Toward Detection of Nocturnal Hypoglycemia in People With Diabetes Using Consumer-Grade Smartwatches and a Machine Learning Approach
Open accessAbstract
Background: Nocturnal hypoglycemia poses significant risks to individuals with insulin-treated diabetes, impacting health and quality of life. Although continuous glucose monitoring (CGM) systems reduce these risks, their poor accuracy at low glucose levels, high cost, and availability limit their use. This study examined physiological biomarkers associated with nocturnal hypoglycemia and evaluated the use of machine learning (ML) to detect hypoglycemia during nighttime sleep using data from consumer-grade smartwatches. Methods: This study analyzed 351 nights of 36 adults with insulin-treated diabetes. Participants wore two smartwatches alongside CGM systems. Linear mixed-effects models compared sleep and vital signs between nights with and without hypoglycemia during early and late sleep. A ML model was trained to detect hypoglycemia solely using smartwatch data. Results: Sixty-six nights with spontaneous hypoglycemia were recorded. Hypoglycemic nights showed increased wake periods, heart rate, stress levels, and activity during early sleep, with weaker effects during late sleep. In nights when hypoglycemia occurred during early sleep, the ML model performed comparable or better than prior studies with an area under the receiver operator curve of 0.78 for level 1 and 0.83 for level 2 hypoglycemia, with sensitivity of 0.78 and 0.89, specificity of 0.64 for both, negative predictive value of 0.94 and 0.99, and positive predictive value of 0.25 and 0.13 for level 1 and level 2 hypoglycemia, respectively. Conclusions: Consumer-grade smartwatches demonstrate promise for detecting nocturnal hypoglycemia, particularly during early sleep. Refining models to reduce false alarms could enhance their clinical utility as low-cost, accessible tools to complement CGM.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Features of heart rate, heart rate variability, movement, and electrodermal activity extracted from wearable data have been applied to detect hypoglycemic events in cohorts ranging from 1 to 36 individuals.
“Features of HR, HRV, movement, and electrodermal activity (EDA; a measure of perspiration) extracted from wearable data have been applied to detect hypoglycemic events in cohorts ranging from 1 to 36 individuals5,26–28.”
From Sleep and temperature data from wearable devices support noninvasive detection of diabetes mellitus in a large-scale, retrospective analysis (2026) · cites this paper as [28]
The Results (Baseline characteristics; page unavailable) report 36 participants, and Methods (Data preprocessing; page unavailable) describe heart-rate, heart-rate-variability, and movement data used for hypoglycemia detection. The Introduction discusses other wearable studies using electrodermal activity, but this study did not use it, and our copy does not establish the lower bound of one individual. (LLM, conf=0.84)
Experimental — accuracy not yet measuredPrior work was generated from small cohorts (the largest cohort of 70 individuals), raising questions of the generalizability of their findings.
“Second, prior work was generated from small cohorts (the largest cohort of 70 individuals), that raise questions of the generalizability of their findings5,22–32.”
From Sleep and temperature data from wearable devices support noninvasive detection of diabetes mellitus in a large-scale, retrospective analysis (2026) · cites this paper as [28]
The Results (Baseline characteristics; page unavailable) report 36 participants in this study’s primary analysis, supporting its inclusion as an example of work based on a small cohort. Our copy does not establish that 70 was the largest cohort across the cited prior work or explicitly link those cohort sizes to concerns about generalizability. (LLM, conf=0.84)
Experimental — accuracy not yet measuredPrior work chiefly assessed statistical aggregation functions of wearable data streams (primary features), which are nonspecific to the effects of DM within data streams that are constantly fluctuating in natural settings.
“Third, prior work chiefly assessed statistical aggregation functions of wearable data streams (primary features)5,26–32, which are nonspecific to the effects of DM within data streams that are constantly fluctuating in natural settings.”
From Sleep and temperature data from wearable devices support noninvasive detection of diabetes mellitus in a large-scale, retrospective analysis (2026) · cites this paper as [28]
The Methods > Data preprocessing section describes derived HRV parameters, accumulated accelerometer activity, and resampling of wearable data into one-minute intervals. That supports the narrower point that this study used processed wearable features. Our copy does not establish that prior work chiefly assessed statistical aggregation functions or that those features are nonspecific to diabetes-related effects in fluctuating natural settings. No page number is available in our copy. (LLM, conf=0.72)
Experimental — accuracy not yet measuredPrior work typically uses less than a day of input.
“First, where prior work typically uses less than a day of input5,22–32, we assess the impact of longitudinal input windows ranging up to multiple weeks and demonstrate improved sensitivity with longer windows (AUROC increases by 0.15 from 1-night to 21-night models, in balanced and imbalanced cohorts).”
