Analytical Performance of Glucose Monitoring Systems at Different Blood Glucose Ranges and Analysis of Outliers in a Clinical Setting
Open accessThe DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
Background: We investigated the analytical accuracy of 27 glucose monitoring systems (GMS) in a clinical setting, using the new ISO accuracy limits. In addition to measuring accuracy at blood glucose (BG) levels < 100 mg/dl and > 100 mg/dl, we also analyzed devices performance with respect to these criteria at 5 specific BG level ranges, making it possible to further differentiate between devices with regard to overall performance. Methods: Carbohydrate meals and insulin injections were used to induce an increase or decrease in BG levels in 37 insulin-dependent patients. Capillary blood samples were collected at 10-minute intervals, and BG levels determined simultaneously using GMS and a laboratory-based method. Results obtained via both methods were analyzed according to the new ISO criteria. Results: Only 12 of 27 devices tested met overall requirements of the new ISO accuracy limits. When accuracy was assessed at BG levels < 100 mg/dl and > 100 mg/dl, criteria were met by 14 and 13 devices, respectively. A more detailed analysis involving 5 different BG level ranges revealed that 13 (48.1%) devices met the required criteria at BG levels between 50 and 150 mg/dl, whereas 19 (70.3%) met these criteria at BG levels above 250 mg/dl. The overall frequency of outliers was low. Conclusions: The assessment of analytical accuracy of GMS at a number of BG level ranges made it possible to further differentiate between devices with regard to overall performance, a process that is of particular importance given the user-centered nature of the devices’ intended use.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Nevertheless, several evaluation or surveillance studies within recent years repeatedly revealed that a non-negligible number of CE-marked BGMS, which do not fulfill the minimum accuracy criteria of ISO 15197, had been or currently still are on sale on the European market.12–14”
From System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015 (2020) · cites this paper as [14]
Citations
Cites
International Organization for Standardization
(2015) · Encyclopedia of Astrobiology · doi:10.1007/978-3-662-44185-5_100574 · Cited by 6,716
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348 · cited 1×
“Inaccuracies might furthermore be explained by errors and differences resulting from the type of reference method used for device calibration.(17,18) The current study’s clinical setting might be regarded as a limiting factor since accuracy testing is usually conducted in accordance with the relevant ISO standard, and…”
MethodFreckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237 · cited 2×
“9%) reached the minimum accuracy requirement if the new ISO criteria were applied.(12) In our study using a clinical setting, only 12 devices (44%) met these requirements, which is in line with Freckmann et al’s results and confirms our previous study with a smaller number of devices.”
ResultBaumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., & Freckmann, G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600511 · Cited by 90
Accuracy of home blood glucose meters during hypoglycemia
Trajanoski, Z., Brunner, G., Gfrerer, R., Wach, P., & Pieber, T. (1996) · Diabetes Care · doi:10.2337/diacare.19.12.1412 · Cited by 85 · cited 1×
“This lack of accuracy in the lower BG ranges has been found also in previous studies.(14,15) Given that all testing was conducted by trained staff, patient error can be excluded as a contributory factor.”
ResultPlasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 74 · cited 1×
“7% higher than values measured by the glucose oxidase method.(10,11) For the assessment of the Accu-Chek devices, BG values of the reference method were therefore adjusted by +5%.”
MethodKlonoff, D. C. (2004) · Diabetes Care · doi:10.2337/diacare.27.3.834 · Cited by 56 · cited 1×
“It has, therefore, been suggested that the analytical accuracy of GMS should be tested not only in relation to BG level ranges below and above 100 mg/dL but also in relation to different glycemic ranges, as already previously proposed for continuous glucose monitoring systems.(7) The aim of the present study was to an…”
MethodKlonoff, D. C. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400301 · Cited by 37 · cited 8×
“472 Journal of Diabetes Science and Technology 8(3)”
BackgroundMethodInfluential(1994) · doi:10.2337/diacare.17.6.595 · Cited by 32
Hasslacher, C., Kulozik, F., & Platten, I. (2013) · Clinical Laboratory · doi:10.7754/clin.lab.2012.120710 · Cited by 26 · cited 1×
“In our study using a clinical setting, only 12 devices (44%) met these requirements, which is in line with Freckmann et al’s results and confirms our previous study with a smaller number of devices.(13) Responses to the consultation process on the new ISO criteria included the suggestion that instead of testing accura…”
ResultSelf-citation
Research works printed but not linked: 2
- International Organization for Standardization, 2003, DIN EN ISO 15197
- Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen, 2008, Dtsch Arztebl, 105, A341
Cited by
Bailey, T. S., Grunberger, G., Bode, B. W., Handelsman, Y., Hirsch, I. B., Jovanovič, L., Roberts, V. L., Rodbard, D., Tamborlane, W. V., & Walsh, J. S. A. (2016) · Endocrine Practice · doi:10.4158/ep151124.cs · Cited by 151
Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815580160 · Cited by 70
Pleus, S., Baumstark, A., Jendrike, N., Mende, J., Link, M., Zschornack, E., Haug, C., & Freckmann, G. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-001067 · Cited by 63
“14 Regarding system accuracy, the minimal demands of ISO 15197 are at least 95% of a system’s measurement results shall be within ±15 mg/dL (0.83 mmol/L) of the reference measurement procedure’s results at BG concentrations 100 mg/dL (5.55 mmol/L) and within 15% at BG concentrations ≥ 100 mg/dL…”
InfluentialPerformance of Cleared Blood Glucose Monitors
Klonoff, D. C., & Prahalad, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815584797 · Cited by 59
Comparative assessment of blood glucose monitoring techniques: a review
Ahmadian, N., Manickavasagan, A., & Ali, A. (2022) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2022.2100496 · Cited by 51
“Regular monitoring and evaluation of these blood glucometers should therefore be performed to maximise the patient’s safety [15,16] (Tables 1 and 2).”
