Technologies for Continuous Glucose Monitoring: Current Problems and Future Promises
Open accessAbstract
Devices for continuous glucose monitoring (CGM) are currently a major focus of research in the area of diabetes management. It is envisioned that such devices will have the ability to alert a diabetes patient (or the parent or medical care giver of a diabetes patient) of impending hypoglycemic/hyperglycemic events and thereby enable the patient to avoid extreme hypoglycemic/hyperglycemic excursions as well as minimize deviations outside the normal glucose range, thus preventing both life-threatening events and the debilitating complications associated with diabetes. It is anticipated that CGM devices will utilize constant feedback of analytical information from a glucose sensor to activate an insulin delivery pump, thereby ultimately realizing the concept of an artificial pancreas. Depending on whether the CGM device penetrates/breaks the skin and/or the sample is measured extracorporeally, these devices can be categorized as totally invasive, minimally invasive, and noninvasive. In addition, CGM devices are further classified according to the transduction mechanisms used for glucose sensing (i.e., electrochemical, optical, and piezoelectric). However, at present, most of these technologies are plagued by a variety of issues that affect their accuracy and long-term performance. This article presents a critical comparison of existing CGM technologies, highlighting critical issues of device accuracy, foreign body response, calibration, and miniaturization. An outlook on future developments with an emphasis on long-term reliability and performance is also presented.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Miniaturization of optical glucose monitoring instrumentation is not currently possible, preventing personalized continuous monitoring devices.
“Moreover, miniaturization of the instrumentation is not currently possible thus precluding personalized devices for continuous monitoring.19”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
Microdialysis avoids direct sensor implantation, but in vivo glucose recovery is erratic because of the foreign body response.
“Microdialysis avoids direct implantation of a sensor but the glucose measurement (i.e., recovery) remains erratic in vivo due to the foreign body response.19,24–25”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
Near-infrared spectroscopy allows 90-95% of light to pass through the stratum corneum and epidermis into subcutaneous space with minimal tissue adsorption independent of skin pigmentation.
“Optical detection of glucose via near infrared (NIR) spectroscopy enables 90–95% of the light to pass through the human stratum corneum and epidermis into subcutaneous space with minimal tissue adsorption independent of skin pigmentation.19,40–41”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
Photobleaching of fluorophores and tissue scattering are significant drawbacks of optical glucose detection.
“Photobleaching of the fluorophore and scattering in tissue are significant drawbacks.19,35–36”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
Most electron mediators used in second-generation glucose biosensors are toxic, and mediator-oxygen competition persists.
“Unfortunately, most mediators are toxic and competition between the mediator and oxygen still exists.19,22”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
The electrode potential used to monitor hydrogen peroxide also oxidizes endogenous electroactive species and produces high current densities.
“Additionally, the working electrode potential required to monitor hydrogen peroxide (i.e., ~+0.6 V vs. Ag/AgCl) will also oxidize electroactive endogenous species (e.g., ascorbic acid and acetaminophen) and create high current densities.19”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
Low oxygen concentrations cause problems with biosensor response to glucose, particularly dynamic range, due to stoichiometric imbalance between oxygen and glucose.
“Indeed, low concentrations of oxygen lead to problems with the biosensor response to glucose (particularly dynamic range) due to stoichiometric imbalance between the two cofactors.19,63”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [19]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
International Diabetes Federation 2017
Carracher, A. M., Marathe, P. H., & Close, K. L. (2018) · Journal of Diabetes · doi:10.1111/1753-0407.12644 · Cited by 13,369 · see citing works
International Organization for Standardization
(2011) · Springer eBooks · doi:10.1007/springerreference_16847 · Cited by 12,502 · see citing works
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · see citing works
Home blood glucose biosensors: a commercial perspective
Newman, J. D., & Turner, A. P. F. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.11.012 · Cited by 901 · see citing works
Nanostructures for enzyme stabilization
Kim, J., Grate, J. W., & Wang, P. (2005) · Chemical Engineering Science · doi:10.1016/j.ces.2005.05.067 · Cited by 843 · see citing works
Glucose Biosensors: 40 Years of Advances and Challenges
Wang, J. (2001) · Electroanalysis · doi:10.1002/1521-4109(200108)13:12 · Cited by 765 · see citing works
Charge- and size-based separation of macromolecules using ultrathin silicon membranes
Striemer, C. C., Gaborski, T. R., McGrath, J. L., & Fauchet, P. M. (2007) · Nature · doi:10.1038/nature05532 · Cited by 753 · see citing works
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660 · see citing works
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620 · see citing works
Fluorescence-based glucose sensors
Pickup, J. C., Hussain, F., Evans, N. D., Rolinski, O. J., & Birch, D. J. S. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.10.002 · Cited by 616 · see citing works
Glucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584 · see citing works
Mano, N., Mao, F., & Heller, A. (2003) · Journal of the American Chemical Society · doi:10.1021/ja0346328 · Cited by 574 · see citing works
Degani, Y., & Heller, A. (1987) · The Journal of Physical Chemistry · doi:10.1021/j100290a001 · Cited by 574 · see citing works
Hyponatremia: evaluating the correction factor for hyperglycemia
Hillier, T. A., Abbott, R. D., & Barrett, E. J. (1999) · The American Journal of Medicine · doi:10.1016/s0002-9343(99)00055-8 · Cited by 567 · see citing works
Popat, K. C., Eltgroth, M., LaTempa, T. J., et al. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.07.037 · Cited by 561 · see citing works
