The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: A single-centre, retrospective analysis
Open accessAbstract
Summary Pulse oximetry is used widely to titrate oxygen therapy and for triage in patients who are critically ill. However, there are concerns regarding the accuracy of pulse oximetry in patients with COVID‐19 pneumonitis and in patients who have a greater degree of skin pigmentation. We aimed to determine the impact of patient ethnicity on the accuracy of peripheral pulse oximetry in patients who were critically ill with COVID‐19 pneumonitis by conducting a retrospective observational study comparing paired measurements of arterial oxygen saturation measured by co‐oximetry on arterial blood gas analysis (SaO 2 ) and the corresponding peripheral oxygenation saturation measured by pulse oximetry (S p O 2 ). Bias was calculated as the mean difference between SaO 2 and S p O 2 measurements and limits of agreement were calculated as bias ±1.96 SD. Data from 194 patients (135 White ethnic origin, 34 Asian ethnic origin, 19 Black ethnic origin and 6 other ethnic origin) were analysed consisting of 6216 paired SaO 2 and S p O 2 measurements. Bias (limits of agreement) between SaO 2 and S p O 2 measurements was 0.05% (−2.21–2.30). Patient ethnicity did not alter this to a clinically significant degree: 0.28% (1.79–2.35), −0.33% (−2.47–2.35) and −0.75% (−3.47–1.97) for patients of White, Asian and Black ethnic origin, respectively. In patients with COVID‐19 pneumonitis, S p O 2 measurements showed a level of agreement with SaO 2 values that was in line with previous work, and this was not affected by patient ethnicity.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Recent studies have shown a positive bias in oxygen saturation measurements in patients with darkly pigmented skin, particularly under low saturation conditions.
“Recent studies have shown a positive bias in oxygen saturation measurements in patients with darkly pigmented skin, particularly under low saturation conditions.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [17]
More recent studies have revealed that dark skin pigmentation influences the accuracy and performance of pulse oximeter devices, resulting in overestimations, with increased incidence in the risk for occult hypoxemia (SO2<88% despite normal SpO2>92%).
“More recent studies have also revealed that dark skin pigmentation influences the accuracy and performance of pulse oximeters devices, resulting in overestimations [17,49,50,51,52,53] with increased incidence in the risk for occult hypoxemia (SO2<88% despite normal SpO2>92%) [54,55].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [17]
The main reason for the upsurge of publications in 2021 may be attributed to the increase of pulse oximetry for triage in patients because of the COVID-19 pandemic and recent concern about accuracy across all skin types.
“The main reason for this surge may be attributed to the increase of this technique for triage in patients because of the current COVID-19 pandemic [12,13,14,15,16,17], and also because of the recent concern about the accuracy of these devices across all skin types [20,57,58,59,60,61].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [17]
Ancestry descriptors were used in 11 studies (29.73%).
“On the other hand, ancestry descriptors were used in 11 studies (29.73%) [17,21,23,24,41,51,53,61,75,82,93]”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [17]
A study in patients with acute COVID-19 pneumonia found a small bias between arterial blood gas analysis SaO2 and SpO2 measurements of 0.28%, −0.33% and −0.75% for 194 patients of "White" (135), "Asian" (34) and "Black" (19) ethnic origin, respectively.