From Sleep and temperature data from wearable devices support noninvasive detection of diabetes mellitus in a large-scale, retrospective analysis (2026) · cites this paper as [28]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069 · see citing works
Machine learning algorithm validation with a limited sample size
Vabalas, A., Gowen, E., Poliakoff, E., & Casson, A. J. (2019) · PLOS ONE · doi:10.1371/journal.pone.0224365 · Cited by 1,901 · see citing works
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831 · see citing works
Epidemiology of severe hypoglycemia in the diabetes control and complications trial
(1991) · The American Journal of Medicine · doi:10.1016/0002-9343(91)80085-z · Cited by 807 · see citing works
Heinemann, L., Freckmann, G., Ehrmann, D., Faber-Heinemann, G., Guerra, S., Waldenmaier, D., & Hermanns, N. (2018) · The Lancet · doi:10.1016/s0140-6736(18)30297-6 · Cited by 522 · see citing works
Continuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458 · see citing works
Decreased Epinephrine Responses to Hypoglycemia during Sleep
Jones, T. W., Porter, P., Sherwin, R. S., Davis, E. A., O'Leary, P. C., Frazer, F., Byrne, G. C., Stick, S. M., & Tamborlane, W. V. (1998) · New England Journal of Medicine · doi:10.1056/nejm199806043382303 · Cited by 298 · see citing works
(1991) · PubMed · Cited by 245 · see citing works
Induction of Hypoglycemia Unawareness by Asymptomatic Nocturnal Hypoglycemia
Veneman, T. F., Mitrakou, A., Mokáň, M., Cryer, P., & Gerich, J. E. (1993) · Diabetes · doi:10.2337/diab.42.9.1233 · Cited by 211 · see citing works
Continuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176 · see citing works
“Turn it Off!”: Diabetes Device Alarm Fatigue Considerations for the Present and the Future
Shivers, J. P., Mackowiak, L., Anhalt, H., & Zisser, H. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700324 · Cited by 159 · see citing works
Robinson, R. T. C. E., Harris, N. D., Ireland, R. H., Macdonald, I. A., & Heller, S. R. (2004) · Diabetologia · doi:10.1007/s00125-003-1292-4 · Cited by 137 · see citing works
Beregszàszi, M., Tubiana-Rufi, N., Benali, K., Noël, M., Bloch, J., & Czernichow, P. (1997) · The Journal of Pediatrics · doi:10.1016/s0022-3476(97)70121-5 · Cited by 122 · see citing works
DeSalvo, D. J., Noor, N., Xie, C., Corathers, S. D., Majidi, S., McDonough, R. J., Polsky, S., Izquierdo, R., Rioles, N., Weinstock, R., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211049783 · Cited by 113 · see citing works
Lee, T., Cho, Y., Cha, K. S., Jung, J., Cho, J., Kim, H., Kim, D., Hong, J., Lee, D., Keum, M., et al. (2023) · JMIR mhealth and uhealth · doi:10.2196/50983 · Cited by 113 · see citing works
The Future of Stress Management: Integration of Smartwatches and HRV Technology
Jerath, R., Syam, M., & Ahmed, S. (2023) · Sensors · doi:10.3390/s23177314 · Cited by 96 · see citing works
Cardiac implications of hypoglycaemia in patients with diabetes – a systematic review
Hanefeld, M., Dütting, E. D., & Bramlage, P. (2013) · Cardiovascular Diabetology · doi:10.1186/1475-2840-12-135 · Cited by 93 · see citing works
Cichosz, S. L., Frystyk, J., Hejlesen, O. K., Tarnow, L., & Fleischer, J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814528838 · Cited by 88 · see citing works
The risks of nocturnal hypoglycaemia in insulin-treated diabetes
Graveling, A. J., & Frier, B. M. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.012 · Cited by 73 · see citing works
FLIRT: A feature generation toolkit for wearable data
Föll, S., Maritsch, M., Spinola, F., Mishra, V., Barata, F., Kowatsch, T., Fleisch, E., & Wortmann, F. (2021) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2021.106461 · Cited by 72 · see citing works
Characterizing Accuracy and Precision of Glucose Sensors and Meters
Rodbard, D. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814541810 · Cited by 65 · see citing works
Non-invasive hypoglycemia monitoring system using extreme learning machine for Type 1 diabetes
Ling, S. H., San, P. P., & Nguyen, H. T. (2016) · ISA Transactions · doi:10.1016/j.isatra.2016.05.008 · Cited by 62 · see citing works
Continuous glucose monitoring: quality of hypoglycaemia detection
Zijlstra, E., Heise, T., Nosek, L., Heinemann, L., & Heckermann, S. (2012) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12001 · Cited by 61 · see citing works
Olde Bekkink, M., Koeneman, M., de Galan, B. E., & Bredie, S. J. (2019) · Diabetes Care · doi:10.2337/dc18-1843 · Cited by 58 · see citing works
Henriksen, M. M., Andersen, H. U., Thorsteinsson, B., & Pedersen-Bjergaard, U. (2018) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2018-00142 · Cited by 57 · see citing works
Psychosocial Aspects of Continuous Glucose Monitoring
Kubiak, T., Mann, C. G., Barnard, K., & Heinemann, L. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816651450 · Cited by 49 · see citing works
Forlenza, G. P., Argento, N. B., & Laffel, L. M. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0034 · Cited by 47 · see citing works
Noninvasive Hypoglycemia Detection in People With Diabetes Using Smartwatch Data
Lehmann, V., Föll, S., Maritsch, M., van Weenen, E., Kraus, M., Lagger, S., Odermatt, K., Albrecht, C., Fleisch, E., Zueger, T., et al. (2023) · Diabetes Care · doi:10.2337/dc22-2290 · Cited by 47 · see citing works
Sleep and the response to hypoglycaemia.