Nonenzymatic Electrochemical Glucose Sensor Based on Nanoporous PtPb Networks
Wang, J., Thomas, D. F., & Chen, A. (2008) · Analytical Chemistry · doi:10.1021/ac701790z · Cited by 534 · see citing works
Enzyme immobilization on electrospun polymer nanofibers: An overview
Wang, Z., Wan, L., Liu, Z.-M., Huang, X., & Xu, Z. (2008) · Journal of Molecular Catalysis B Enzymatic · doi:10.1016/j.molcatb.2008.05.005 · Cited by 523 · see citing works
Neovascularization of synthetic membranes directed by membrane microarchitecture
Brauker, J. H., Carr‐Brendel, V. E., Martinson, L. A., Crudele, J., Johnston, W. D., & Johnson, R. C. (1995) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820291208 · Cited by 517 · see citing works
Carbon Nanotubes for Electronic and Electrochemical Detection of Biomolecules
Kim, S. N., Rusling, J. F., & Papadimitrakopoulos, F. (2007) · Advanced Materials · doi:10.1002/adma.200700665 · Cited by 499 · see citing works
Self-citationSuction Blister Device for Separation of Viable Epidermis from Dermis*
Kiistala, U. (1968) · Journal of Investigative Dermatology · doi:10.1038/jid.1968.15 · Cited by 457 · see citing works
Highly Ordered Platinum‐Nanotubule Arrays for Amperometric Glucose Sensing
Yuan, J., Wang, K., & Xia, X. (2005) · Advanced Functional Materials · doi:10.1002/adfm.200400321 · Cited by 433 · see citing works
Improvement in Glycemic Excursions With a Transcutaneous, Real-Time Continuous Glucose Sensor
Garg, S. K., Zisser, H., Schwartz, S. L., Bailey, T. S., Kaplan, R. A., Ellis, S. L., & Jovanovič, L. (2006) · Diabetes Care · doi:10.2337/diacare.29.01.06.dc05-1686 · Cited by 429 · see citing works
Chen, T., Barton, S. C., Binyamin, G. N., Gao, Z., Zhang, Y., Kim, H.-H., & Heller, A. (2001) · Journal of the American Chemical Society · doi:10.1021/ja0163164 · Cited by 427 · see citing works
Onuki, Y., Bhardwaj, U., Papadimitrakopoulos, F., & Burgess, D. J. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200610 · Cited by 417 · see citing works
Self-citationBoyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404 · see citing works
Partitioning glucose distribution/transport, disposal, and endogenous production during IVGTT
Hovorka, R., Shojaee-Moradie, F., Carroll, P. V., Chassin, L. J., Gowrie, I. J., Jackson, N. C., Tudor, R. S., Umpleby, A. M., & Jones, R. H. (2002) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.00304.2001 · Cited by 393 · see citing works
Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390 · see citing works
Emerging synergy between nanotechnology and implantable biosensors: A review
Vaddiraju, S., Tomazos, I., Burgess, D. J., Jain, F., & Papadimitrakopoulos, F. (2009) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2009.12.001 · Cited by 370 · see citing works
Self-citationPhotonic Crystal Glucose-Sensing Material for Noninvasive Monitoring of Glucose in Tear Fluid
Alexeev, V. L., Das, S., Finegold, D. N., & Asher, S. A. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2004.039701 · Cited by 370 · see citing works
Optical glucose sensing in biological fluids: an overview
McNichols, R. J., & Coté, G. L. (2000) · Journal of Biomedical Optics · doi:10.1117/1.429962 · Cited by 369 · see citing works
Russell, R. J., Pishko, M. V., Gefrides, C. C., McShane, M. J., & Coté, G. L. (1999) · Analytical Chemistry · doi:10.1021/ac990060r · Cited by 354 · see citing works
Spectroscopic and Clinical Aspects of Noninvasive Glucose Measurements
Khalil, O. S. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.2.165 · Cited by 331 · see citing works
WEARABLE ARTIFICIAL ENDOCRINE PANCREAS WITH NEEDLE-TYPE GLUCOSE SENSOR
Shichiri, M., Yamasaki, Y., Kawamori, R., Hakui, N., & Abe, H. (1982) · The Lancet · doi:10.1016/s0140-6736(82)92788-x · Cited by 314 · see citing works
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312 · see citing works
A Miniature Biofuel Cell Operating in A Physiological Buffer
Mano, N., Mao, F., & Heller, A. (2002) · Journal of the American Chemical Society · doi:10.1021/ja028514g · Cited by 307 · see citing works
Noninvasive Blood Glucose Monitoring With Optical Coherence Tomography
Larin, K. V., Eledrisi, M. S., Motamedi, M., & Esenaliev, R. O. (2002) · Diabetes Care · doi:10.2337/diacare.25.12.2263 · Cited by 293 · see citing works
Burmeister, J. J., & Gerhardt, G. A. (2001) · Analytical Chemistry · doi:10.1021/ac0010429 · Cited by 289 · see citing works
Sensors for glucose monitoring: technical and clinical aspects
Koschinsky, T., & Heinemann, L. (2001) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.188 · Cited by 288 · see citing works
Tura, A., Maran, A., & Pacini, G. (2007) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2006.10.027 · Cited by 283 · see citing works
Barone, P. W., Parker, R. S., & Strano, M. S. (2005) · Analytical Chemistry · doi:10.1021/ac0511997 · Cited by 279 · see citing works
Raman spectroscopy for noninvasive glucose measurements
Enejder, A. M. K., Scecina, T. G., Oh, J., Hunter, M., Shih, W., Šašić, S., Horowitz, G. L., & Feld, M. S. (2005) · Journal of Biomedical Optics · doi:10.1117/1.1920212 · Cited by 272 · see citing works
Microvoltammetric techniques and sensors for monitoring neurochemical dynamics in vivo. A review
O'Neill, R. D. (1994) · The Analyst · doi:10.1039/an9941900767 · Cited by 270 · see citing works
Current development in non-invasive glucose monitoring
Ferrante do Amaral, C. E., & Wolf, B. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.06.003 · Cited by 269 · see citing works
Hickey, T., Kreutzer, D., Burgess, D. J., & Moussy, F. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00291-5 · Cited by 268 · see citing works
Self-citationAmperometric Biosensors Based on Redox Polymer−Carbon Nanotube−Enzyme Composites
Joshi, P. P., Merchant, S. A., Wang, Y., et al. (2005) · Analytical Chemistry · doi:10.1021/ac0484169 · Cited by 262 · see citing works
Non-Invasive Glucose Measurement Technologies: An Update from 1999 to the Dawn of the New Millennium
Khalil, O. S. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.660 · Cited by 256 · see citing works
Transdermal monitoring of glucose and other analytes using ultrasound
Kost, J., Mitragotri, S., Gabbay, R. A., Pishko, M. V., & Langer, R. (2000) · Nature Medicine · doi:10.1038/73213 · Cited by 253 · see citing works
Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 252 · see citing works
Can continuous glucose monitoring be used for the treatment of diabetes?
