Exhaled breath condensate: methodological recommendations and unresolved questions
Open accessThe DOI matches its Crossref record; 281 of the 281 references it cites were found in a registry or an index.
Abstract
Collection of exhaled breath condensate (EBC) is a noninvasive method for obtaining samples from the lungs. EBC contains large number of mediators including adenosine, ammonia, hydrogen peroxide, isoprostanes, leukotrienes, nitrogen oxides, peptides and cytokines. Concentrations of these mediators are influenced by lung diseases and modulated by therapeutic interventions. Similarly EBC pH also changes in respiratory diseases. The aim of the American Thoracic Society/European Respiratory Society Task Force on EBC was to identify the important methodological issues surrounding EBC collection and assay, to provide recommendations for the measurements and to highlight areas where further research is required. Based on the currently available evidence and the consensus of the expert panel for EBC collection, the following general recommendations were put together for oral sample collection: collect during tidal breathing using a noseclip and a saliva trap; define cooling temperature and collection time (10 min is generally sufficient to obtain 1–2 mL of sample and well tolerated by patients); use inert material for condenser; do not use resistor and do not use filter between the subject and the condenser. These are only general recommendations and certain circumstances may dictate variation from them. Important areas for future research involve: ascertaining mechanisms and site of exhaled breath condensate particle formation; determination of dilution markers; improving reproducibility; employment of EBC in longitudinal studies; and determining the utility of exhaled breath condensate measures for the management of individual patients. These studies are required before recommending this technique for use in clinical practice.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“EBC was obtained approximately 70 min before swimming began and 35 min after swimming ended using an EcoScreen condenser (Jaeger GmbH, Würzburg, Germany) following American Thoracic Society/European Respiratory Society Task Force recommendations (Horvath et al. 2005).”
From Short-Term Changes in Respiratory Biomarkers after Swimming in a Chlorinated Pool (2010) · cites this paper as Horvath et al. 2005
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Brainard, D. H. (1997) · Spatial Vision · doi:10.1163/156856897x00357 · Cited by 19,702
The VideoToolbox software for visual psychophysics: transforming numbers into movies
Pelli, D. G. (1997) · Spatial Vision · doi:10.1163/156856897x00366 · Cited by 11,098
Mixed-Effects Models in S and S-PLUS
(2001) · Technometrics · doi:10.1198/tech.2001.s574 · Cited by 9,278
Best Linear Unbiased Estimation and Prediction under a Selection Model
Henderson, C. R. (1975) · Biometrics · doi:10.2307/2529430 · Cited by 2,497
Metabonomics: a platform for studying drug toxicity and gene function
Nicholson, J. K., Connelly, J., Lindon, J. C., & Holmes, E. (2002) · Nature Reviews Drug Discovery · doi:10.1038/nrd728 · Cited by 1,914
Nonlinear Mixed Effects Models for Repeated Measures Data
Lindstrom, M. J., & Bates, D. M. (1990) · Biometrics · doi:10.2307/2532087 · Cited by 1,795
That BLUP is a Good Thing: The Estimation of Random Effects
Robinson, G. K. (1991) · Statistical Science · doi:10.1214/ss/1177011926 · Cited by 1,686
Effects of antioxidant enzymes in the molecular control of reactive oxygen species toxicology
Matés, >. M. (2000) · Toxicology · doi:10.1016/s0300-483x(00)00306-1 · Cited by 1,518
Peroxynitrite-mediated tyrosine nitration catalyzed by superoxide dismutase
Ischiropoulos, H., Zhu, L., Chen, J., Tsai, M., Martin, J. C., Smith, C. D., & Beckman, J. S. (1992) · Archives of Biochemistry and Biophysics · doi:10.1016/0003-9861(92)90431-u · Cited by 1,495
Research works printed but not linked: 40
- 2002, Am J Respir Crit Care Med, 65, 663
- 1999, Eur Respir Rev, 68, 246
- 1985, Am Rev Respir Dis, 131, 894
- 2002, Eur J Med Res, 7, 353
- 2003, Eur J Med Res, 8, 505
- 1998, Monaldi Arch Chest Dis, 53, 268
- 1999, Eur Respir J, 14, 245
- 10.1002/1615-9861(200206)2:6<690::aid-prot690>3.0.co;2-6 · doi:10.1002/1615-9861(200206)2:6<690::aid-prot690>3.0.co;2-6
- 2000, Monaldi Arch Chest Dis, 55, 185
- 2003, Clin Lab, 49, 637
- 1997, Eur Respir J, 10, 1231
- 1986, Lancet, I, 11
- 1993, Anesth Analg, 77, 963