“Besides, another study was performed in patients with acute COVID-19 pneumonia to find out a relationship between patient ancestry and the accuracy of pulse oximetry found a small bias between arterial blood gas analysis SaO2 and SpO2 measurements of 0.28%, −0.33% and −0.75% for 194 patients of “White” (135), “Asian” (34) and “Black” (19) ethnic origin, respectively, [17].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [17]
Cites
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · see citing works
Racial and Ethnic Health Disparities Related to COVID-19
Lopez, L., Hart, L. H., & Katz, M. H. (2021) · JAMA · doi:10.1001/jama.2020.26443 · Cited by 712 · see citing works
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · see citing works
Accuracy of pulse oximetry in the intensive care unit
de Louw, A. V., Cracco, C., Cerf, C., HARF, A., Duvaldestin, P., Lemaire, F., & Brochard, L. J. (2001) · Intensive Care Medicine · doi:10.1007/s001340101064 · Cited by 307 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Trends in Intensive Care for Patients with COVID-19 in England, Wales, and Northern Ireland
Doidge, J., Gould, D. W., Ferrando-Vivas, P., Mouncey, P. R., Thomas, K., Shankar‐Hari, M., Harrison, D. A., & Rowan, K. (2020) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.202008-3212oc · Cited by 169 · see citing works
Wilson, B., Cowan, H. J., Lord, J. A., Zuege, D. J., & Zygun, D. A. (2010) · BMC Emergency Medicine · doi:10.1186/1471-227x-10-9 · Cited by 146 · see citing works
Perkins, G. D., McAuley, D. F., Giles, S., Routledge, H., & Gao, F. (2003) · Critical Care · doi:10.1186/cc2339 · Cited by 136 · see citing works
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102 · see citing works
Ebmeier, S. J., Barker, M., Bacon, M., Beasley, R. C., Bellomo, R., Chong, C. K., Eastwood, G. M., Gilchrist, J., Kagaya, H., Pilcher, J., et al. (2018) · Anaesthesia and Intensive Care · doi:10.1177/0310057x1804600307 · Cited by 92 · see citing works
The appropriateness of Bland-Altman’s approximate confidence intervals for limits of agreement
Shieh, G. (2018) · BMC Medical Research Methodology · doi:10.1186/s12874-018-0505-y · Cited by 44 · see citing works
Philip, K. E. J., Bennett, B., Fuller, S., Lonergan, B., McFadyen, C., Burns, J., Tidswell, R., & Vlachou, A. (2020) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2020-000778 · Cited by 35 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Fawzy, A., Wu, T. D., Wang, K., Robinson, M. L., Farha, J., Bradke, A., Golden, S. H., Xu, Y., & Garibaldi, B. T. (2022) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2022.1906 · Cited by 326 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
Henry, N. R., Hanson, A. C., Schulte, P. J., Warner, N. S., Manento, M. N., Weister, T. J., & Warner, M. A. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005394 · Cited by 133 · see citing works
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119 · see citing works
Racial Disparity in Oxygen Saturation Measurements by Pulse Oximetry: Evidence and Implications
Jamali, H., Castillo, L. T., Morgan, C. C., Coult, J., Muhammad, J. L., Osobamiro, O. O., Parsons, E. C., & Adamson, R. (2022) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202203-270cme · Cited by 83 · see citing works
Farooq, A. S., & Zhang, P. (2022) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2022.113715 · Cited by 79 · see citing works
Martín, D., Johns, C., Sorrell, L., Healy, E., Phull, M., Olusanya, O., Peters, M. J., & Fabes, J. (2024) · British Journal of Anaesthesia · doi:10.1016/j.bja.2024.01.023 · Cited by 80 · see citing works