Reach, G., & Wilson, G. S. (1992) · Analytical Chemistry · doi:10.1021/ac00030a001 · Cited by 239 · see citing works
Long-term engraftment of hepatocytes transplanted on biodegradable polymer sponges
Mooney, D. J., Sano, K., Kaufmann, P. M., Majahod, K., Schloo, B. L., Vacanti, J. P., & Langer, R. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19971205)37:3 · Cited by 233 · see citing works
Harrison, D. J., Turner, R. F. B., & Baltes, H. P. (1988) · Analytical Chemistry · doi:10.1021/ac00170a003 · Cited by 232 · see citing works
Concurrent delivery of dexamethasone and VEGF for localized inflammation control and angiogenesis
Patil, S. D., Papadmitrakopoulos, F., & Burgess, D. J. (2007) · Journal of Controlled Release · doi:10.1016/j.jconrel.2006.10.013 · Cited by 230 · see citing works
Self-citationAccuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R., Bugler, J., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228 · see citing works
Designing synthetic materials to control stem cell phenotype
Saha, K., Pollock, J. F., Schaffer, D. V., & Healy, K. E. (2007) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2007.05.030 · Cited by 219 · see citing works
Implanted Electrochemical Glucose Sensors for the Management of Diabetes
Heller, A. (1999) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.1.1.153 · Cited by 215 · see citing works
Hickey, T., Kreutzer, D. L., Burgess, D. J., & Moussy, F. G. (2002) · Journal of Biomedical Materials Research · doi:10.1002/jbm.10016 · Cited by 187 · see citing works
Self-citationAdvances in Photoacoustic Noninvasive Glucose Testing
MacKenzie, H. A., Ashton, H., Spiers, S., Shen, Y., Freeborn, S. S., Hannigan, J., Lindberg, J. M., & Rae, P. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.9.1587 · Cited by 186 · see citing works
Rabinovitch, B., March, W. F., & Adams, R. L. (1982) · Diabetes Care · doi:10.2337/diacare.5.3.254 · Cited by 180 · see citing works
Ward, W. K., Slobodzian, E. P., Tiekotter, K. L., & Wood, M. D. (2002) · Biomaterials · doi:10.1016/s0142-9612(02)00160-6 · Cited by 176 · see citing works
Application of Chronic Intravascular Blood Glucose Sensor in Dogs
Armour, J. C., Lucisano, J. Y., Bd, M., & Gough, D. A. (1990) · Diabetes · doi:10.2337/diab.39.12.1519 · Cited by 175 · see citing works
Accuracy of Perceiving Blood Glucose in IDDM
Cox, D. J., Clarke, W. L., Gonder-Frederick, L. A., Pohl, S. L., Hoover, C. W., Snyder, A. L., Zimbelman, L. K., Carter, W. R., Bobbitt, S. A., & Pennebaker, J. W. (1985) · Diabetes Care · doi:10.2337/diacare.8.6.529 · Cited by 169 · see citing works
Nanoporous anti-fouling silicon membranes for biosensor applications
Desai, T. A., Hansford, D. J., Leoni, L., Essenpreis, M., & Mauro Ferrari, B. (2000) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(00)00088-9 · Cited by 163 · see citing works
Gold nanoparticles integrated in a nanotube array for electrochemical detection of glucose
Zhou, Y., Yang, S. C., Qian, Q., et al. (2008) · Electrochemistry Communications · doi:10.1016/j.elecom.2008.11.010 · Cited by 160 · see citing works
Moussy, F. G., Harrison, D. J., O'Brien, D. W., & Rajotte, R. V. (1993) · Analytical Chemistry · doi:10.1021/ac00063a023 · Cited by 158 · see citing works
Gorton, L. G. O., Bremle, G., Csöregi, E., Jönsson–Pettersson, G., & Persson, B. (1991) · Analytica Chimica Acta · doi:10.1016/0003-2670(91)87006-s · Cited by 157 · see citing works
Caduff, A., Dewarrat, F., Talary, M., Stalder, G., Heinemann, L., & Feldman, Y. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.01.031 · Cited by 155 · see citing works
Fluorescence Glucose Detection: Advances Toward the Ideal In Vivo Biosensor
Moschou, E. A., Sharma, B. V., Deo, S. K., & Daunert, S. (2004) · Journal of Fluorescence · doi:10.1023/b:jofl.0000039341.64999.83 · Cited by 154 · see citing works
Reversible Control of Carbon Nanotube Aggregation for a Glucose Affinity Sensor
Barone, P. W., & Strano, M. S. (2006) · Angewandte Chemie International Edition · doi:10.1002/anie.200603138 · Cited by 152 · see citing works
Continuous Glucose Sensing with a Fluorescent Thin‐Film Hydrogel
Suri, J. T., Cordes, D. B., Cappuccio, F. E., et al. (2003) · Angewandte Chemie International Edition · doi:10.1002/anie.200352405 · Cited by 149 · see citing works
A Miniature Optical Glucose Sensor Based on Affinity Binding
Mansouri, S., & Schultz, J. S. (1984) · Nature Biotechnology · doi:10.1038/nbt1084-885 · Cited by 148 · see citing works
In Vivo Chemical Sensors: Tackling Biocompatibility
Frost, M. C., & Meyerhoff, M. E. (2006) · Analytical Chemistry · doi:10.1021/ac069475k · Cited by 145 · see citing works
Sachedina, N., & Pickup, J. C. (2003) · Diabetic Medicine · doi:10.1046/j.1464-5491.2003.01037.x · Cited by 142 · see citing works
An Oxygen-Insensitive Reagentless Glucose Biosensor Based on Osmium-Complex Modified Polypyrrole
