Evaluation of a Minimally Invasive System for Measuring Glucose Area under the Curve during Oral Glucose Tolerance Tests: Usefulness of Sweat Monitoring for Precise Measurement

Open access
Sakaguchi, Kazuhiko, Hirota, Yushi, Hashimoto, Naoko, Ogawa, Wataru, Hamaguchi, Tomoya, Matsuo, Toshihiro, Miyagawa, Jun-Ichiro, Namba, Mitsuyoshi +9 more·2013·Journal of Diabetes Science and Technology

Abstract

Aims: We developed a system for measuring glucose area under the curve (AUC) using minimally invasive interstitial fluid extraction technology (MIET). Sweat contamination during interstitial fluid glucose (IG) extraction affects the accuracy of glucose AUC measurement, because this technology uses extracted sodium ion levels as an internal standard. Therefore, we developed a sweat monitoring patch to reduce this effect and investigated its efficacy in volunteers undergoing oral glucose tolerance tests (OGTTs). Materials and Methods: Fifty diabetes mellitus inpatients and 10 healthy subjects undergoing the 75 g OGTT were included. Two sites on the forearm were pretreated with microneedle arrays, then hydrogels for interstitial fluid extraction were placed on the treated sites. Simultaneously, hydrogels for sweat monitoring were placed on untreated sites near the treated sites. Plasma glucose (PG) levels were measured every 30 min for 2 h to calculate reference AUC values. Using MIET, IG AUC was calculated from extracted glucose and sodium ion levels after attachment of the hydrogel for 2 h. Results: Good correlation between IG AUC measurements using MIET and reference AUCs measured using PG levels was confirmed over a wide AUC range (202–610 mg/h/dl) after correction for the sweat-induced error detected by the hydrogel patches on the nonpretreated skin. Strong correlation between IG AUC and peak glucose levels indicates that glucose spikes can be easily detected by this system. Conclusion: We confirmed the effectiveness of a sweat monitoring patch for precise AUC measurement using MIET. This novel, easy-to-use system has potential for glucose excursion evaluation in daily clinical practice.

Integrity signals

Registry

Crossref · 10.1177/193229681300700313

Cites

26 of 26 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Although several studies have indicated that sweat glucose concentrations can correlate with concurrent blood levels, the limitations of uncontrolled conditions, contamination, irregular sampling, and low volumes can significantly impact the accuracy of collected data.

    “Although several studies have indicated that sweat glucose concentrations can correlate with concurrent blood levels, these limitations can significantly impact the accuracy of collected data 64, 76,77.”

    From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [77]

  • Accurate measurement of physiologically relevant sweat glucose concentrations with epidermal biosensors faces major challenges related to uncontrolled operational conditions, glucose contamination, irregular sampling rates, and low sampling volumes.

    “The accurate measurement of physiologically relevant sweat glucose concentrations with epidermal biosensors faces several major challenges related to uncontrolled operational conditions (e.g., varying temperature and pH), glucose contamination from various sources, irregular sampling rates, and low sampling volumes.”

    From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [77]

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Sakaguchi, K., Hirota, Y., Hashimoto, N., Ogawa, W., Hamaguchi, T., Matsuo, T., Miyagawa, J.-I., Namba, M., Sato, T., Okada, S., Tomita, K., Matsuhisa, M., Kaneto, H., Kosugi, K., Maegawa, H., Nakajima, H., & Kashiwagi, A.
Affiliations
  • Kobe University (JP)
  • Hyogo Medical University (JP)
  • Sysmex (Japan) (JP)
  • Osaka International Cancer Institute (JP)
  • Diabetes Therapeutics and Research Center, Tokushima University (JP)
  • The University of Osaka (JP)
  • Osaka Police Hospital (JP)
  • Shiga University of Medical Science (JP)
Corresponding author
Sakaguchi, Kazuhiko
Venue
Journal of Diabetes Science and Technology
Published
2013-05
Volume
7
Issue
3
Pages
678-688
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2116321774
Cited by
50
References
26
Citations by year
2026: 2 · 2025: 3 · 2024: 5 · 2023: 6 · 2022: 1 · 2021: 5 · 2020: 3 · 2019: 6 · 2018: 5 · 2017: 7 · 2016: 5 · 2015: 1 · 2014: 1
FWCI
1.84
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • Europe PMC (PubMed Central) (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Pharmaceutical Science › Advancements in Transdermal Drug Delivery
  • Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
Keywords
glucose area under the curve, oral glucose tolerance test, interstitial glucose, microneedle arrays, hydrogel patch, glycemic excursions, diabetes mellitus
MeSH terms
Diabetes Mellitus, Female, Glucose, Glucose Tolerance Test, Humans, Male, Middle Aged, Monitoring, Physiologic, Self Care, Sweat, Area Under Curve, Hydrogels, Extracellular Fluid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review