Sudat, S. E. K., Wesson, P., Rhoads, K. F., Brown, S., Aboelata, N., Pressman, A. R., Mani, A., & Azar, K. M. J. (2022) · American Journal of Epidemiology · doi:10.1093/aje/kwac164 · Cited by 58 · see citing works
Honarmand, K., Wax, R. S., Penoyer, D., Lighthall, G., Danesh, V. C., Rochwerg, B. N., Cheatham, M. L., Davis, D. P., DeVita, M. A., Downar, J., et al. (2024) · Critical Care Medicine · doi:10.1097/ccm.0000000000006072 · Cited by 57 · see citing works
Fawzy, A., Wu, T. D., Wang, K., Sands, K. E., Fisher, A. M., Arnold Egloff, S. A., DellaVolpe, J. D., Iwashyna, T. J., Xu, Y., & Garibaldi, B. T. (2023) · JAMA Network Open · doi:10.1001/jamanetworkopen.2023.30856 · Cited by 50 · see citing works
Melanometry for objective evaluation of skin pigmentation in pulse oximetry studies
Vasudevan, S., Vogt, W. C., Weininger, S., & Pfefer, T. J. (2024) · Communications Medicine · doi:10.1038/s43856-024-00550-7 · Cited by 46 · see citing works
Marinari, S., Volpe, P., Simoni, M., Aventaggiato, M., Benedetto, F. D., Nardini, S. G., Sanguinetti, C. M., & Palange, P. (2022) · Sensors · doi:10.3390/s22135031 · Cited by 38 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialRacial effects on Masimo pulse oximetry: a laboratory study
Barker, S. J., & Wilson, W. C. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00927-w · Cited by 36 · see citing works
Skin Pigmentation and Pulse Oximeter Accuracy in the Intensive Care Unit: A Pilot Prospective Study
Fawzy, A., Ali, H., Dziedzic, P. H., Potu, N., Calvillo, E., Golden, S. H., Iwashyna, T. J., Suárez, J. I., Hager, D. N., & Garibaldi, B. T. (2024) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.202401-0036le · Cited by 28 · see citing works
Cabanas, A. M., Martín-Escudero, P., & Shelley, K. H. (2023) · British Journal of Anaesthesia · doi:10.1016/j.bja.2023.07.005 · Cited by 28 · see citing works
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmüller, K. (2022) · Anaesthesia · doi:10.1111/anae.15656 · Cited by 26 · see citing works
Self-citationKyriacou, P. A., Charlton, P. H., Al-Halawani, R., & Shelley, K. H. (2022) · British Journal of Anaesthesia · doi:10.1016/j.bja.2022.03.011 · Cited by 27 · see citing works
Utility of Skin Tone on Pulse Oximetry in Critically Ill Patients: A Prospective Cohort Study
Hao, S., Dempsey, K., de Matos, J. C. R. G., Cox, C. E., Rotemberg, V. M., Gichoya, J. W., Kibbe, W. A., Hong, C., & Wong, A.-K. I. (2024) · Critical Care Explorations · doi:10.1097/cce.0000000000001133 · Cited by 18 · see citing works
Chew, R., Zhang, M., Chandna, A., & Lubell, Y. (2021) · BMJ Global Health · doi:10.1136/bmjgh-2021-007282 · Cited by 15 · see citing works
Nachman, D., Eisenkraft, A., Goldstein, N., Ishay, A. B., Fons, M., Merin, R., & Gepner, Y. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.911544 · Cited by 16 · see citing works
The Big Consequences of Small Discrepancies: Why Racial Differences in Pulse Oximetry Errors Matter
Holder, A. L., & Wong, A.-K. I. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005447 · Cited by 13 · see citing works
Norton, H. (2021) · Anaesthesia · doi:10.1111/anae.15620 · Cited by 12 · see citing works
Bradley, M., Barrett, S., McKelvey, T., Carpenter, J., & Dean, D. M. D. (2025) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2025.100876 · Cited by 7 · see citing works
Monte Carlo Simulation of the Effect of Human Skin Melanin in Light-Tissue Interactions
Al-Halawani, R., Chatterjee, S., & Kyriacou, P. A. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871350 · Cited by 7 · see citing works
James, A., Petit, M., Biancale, F., Bouglé, A., Degos, V., Monsel, A., Vieillard‐Baron, A., & Constantin, J. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00959-2 · Cited by 7 · see citing works
Racial discrepancies in oximetry: where do we stand?