Habermüller, K., Ramanavičius, A., Laurinavičius, V., et al. (2000) · Electroanalysis · doi:10.1002/1521-4109(200011)12:17 · Cited by 140 · see citing works
Progress Toward an In Vivo Surface-Enhanced Raman Spectroscopy Glucose Sensor
Lyandres, O., Yuen, J. M., Shah, N. C., Duyne, R. P. V., Walsh, J. T., & Glucksberg, M. R. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0288 · Cited by 135 · see citing works
Norton, L. W., Koschwanez, H. E., Wisniewski, N. A., Klitzman, B., & Reichert, W. M. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31088 · Cited by 128 · see citing works
Fiber-optic biosensors based on fluorescence energy transfer
Meadows, D. L. (1988) · Talanta · doi:10.1016/0039-9140(88)80053-5 · Cited by 127 · see citing works
Reverse Iontophoresis: Development of a Noninvasive Approach for Glucose Monitoring
Rao, G., Glikfeld, P., & Guy, R. H. (1993) · Pharmaceutical Research · doi:10.1023/a:1018926215306 · Cited by 126 · see citing works
Performance of subcutaneously implanted glucose sensors for continuous monitoring
Gerritsen, M. (1999) · The Netherlands Journal of Medicine · doi:10.1016/s0300-2977(99)00006-6 · Cited by 122 · see citing works
Yuhashi, N., Tomiyama, M., Okuda‐Shimazaki, J., et al. (2004) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.08.017 · Cited by 117 · see citing works
*The Diabetes Research in Children Network (DirecNet) Study Group (2004) · Diabetes Care · doi:10.2337/diacare.27.3.722 · Cited by 114 · see citing works
Caduff, A., Talary, M. S., Mueller, M. C., Dewarrat, F., Klisic, J., Donath, M. Y., Heinemann, L., & Stahel, W. A. (2009) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2009.02.001 · Cited by 109 · see citing works
Study of a reagentless lactate electrode
Blaedel, W. J., & Jenkins, R. A. (1976) · Analytical Chemistry · doi:10.1021/ac50002a045 · Cited by 107 · see citing works
A glucose-sensing contact lens: from bench top to patient
Badugu, R., Lakowicz, J. R., & Geddes, C. D. (2005) · Current Opinion in Biotechnology · doi:10.1016/j.copbio.2004.12.007 · Cited by 105 · see citing works
Myung, Y., Jang, D. M., Cho, Y. J., et al. (2009) · The Journal of Physical Chemistry C · doi:10.1021/jp806633j · Cited by 101 · see citing works
Bhardwaj, U., Sura, R., Papadimitrakopoulos, F., & Burgess, D. J. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100103 · Cited by 99 · see citing works
Self-citationPishko, M. V., Katakis, I., Lindquist, S., Ye, L., Gregg, B. A., & Heller, A. (1990) · Angewandte Chemie International Edition in English · doi:10.1002/anie.199000821 · Cited by 99 · see citing works
A Novel Noninvasive Blood Glucose Monitor
Malchoff, C. D., Shoukri, K., Landau, J. I., & Buchert, J. M. (2002) · Diabetes Care · doi:10.2337/diacare.25.12.2268 · Cited by 96 · see citing works
Amperometric Glucose Sensors: Sources of Error and Potential Benefit of Redundancy
Castle, J. R., & Ward, W. K. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400127 · Cited by 95 · see citing works
Evaluation of a Subcutaneous Glucose Sensor out to 3 Months in a Dog Model
GILLIGAN, B. J., Shults, M. C., Rhodes, R. K., & Updike, S. J. (1994) · Diabetes Care · doi:10.2337/diacare.17.8.882 · Cited by 90 · see citing works
Pendra goes Dutch: lessons for the CE mark in Europe
Wentholt, I. M. E., Hoekstra, J. B. L., Zwart, A., & DeVries, J. H. (2005) · Diabetologia · doi:10.1007/s00125-005-1754-y · Cited by 90 · see citing works
Amperometric glucose sensor with ferrocene as an electron transfer mediator
Dong, S., Wang, B., & Liu, B. (1992) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(92)87018-k · Cited by 87 · see citing works
Weiss, R., Yegorchikov, Y., Shusterman, A., & Raz, I. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0059 · Cited by 86 · see citing works
Harman‐Boehm, I., Gal, A., Raykhman, A. M., Naidis, E., & Mayzel, Y. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400312 · Cited by 85 · see citing works
Percutaneous fiber-optic sensor for chronic glucose monitoring in vivo
Liao, K.-C., Hogen‐Esch, T. E., RICHMOND, F. J. R., et al. (2008) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2008.01.012 · Cited by 82 · see citing works
Initial Clinical Testing of a Holographic Non-Invasive Contact Lens Glucose Sensor
Domschke, A., March, W. F., Kabilan, S., & Lowe, C. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.89 · Cited by 81 · see citing works
In vivo glucose monitoring: Towards ‘Sense and Act’ feedback-loop individualized medical systems
Wang, J. (2007) · Talanta · doi:10.1016/j.talanta.2007.10.023 · Cited by 81 · see citing works
Kim, T.-K., & Burgess, D. J. (2002) · Journal of Pharmacy and Pharmacology · doi:10.1211/002235702760089009 · Cited by 79 · see citing works
Self-citationBone morphogenetic protein encapsulated with a biodegradable and biocompatible polymer
