., et al. (2015) · PubMed · Cited by 30 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K. (2022) · Anaesthesia · doi:10.1111/anae.15581 · Cited by 45 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
- Authors
- Ebmeier, S. J., Barker, M., Bacon, M., Beasley, R. C., Bellomo, R., Chong, C. K., Eastwood, G. M., Gilchrist, J., Kagaya, H., Pilcher, J., Reddy, S. K., Ridgeon, E., Sarma, N., Spragas, S., Tanaka, A., Tweedie, M., Weatherall, M., & Young, P. J.
- Affiliations
- Medical Research Institute of New Zealand (NZ)
- Austin Hospital (AU)
- University of Otago (NZ)
- Wellington Hospital (NZ)
- Corresponding author
- Young, P. J.
- ORCID
- Venue
- Anaesthesia and Intensive Care
- Publisher
- SAGE Publications
- Published
- 2018-05
- Volume
- 46
- Issue
- 3
- Pages
- 297-303
- ISSN
- 0310-057X (ISSN-L)
- 1448-0271
- OpenAlex
- W2889446176
- Cited by
- 92
- References
- 16
- Citations by year
- 2026: 6 · 2025: 8 · 2024: 16 · 2023: 7 · 2022: 38 · 2021: 10 · 2020: 2 · 2019: 4 · 2018: 1
- FWCI
- 1.70
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Anaesthesia and Intensive Care (publishedVersion)
- PubMed (publishedVersion)
- Own your potential (DEAKIN) (submittedVersion)
- Figshare (submittedVersion, cc-by)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Respiratory Support and Mechanisms
- Keywords
- pulse oximetry, arterial oxygen saturation, measurement bias, limits of agreement, critically ill patients, intensive care unit, body temperature, skin colour, vasopressors, inotropic agents, capillary refill time
- MeSH terms
- Aged, Female, Humans, Intensive Care Units, Male, Middle Aged, New Zealand, Oximetry, Oxygen, Prospective Studies, Pulmonary Gas Exchange
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review