Knight, M., Subbe, C. P., & Inada-Kim, M. (2021) · Anaesthesia · doi:10.1111/anae.15635 · Cited by 7 · see citing works
Wang, Q., Carson, A. L., & Zheng, S. Y. (2024) · Production and Operations Management · doi:10.1177/10591478241270115 · Cited by 3 · see citing works
Racial Disparities in Pulse Oximetry, in COVID-19 and ICU Settings
Reep, C. A. T., Fleuren, L. M., Heunks, L. M. A., & Wils, E. (2024) · Critical Care Explorations · doi:10.1097/cce.0000000000001132 · Cited by 3 · see citing works
Marlar, A. I., Knabe, B. K., Taghikhan, Y., Applegate, R. L., & Fleming, N. (2024) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-024-01211-9 · Cited by 3 · see citing works
Chen, Q.-C., Shen, J., Huang, Y.-L., Kong, R., Xie, Y.-M., & Wang, S. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e28336 · Cited by 1 · see citing works
Limitations in the Use of Pulse Oximetry in Patients of Color
Pal, A. D. (2023) · Journal of PeriAnesthesia Nursing · doi:10.1016/j.jopan.2023.08.018 · Cited by 2 · see citing works
Survey in emergency medicine of skin pigmentation bias in pulse oximeters
Maher, P. J., Goel, N. N., Horowitz, C. R., Rea, M. S., Figueiro, M. G., & Richardson, L. D. (2024) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2024.03.002 · Cited by 2 · see citing works
Discordance Between Invasive and NonInvasive Oxygen Saturation in Critically Ill COVID-19 Patients
Fox, T. H., Mazalewski, W., Tran, H. V., Lindsey, T., Krishnan, A., Kantrow, S. P., Happel, K. I., Janz, D. R., deBoisblanc, B. P., & Lammi, M. R. (2023) · Journal of Intensive Care Medicine · doi:10.1177/08850666231186947 · Cited by 2 · see citing works
Evrard, C., Attaoui, A. E., Pistea, C., Enache, I., Marriott, M., Mayaud, L., Charloux, A., & Gény, B. (2024) · Sensors · doi:10.3390/s25010127 · Cited by 1 · see citing works
Malíčková, L., Daneshjo, N., Daňo, F., & Ševčíková, R. (2023) · TEM Journal · doi:10.18421/tem122-43 · Cited by 1 · see citing works
How to write a Science Letter for Anaesthesia
Bailey, C. R., & El‐Boghdadly, K. (2022) · Anaesthesia · doi:10.1111/anae.15830 · Cited by 1 · see citing works
The potential for unnecessary medical interventions due to inaccurate pulse oximetry measurements
Jacobs, J. W., & Abels, E. A. (2022) · Heart & Lung · doi:10.1016/j.hrtlng.2022.09.016 · Cited by 1 · see citing works
Saman, J. A., Rahmatinejad, M., Roodabr, G., Amini-Saman, Y., Kazazi, S., Evani, N., & Rahmani, S. (2023) · Journal of Clinical Research in Paramedical Sciences · doi:10.5812/jcrps-135832
Nelson, B. W., Singh, S., Bennett, M., Chen, C., Shin, S., & Ghanbari, H. (2024) · Research Square · doi:10.21203/rs.3.rs-3882498/v1
Pulse Oximetry: A Vital Tool with Important Functional Limitations for Patients with Dark Skin Color
Harris, D. E. (2024) · Universal Library of Clinical Nursing · doi:10.70315/uloap.ulcnu.2024.0101005
Accuracy of fingertip pulse oximeters: a device accuracy studyPreprint
Fleming, S., Titterington, M., Ashdown, H. F., & Rahman, N. M. (2025) · medRxiv · doi:10.64898/2025.12.05.25341372
Villavicencio, R. E. A., Panata, S. A. T., & Benavides, J. A. C. (2026) · Revista Científica Multidisciplinar G-nerando · doi:10.66473/rcmg.v7i2.1274
- Authors
- Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K.
- Affiliations
- Sheffield Teaching Hospitals NHS Foundation Trust (GB)
- University of Nottingham (GB)
- Department of Critical Care Sheffield Teaching Hospitals NHS Foundation Trust Sheffield UK
- Corresponding author
- Wiles, M. D.
- Venue
- Anaesthesia
- Publisher
- Wiley
- Published
- 2021-09-20
- Volume
- 77
- Issue
- 2
- Pages
- 143-152
- ISSN
- 0003-2409 (ISSN-L)
- 1365-2044
- PMC
- PMC8653100
- OpenAlex
- W3200483630
- Cited by
- 45
- References
- 13
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 14 · 2023: 8 · 2022: 17 · 2021: 3
- FWCI
- 3.19
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Anaesthesia (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Infectious Diseases › COVID-19 Clinical Research Studies
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- pulse oximetry accuracy, COVID-19 pneumonitis, patient ethnicity, skin pigmentation, arterial oxygen saturation, co-oximetry, arterial blood gases, critically ill patients
- MeSH terms
- COVID-19, SARS-CoV-2, Oxygen Saturation, Critical Care, Ethnicity, Female, Humans, Male, Middle Aged, Oximetry, Oxygen Inhalation Therapy, Retrospective Studies, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review