Isobe, M., Amagasa, T., Oida, S., Yamazaki, Y., Ishihara, K., & Nakabayashi, N. (1996) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(199611)32:3 · Cited by 76 · see citing works
In vivo evaluation of an electroenzymatic glucose sensor implanted in subcutaneous tissue
Johnson, K. W., Mastrototaro, J. J., Howey, D. C., Brunelle, R. L., Burden-Brady, P. L., Bryan, N. A., Andrew, C. C., Rowe, H., Allen, D. J., Noffke, B. W., et al. (1992) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(92)85053-d · Cited by 73 · see citing works
Continuous Glucose Sensors: Continuing Questions about Clinical Accuracy
Clarke, W. L., & Kovatchev, B. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100510 · Cited by 71 · see citing works
Immobilization Techniques to Avoid Enzyme Loss from Oxidase-Based Biosensors: A One-Year Study
House, J. L., Anderson, E. M., & Ward, W. K. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100104 · Cited by 71 · see citing works
Gebhart, S. S. P., Faupel, M. L., Fowler, R. T., Kapsner, C., Lincoln, D., McGee, V., Pasqua, J. J., Steed, L., Wangsness, M., Xu, F., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503321827812 · Cited by 69 · see citing works
New technology for multi-sensor silicon needles for biomedical applications
Errachid, A., Ivorra, A., Aguiló, J., et al. (2001) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(01)00826-7 · Cited by 66 · see citing works
Polymer−Enzyme Composite Biosensor with High Glutamate Sensitivity and Low Oxygen Dependence
McMahon, C. P., & O'Neill, R. D. (2005) · Analytical Chemistry · doi:10.1021/ac048686r · Cited by 65 · see citing works
Analysis of ultrasonically extracted interstitial fluid as a predictor of blood glucose levels
Mitragotri, S., Coleman, M. D., Kost, J., & Langer, R. (2000) · Journal of Applied Physiology · doi:10.1152/jappl.2000.89.3.961 · Cited by 65 · see citing works
Continuous Glucose Monitoring with Glucose Sensors: Calibration and Assessment Criteria
Lodwig, V., & Heinemann, L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250596 · Cited by 64 · see citing works
Galeska, I., Hickey, T., Moussy, F. G., Kreutzer, D. L., & Papadimitrakopoulos, F. (2001) · Biomacromolecules · doi:10.1021/bm010112y · Cited by 62 · see citing works
Self-citationSingle Walled Carbon Nanotubes as Reporters for the Optical Detection of Glucose
Barone, P. W., & Strano, M. S. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300204 · Cited by 61 · see citing works
Bioelectrochemical sensor based on PQQ-dependent glucose dehydrogenase
Malinauskas, A., Kuzmarskyt≐, J., Meškys, R., et al. (2004) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2004.02.006 · Cited by 61 · see citing works
Gourzi, M., Rouane, A., Guelaz, R., Alavi, M., McHugh, M. B., Nadi, M., & Roth, P. (2004) · Journal of Medical Engineering & Technology · doi:10.1080/03091900410001720247 · Cited by 60 · see citing works
Extended-Range Glucose Sensor Employing Engineered Glucose Dehydrogenases
Yamazaki, T., Kojima, K., & Sode, K. (2000) · Analytical Chemistry · doi:10.1021/ac000151k · Cited by 60 · see citing works
Yu, X., Sotzing, G. A., Papadimitrakopoulos, F., et al. (2003) · Analytical Chemistry · doi:10.1021/ac034188r · Cited by 58 · see citing works
Self-citationEnzymatic glucose sensors. Improved long-term performance in vitro and in vivo.
Updike, S. J., Shults, M. C., Rhodes, R. K., GILLIGAN, B. J., LUEBOW, J. O., & von Heimburg, D. (1994) · PubMed · Cited by 57 · see citing works
Pfützner, A., Caduff, A., Larbig, M., Schrepfer, T., & Forst, T. M. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/1520915041705839 · Cited by 55 · see citing works
Preparation and characterization of implantable sensors with nitric oxide release coatings
Frost, M. C., Batchelor, M. M., Lee, Y.-M., Zhang, H., Kang, Y., Oh, B. K., Wilson, G. S., Gifford, R., Rudich, S. M., & Meyerhoff, M. E. (2003) · Microchemical Journal · doi:10.1016/s0026-265x(03)00033-x · Cited by 55 · see citing works
Wentholt, I. M. E., Hart, A. A. M., Hoekstra, J. B. L., & DeVries, J. H. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0007 · Cited by 54 · see citing works
Evaluation of silicon nanoporous membranes and ECM-based microenvironments on neurosecretory cells
A, C. A. L., Fleischman, A. J., Roy, S., & Desai, T. A. (2006) · Biomaterials · doi:10.1016/j.biomaterials.2005.12.017 · Cited by 54 · see citing works
Updike, S. J., Shults, M. C., Rhodes, R. K., GILLIGAN, B. J., LUEBOW, J. O., & von Heimburg, D. (1994) · ASAIO Journal · doi:10.1097/00002480-199404000-00007 · Cited by 52 · see citing works
McMahon, C. P., Killoran, S. J., & O'Neill, R. D. (2005) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2005.03.026 · Cited by 50 · see citing works
Glucose sensor with improved haemocompatibilty
