Lot-to-Lot Variability of Test Strips and Accuracy Assessment of Systems for Self-Monitoring of Blood Glucose according to ISO 15197
Open accessThe DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
Background: Accurate and reliable blood glucose (BG) measurements require that different test strip lots of the same BG monitoring system provide comparable measurement results. Only a small number of studies addressing this question have been published. Methods: In this study, four test strip lots for each of five different BG systems [Accu-Chek® Aviva (system A), FreeStyle Lite® (system B), GlucoCheck XL (system C), Pura™/mylife™ Pura (system D), and OneTouch® Verio ™ Pro (system E)] were evaluated with procedures according to DIN EN ISO 15197:2003. The BG system measurement results were compared with the manufacturer's measurement procedure (glucose oxidase or hexokinase method). Relative bias according to Bland and Altman and system accuracy according to ISO 15197 were analyzed. A BG system consists of the BG meter itself and the test strips. Results: The maximum lot-to-lot difference between any two of the four evaluated test strip lots per BG system was 1.0% for system E, 2.1% for system A, 3.1% for system C, 6.9% for system B, and 13.0% for system D. Only two systems (systems A and B) fulfill the criteria of DIN EN ISO 15197:2003 with each test strip lot. Conclusions: Considerable lot-to-lot variability between test strip lots of the same BG system was found. These variations add to other sources of inaccuracy with the specific BG system. Manufacturers should regularly and effectively check the accuracy of their BG meters and test strips even between different test strip lots to minimize risk of false treatment decisions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Global guideline for type 2 diabetes: recommendations for standard, comprehensive, and minimal care
IDF Clinical Guidelines Task Force (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2006.01918.x · Cited by 433
Polonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Petersen, B., Schweitzer, M., & Wagner, R. S. (2011) · Diabetes Care · doi:10.2337/dc10-1732 · Cited by 429
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Martin, S., Schneider, B., Heinemann, L., Lodwig, V., Kurth, H.-J., Kolb, H., Scherbaum, W. A., & for the ROSSO Study Group (2006) · Diabetologia · doi:10.1007/s00125-005-0083-5 · Cited by 284
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
Self-citationSystem Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195
Self-citationSkeie, S., Thue, G., Nerhus, K., & Sandberg, S. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.7.994 · Cited by 131
Self-Monitoring of Blood Glucose: The Use of the First or the Second Drop of Blood
Hortensius, J., Slingerland, R. J., Kleefstra, N., Logtenberg, S. J. J., Groenier, K. H., Houweling, S. T., & Bilo, H. J. G. (2011) · Diabetes Care · doi:10.2337/dc10-1694 · Cited by 76
Research works printed but not linked: 15
- 10.1056/NEJM199309303291401 · DOI · Paper page
- 10.1016/j.amjmed.2005.07.053 · DOI
- 10.4158/EP.13.S1.1 · DOI · Paper page
- 10.1016/j.amjmed.2005.07.055 · DOI
- 10.1093/clinchem/44.9.2056 · doi:10.1093/clinchem/44.9.2056
- 10.1016/j.cca.2005.07.027 · DOI
- 10.1089/dia.2008.0034 · DOI · Paper page
- 10.1373/clinchem.2005.049080 · DOI · Paper page
- 10.1016/S0009-9120(96)00092-6 · DOI
- 10.1016/0009-9120(95)02010-1 · DOI
- 10.1136/jcp.2003.013417 · DOI · Paper page
- 10.1089/dia.2010.0255 · DOI
- 10.2337/dc10-1705 · DOI · Paper page
- 10.1089/dia.2010.0223 · DOI · Paper page
- Essack Y, 2009, JEMDSA, 14, 102
Cited by
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816641433 · Cited by 115
Self-citationFreckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815580160 · Cited by 70
Self-citationBrazg, R. L., Klaff, L. J., & Parkin, C. G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700117 · Cited by 68
Pleus, S., Baumstark, A., Jendrike, N., Mende, J., Link, M., Zschornack, E., Haug, C., & Freckmann, G. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-001067 · Cited by 63
Self-citation
Metadata
- Authors
- Baumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., & Freckmann, G.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2012-09
- Volume
- 6
- Issue
- 5
- Pages
- 1076-1086
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2121941402
- Cited by
- 90
- References
- 9
- FWCI
- 9.20
- Type
- journal article
- Language
- en
- Open access
- bronze
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Lot-to-Lot Variability of Test Strips and Accuracy Assessment of Systems for Self-Monitoring of Blood Glucose according to ISO 15197. https://app.cochat.ai/papers/10.1177/193229681200600511
- The paper
- Baumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., & Freckmann, G. (2012). Lot-to-Lot Variability of Test Strips and Accuracy Assessment of Systems for Self-Monitoring of Blood Glucose according to ISO 15197. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229681200600511
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.