- 2003, Physiol Res, 52, 327
- 1995, Lancet, 346, 1235
- 10.1053/rmed.2001.1117 · DOI
- 1993, J Pharmacol Exp Ther, 267, 1529
- 1993, Am Rev Respir Dis, 15, 161
- 2002, Int Marit Health, 53, 93
- 1995, Appl Cardiopulmon Pathophysiol, 5, 215
- 10.1164/rccm.200206-507oc · DOI · Paper page
- 2003, Am J Respir Crit Care Med, 167, 91
- 2002, Int J Biol Markers, 17, 141
- 2001, J Heart Lung Transplant, 20, 184
- 1998, Pneumologie, 52, 426
- 2002, Eur J Med Res, 7, 290
- 2002, Curr Opin Pulm Med, 1, 25
- 2002, Int J Occup Med Environ Health, 15, 317
- 2003, Am J Respir Crit Care Med, 168, 1129
- 2004, Undersea Hyperb Med, 31, 217
- 2004, Physiol Res, 53, 561
- 2004, J Chromatogr B Analyt Technol Biomed Life Sci, 809, 199, 10.1016/s1570-0232(04)00513-6 · doi:10.1016/s1570-0232(04)00513-6
- 10.1007/s00408-004-2506-3 · DOI
- 10.1136/vr.155.23.744 · doi:10.1136/vr.155.23.744
- 2004, Thorax, 59, 1005
- 10.1002/rcm.1682 · DOI
- 2004, Physiol Res, 53, 561
- 10.1016/j.jfms.2003.12.002 · DOI
- 2004, Environ Health Perspect, 112, 1293, 10.1289/ehp.7108 · DOI
- 10.1016/s0956-5663(02)00083-0 · DOI · Paper page
Cited by
Items 501–600 of 1,000
Sun, C., Zhou, T., Xie, G., Fu, S., Gao, L., Liao, J., Wu, Y., & Wang, G. (2019) · Journal of Proteomics · doi:10.1016/j.jprot.2019.103392 · Cited by 26
Exhaled breath analysis in the diagnosis of head and neck cancer
Mäkitie, A. A., Almangush, A., Youssef, O., Metsälä, M., Renkonen, S., Nixon, I. J., Haigentz, M., Rodrigo, J. P., Saba, N. F., Poorten, V. V., et al. (2019) · Head & Neck · doi:10.1002/hed.26043 · Cited by 27
“Additionally, the american thoracic society/european respiratory society Task Force on EBC recommend using a nasal clip to minimize the entry of nasal airway lining fluid into the sampled air and to ensure that all exhaled air will pass through the mouth.(18) Oral bacteria and oral enzymatic activity can also influenc…”
BackgroundChen, Y.-C., & O’Hare, D. (2020) · The Analyst · doi:10.1039/c9an02438g · Cited by 26
Exhaled breath condensate pH in patients with cystic fibrosis
Antus, B., Barta, I., Csiszér, E., & Kelemen, K. (2012) · Inflammation Research · doi:10.1007/s00011-012-0508-9 · Cited by 26
Exhaled Breath Condensate pH Assays
Davis, M. D., & Hunt, J. (2012) · Immunology and Allergy Clinics of North America · doi:10.1016/j.iac.2012.06.003 · Cited by 26
“As part of the airway, the oropharynx likely can contribute to EBC pH when the sample is collected orally(22), but it does not in general Hunt Page 3”
BackgroundSelf-citationThe effect of allergic rhinitis on adenosine concentration in exhaled breath condensate
Vass, G., Huszár, É., Augusztinovicz, M., Baktai, G., Barát, E., Herjavecz, I., & Horváth, I. (2006) · Clinical & Experimental Allergy · doi:10.1111/j.1365-2222.2006.02496.x · Cited by 28
Self-citationAdvanced strategy for cancer detection based on volatile organic compounds in breath
Jia, Z., Jiang, Y., Shang, T., Cao, H., Li, J., Lin, C., Pu, P., Xu, H., Liu, Y., Huang, Y., et al. (2025) · Journal of Nanobiotechnology · doi:10.1186/s12951-025-03526-4 · Cited by 25
Lessons learned from metabolomics in cystic fibrosis
Muhlebach, M. S., & Sha, W. (2015) · Molecular and Cellular Pediatrics · doi:10.1186/s40348-015-0020-8 · Cited by 26
“potential influence of different collection devices [18].”
Sputum Eicosanoid Profiling in Exacerbations of Chronic Obstructive Pulmonary Disease
Drozdovszky, O., Barta, I., & Antus, B. (2014) · Respiration · doi:10.1159/000358099 · Cited by 26
Lin, J.-L., Bonnichsen, M. H., & Thomas, P. S. (2011) · Journal of Breath Research · doi:10.1088/1752-7155/5/1/011001 · Cited by 25
“determination of dilution markers [11].”
Acute changes in lung function associated with proximity to a steel plant: A randomized study
Dales, R., Kauri, L. M., Cakmak, S., Mahmud, M., Weichenthal, S., Ryswyk, K. V., Kumarathasan, P., Thomson, E. M., Vincent, R., Broad, G., et al. (2013) · Environment International · doi:10.1016/j.envint.2013.01.014 · Cited by 25
Increased of exhaled breath condensate neutrophil chemotaxis in acute exacerbation of COPD
Corhay, J. L., Moermans, C., Henket, M. A., Dang, D. N., Duysinx, B., & Louis, R. E. (2014) · Respiratory Research · doi:10.1186/s12931-014-0115-0 · Cited by 25
“Exhaled breath condensate (EBC) was collected by using a commercially available condenser (EcoScreen®, Erich Jaeger Viasys GmbH, Hoechberg, Germany), according to ATS/ERS recommendations [9], which yielded nongaseous components of the expiratory air.”