Yang, Y. (2000) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(00)00082-8 · Cited by 49 · see citing works
Enhanced Glucose Sensor Linearity Using Poly(Vinyl Alcohol) Hydrogels
Vaddiraju, S., Singh, H., Burgess, D. J., Jain, F., & Papadimitrakopoulos, F. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300434 · Cited by 49 · see citing works
Self-citationCalcification-Resistant Nafion/Fe3+ Assemblies for Implantable Biosensors
Galeska, I., Chattopadhyay, D., Moussy, F. G., & Papadimitrakopoulos, F. (2000) · Biomacromolecules · doi:10.1021/bm0002813 · Cited by 45 · see citing works
Self-citationO’Brien, K. B., Killoran, S. J., O'Neill, R. D., & Lowry, J. P. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.12.020 · Cited by 45 · see citing works
Winet, H., & Bao, J. Y. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19980615)40:4 · Cited by 44 · see citing works
Sources of instability of ‘wired’ enzyme anodes in serum: urate and transition metal ions
Binyamin, G. N., Chen, T., & Heller, A. (2001) · Journal of Electroanalytical Chemistry · doi:10.1016/s0022-0728(00)00267-9 · Cited by 44 · see citing works
Platinum nanoparticle-modified carbon fiber ultramicroelectrodes for mediator-free biosensing
Wu, Z., Chen, L., Shen, G., et al. (2006) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2005.12.023 · Cited by 44 · see citing works
Oxygen Insensitive Glucose Biosensor Based on PQQ-Dependent Glucose Dehydrogenase
Laurinavičius, V., Kurtinaitienė, B., Liauksminas, V., et al. (1999) · Analytical Letters · doi:10.1080/00032719908542822 · Cited by 43 · see citing works
Development of a Real-Time Corneal Birefringence Compensated Glucose Sensing Polarimeter
Cameron, B. D., & Anumula, H. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.156 · Cited by 42 · see citing works
Lutgers, H. L. (2000) · The Netherlands Journal of Medicine · doi:10.1016/s0300-2977(00)00022-x · Cited by 42 · see citing works
Coil-type implantable glucose biosensor with excess enzyme loading
Yu, B. (2005) · Frontiers in bioscience · doi:10.2741/1547 · Cited by 42 · see citing works
The SCGM1 System: Subcutaneous Continuous Glucose Monitoring Based on Microdialysis Technique
Schoemaker, M., Andreis, E., Röper, J., Kotulla, R., Lodwig, V., Obermaier, K., Stephan, P., Reuschling, W., Rutschmann, M., Schwaninger, R., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250613 · Cited by 41 · see citing works
An electroenzymatic glucose sensor fabricated on a flexible substrate
Mastrototaro, J. J., Johnson, K. W., Morff, R. J., et al. (1991) · Sensors and Actuators B Chemical · doi:10.1016/0925-4005(91)80234-b · Cited by 40 · see citing works
Layer-by-Layer Assembled Semipermeable Membrane for Amperometric Glucose Sensors
Tipnis, R., Vaddiraju, S., Jain, F., Burgess, D. J., & Papadimitrakopoulos, F. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100209 · Cited by 39 · see citing works
Self-citationThe role of H2O2 outer diffusion on the performance of implantable glucose sensors
Vaddiraju, S., Burgess, D. J., Jain, F., & Papadimitrakopoulos, F. (2008) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2008.08.015 · Cited by 39 · see citing works
Self-citationKvist, P. H., Iburg, T. M., Aalbæk, B., Gerstenberg, M., Schoier, C., Kaastrup, P., Buch‐Rasmussen, T., Hasselager, E., & Jensen, H. E. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.546 · Cited by 38 · see citing works
Sung, W. J., Na, K., & Bae, Y. H. (2004) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2003.12.005 · Cited by 37 · see citing works
Free-Standing Nanogold Membranes as Scaffolds for Enzyme Immobilization
Phadtare, S., Vathipadiekal, V., Wadgaonkar, P. P., Rao, M., & Sastry, M. (2004) · Langmuir · doi:10.1021/la035870j · Cited by 37 · see citing works
Nanostructured Antifouling Poly(ethylene glycol) Films for Silicon-Based Microsystems
Sharma, S., & Desai, T. (2005) · Journal of Nanoscience and Nanotechnology · doi:10.1166/jnn.2005.030 · Cited by 37 · see citing works
Anti‐inflammatory properties of micropatterned titanium coatings
Sahlin, H., Contreras, R., Gaskill, D. F., Bjursten, L. M., & Frangos, J. A. (2005) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.30642 · Cited by 36 · see citing works
Continuous Monitoring of Blood Glucose
Robert, J. (2002) · Hormone Research in Paediatrics · doi:10.1159/000053321 · Cited by 35 · see citing works
Implantable Continuous Glucose Sensors
Renard, É. (2008) · Current Diabetes Reviews · doi:10.2174/157339908785294406 · Cited by 34 · see citing works
Structure of the Pyrroloquinoline Quinone Radical in Quinoprotein Ethanol Dehydrogenase
Kay, C. W. M., Mennenga, B., Görisch, H., et al. (2005) · Journal of Biological Chemistry · doi:10.1074/jbc.m511132200 · Cited by 34 · see citing works
Bench-to-bedside review: microdialysis in intensive care medicine.
