Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison

Open access
Freckmann, Guido, Link, Manuela, Kamecke, Ulrike, Haug, Cornelia, Baumgartner, Bernhard, Weitgasser, Raimund · 2019 · Journal of Diabetes Science and Technology
Cited by 35

The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.

Abstract

Background: To be able to compare continuous glucose monitoring (CGM) systems, they have to be worn in parallel by the same subjects. This study evaluated the performance and usability of three different CGM systems in direct comparison. Method: In this open, prospective study at two sites, 54 patients with diabetes wore three CGM systems each (Dexcom G5™ Mobile CGM system [DG5], Guardian™ Connect system [GC], and a Roche CGM system [RCGM]) in parallel for 6 or 7 days in a mixed inpatient and outpatient setting. Capillary comparison measurements were performed using a self-monitoring of blood glucose (SMBG) system. During study site visits, glucose excursions were induced. Performance of the systems was evaluated by calculating mean absolute relative differences (MARD, calculated as absolute differences for glucose concentrations <100 mg/dL and as relative differences for glucose concentrations ≥100 mg/dL), and mean relative differences (MRD, bias) between CGM and SMBG results. In addition, usability of the systems was assessed. Results: Overall MARD was 10.1 ± 2.1 for DG5, 11.5 ± 4.2 for GC, and 11.9 ± 5.6 for RCGM. Performance improved in all systems after the first day of use. All systems showed >99% of values within zones A and B of the consensus error grid. Overall, all CGM systems showed a small negative bias compared to SMBG. Usability of the systems differed regarding patch adhesion rate, failure rate, and patient rating. Most patients preferred GC, but in general all systems were rated positively. Conclusion: All three CGM systems showed similar overall accuracy in this direct comparison, but small differences were observed with regard to specific glucose ranges and usability aspects.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/1932296819826965

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 7 references

Cited by

See all 35

Metadata

Authors
Freckmann, G., Link, M., Kamecke, U., Haug, C., Baumgartner, B., & Weitgasser, R.
Venue
Journal of Diabetes Science and Technology
Published
2019-02-07
Volume
13
Issue
5
Pages
890-898
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2911681618
Cited by
35
References
7
FWCI
3.10
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison. https://app.cochat.ai/papers/10.1177/1932296819826965
The paper
Freckmann, G., Link, M., Kamecke, U., Haug, C., Baumgartner, B., & Weitgasser, R. (2019). Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296819826965

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.