BackgroundMethodInfluentialMarek, E., Volke, J., Hawener, I., Platen, P., Mückenhoff, K., & Marek, W. (2010) · Journal of Breath Research · doi:10.1088/1752-7155/4/1/017105 · Cited by 25
Ferdinands, J. M., Crawford, C. A. G., Greenwald, R., Sickle, D. V., Hunter, E., & Teague, W. G. (2008) · Environmental Health · doi:10.1186/1476-069x-7-10 · Cited by 28
Hoffmeyer, F., Raulf, M., Lehnert, M., Kendzia, B., Bernard, S., Berresheim, H., Düser, M., Henry, J., Weiß, T., Koch, H. M., et al. (2012) · Journal of Toxicology and Environmental Health · doi:10.1080/15287394.2012.675303 · Cited by 25
“Collection of EBC General methodological recommendations on the collection and analysis of EBC were followed (Horvath et al. 2005).”
BackgroundMethodBime, C., Nguyen, J. M., & Wise, R. A. (2011) · Current Opinion in Pulmonary Medicine · doi:10.1097/mcp.0b013e32834db0f2 · Cited by 33
Fireman, E. M., Shtark, M., Priel, I. E., Shiner, R. J., Mor, R., Kivity, S., & Fireman, Z. (2007) · Inflammation · doi:10.1007/s10753-007-9020-8 · Cited by 25
Noninvasive monitoring of airway inflammation and steroid reduction in children with asthma
Zacharasiewicz, A. S., Erin, E. M., & Bush, A. (2006) · Current Opinion in Allergy and Clinical Immunology · doi:10.1097/01.all.0000225152.37403.21 · Cited by 25
Jouyban, A., Farajzadeh, M. A., Khoubnasabjafari, M., Jouyban-Gharamaleki, V., & Afshar Mogaddam, M. R. (2019) · Analytical Methods · doi:10.1039/c8ay02279h · Cited by 24
Rindlisbacher, B., Strebel, C., Guler, S. A., Kollár, A., Geiser, T., Fiedler, M., Leichtle, A. B., Bovet, C., & Funke-Chambour, M. (2017) · Journal of Breath Research · doi:10.1088/1752-7163/aa840a · Cited by 24
“EBC collection EBC was collected using the Ecoscreen EBC collecting device from Erich Jaeger, Hoechberg, Germany, according to the manufacturer’s instructions and the ATS/ERS methodological recommendations for EBC collection [9].”
BackgroundMethodUltrafine particles in airways: a novel marker of COPD exacerbation risk and inflammatory status
Klein, E. F., Adir, Y., Krencel, A., Peri, R., Vasserman, B., Fireman, E. M., & Kessel, A. (2019) · International Journal of COPD · doi:10.2147/copd.s187560 · Cited by 30
“All subjects were asked to avoid smoking and any type of inhaled medication prior to EBC collection, as recommended by EBC guidelines.(10) All subjects signed informed consent, and the study was approved by the Helsinki committees of the Bnai Zion and Carmel Medical Centers (Nos.”
MethodA novel system for the comprehensive collection of nonvolatile molecules from human exhaled breath
Chen, D., Bryden, W. A., & McLoughlin, M. J. (2020) · Journal of Breath Research · doi:10.1088/1752-7163/abba87 · Cited by 24
Hillas, G., Κostikas, Κ., Mantzouranis, K., BESSA, V., Kontogianni, K., Papadaki, G. F., Papiris, S. A., Alchanatis, M., Loukides, S., & Bakakos, P. (2011) · Respirology · doi:10.1111/j.1440-1843.2011.01984.x · Cited by 26
Ahmadzai, H., Cameron, B., Chui, J. J. Y., Lloyd, A. R., Wakefield, D., & Thomas, P. S. (2013) · Journal of Breath Research · doi:10.1088/1752-7155/7/4/046003 · Cited by 24
Analytical methods based on exhaled breath for early detection of lung cancer
de Castro, M. D. L., & Fernández-Peralbo, M. A. (2012) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2012.03.018 · Cited by 24
de la Riva-Velasco, E., Krishnan, S., & Dozor, A. J. (2012) · Journal of Asthma · doi:10.3109/02770903.2012.701363 · Cited by 26
Núñez‐Naveira, L., Mariñas‐Pardo, L., & Montero-Martínez, C. (2019) · Lung · doi:10.1007/s00408-019-00238-z · Cited by 24
IL-17 Promotes Nitric Oxide Production in Non-Small-Cell Lung Cancer
Nicola, S., Ridolfi, I., Rolla, G., Filosso, P. L., Giobbe, R., Boita, M., Culla, B., Bucca, C. B., Solidoro, P., & Brussino, L. (2021) · Journal of Clinical Medicine · doi:10.3390/jcm10194572 · Cited by 26
Real-Time High-Resolution Tandem Mass Spectrometry Identifies Furan Derivatives in Exhaled Breath
García‐Gómez, D., Bregy, L., Barrios‐Collado, C., Vidal‐de‐Miguel, G., & Zenobi, R. (2015) · Analytical Chemistry · doi:10.1021/acs.analchem.5b01509 · Cited by 24
Biomarkers in Paediatric Cystic Fibrosis Lung Disease
Ramsey, K. A., Schultz, A., & Stick, S. M. (2015) · Paediatric Respiratory Reviews · doi:10.1016/j.prrv.2015.05.004 · Cited by 29
Do, R., Bartlett, K. H., Chu, W., Dimich‐Ward, H., & Kennedy, S. (2007) · Respiratory Medicine · doi:10.1016/j.rmed.2007.10.002 · Cited by 24
IL-8 and pH Values in Exhaled Condensate after Antibiotics in Cystic Fibrosis Children
Bodini, A., D’Orazio, C., Peroni, D. G., Corradi, M., Zerman, L., Folesani, G., Assael, B. M., Boner, A. L., & PIACENTINI, G. L. (2007) · International Journal of Immunopathology and Pharmacology · doi:10.1177/039463200702000305 · Cited by 24
Airborne biomarker localization engine for open-air point-of-care detection
Ma, J., Laune, M., Li, P., Lu, J. N., Yue, J., Yu, Y., Mansur, Y., Ritwik, A. P. D., Peri, S., Cleary, J., et al. (2025) · Nature Chemical Engineering · doi:10.1038/s44286-025-00223-9 · Cited by 23
“Airborne biomarkers from human-exhaled breath condensates (EBCs) contain many physiologically relevant metabolites 7 , but many are very dilute and require sample enrichment for downstream sensors.”