Klaus, S., Heringlake, M., & Bahlmann, L. (2004) · Critical Care · doi:10.1186/cc2882 · Cited by 33 · see citing works
Poly-o-phenylenediamine as redox mediator for laccase
Pałys, B., Bokun, A. A., & Rogalski, J. (2007) · Electrochimica Acta · doi:10.1016/j.electacta.2007.05.029 · Cited by 33 · see citing works
Block, C. E. M. D., Vertommen, J., Manuel‐y‐Keenoy, B., & Gaal, L. V. (2008) · Current Diabetes Reviews · doi:10.2174/157339908785294415 · Cited by 32 · see citing works
Implantable glucose sensors: Present status and future developments
Kerner, W. (2001) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/s-2001-18593 · Cited by 30 · see citing works
Crespi, F. (1991) · Analytical Biochemistry · doi:10.1016/0003-2697(91)90152-j · Cited by 29 · see citing works
In vivo calibration of the subcutaneous amperometric glucose sensors using a non-enzyme electrode
Jeong, R.-A., Hwang, J. Y., Joo, S., Chung, T. D., Park, S., Kang, S. K., Lee, W., & Kim, H. C. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00219-7 · Cited by 28 · see citing works
Novel Hyaluronic Acid Coating for Potential Use in Glucose Sensor Design
Praveen, S., Hanumantha, R., Belovich, J., & Davis, B. L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503765691893 · Cited by 26 · see citing works
(1997) · Diabetes Care · doi:10.2337/diacare.20.1.s1 · Cited by 26 · see citing works
A Wire-Based Dual-Analyte Sensor for Glucose and Lactate: In Vitro and In Vivo Evaluation
Ward, W. K., House, J. L., Birck, J., Anderson, E. M., & Jansen, L. B. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504774198106 · Cited by 25 · see citing works
Evaluation of Kromoscopy: resolution of glucose and urea
Helwig, A. M., Arnold, M. A., & Small, G. W. (2000) · Applied Optics · doi:10.1364/ao.39.004715 · Cited by 24 · see citing works
Rossetti, P., Porcellati, F., Fanelli, C. G., & Bolli, G. B. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.326 · Cited by 23 · see citing works
Clinical Experience of an Iontophoresis Based Glucose Measuring System
Rhee, S. Y., Chon, S., Koh, G., et al. (2007) · Journal of Korean Medical Science · doi:10.3346/jkms.2007.22.1.70 · Cited by 23 · see citing works
Glucose monitoring by microdialysis: performance in a multicentre study
Nielsen, J. K., Freckmann, G., Kapitza, C., Ocvirk, G., Koelker, K.-H., Kamecke, U., Gillen, R. C., Amann-Zalán, I., Jendrike, N. R. M., Christiansen, J. S., et al. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2009.02750.x · Cited by 23 · see citing works
Tamiya, E., Sugiura, Y., Takeuchi, T., et al. (1993) · Sensors and Actuators B Chemical · doi:10.1016/0925-4005(93)87003-8 · Cited by 23 · see citing works
Thermostable Peroxidase−Polylysine Films for Biocatalysis at 90 °C
Guto, P. M., Kumar, C. V., & Rusling, J. F. (2007) · The Journal of Physical Chemistry B · doi:10.1021/jp071525h · Cited by 23 · see citing works
Park, E.-J., Werner, J. R., Beebe, J., et al. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300316 · Cited by 23 · see citing works
Renard, C. B., Askari, B., Suzuki, L. A., et al. (2003) · Diabetologia · doi:10.1007/s00125-003-1247-9 · Cited by 22 · see citing works
Loew, N., Scheller, F. W., & Wollenberger, U. (2004) · Electroanalysis · doi:10.1002/elan.200403005 · Cited by 22 · see citing works
Study of a new electromagnetic sensor for glycaemia measurement:in vitroresults on blood pig
Gourzi, M., Rouane, A., Guelaz, R., Nadi, M., & Jaspard, F. (2003) · Journal of Medical Engineering & Technology · doi:10.1080/0309190031000098845 · Cited by 22 · see citing works
Mechanical and biological properties of nanoporous carbon membranes
Narayan, R. J., Aggarwal, R., Wei, W., et al. (2008) · Biomedical Materials · doi:10.1088/1748-6041/3/3/034107 · Cited by 19 · see citing works
Microdialysis of Interstitial Fluid for Continuous Glucose Measurement
Klonoff, D. C. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250569 · Cited by 19 · see citing works
Performance metrics for continuous interstitial glucose monitoring: approved guideline
Klonoff, D. C., & Clinical And Laboratory Standards Institute (2019) · Harvard Dataverse · doi:10.7910/dvn/rpo6k9 · Cited by 19 · see citing works
Effects of glucose in blood and skin impedance spectroscopy
do Amaral, C. E. F., & Wolf, B. A. (2007) · IEEE AFRICON · doi:10.1109/afrcon.2007.4401489 · Cited by 18 · see citing works
Huber, D., Talary, M. S., Dewarrat, F., & Caduff, A. (2007) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-007-0229-3 · Cited by 16 · see citing works
POLARIMETRIC METHODS FOR MEASUREMENT OF INTRA OCULAR GLUCOSE CONCENTRATION
Rawer, R., Stork, W., & Müller-Glaser, K.-D. (2002) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmte.2002.47.s1a.186 · Cited by 16 · see citing works
McShane, M. J. (2007) · Springer eBooks · doi:10.1007/0-387-33015-1_6 · Cited by 13 · see citing works
Ishihara, K., Nakabayashi, N., Nishida, K., Sakakida, M., & Shichiri, M. (1994) · ACS symposium series · doi:10.1021/bk-1994-0556.ch016 · Cited by 13 · see citing works
APPLICATION OF POLYANION/Fe3 MULTILAYERED MEMBRANES IN PREVENTION OF BIOSENSOR MINERALIZATION
Galeska, I., Chattopadhyay, D., & Papadimitrakopoulos, F. (2002) · Journal of Macromolecular Science Part A · doi:10.1081/ma-120014846 · Cited by 12 · see citing works
Self-citationD’Archangelo, M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200228 · Cited by 10 · see citing works
Frost, M. C., & Meyerhoff, M. E. (2004) · Methods in enzymology on CD-ROM/Methods in enzymology · doi:10.1016/s0076-6879(04)81045-0 · Cited by 9 · see citing works
Amperometric enzyme electrode with the use of dehydrogenase and NAD(P)H oxidase
Mizutani, F., Yabuki, S., & Katsura, T. (1993) · Sensors and Actuators B Chemical · doi:10.1016/0925-4005(93)85097-t · Cited by 9 · see citing works
Pendra goes Dutch; lessons for the CE mark in Europe
DeVries, J. H., Wentholt, I. M. E., Zwart, A., et al. (2006) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(06)70006-7 · Cited by 8 · see citing works
Skoog, M., & Johansson, G. (1991) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(91)87005-v · Cited by 8 · see citing works
Cembrowski, G. S., Smith, B., & O’Malley, E. M. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300407 · Cited by 5 · see citing works
Feng, L., Pan, T., & Sevick-Muraca, E. M. (2005) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.589161 · Cited by 3 · see citing works
[Use of a wearable artificial pancreas].