BackgroundFreckmann, G., Link, M., Schmid, C., Pleus, S., Baumstark, A., & Haug, C. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0085 · Cited by 46
A review of blood glucose monitor accuracy
King, F., Ahn, D., Hsiao, V., Porco, T., & Klonoff, D. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0232 · Cited by 37
(2015) · doi:10.1177/1932296815580161 · Cited by 31
Freckmann, G., Jendrike, N. R. M., Baumstark, A., Pleus, S., Liebing, C., & Haug, C. (2018) · Diabetes Therapy · doi:10.1007/s13300-018-0392-6 · Cited by 25
Freckmann, G., Baumstark, A., Jendrike, N. R. M., Rittmeyer, D., Pleus, S., & Haug, C. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0341 · Cited by 23
Vanderlaan, E. L., Nolan, J. K., Sexton, J., Evans‐Molina, C., Lee, H., & Voytik‐Harbin, S. L. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115409 · Cited by 31
Yu-Fei, W., Wei-Ping, J., Ming-Hsun, W., Miao-O, C., Ming-Chang, H., Chi-Pin, W., & Ming-Shih, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817705143 · Cited by 15
Demircik, F., Klonoff, D. C., Musholt, P. B., Ramljak, S., & Pfützner, A. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0202 · Cited by 7
“…on dynamic elec-trochemistry, such as OneTouch (cid:2) Verio Pro (cid:2) (LifeScan) or BG*Star (cid:2) (Sanofi), were reported with unfavorable accuracy results by one group with a testing protocol developed in-house, 5,6 but with excellent results by our and other groups using the ISO protocol.”
Performance of two updated blood glucose monitoring systems: an evaluation following ISO 15197:2013
Pleus, S., Baumstark, A., Rittmeyer, D., Jendrike, N. R. M., Haug, C., & Freckmann, G. (2016) · Current Medical Research and Opinion · doi:10.1185/03007995.2016.1146666 · Cited by 6
Setford, S. J., Smith, A. F., McColl, D., Grady, M., Koria, K., & Cameron, H. (2015) · Expert Review of Medical Devices · doi:10.1586/17434440.2015.1102049 · Cited by 6
Jeanny, G., & Hope, P. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815608871 · Cited by 5
“This is a 1-time process, and postmarket release evaluations or independent evaluations are not mandatory, despite studies repeatedly showing test strip lots of available systems not adhering with ISO 15197 accuracy criteria.(3-6) In 2 studies, performed in April and May 2013 and in October 2014, the SMBG systems Gluc…”
MethodLuppa, P. B., Zeller, M., Pieper, M., Kaiser, P., Weiss, N., Vierbaum, L., & Freckmann, G. (2024) · Frontiers in Molecular Biosciences · doi:10.3389/fmolb.2024.1371426 · Cited by 4
Pleus, S., Flacke, F., Sieber, J., Haug, C., & Freckmann, G. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817707292 · Cited by 4
Accuracy Evaluation of Five Blood Glucose Monitoring Systems in Patients from Different Departments
Li, X., Zheng, S., Pan, S., Yan, T., Di, J., Yang, Y., & Cheng, J. (2019) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/a-0855-3726 · Cited by 5
Moriuchi, T., Otaki, Y., Satou, H., Chai, F., Hayashida, Y., Aikawa, R., Sugiyama, T., & Sode, K. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211060865 · Cited by 2
Chandnani, S. R., Ramakrishna, C., Dave, B., Kothavade, P. S., & Thakkar, A. S. (2017) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2017/26596.9892 · Cited by 4
Variaciones en la glucemia basal en la aplicación de técnicas osteopáticas.
Villa, C. V. (2017) · RIUCAM
Hypoglycemia Detection in Diabetes
Richardson, J. M., & Shaginian, R. M. (2022) · IntechOpen eBooks · doi:10.5772/intechopen.103137
Baghous, H., Aouiche, S., & AZZOUG, S. (2026) · IOSR Journal of Dental and Medical Sciences · doi:10.9790/0853-2503024960
Metadata
- Authors
- Hasslacher, C., Kulozik, F., & Platten, I.
- Affiliations
- Diabetesinstitut Heidelberg (DE)
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-02-18
- Volume
- 8
- Issue
- 3
- Pages
- 466-472
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4455429
- OpenAlex
- W2163658918
- Cited by
- 22
- References
- 16
- Citations by year
- 2024: 1 · 2023: 1 · 2022: 2 · 2021: 1 · 2020: 1 · 2019: 1 · 2018: 2 · 2017: 4 · 2016: 3 · 2015: 6
- FWCI
- 3.49
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- blood glucose, analytical accuracy, type 1 diabetes, capillary blood glucose, outlier analysis, device performance comparison, hyperglycemia
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Dietary Carbohydrates, Equipment Design, Female, Humans, Hypoglycemic Agents, Insulin, Male, Materials Testing, Middle Aged, Predictive Value of Tests, Time Factors, Blood Glucose Self-Monitoring, Reproducibility of Results, Biomarkers, Treatment Outcome
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.