Spaceflight induced changes in the human proteome
Kononikhin, A. S., Starodubtseva, N. L., Pastushkova, L. K., Kashirina, D. N., Fedorchenko, K. Y., Brhozovsky, A. G., Popov, I. A., Larina, I. M., & Николаев, Е. Н. (2016) · Expert Review of Proteomics · doi:10.1080/14789450.2017.1258307 · Cited by 23
“) are identifiable in EBC, and some of them are established markers of certain human disorders [54,119,120].”
BackgroundFaversani, A., Favero, C., Dioni, L., Pesatori, A. C., Bollati, V., Montoli, M., Musso, V., Terrasi, A., Fusco, N., Nosotti, M., et al. (2021) · Frontiers in Oncology · doi:10.3389/fonc.2021.643280 · Cited by 24
Exhaled breath analysis of non-volatile drugs: Towards clinical applications
Huang, H. M., Yang, J., Chen, T., Hu, L., Huan, T., Zhang, W., Zhang, K., & Li, X. (2024) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2024.117541 · Cited by 24
Hulo, S., Radauceanu, A., Chérot‐Kornobis, N., Howsam, M., Vacchina, V., de Broucker, V., Rousset, D., Grzebyk, M., Dziurla, M., Sobaszek, A., et al. (2015) · International Journal of Hygiene and Environmental Health · doi:10.1016/j.ijheh.2015.08.003 · Cited by 25
Turchiarelli, V., Schinkel, J., Molenkamp, R., Barbaro, M. P. F., Carpagnano, G. E., Spanevello, A., Lutter, R., Bel, E. H., & Sterk, P. J. (2011) · Allergy · doi:10.1111/j.1398-9995.2011.02600.x · Cited by 23
“Exhaled breath condensate was collected by Ecoscreen (Erich Jaeger GmbH, Hochberg, Germany) according to the ERS recommendations (22).”
Debley, J. S., Ohanian, A., Spiekerman, C. F., Aitken, M. L., & Hallstrand, T. S. (2010) · CHEST Journal · doi:10.1378/chest.10-0101 · Cited by 23
Exhaled breath 8-isoprostane as a marker of asthma severity
Piotrowski, W. J., Majewski, S., Marczak, J., Kurmanowska, Z., Górski, P., & Antczak, A. J. (2012) · Archives of Medical Science · doi:10.5114/aoms.2012.28639 · Cited by 23
Independent Information of Nonspecific Biomarkers in Exhaled Breath Condensate
Dressel, H., Müller, F., Fischer, R., Römmelt, H., Hohlfeld, J. M., Behr, J., Huber, R. M., Nowak, D., & Jörres, R. A. (2010) · Respiration · doi:10.1159/000319945 · Cited by 25
“Exhaled breath condensate (EBC) contains numerous biomarkers that have the potential to play a role in the diagnosis and monitoring of disease progression in respiratory disorders or in the evaluation of therapeutic interventions [”
BackgroundAhmadzai, H., Wakefield, D., & Thomas, P. S. (2011) · Inflammopharmacology · doi:10.1007/s10787-011-0079-3 · Cited by 23
Is NMR-based metabolomic analysis of exhaled breath condensate accurate?
Izquierdo-García, J. L., Peces‐Barba, G., Heili-Frades, S. B., Diaz, R., Want, E. J., & Ruíz-Cabello, J. (2011) · European Respiratory Journal · doi:10.1183/09031936.00094010 · Cited by 23
The Value of Sputum 8-Isoprostane in Detecting Oxidative Stress in Mild Asthma
Louhelainen, N., Rytilä, P. H., Obase, Y., Mäkelä, M. J., Haahtela, T., Kinnula, V. L., & Pelkonen, A. S. (2008) · Journal of Asthma · doi:10.1080/02770900701840261 · Cited by 23
“Condensate is still poorly standardized and the samples exhibit high variability in airway diseases and their progression (25, 26).”