Shichiri, M., Kawamori, R., & Yamasaki, Y. (1985) · PubMed · Cited by 1 · see citing works
Flagged first, then by citations
Research works printed but not linked: 13
- [21]21. Newman JD, Turner AP. Home blood glucose biosensors: a commercial perspective. Biosens Bioelectron. 2005;20(12):2435-53.
- [80]80. Ferrante do Amaral CE, Wolf B. Current development in noninvasive glucose monitoring. Med Eng Phys. 2008;30(5):541-9.
- [95]95. Renard E, Shah R, Miller M, Kolopp M, Costalat G, Bringer J. Sustained safety and accuracy of central IV glucose sensors connected to implanted insulin pumps and short-term closed-loop trials in diabetic patients. Diabetologia 2003;46:A47.
- [98]98. Gourzi M, Rouane A, Guelaz R, Alavi MS, McHugh MB, Nadi M, Roth P. Non-invasive glycaemia blood measurements by electromagnetic sensor: study in static and dynamic blood circulation. J Med Eng Technol. 2005;29(1):22-6.
- [113]113. Vaddiraju S, Tomazos I, Burgess DJ, Jain FC, Papadimitrakopoulos F. Emerging synergy between nanotechnology and implantable biosensors: a review. Biosens Bioelectron. 2010;25(7):1553-65.
- [119]119. Dong S, Wang B, Liu B. Amperometric glucose sensor with ferrocene as an electron transfer mediator. Biosens Bioelectron. 1992;7(3):215-22.
- [121]121. Yu X, Sotzing GA, Papadimitrakopoulos F, Rusling JF. Wiring of enzymes to electrodes by ultrathin conductive polyion underlayers: enhanced catalytic response to hydrogen peroxide. Anal Chem. 2003;75(17):4565-71.
- [146]146. Ishihara K, Tanaka S, Furukawa N, Kurita K, Nakabayashi N. Improved blood compatibility of segmented polyurethanes by polymeric additives having phospholipid polar groups. I. Molecular design of polymeric additives and their functions. J Biomed Mater Res. 1996;32(3):391-9.
- [157]157. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. Engineering the tissue which encapsulates subcutaneous implants. I. Diffusion properties. J Biomed Mater Res. 1997;37(3):401-12.
- [158]158. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. Engineering the tissue which encapsulates subcutaneous implants. III. Effective tissue response times. J Biomed Mater Res. 1998;40(4):598-605.
- [159]159. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. Engineering the tissue which encapsulates subcutaneous implants. II. Plasmatissue exchange properties. J Biomed Mater Res. 1998;40(4):586-97.
- [174]174. Guto PM, Kumar CV, Rusling JF. Thermostable peroxidasepolylysine films for biocatalysis at 90 degrees C. J Phys Chem B. 2007;111(30):9125-31.
- [187]187. Renard E, Shah R, Miller M, Kolopp M, Costalat G, Bringer J. Sustained safety and accuracy of central IV glucose sensors connected to implanted insulin pumps and short-term closed-loop trials in diabetic patients. Diabetologia. 2003;46:A47.
Cited by
Sorted by citations
Yetisen, A. K., Martinez‐Hurtado, J. L., Ünal, B., Khademhosseini, A., & Butt, H. (2018) · Advanced Materials · doi:10.1002/adma.201706910 · Cited by 569 · see citing works
Cappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Continuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320 · see citing works
Smart Devices for Older Adults Managing Chronic Disease: A Scoping Review
Kim, B. Y., & Lee, J. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.7141 · Cited by 217 · see citing works
Xie, X., Doloff, J. C., Yesilyurt, V., Sadraei, A., McGarrigle, J. J., Omami, M., Veiseh, O., Farah, S., Isa, D., Ghani, S., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0273-3 · Cited by 235 · see citing works
Munje, R. D., Muthukumar, S., & Prasad, S. (2017) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2016.07.088 · Cited by 190 · see citing works
Device and method for noninvasive glucose assessment
Segman, Y. (. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818763457 · Cited by 55 · see citing works
Sakaguchi, K., Hirota, Y., Hashimoto, N., Ogawa, W., Hamaguchi, T., Matsuo, T., Miyagawa, J.-I., Namba, M., Sato, T., Okada, S., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700313 · Cited by 50 · see citing works
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46 · see citing works
A Head-to-Head Comparison Study of the First-Day Performance of Two Factory-Calibrated CGM Systems
Denham, D. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819895505 · Cited by 18 · see citing works
- Authors
- Vaddiraju, S., Burgess, D. J., Tomazos, I., Jain, F. C., & Papadimitrakopoulos, F.
- Affiliations
- University of Connecticut (US)
- Corresponding author
- Papadimitrakopoulos, Fotios
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-11
- Volume
- 4
- Issue
- 6
- Pages
- 1540-1562
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3005068
- OpenAlex
- W2161313513
- Cited by
- 285
- References
- 181
- Citations by year
- 2026: 19 · 2025: 14 · 2024: 9 · 2023: 17 · 2022: 12 · 2021: 15 · 2020: 8 · 2019: 24 · 2018: 25 · 2017: 19 · 2016: 27 · 2015: 24 · 2014: 22 · 2013: 25 · 2012: 14
- FWCI
- 12.10
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Heart, Lung, and Blood Institute (1-R21-HL090458-01, R21HL090458)
- National Institute of Environmental Health Sciences (ES013557, R01ES013557)
- National Institute of Biomedical Imaging and Bioengineering (R43EB011886)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Surgery › Pancreatic function and diabetes
- Keywords
- continuous glucose monitoring, diabetes management, artificial pancreas, glucose sensor, hypoglycemia detection, hyperglycemia, non-invasive glucose monitoring, electrochemical glucose sensor, optical glucose sensor, piezoelectric sensors, foreign body reaction, sensor calibration, device miniaturization, sensor accuracy, long-term reliability
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Equipment Design, Forecasting, Humans, Insulin Infusion Systems, Miniaturization, Predictive Value of Tests, Signal Processing, Computer-Assisted, Time Factors, Reproducibility of Results, Biomarkers, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review