Technologies for Continuous Glucose Monitoring: Current Problems and Future Promises

Open access
Vaddiraju, Santhisagar, Burgess, Diane J., Tomazos, Ioannis, Jain, Faquir C., Papadimitrakopoulos, Fotios·2010·Journal of Diabetes Science and Technology

Abstract

Devices for continuous glucose monitoring (CGM) are currently a major focus of research in the area of diabetes management. It is envisioned that such devices will have the ability to alert a diabetes patient (or the parent or medical care giver of a diabetes patient) of impending hypoglycemic/hyperglycemic events and thereby enable the patient to avoid extreme hypoglycemic/hyperglycemic excursions as well as minimize deviations outside the normal glucose range, thus preventing both life-threatening events and the debilitating complications associated with diabetes. It is anticipated that CGM devices will utilize constant feedback of analytical information from a glucose sensor to activate an insulin delivery pump, thereby ultimately realizing the concept of an artificial pancreas. Depending on whether the CGM device penetrates/breaks the skin and/or the sample is measured extracorporeally, these devices can be categorized as totally invasive, minimally invasive, and noninvasive. In addition, CGM devices are further classified according to the transduction mechanisms used for glucose sensing (i.e., electrochemical, optical, and piezoelectric). However, at present, most of these technologies are plagued by a variety of issues that affect their accuracy and long-term performance. This article presents a critical comparison of existing CGM technologies, highlighting critical issues of device accuracy, foreign body response, calibration, and miniaturization. An outlook on future developments with an emphasis on long-term reliability and performance is also presented.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/193229681000400632

Cites

181 of 181 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Miniaturization of optical glucose monitoring instrumentation is not currently possible, preventing personalized continuous monitoring devices.

    “Moreover, miniaturization of the instrumentation is not currently possible thus precluding personalized devices for continuous monitoring.19”
  • Microdialysis avoids direct sensor implantation, but in vivo glucose recovery is erratic because of the foreign body response.

    “Microdialysis avoids direct implantation of a sensor but the glucose measurement (i.e., recovery) remains erratic in vivo due to the foreign body response.19,24–25”
  • Near-infrared spectroscopy allows 90-95% of light to pass through the stratum corneum and epidermis into subcutaneous space with minimal tissue adsorption independent of skin pigmentation.

    “Optical detection of glucose via near infrared (NIR) spectroscopy enables 90–95% of the light to pass through the human stratum corneum and epidermis into subcutaneous space with minimal tissue adsorption independent of skin pigmentation.19,40–41”
  • Photobleaching of fluorophores and tissue scattering are significant drawbacks of optical glucose detection.

    “Photobleaching of the fluorophore and scattering in tissue are significant drawbacks.19,35–36”
  • Most electron mediators used in second-generation glucose biosensors are toxic, and mediator-oxygen competition persists.

    “Unfortunately, most mediators are toxic and competition between the mediator and oxygen still exists.19,22”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 13

  1. [21]21. Newman JD, Turner AP. Home blood glucose biosensors: a commercial perspective. Biosens Bioelectron. 2005;20(12):2435-53.
  2. [80]80. Ferrante do Amaral CE, Wolf B. Current development in noninvasive glucose monitoring. Med Eng Phys. 2008;30(5):541-9.
  3. [95]95. Renard E, Shah R, Miller M, Kolopp M, Costalat G, Bringer J. Sustained safety and accuracy of central IV glucose sensors connected to implanted insulin pumps and short-term closed-loop trials in diabetic patients. Diabetologia 2003;46:A47.
  4. [98]98. Gourzi M, Rouane A, Guelaz R, Alavi MS, McHugh MB, Nadi M, Roth P. Non-invasive glycaemia blood measurements by electromagnetic sensor: study in static and dynamic blood circulation. J Med Eng Technol. 2005;29(1):22-6.
  5. [113]113. Vaddiraju S, Tomazos I, Burgess DJ, Jain FC, Papadimitrakopoulos F. Emerging synergy between nanotechnology and implantable biosensors: a review. Biosens Bioelectron. 2010;25(7):1553-65.
  6. [119]119. Dong S, Wang B, Liu B. Amperometric glucose sensor with ferrocene as an electron transfer mediator. Biosens Bioelectron. 1992;7(3):215-22.
  7. [121]121. Yu X, Sotzing GA, Papadimitrakopoulos F, Rusling JF. Wiring of enzymes to electrodes by ultrathin conductive polyion underlayers: enhanced catalytic response to hydrogen peroxide. Anal Chem. 2003;75(17):4565-71.
  8. [146]146. Ishihara K, Tanaka S, Furukawa N, Kurita K, Nakabayashi N. Improved blood compatibility of segmented polyurethanes by polymeric additives having phospholipid polar groups. I. Molecular design of polymeric additives and their functions. J Biomed Mater Res. 1996;32(3):391-9.
  9. [157]157. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. Engineering the tissue which encapsulates subcutaneous implants. I. Diffusion properties. J Biomed Mater Res. 1997;37(3):401-12.
  10. [158]158. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. Engineering the tissue which encapsulates subcutaneous implants. III. Effective tissue response times. J Biomed Mater Res. 1998;40(4):598-605.
  11. [159]159. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. Engineering the tissue which encapsulates subcutaneous implants. II. Plasmatissue exchange properties. J Biomed Mater Res. 1998;40(4):586-97.
  12. [174]174. Guto PM, Kumar CV, Rusling JF. Thermostable peroxidasepolylysine films for biocatalysis at 90 degrees C. J Phys Chem B. 2007;111(30):9125-31.
  13. [187]187. Renard E, Shah R, Miller M, Kolopp M, Costalat G, Bringer J. Sustained safety and accuracy of central IV glucose sensors connected to implanted insulin pumps and short-term closed-loop trials in diabetic patients. Diabetologia. 2003;46:A47.

Cited by

Sorted by citations

Authors
Vaddiraju, S., Burgess, D. J., Tomazos, I., Jain, F. C., & Papadimitrakopoulos, F.
Affiliations
University of Connecticut (US)
Corresponding author
Papadimitrakopoulos, Fotios
Venue
Journal of Diabetes Science and Technology
Published
2010-11
Volume
4
Issue
6
Pages
1540-1562
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2161313513
Cited by
285
References
181
Citations by year
2026: 19 · 2025: 14 · 2024: 9 · 2023: 17 · 2022: 12 · 2021: 15 · 2020: 8 · 2019: 24 · 2018: 25 · 2017: 19 · 2016: 27 · 2015: 24 · 2014: 22 · 2013: 25 · 2012: 14
FWCI
12.10
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • National Heart, Lung, and Blood Institute (1-R21-HL090458-01, R21HL090458)
  • National Institute of Environmental Health Sciences (ES013557, R01ES013557)
  • National Institute of Biomedical Imaging and Bioengineering (R43EB011886)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
  • Health Sciences › Medicine › Surgery › Pancreatic function and diabetes
Keywords
continuous glucose monitoring, diabetes management, artificial pancreas, glucose sensor, hypoglycemia detection, hyperglycemia, non-invasive glucose monitoring, electrochemical glucose sensor, optical glucose sensor, piezoelectric sensors, foreign body reaction, sensor calibration, device miniaturization, sensor accuracy, long-term reliability
MeSH terms
Blood Glucose, Diabetes Mellitus, Equipment Design, Forecasting, Humans, Insulin Infusion Systems, Miniaturization, Predictive Value of Tests, Signal Processing, Computer-Assisted, Time Factors, Reproducibility of Results, Biomarkers, Monitoring, Ambulatory

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review