BackgroundElevated exhaled leukotriene B4 in the small airway compartment in children with asthma
Trischler, J., Müller, C.-M., Könitzer, S., Prell, E., Korten, I., Unverzagt, S., & Lex, C. (2015) · Annals of Allergy Asthma & Immunology · doi:10.1016/j.anai.2014.11.022 · Cited by 23
Gomà, A., de Lluis, R., Roca-Ferrer, J., Lafuente, J., & Picado, C. (2017) · Environmental Research · doi:10.1016/j.envres.2016.10.030 · Cited by 23
A Technology Roadmap of Smart Biosensors From Conventional Glucose Monitoring Systems
Shende, P. K., Sahu, P., & Gaud, R. S. (2017) · Therapeutic Delivery · doi:10.4155/tde-2017-0012 · Cited by 24
Effect of temperature control on the metabolite content in exhaled breath condensate
Zamuruyev, K. O., Borras, E., Pettit, D. R., Aksenov, A. A., Simmons, J. D., Weimer, B. C., Schivo, M., Kenyon, N. J., Delplanque, J., & Davis, C. E. (2017) · Analytica Chimica Acta · doi:10.1016/j.aca.2017.12.025 · Cited by 25
Stachowiak, Z., Wojsyk‐Banaszak, I., Jończyk‐Potoczna, K., Narożna, B., Langwiński, W., Kycler, Z., Sobkowiak, P., Bręborowicz, A., & Szczepankiewicz, A. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9061887 · Cited by 23
“We used the Turbo DECCS device (Medivac) to collect EBC according to ERS recommendation [25].”
MethodIdrose, N. S., Tham, R., Lodge, C., Lowe, A. J., Bui, D., Perret, J. L., Vicendese, D., Newbigin, E., Tang, M. L. K., Aldakheel, F. M., et al. (2020) · Allergy · doi:10.1111/all.14566 · Cited by 23
“The median (range) age of the study participants was 21 years (6-71).”
BackgroundExhaled breath condensate pH is influenced by respiratory droplet dilution
Bikov, A., Gálffy, G., Tamási, L., Lázár, Z., Losonczy, G., & Horváth, I. (2012) · Journal of Breath Research · doi:10.1088/1752-7155/6/4/046002 · Cited by 23
“Exhaled breath condensate (EBC) is a completely noninvasive method for airway sampling [6].”
MethodSelf-citationLozano, J., Cruz, M. J., Piquer, M., Giner, M. T., & Plaza, A. M. (2014) · International Archives of Allergy and Immunology · doi:10.1159/000368832 · Cited by 23
Heinicke, I., Boehler, A., Rechsteiner, T. J., Bogdanova, A. Y., JELKMANN, W. E. B., Höfer, M., Rawlings, P., Araneda, O. F., Behn, C., Gassmann, M., et al. (2009) · European Journal of Applied Physiology · doi:10.1007/s00421-009-1014-9 · Cited by 24
“Collection of exhaled breath condensate (EBC) offers the opportunity to measure these markers in the lower respiratory tract in a noninvasive manner (Horvath et al. 2005).”
BackgroundExhaled breath condensate pH in infants and children with acute and recurrent wheezy bronchitis
Vogelberg, C., Würfel, C., Knoetzsch, A., Kahlert, A., Range, U., & Leupold, W. (2007) · Pediatric Pulmonology · doi:10.1002/ppul.20712 · Cited by 23
The effect of gas standardisation on exhaled breath condensate pH: Fig. 1—
Borrill, Z., Smith, J. A., Naylor, J., Woodcock, A. A., & Singh, D. (2006) · European Respiratory Journal · doi:10.1183/09031936.06.00026706 · Cited by 23
Catching breath: monitoring airway inflammation using exhaled breath condensate
Holz, O. (2005) · European Respiratory Journal · doi:10.1183/09031936.05.00071305 · Cited by 24
Idiopathic pulmonary fibrosis (IPF): Diagnostic routes using novel biomarkers
Bartold, K., Iskierko, Z., Sharma, P. S., Lin, H., & Kutner, W. (2024) · Biomedical Journal · doi:10.1016/j.bj.2024.100729 · Cited by 27
“The correlation between EBC biomarkers and pulmonary inflammation has been widely reported [61,62].”
BackgroundIdentification of Intrinsic Airway Acidification in Pulmonary Tuberculosis
Ngamtrakulpanit, L., Yu, Y., Adjei, A. A., Amoah, G., Gaston, B., & Hunt, J. F. (2010) · Global Journal of Health Science · doi:10.5539/gjhs.v2n1p106 · Cited by 22
Self-citationZhou, F., Chen, J., Tao, G., Zhu, M.-X., Xie, W., & Cao, X. (2014) · Respiratory Medicine · doi:10.1016/j.rmed.2014.08.003 · Cited by 22
“Various biomarkers in EBC have been identified and evaluated in inflammation and oxidative stress [5] and [6].”
BackgroundInduced Sputum, Exhaled NO, and Breath Condensate in Occupational Medicine
Chérot‐Kornobis, N., Hulo, S., de Broucker, V., Hassoun, S. M., Lepage, N., Edmé, J. L., & Sobaszek, A. (2012) · Journal of Occupational and Environmental Medicine · doi:10.1097/jom.0b013e31825a34b3 · Cited by 24
Lim, M. Y., & Thomas, P. S. (2013) · International Journal of Chronic Diseases · doi:10.1155/2013/578613 · Cited by 27
“9% of EBC comprising water vapour [67], concentrations of the mediators of interest can sometimes be close to or below the detection limit of the appropriate assays; thus, assays of sufficient sensitivity are needed to effectively measure biomarkers in EBC [71, 77].”
