Time Delay of CGM Sensors: Relevance, Causes, and Countermeasures
Open accessThe DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Abstract
Background: Continuous glucose monitoring (CGM) is a powerful tool to support the optimization of glucose control of patients with diabetes. However, CGM systems measure glucose in interstitial fluid but not in blood. Rapid changes in one compartment are not accompanied by similar changes in the other, but follow with some delay. Such time delays hamper detection of, for example, hypoglycemic events. Our aim is to discuss the causes and extent of time delays and approaches to compensate for these. Methods: CGM data were obtained in a clinical study with 37 patients with a prototype glucose sensor. The study was divided into 5 phases over 2 years. In all, 8 patients participated in 2 phases separated by 8 months. A total number of 108 CGM data sets including raw signals were used for data analysis and were processed by statistical methods to obtain estimates of the time delay. Results: Overall mean (SD) time delay of the raw signals with respect to blood glucose was 9.5 (3.7) min, median was 9 min (interquartile range 4 min). Analysis of time delays observed in the same patients separated by 8 months suggests a patient dependent delay. No significant correlation was observed between delay and anamnestic or anthropometric data. The use of a prediction algorithm reduced the delay by 4 minutes on average. Conclusions: Prediction algorithms should be used to provide real-time CGM readings more consistent with simultaneous measurements by SMBG. Patient specificity may play an important role in improving prediction quality.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Coherence and time delay estimation
Carter, G. C. (1987) · Proceedings of the IEEE · doi:10.1109/proc.1987.13723 · Cited by 1,200 · cited 1×
“Other such procedures aiming for the same goal have been proposed; however, their major disadvantage is that they are based on postprocessing of measured data and do not provide any help with real-time data.(21) This would, however, be necessary for clinical practice.”
BackgroundDiscrete-time signal processing
Poularikas, A. D., & Ramadan, Z. M. (2017) · Adaptive Filtering Primer With Matlab® · doi:10.1201/9781315221946-2 · Cited by 591
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
Continuous Glucose Monitoring Systems: A Review
Vashist, S. (2013) · Diagnostics · doi:10.3390/diagnostics3040385 · Cited by 261 · cited 1×
“In addition, CGM systems help avoid hypoglycemic events by providing alarms when a predefined glucose threshold is exceeded or even prior to this if a pending hypoglycemia is predicted.(1,2) One reason why CGM data should not be used for determining insulin doses by patients with diabetes, according to the officially…”
BackgroundPhysiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 251
Time Lag of Glucose From Intravascular to Interstitial Compartment in Humans
Basu, A., Dube, S., Slama, M., Errazuriz, I., Amezcua, J. C., Kudva, Y. C., Peyser, T., Carter, R. E., Cobelli, C., & Basu, R. (2013) · Diabetes · doi:10.2337/db13-1132 · Cited by 205 · cited 1×
“A more recent estimation based on glucose tracer experiments with healthy humans using a microdialysis technique resulted in an average physiologic time delay of 5-6 minutes.(7) Other estimates using the same approach for a very small study population report a similar value for patients with type 1 diabetes.”
MethodDelays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300528 · Cited by 198 · cited 2×
“Here again only little information is available about the impact of all these time-consuming factors on the overall time delay of commercial CGM systems; thus far, values of 3 to 12 minutes have been reported.(9,16) The trade-off between variance of the sensor readings and time delay introduced by filtering can be app…”
BackgroundMethodContinuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
Bequette, B. W. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400222 · Cited by 192 · cited 1×
“Filtering is another possibility for reducing the noise of the sensor signal and use of additional information increases the speed of this compensation.(11,12) However, mathematical filters applied in real time to provide reliable glucose information unavoidably generate an algorithmic time delay (since real time filt…”
BackgroundTime Lag of Glucose From Intravascular to Interstitial Compartment in Type 1 Diabetes
Basu, A., Dube, S., Veettil, S., Slama, M., Kudva, Y. C., Peyser, T., Carter, R. E., Cobelli, C., & Basu, R. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814554797 · Cited by 164 · cited 1×
“Other estimates using the same approach for a very small study population report a similar value for patients with type 1 diabetes.(8) There is strong evidence for a great variability of the time delay between individual subjects even under reproducible experimental conditions and independently of a specific CGM syste…”
ResultRebrin, K., Sheppard, N. F., & Steil, G. M. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400507 · Cited by 152 · cited 1×
“Here again only little information is available about the impact of all these time-consuming factors on the overall time delay of commercial CGM systems; thus far, values of 3 to 12 minutes have been reported.(9,16) The trade-off between variance of the sensor readings and time delay introduced by filtering can be app…”
MethodUniversal Glucose Models for Predicting Subcutaneous Glucose Concentration in Humans
Gani, A., Gribok, A. V., Lu, Y., Ward, W. K., Vigersky, R. A., & Reifman, J. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2034141 · Cited by 124
Development of a Neural Network for Prediction of Glucose Concentration in Type 1 Diabetes Patients
Pappada, S. M., Cameron, B. D., & Rosman, P. M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200507 · Cited by 124
Graphical and Numerical Evaluation of Continuous Glucose Sensing Time Lag
Kovatchev, B. P., Shields, D., & Breton, M. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0044 · Cited by 91 · cited 1×