Jauch-Chara, K., & Schultes, B. (2010) · Best practice & research. Clinical endocrinology & metabolism · doi:10.1016/j.beem.2010.07.006 · Cited by 46 · see citing works
Koivikko, M. L., Tulppo, M. P., Kiviniemi, A. M., Kallio, M., Perkiömäki, J. S., Salmela, P. I., Airaksinen, J. K. E., & Huikuri, H. V. (2012) · Diabetes Care · doi:10.2337/dc11-2120 · Cited by 46 · see citing works
Divilly, P., Martine‐Edith, G., Zaremba, N., Søholm, U., MAHMOUDI, Z., Cigler, M., Ali, N., Abbink, E. J., Brøsen, J. M. B., de Galan, B. E., et al. (2024) · Diabetes Care · doi:10.2337/dc23-2332 · Cited by 42 · see citing works
Awakening and Counterregulatory Response to Hypoglycemia during Early and Late Sleep
Jauch-Chara, K., Hallschmid, M., Gais, S., Oltmanns, K. M., Peters, A., Born, J., & Schultes, B. (2007) · Diabetes · doi:10.2337/db07-0044 · Cited by 35 · see citing works
Kruger, D. F., & Anderson, J. E. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0193 · Cited by 24 · see citing works
Siamashvili, M., Davis, H. A., & Davis, S. N. (2021) · Expert Review of Endocrinology & Metabolism · doi:10.1080/17446651.2021.1979391 · Cited by 23 · see citing works
(4461955), H. C., & Nagai, Y. (2012) · Springer eBooks · doi:10.1007/978-1-4419-1005-9_13 · Cited by 23 · see citing works
Avoiding Bias in Classification Accuracy - A Case Study for Activity Recognition
Koskimäki, H. (2015) · IEEE Symposium Series on Computational Intelligence (SSCI) · doi:10.1109/ssci.2015.52 · Cited by 15 · see citing works
Farrell, C. M., McNeilly, A. D., Hapca, S. M., & McCrimmon, R. J. (2020) · Diabetes Care · doi:10.2337/dc20-0882 · Cited by 15 · see citing works
Radan, I., Rajer, E., Bratina, N., Neubauer, D., Kržišnik, C., & Battelino, T. (2004) · Acta Diabetologica · doi:10.1007/s00592-004-0141-3 · Cited by 14 · see citing works
Bachmann, S., Auderset, A., Burckhardt, M., Szinnai, G., Hess, M., Zumsteg, U., Denhaerynck, K., & Donner, B. C. (2021) · Pediatric Diabetes · doi:10.1111/pedi.13262 · Cited by 13 · see citing works
Martine‐Edith, G., Divilly, P., Zaremba, N., Søholm, U., Broadley, M. M., Baumann, P. M., MAHMOUDI, Z., Gomes, M., Ali, N., Abbink, E. J., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0522 · Cited by 11 · see citing works
Smartwatches for non‐invasive hypoglycaemia detection during cognitive and psychomotor stress
Maritsch, M., Föll, S., Lehmann, V. F., Styger, N., Bérubé, C., Kraus, M., Feuerriegel, S., Kowatsch, T., Züger, T., Fleisch, E., et al. (2023) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15402 · Cited by 8 · see citing works
Kladov, D. E., Berikov, V., Semenova, J. F., & Klimontov, V. V. (2023) · Journal of Personalized Medicine · doi:10.3390/jpm13101454 · Cited by 6 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- [39]Ke G Meng Q Finley T, et al. LightGBM: a highly efficient gradient boosting decision tree. In: NIPS’17: Proceedings of the 31st International Conference on Neural Information Processing Systems. New York, NY: ACM; 2017: 3149- 3157.