BackgroundInfluentialTrischler, J., Merkel, N., Könitzer, S., Müller, C.-M., Unverzagt, S., & Lex, C. (2012) · Respiratory Research · doi:10.1186/1465-9921-13-14 · Cited by 22
“Exhaled breath condensate (EBC) is a well-known method to collect inflammation mediators and other soluble particles in exhaled breath [10].”
MethodPharmacotherapy of Patients with Mild Persistent Asthma: Strategies and Unresolved Issues
Montuschi, P. (2011) · Frontiers in Pharmacology · doi:10.3389/fphar.2011.00035 · Cited by 23
Analysis of Airborne Biomarkers for Point-of-Care Diagnostics
Fung, A. O., & Mykhaylova, N. (2014) · SLAS TECHNOLOGY · doi:10.1177/2211068213517119 · Cited by 22
Biomarkers of oxidative-stress and inflammation in exhaled breath condensate from hospital cleaners
Casimirri, E., Stendardo, M., Bonci, M., Andreoli, R., Bottazzi, B., Leone, R., Schito, M., Vaccari, A., Papi, A. A., Contoli, M., et al. (2015) · Biomarkers · doi:10.3109/1354750x.2015.1118541 · Cited by 22
Aliberti, S., Morlacchi, L. C., Faverio, P., Fernandez-Botrán, R., Cosentini, R., Mantero, M., Peyrani, P., Ramírez, J. A., Bordón, J., & Blasi, F. (2016) · Pneumonia · doi:10.1186/s41479-016-0009-7 · Cited by 22
Kruizinga, M. D., Birkhoff, W. A. J., van Esdonk, M. J., Klarenbeek, N. B., Cholewinski, T., Nelemans, T., Dröge, M. J., Cohen, A. F., & Zuiker, R. G. J. A. (2019) · British Journal of Clinical Pharmacology · doi:10.1111/bcp.14156 · Cited by 22
Analytical Methods for Breath Investigation
Ligor, T. (2009) · Critical Reviews in Analytical Chemistry · doi:10.1080/10408340802569498 · Cited by 22
“After condensation at sub-ambient temperature, exhaled breath condensate has to be analyzed by means of LC/MS or immuno assays (15).”
MethodWang, C.-J., Yang, N.-H., Liou, S., & Lee, H. (2010) · Talanta · doi:10.1016/j.talanta.2010.07.015 · Cited by 22
Ziętkowski, Z., Skiepko, R., Tomasiak, M. M., Lenczewska, D., & Bodzenta‐Łukaszyk, A. (2010) · International Archives of Allergy and Immunology · doi:10.1159/000319205 · Cited by 22
“Forced expiratory volume in 1 s (FEV 1 ) was evaluated. Before the examination the patients did not take any medications that could change spirometry results. The highest value of three technically satisfactory attempts was recorded.EBC was collected by using a commercially available condenser (EcoScreen; Erich Jaeger…”
MethodThe effect of temperature on exhaled breath condensate collection
Vyas, A., Zhang, Q., Gunaratne, S., Lee, W., Lin, J.-L., Lin, J.-S., Warwick, G., & Thomas, P. S. (2012) · Journal of Breath Research · doi:10.1088/1752-7155/6/3/036002 · Cited by 22
Exhaled eicosanoid profiles in children with atopic asthma and healthy controls
Głowacka, E., Jedynak‐Wąsowicz, U., Sanak, M., & LIS, G. J. (2012) · Pediatric Pulmonology · doi:10.1002/ppul.22615 · Cited by 22
“EBC collections were made in accordance with the standards of ATS/ERS using an Eco Screen I Jaeger (Hoechberg GmbH, Wuerzburg, Germany) condenser equipped with a saliva trap.(15) The examined patients were instructed to breathe at tidal volume for 20 min through a mouthpiece connected to a set of one-way Fig.”
MethodFerrari, P. A., Salis, C. B., & Macciò, A. (2025) · Life · doi:10.3390/life15050683 · Cited by 22
“Various methodologies have been implemented to prevent contamination, including using saliva traps, swan neck-shaped collection tubes, pre-collection mouth rinsing with 4.5% sodium bi-carbonate solution, and encouraging subjects to perform periodic swallowing during sampling [75,102].”
Influentialvon Jagwitz, M., Peßler, F., Akmatov, M. K., Li, J., Range, U., & Vogelberg, C. (2011) · Journal of Allergy and Clinical Immunology · doi:10.1016/j.jaci.2011.03.005 · Cited by 21
Severe bronchial asthma in children: a review of novel biomarkers used as predictors of the disease
Uwaezuoke, S. N., Ayuk, A. C., & Eze, J. N. (2018) · Journal of Asthma and Allergy · doi:10.2147/jaa.s149577 · Cited by 26
“EBC is obtained after condensation of EB with a commercially available refrigerating device based on American Thoracic Society/European Respiratory Society guidelines.(97) Generally, it has been documented that the following molecules in EBC are elevated in severe asthma: leukotrienes, lipoxins, arachidonic acid metab…”
Exhaled Breath Condensate in Childhood Asthma: A Review and Current Perspective
Bannier, M. A. G. E., Rosias, P. P. R., Jöbsis, Q., & Dompeling, E. (2019) · Frontiers in Pediatrics · doi:10.3389/fped.2019.00150 · Cited by 21
Turcu, V., Wild, P. P., Hemmendinger, M., Sauvain, J., Bergamaschi, E., Hopf, N. B., & Canu, I. G. (2022) · Toxics · doi:10.3390/toxics10050258 · Cited by 21
“Although a collection time of 10 min has been recommended, further research on the effect of collection time and temperature versus expiratory volume is needed [17].”