“The number of publications in which data about the time delays observed with different CGM systems are reported, favorably from head-to-headcomparisons, is quite limited.(17,18) For patients with diabetes who use a given CGM system, the combined effect of time delays and measurement errors is of concern as both factor…”
BackgroundPalerm, C. C., & Bequette, B. W. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100505 · Cited by 89 · cited 1×
“In addition, CGM systems help avoid hypoglycemic events by providing alarms when a predefined glucose threshold is exceeded or even prior to this if a pending hypoglycemia is predicted.(1,2) One reason why CGM data should not be used for determining insulin doses by patients with diabetes, according to the officially…”
BackgroundA meta-learning approach to the regularized learning—Case study: Blood glucose prediction
Naumova, V., Pereverzev, S. V., & Sivananthan, S. (2012) · Neural Networks · doi:10.1016/j.neunet.2012.05.004 · Cited by 73
Metabolic Biofouling of Glucose Sensors in Vivo : Role of Tissue Microhemorrhages
Klueh, U., Liu, Z., Feldman, B., Henning, T. P., Cho, B., Ouyang, T., & Kreutzer, D. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500313 · Cited by 63
Favero, S. D., Facchinetti, A., Sparacino, G., & Cobelli, C. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2293531 · Cited by 60 · cited 1×
“In addition, algorithms are employed to smooth the glucose profile over time and to reduce and/or eliminate artifacts (socalled smart sensors).(14,15) Here again only little information is available about the impact of all these time-consuming factors on the overall time delay of commercial CGM systems; thus far, valu…”
MethodFacchinetti, A., Sparacino, G., & Cobelli, C. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2161083 · Cited by 58 · cited 1×
“In addition, algorithms are employed to smooth the glucose profile over time and to reduce and/or eliminate artifacts (socalled smart sensors).(14,15) Here again only little information is available about the impact of all these time-consuming factors on the overall time delay of commercial CGM systems; thus far, valu…”
MethodWentholt, I. M. E., Hart, A. A. M., Hoekstra, J. B. L., & Devries, J. H. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0007 · Cited by 54
Evaluation of the Performance of a Novel System for Continuous Glucose Monitoring
Zschornack, E., Schmid, C., Pleus, S., Link, M., Klötzer, H.-M., Obermaier, K., Schoemaker, M., Strasser, M., Frisch, G., Schmelzeisen-Redeker, G., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700403 · Cited by 54 · cited 2×
“Details of the study design have been reported elsewhere.(19) A total of 37 patients with type 1 diabetes participated in the clinical study (see Table 1 for patient characteristics).”
BackgroundMethodSelf-citationNielsen, J. K., Djurhuus, C. B., Gravholt, C. H., Carus, A. C., Granild-Jensen, J., Ørskov, H., & Christiansen, J. S. (2005) · Diabetes · doi:10.2337/diabetes.54.6.1635 · Cited by 53 · cited 1×
“ISF glucose levels are closer to those in the brain than the results of BG measurements.(22) As the brain is the clinically relevant organ (glucose levels in the brain are critical for preventing, eg, hypoglycemic symptoms), it represents the ultimate target for blood glucose control.”
ResultContinuous-time interval model identification of blood glucose dynamics for type 1 diabetes
Kirchsteiger, H., Johansson, R., Renard, É., & del Re, L. (2014) · International Journal of Control · doi:10.1080/00207179.2014.897004 · Cited by 44
Self-citationOverview of a Novel Sensor for Continuous Glucose Monitoring
Schmelzeisen-Redeker, G., Staib, A., Strasser, M., Müller, U., & Schoemaker, M. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700402 · Cited by 43
Self-citationIdentification of empirical dynamic models from type 1 diabetes subject data
Finan, D. A., Palerm, C. C., Doyle, F. J., Zisser, H., Jovanovič, L., Bevier, W. C., & Seborg, D. E. (2008) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2008.4586802 · Cited by 28
Scuffi, C., Lucarelli, F., & Valgimigli, F. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600618 · Cited by 27 · cited 3×
“A related analysis of the effect of time delays on the CGM performance was also presented in Scuffi et al.6 To separate the effects of time delays from that of analytical error, 1 guideline for the evaluation of CGM systems issued by the Clinical and Laboratory Standard Institute (CLSI)20 suggests procedures to retros…”
MethodInfluentialWard, W. K., Engle, J. M., Branigan, D. L., Youssef, J. E., Massoud, R. G., & Castle, J. R. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03545.x · Cited by 20
Cited by
Lin, S., Cheng, X., Zhu, J., Wang, B., Jelinek, D., Zhao, Y., Wu, T.-Y., Horrillo, A., Tan, J., Yeung, J., et al. (2022) · Science Advances · doi:10.1126/sciadv.abq4539 · Cited by 255
Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202
Self-citationHeinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
“Not much is known about the impact of algorithms and filtering on the MARD as the manufacturers of these devices regard these data and computational improvement opportunities as proprietary intellectual property, and the raw data, the sensor current, is rarely if ever available to outside parties or the end users.(3)…”
BackgroundMethodSelf-citationLongo, R. R., Elias, H., Khan, M., & Seley, J. J. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211008446 · Cited by 89
Standardization process of continuous glucose monitoring: Traceability and performance
Freckmann, G., Nichols, J. H., Hinzmann, R., Klonoff, D. C., Ju, Y., Diem, P., Makris, K., & Slingerland, R. J. (2021) · Clinica Chimica Acta · doi:10.1016/j.cca.2020.12.025 · Cited by 62
Self-citation
Metadata
- Authors
- Schmelzeisen-Redeker, G., Schoemaker, M., Kirchsteiger, H., Freckmann, G., Heinemann, L., & del Re, L.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2015-08-04
- Volume
- 9
- Issue
- 5
- Pages
- 1006-1015
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W1695862053
- Cited by
- 161
- References
- 26
- FWCI
- 6.58
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.