- [40]Pedregosa F Varoquaux G Gramfort A, et al. Scikit-learn: machine learning in Python. Mach Learn Python. 2011; 12: 2825- 2830.
Cited by
Sorted by citations
A catecholamine-independent pathway controlling adaptive adipocyte lipolysis
Zhang, X., Panicker, S. S., Bollinger, J. M., Majumdar, A., Kheireddine, R., Dabill, L. F., Kim, C. T., Kleiboeker, B. A., Zhang, F., Chen, Y., et al. (2026) · Nature Metabolism · doi:10.1038/s42255-025-01424-5 · Cited by 9 · see citing works
Piersanti, A., Littero, M., Del Giudice, L. L., Marcantoni, I., Burattini, L., Tura, A., & Morettini, M. (2025) · Sensors · doi:10.3390/s25134038 · Cited by 5 · see citing works
Toward Affordable and Non-Invasive Detection of Hypoglycemia: A Machine Learning Approach
Obiuwevwi, L., Rechowicz, K. J., Ashok, V. G., & Jayarathna, S. (2025) · IEEE International Conference on Information Reuse and Integration (IRI) · doi:10.1109/iri66576.2025.00036 · Cited by 3 · see citing works
Hypoglycaemia and Cardiac Arrhythmias in Type 1 Diabetes Mellitus: A Mechanistic Review
Mavromoustakou, K., Fragoulis, C., Cholidou, K. G., Sotiropoulou, Z., Anagnostopoulos, N., Gastouniotis, I., Lontou, S.-P., Dimitriadis, K. S., Thanopoulou, A. C., Chrysohoou, C. A., et al. (2026) · Journal of Personalized Medicine · doi:10.3390/jpm16010045 · Cited by 3 · see citing works
Radar-Based Monitoring for Non-Contact Detection of Nocturnal Hypoglycemia in Diabetes: A Review
Ravaioli, C., Mendez, C., Gantner, S., Bader, S. B., & Witthauer, L. (2026) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2026.3663452 · Cited by 1 · see citing works
Self-citationHidden in the Night: Wearable Sleep Assessment of Nocturnal Hypoglycaemia in Type 1 DiabetesPreprint
Alsuhaymi, A., Nutter, P. W., Thabit, H., & Harper, S. (2026) · medRxiv · doi:10.64898/2026.01.22.26344161 · Cited by 1 · see citing works
Viswanath, V. K., Navaneethan, S., Burks, J. H., Soltani, S., Kasl, P., Hartogensis, W., Dilchert, S., Hecht, F. M., Mason, A. E., Wang, E. J., et al. (2026) · Communications Medicine · doi:10.1038/s43856-026-01501-0
ECGluFormer: glucose prediction from ECG via multi-loss, transformer-based aggregation
Tseng, M.-R., & Gutiérrez‐Osuna, R. (2025) · IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI ...) · doi:10.1109/bhi67747.2025.11269483
Artificial Intelligence-Based Depression Detection
Kiss, G. J., & Viktor, P. (2026) · Sensors · doi:10.3390/s26020748
Braun, M., Schoebel, C., & Dietz-Terjung, S. (2026) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-026-00547-5
- Authors
- Mendez, C., Kaykayoglu, C. A., Bähler, T., Künzler, J., Lizoain, A., Rothenbühler, M., Schmidt, M. H., Laimer, M., & Witthauer, L.
- Affiliations
- University of Bern (CH)
- Diabetes Center Berne (CH)
- Corresponding author
- Witthauer, Lilian
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2025-02-25
- Volume
- 20
- Issue
- 4
- Pages
- 1266-1275
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC11851596
- OpenAlex
- W4407923260
- Cited by
- 10
- References
- 43
- Citations by year
- 2026: 7 · 2025: 3
- FWCI
- 6.82
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Journal of Diabetes Science and Technology (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Open Access CRIS of the University of Bern
- Funders
- Vontobel-Stiftung (0847/2002)
- Diabetes Center Berne
- University of Bern
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Keywords
- nocturnal hypoglycemia, diabetes mellitus, insulin-treated diabetes, hypoglycemia detection, machine learning, sleep, heart rate, stress, physiological biomarkers, continuous glucose monitoring, low-cost monitoring
- MeSH terms
- Machine Learning, Continuous Glucose Monitoring, Digital Health, Predictive Learning Models, Adult, Aged, Blood Glucose, Diabetes Mellitus, Type 1, Female, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Male, Middle Aged, Sleep, Blood Glucose Self-Monitoring, Biomarkers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review