BackgroundSize-Resolved Elemental Composition of Respiratory Particles in Three Healthy Subjects
Prussin, A. J., Cheng, Z., Leng, W., China, S., & Marr, L. C. (2023) · Environmental Science & Technology Letters · doi:10.1021/acs.estlett.3c00156 · Cited by 22
Next Generation Sequencing for miRNA Detection on the Exhaled Breath Condensate: A Pilot Study
Cherchi, R., Cusano, R., Orrù, S., Ferrari, P. A., Massidda, M., Fotia, G., Matteis, S. D., & Cocco, P. (2023) · Epigenetics Insights · doi:10.1177/25168657231160985 · Cited by 21
IL-5, IL-8 and MMP -9 levels in exhaled breath condensate of atopic and nonatopic asthmatic children
Türkeli, A., Yılmaz, Ö., Taneli, F., Dinç, G., Kanık, E. T., Kızılkaya, M. H., Gözükara, C., & Yüksel, H. (2015) · Respiratory Medicine · doi:10.1016/j.rmed.2015.04.004 · Cited by 21
Muñoz, X., Bustamante, V., López-Campos, J. L., Cruz, M. J., & Barreiro, E. (2015) · International Archives of Allergy and Immunology · doi:10.1159/000371849 · Cited by 21
Asano, T., Takemura, M., Kanemitsu, Y., Yokota, M., Fukumitsu, K., Takeda, N., Ichikawa, H., Hijikata, H., Uemura, T., Takakuwa, O., et al. (2017) · Journal of Asthma · doi:10.1080/02770903.2017.1332203 · Cited by 21
“To assess lower airway inflammation, several methods, including an induced sputum [8], bronchoscopic mucosal biopsy, bronchoalveolar lavage [9], exhaled breath condensate [10] and fractional exhaled nitric oxide (FeNO50) measurements [11] are available.”
MethodPortable exhaled breath condensate metabolomics for daily monitoring of adolescent asthma
Schmidt, A. J., Borras, E., Nguyen, A. P., Yeap, D., Kenyon, N. J., & Davis, C. E. (2019) · Journal of Breath Research · doi:10.1088/1752-7163/ab35b5 · Cited by 21
“EBC is considered a filtrate of blood, reflects the lung epithelia [1], and has been used for diseases such as asthma [5], chronic obstructive pulmonary disease (COPD) [6], pneumonia [7] or cystic fibrosis [8].”
BackgroundProteome Profiling of the Exhaled Breath Condensate after Long-Term Spaceflights
Kononikhin, A. S., Brzhozovskiy, A. G., Ryabokon, A. M., Fedorchenko, K. Y., Zakharova, N. V., Спасский, А. И., Popov, I. A., Ilyin, V., Solovyeva, Z., Pastushkova, L. K., et al. (2019) · International Journal of Molecular Sciences · doi:10.3390/ijms20184518 · Cited by 24
Application of Proteomics in Sarcoidosis
Guerrero, C. R., Maier, L. A., Griffin, T. J., Higgins, L., Najt, C. P., Perlman, D., & Bhargava, M. (2020) · American Journal of Respiratory Cell and Molecular Biology · doi:10.1165/rcmb.2020-0070ps · Cited by 23
Cheng, S.-L. (2021) · Diagnostics · doi:10.3390/diagnostics11010083 · Cited by 24
“However, the measurement of mediators within exhaled breath condensate is not very reproducible, and the technique warrants improvement [56].”
BackgroundSantos-Rivera, M., Woolums, A. R., Thoresen, M., Meyer, F., & Vance, C. K. (2022) · Molecules · doi:10.3390/molecules27020549 · Cited by 23
“While the condensing water absorbs water-soluble volatile compounds, respiratory droplets containing non-volatile molecules compose a minor fraction of the condensate [29,30].”
BackgroundInfluentialSin, D. D., Man, S. P., Marciniuk, D. D., Ford, G. T., Fitzgerald, M., Wong, E. Y. L., York, E. L., Mainra, R., Ramesh, W., Melenka, L., et al. (2006) · BMC Pulmonary Medicine · doi:10.1186/1471-2466-6-3 · Cited by 22
Biomarkers of lung injury after one‐lung ventilation for lung resection
Bastin, A. J., Sato, H., Davidson, S. J., Quinlan, G. J., & Griffiths, M. (2010) · Respirology · doi:10.1111/j.1440-1843.2010.01870.x · Cited by 21
A mobile instrumentation platform to distinguish airway disorders
Schivo, M., Seichter, F., Aksenov, A. A., Pasamontes, A., Peirano, D. J., Mizaikoff, B., Kenyon, N. J., & Davis, C. E. (2013) · Journal of Breath Research · doi:10.1088/1752-7155/7/1/017113 · Cited by 22
Sawyer, K., Samet, J. M., Ghio, A. J., Pleil, J. D., & Madden, M. C. (2008) · Journal of Breath Research · doi:10.1088/1752-7155/2/3/037019 · Cited by 23
“Noseclips were not used during the 15 min collection period as per some recommendations (Horvath et al 2005) partially due to possible subject discomfort which in turn can potentially affect how rapidly subjects achieve normal TV and breath frequency (BF).”
MethodGoldoni, M., Corradi, M., Mozzoni, P., Folesani, G., Alinovi, R., Pinelli, S., Andreoli, R., Pigini, D., Tillo, R., Filetti, A., et al. (2013) · Journal of Breath Research · doi:10.1088/1752-7155/7/1/017101 · Cited by 27
“Despite the opinion of ATS/ERS Task Force [27], the deaeration of a sample cannot completely ensure homogeneous removal of CO2, thereby contributing to inter-subject variability of pH.”
BackgroundAmbient temperature impacts on pH of exhaled breath condensate
Koczulla, A. R., Noeske, S., Herr, C., Dette, F. G., Pinkenburg, O., Schmid, S., Jörres, R. A., Vogelmeier, C. F., & Bals, R. (2009) · Respirology · doi:10.1111/j.1440-1843.2009.01664.x · Cited by 21
Exhaled cytokines in systemic lupus erythematosus with lung involvement
Nielepkowicz‐Goździńska, A., Fendler, W. M., Robak, E., Kulczycka, L., Górski, P., Pietras, T., Brzeziańska‐Lasota, E., & Antczak, A. J. (2013) · Polskie Archiwum Medycyny Wewnętrznej · doi:10.20452/pamw.1676 · Cited by 21
Transition and post-transition metals in exhaled breath condensate
Ghio, A. J., Madden, M. C., & Esther, C. R. (2018) · Journal of Breath Research · doi:10.1088/1752-7163/aaa214 · Cited by 21
“The EBC is a more dilute sampling of the epithelial lining fluid [27].”
BackgroundThompson, C. M., & Grafström, R. (2007) · Journal of Toxicology and Environmental Health · doi:10.1080/15287390701598259 · Cited by 22
“Recent reviews of exhaled breath markers indicate that many compounds are currently being explored as markers of airway physiology and pathophysiology—including nitric oxide derivatives, thiols, S -nitrosothiols, leukotrienes, interleukins, peroxides, and others ( Horvath et al, 2005 ; Liu & Thomas, 2005 ).”
BackgroundMethodInfluentialEvidence for primary laryngeal inhalant allergy: a randomized, double‐blinded crossover study
Roth, D. F., Abbott, K. V., Carroll, T. L., & Ferguson, B. J. (2013) · International Forum of Allergy & Rhinology · doi:10.1002/alr.21051 · Cited by 20
Zhou, X., Qi, M., Li, K., Xue, Z., & Wang, T. (2023) · Cell Reports Physical Science · doi:10.1016/j.xcrp.2023.101678 · Cited by 22
The Future of Pulmonary Function Testing
MacIntyre, N. R. (2012) · Respiratory Care · doi:10.4187/respcare.01422 · Cited by 21
Möller, W., Heimbeck, I., Weber, N., Saba, G. K., Körner, B., Neiswirth, M., & Kohlhäufl, M. J. (2010) · Journal of Aerosol Medicine and Pulmonary Drug Delivery · doi:10.1089/jamp.2009.0764 · Cited by 21
Metadata
- Authors
- Horváth, I., Hunt, J., & Barnes, P. J.
- Affiliations
- Országos Korányi Tbc és Pulmonológiai Intézet (HU)
- University of Virginia (US)
- Institute of Cancer Research (GB)
- Venue
- European Respiratory Journal
- Publisher
- European Respiratory Society (ERS)
- Published
- 2005-09
- Volume
- 26
- Issue
- 3
- Pages
- 523-548
- ISSN
- 0903-1936 (ISSN-L)
- 1399-3003
- OpenAlex
- W1989935661
- Cited by
- 1,284
- References
- 280
- Citations by year
- 2026: 43 · 2025: 28 · 2024: 35 · 2023: 30 · 2022: 51 · 2021: 43 · 2020: 47 · 2019: 50 · 2018: 53 · 2017: 61 · 2016: 60 · 2015: 68 · 2014: 76 · 2013: 82 · 2012: 107
- FWCI
- 48.86
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- European Respiratory Journal (publishedVersion)
- PubMed (publishedVersion)
- Research Publications (Maastricht University) (publishedVersion, other-oa)
- Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion)
- Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
- Radboud Repository (Radboud University) (submittedVersion)
- OpenAgrar (submittedVersion)
- Padua Research Archive (University of Padova) (submittedVersion)
- Research Publications (Maastricht University) (submittedVersion, other-oa)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Chemical Sensor Technologies
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Asthma and respiratory diseases
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Chronic Obstructive Pulmonary Disease (COPD) Research
- Keywords
- exhaled breath condensate, adenosine, hydrogen peroxide, isoprostanes, leukotrienes, nitrogen oxides, cytokines, reproducibility, longitudinal study, lung disease
- MeSH terms
- Breath Tests, Humans, Lung Diseases, Reproducibility of Results, Biomarkers, Oxidative Stress
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.